| Device | INTEGRA ARTIFICAL SKIN DERMAL REGENERATION TEMPLATE |
| Generic Name | Dressing, wound and burn, interactive |
| Applicant | Integra Lifesciences Corp. 105 Morgan Ln. Plainsboro, NJ 08536 |
| PMA Number | P900033 |
| Supplement Number | S056 |
| Date Received | 08/19/2016 |
| Decision Date | 09/15/2016 |
| Product Code |
MGR |
| Advisory Committee |
General & Plastic Surgery |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Replacement of a portion of the water deionization (DI) system, located in the Collagen Manufacturing Center. Integra will replace a portion of the DI system with a new Evoqua IonRight GEN3 system. The change is being made to lower operating expenses, as well as reduce the frequency of service of the system by reducing the frequency at which the vessels are shipped off site for regeneration. |