| Device | MEDTRONIC(R) CAPSURE(R) SP TRANSVENOUS PACING LEAD |
| Generic Name | permanent pacemaker Electrode |
| Regulation Number | 870.3680 |
| Applicant | Medtronic Cardiac Rhythm Disease Management 8200 Coral Sea St. N.E Mounds View, MN 55112 |
| PMA Number | P830061 |
| Supplement Number | S025 |
| Date Received | 01/30/1995 |
| Decision Date | 10/31/1995 |
| Product Code |
DTB |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement APPROVAL FOR A CHANGE IN THE STERILIZATION PROCESS FROM THE USE OF 12 PERCENT ETHYLENE OXIDE (ET0), 88 PERCENT CFC, TO 100 PERCENT ETO |