Device | InterStim Therapy System and Verify Evaluation System (SNS Urinary) |
Generic Name | Stimulator, electrical, implantable, for incontinence |
Regulation Number | 876.5270 |
Applicant | MEDTRONIC NEUROMODULATION 7000 CENTRAL AVENUE, N.E. MINNEAPOLIS, MN 55432 |
PMA Number | P970004 |
Supplement Number | S329 |
Date Received | 01/25/2021 |
Decision Date | 08/30/2021 |
Product Code |
EZW |
Advisory Committee |
Gastroenterology/Urology |
Supplement Type | 135 Review Track For 30-Day Notice |
Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
Expedited Review Granted? | No |
Combination Product | No |
Approval Order Statement Approval for a cleaning process change at DSM Biomedical in Berkeley California, a supplier to Medtronic Neuromodulation, that involves the addition of an option to replace solvent cleaning the equipment between material lots with a process that uses the next lot of material to purge the system (the first 30 seconds of material coming out of this portion of the manufacturing line is discarded). The change involves the continuous reactor system to be purged between production batches of same hardness grade and purged and cleaned between production batches of different hardness. |