| Device | DIASORIN ETI-HA-IGMK PLUS |
| Generic Name | Hepatitis a test (antibody and igm antibody) |
| Regulation Number | 866.3310 |
| Applicant | DiaSorin, Inc. 1951 Northwestern Ave. Stillwater, MN 55082 |
| PMA Number | P890014 |
| Supplement Number | S002 |
| Date Received | 11/24/1997 |
| Decision Date | 09/16/1998 |
| Reclassified Date | 02/09/2006 |
| Product Code |
LOL |
| Advisory Committee |
Microbiology |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for a change in the Enzyme tracer and Tracer Diluent from a IgG (sheep) anti-HAV-HRP to a Fab anti-HAV (Mab)-HRP; a chang ein the Positive Control material to a human source sera; the elimination of the mercury preservative in the reagents; a change in the incubation times from an overnight procedure to a four and one-half hour incubation; the addition of an indicator dye; and the addition of a Calibrator reagent. The device, as modified, will be marketed under the trade name DiaSorin ETI-HA-IGMK PLUS and is intended for use in the qualitative determination of IgM antibodies to hepatitis A virus (IgM anti-HAV) in human serum or plasma. This assay is indicated for use as an aid in the diagnosis and monitoring of acute or recent hepatitis A virus infection. |