| Device | MEDSTREAM AND CODMAN 3000 IMPLANTABLE INFUSION PUMP |
| Generic Name | Pump, infusion, implanted, programmable |
| Applicant | Intera Oncology 180 Wells Ave., Suite 300a Newton, MA 02459 |
| PMA Number | P890055 |
| Supplement Number | S061 |
| Date Received | 09/08/2015 |
| Decision Date | 10/08/2015 |
| Product Code |
LKK |
| Advisory Committee |
General Hospital |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement CHANGES TO THE STERILIZATION PROCESS: 1) STEAM INJECTION RATE CURRENTLY AT 20 MBAR/MIN WILL CHANGE TO 5 MBAR/MIN; 2) GAS INJECTION RATE CURRENTLY AT 25 MBAR/MIN WILL CHANGE TO 20 MBAR/MIN; 3) MINIMUM PRIMARY DEGASSING TIME OF 10 HOURS WILL CHANGE TO 8 HOURS; 4) THE HANDLES FROM THE STERILIZATION SHIPPER, NUMBER 197548018, WILL BE REMOVED; 5) ADD A SECOND ETHYLENE OXIDE (EO) STERILIZATION CHAMBER (CHAMBER #4; 6) A MINIMUM EO STERILIZATION LOAD IN CHAMBER #3 (EXISTING CHAMBER) AND CHAMBER #4; AND 7) A 2X EO STERILIZATION FOR BOTH CHAMBER #3 AND #4. |
|
|