| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | Activa Deep Brain Stimulation Therapy System |
| Generic Name | Stimulator, electrical, implanted, for parkinsonian tremor |
| Applicant | Medtronic, Inc. 7000 Central Ave., NE Minneapolis, MN 55432 |
| PMA Number | P960009 |
| Supplement Number | S284 |
| Date Received | 06/07/2017 |
| Decision Date | 09/01/2017 |
| Product Code |
MHY |
| Advisory Committee |
Neurology |
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for a minor design change (lowering values of resistances R54 and R56 in the analog board of the INSR) to the Implantable Neurostimulator Recharger (INSR) used in the Restore Spinal Cord Stimulation and Activa Deep Brain Stimulation therapy systems. |
|
|