Device | SWEETIP RX/SWEET PICOTIP RX/FLEXTEND |
Generic Name | permanent pacemaker Electrode |
Regulation Number | 870.3680 |
Applicant | BOSTON SCIENTIFIC 4100 Hamline Avenue North St. Paul, MN 55112 |
PMA Number | P960006 |
Supplement Number | S007 |
Date Received | 11/13/2002 |
Decision Date | 12/13/2002 |
Product Code |
DTB |
Advisory Committee |
Cardiovascular |
Supplement Type | 30-Day Notice |
Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
Expedited Review Granted? | No |
Combination Product | No |
Approval Order Statement CHANGE IN THE MANUFACTURING PROCESS TO THE PARAMETRIC RELEASE METHOD, WHICH INVOLVES STERILE PRODUCT RELEASE BASED ON CONFORMANCE TO PRE-DEFINED STERILIZATION PARAMETERS RATHER THAN A RELEASE BASED ON DEMONSTRATING NO GROWTH IN BIOLOGICAL INDICATOR. |