| Device | LifePulse High Frequency Ventilator (LifePulse HFV) Model 204 |
| Generic Name | VENTILATOR, HIGH FREQUENCY |
| Applicant | Bunnell, Inc. 436 Lawndale Dr. Salt Lake City, UT 84115 |
| PMA Number | P850064 |
| Supplement Number | S058 |
| Date Received | 04/13/2026 |
| Decision Date | 05/07/2026 |
| Product Code |
LSZ |
| Advisory Committee |
Anesthesiology |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement This is a 30-DN to provide the change in power cord testing as a result of a CAPA investigation into a miswiring of one single power cord. the testing is adequate to test the incoming inspection for cross-wired power cords and not as a complete safety check. |