Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceSCIMED COYOTE PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA) CATHETER
Generic NameCatheters, transluminal coronary angioplasty, percutaneous
Regulation Number870.5100
ApplicantScimed Life Systems, Inc.
One Scimed Pl.
Maple Grove, MN 55311
PMA NumberP860019
Supplement NumberS151
Date Received12/14/1998
Decision Date01/11/1999
Reclassified Date10/08/2010
Product Code LOX 
Advisory Committee Cardiovascular
Supplement TypeReal-Time Process
Supplement Reason Labeling Change - Indications/instructions/shelf life/tradename
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
Approval for the reduction in the rated burst pressure from 16 to 14 atm for all available lengths (20, 30, and 40mm) of the 2.0mm and 2.5mm diameter balloon sizes of the COYOTE(TM) PTCA catheter.
-
-