Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceCOLLASTAT ABSORABLE COLLAGEN HEMOSTATIC SPONGE & MICROFIBRILLAR FORM
Generic NameAgent, absorbable hemostatic, collagen based
Regulation Number878.4490
ApplicantIntegra LifeSciences Corporation
311 Enterprise Dr.
Planisboro, NJ 08536
PMA NumberP810006
Supplement NumberS047
Date Received08/14/2013
Decision Date04/04/2014
Product Code LMF 
Advisory Committee General & Plastic Surgery
Supplement Type135 Review Track For 30-Day Notice
Supplement Reason Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
APPROVAL FOR THE FOLLOWING PROPOSED CHANGES: 1) CHANGE IN CLEANING DETERGENT FROM METRIZYME (FORMULATED ENZYMATIC CLEANER) TO CIP-100 (FORMULATED ALKALINE CLEANER); 2) ESTABLISHMENT OF AN EQUIPMENT DIRTY HOLD TIME (DHT) OF 24-HOURS; 3) UTILIZATION OF 70% ISOPROPYL ALCOHOL (IPA), AS REQUIRED; 4) UTILIZATION OF SODIUM LAURYL SULFATE (SLS), AS REQUIRED; AND5) ESTABLISHING THE USE OF TOTAL ORGANIC CARBONS (TOCS) AS A METHOD TO TEST FOR RESIDUES.
-
-