| Device | CHILLI COOLED ABLATION SYSTEM |
| Generic Name | Cardiac ablation percutaneous catheter |
| Applicant | Boston Scientific Corp 150 Baytech Dr. San Jose, CA 95134 |
| PMA Number | P980003 |
| Supplement Number | S025 |
| Date Received | 10/18/2010 |
| Decision Date | 04/08/2011 |
| Withdrawal Date | 02/28/2024 |
| Product Code |
LPB |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement APPROVAL FOR TEN DESIGN CHANGES TO THE CIRCUCOOL MODEL 8005 PUMP SYSTEM TO ADDRESS STOPPAGE AND MOTOR NOISE COMPLAINTS. THESE CHANGES INCLUDE: 1) CHANGING A SCHOTKY DIODE FROM ON SEMICONDUCTOR¿S 1N5822 (40V MAXIMUM REVERSE VOLTAGE) TO ON SEMICONDUCTOR¿S MBR360 (60V MAXIMUM REVERSE VOLTAGE). 2) ADDITION OF A 33.3V VOLTAGE SUPPRESSOR (ON SEMICONDUCTOR P6KE39AG) ACROSS MOTOR TERMINALS AT THE MOTOR. 3) DIMENSIONAL AND TOLERANCE REVISIONS TO THE ROLLER SLEEVE, SHAFT ROTOR AND ROLLER COMPONENTS. |