| Device | MEDTRONIC ICD'S |
| Generic Name | Implantable cardioverter defibrillator (non-CRT) |
| Applicant | Medtronic Cardiac Rhythm Disease Management 8200 Coral Sea St. N.E Ms Mvs11 Mounds View, MN 55112 |
| PMA Number | P980016 |
| Supplement Number | S115 |
| Date Received | 11/16/2007 |
| Decision Date | 12/04/2007 |
| Product Code |
LWS |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement APPROVAL FOR MINOR DESIGN CHANGES THAT ADD TWO ALTERNATE TANTALUM CAPACITORS (XTC017, XTC018) WITH INCREASED VOLTAGE RATINGS TO APPROVED DEVICE COMPONENTS. THE DESIGN MODIFICATIONS ARE INTENDED TO REDUCE MANUFACTURING YIELD LOSS AND ASSOCIATED REWORK. |