| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | LCS(R) TOTAL KNEE SYSTEM DESIGN MODIFICATION TO ROTATING PLATFORM BEARINGS |
| Generic Name | Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing |
| Applicant | Depuy, Inc. P.O. Box 988 700 Orthopaedic Dr. Warsaw, IN 46581 |
| PMA Number | P830055 |
| Supplement Number | S069 |
| Date Received | 02/14/2000 |
| Decision Date | 03/16/2000 |
| Product Code |
NJL |
| Advisory Committee |
Orthopedic |
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for a design modification to the proximal articular surface of the LCS Rotating Platform Bearings to match the geometry of the PFC sigma femoral components that are implanted using slightly different distal cuts than the current LCS femoral components. |
|
|