 Class 2 Device Recall BD MAX GBSsee related information
Date Initiated by FirmMay 24, 2011
Date PostedSeptember 16, 2011
Recall Status1 Terminated 3 on October 12, 2011
Recall NumberZ-3220-2011
Recall Event ID 59359
Product Classification System, nucleic acid amplification test, dna, methicillin resistant staphylococcus aureus, direct sp - Product Code NQX
ProductBD MAX GBS Catalog Number 441772 For In Vitro Diagnostic Use, For use with the BD MAX system.The Sample Preparation Reagent kit consists of 24 (2 x 12) 2D barcode tubes which verifies against the number entered in the worklist with the handheld barcode reader. The BD MAX System is a qualitative in vitro diagnostic test obtained from vaginal and rectal swab specimens from antepartum pregnant women designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures after incubation for greater than or equal to 18 hours. The BD MAX DNA extraction kit is used with the BD MAX System to extract DNA from Gram-negative bacteria and viruses, which may be present in clinical specimens.
Code Information Lot numbers 103271, 103272, 110071, 110072, 110391, 110401, 110521, 110522, 110561, 110562, 110181, 110611, 110941, 110942, 111111
Recalling Firm/
Manufacturer
Becton Dickinson & Co.
BD Diagnostic Systems
7 Loveton Circle
Sparks MD 21152-0999
For Additional Information ContactJohn R. Gerlich
410-316-4000
Manufacturer Reason
for Recall
Medical device bar code reader does not function as intended and could result in associating clinical test results to wrong patient.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionBD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated May 2011, to all affected customers. The letter identified the products, the problem, and the action to be taken by the customers. The letter advised customers to discontinue use of the Sample Preparation Reagent tubes in the referenced kits and discard any remaining products. Customers were requested to complete an attached reply form to acknowledge their receipt of the notification and fax it to (410) 316-4258. For further assistance regarding replacements contact BD Customer Service Dept. 1-800-675-0908. For all other inquiries, contact BD Technical Services Dept. 1-800-638-8663.
Quantity in Commerce61 kits
DistributionWorldwide Distribution-USA (nationwide) including the states of CA, KY, MD, MI, NC, NY, TX, and WI and the country Belgium.