|
Class 2 Device Recall BD MAX GBS |
|
Date Initiated by Firm |
May 24, 2011 |
Date Posted |
September 16, 2011 |
Recall Status1 |
Terminated 3 on October 12, 2011 |
Recall Number |
Z-3220-2011 |
Recall Event ID |
59359 |
Product Classification |
System, nucleic acid amplification test, dna, methicillin resistant staphylococcus aureus, direct sp - Product Code NQX
|
Product |
BD MAX GBS Catalog Number 441772 For In Vitro Diagnostic Use, For use with the BD MAX system.The Sample Preparation Reagent kit consists of 24 (2 x 12) 2D barcode tubes which verifies against the number entered in the worklist with the handheld barcode reader. The BD MAX System is a qualitative in vitro diagnostic test obtained from vaginal and rectal swab specimens from antepartum pregnant women designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures after incubation for greater than or equal to 18 hours.
The BD MAX DNA extraction kit is used with the BD MAX System to extract DNA from Gram-negative bacteria and viruses, which may be present in clinical specimens.
|
Code Information |
Lot numbers 103271, 103272, 110071, 110072, 110391, 110401, 110521, 110522, 110561, 110562, 110181, 110611, 110941, 110942, 111111 |
Recalling Firm/ Manufacturer |
Becton Dickinson & Co. BD Diagnostic Systems 7 Loveton Circle Sparks MD 21152-0999
|
For Additional Information Contact |
John R. Gerlich 410-316-4000
|
Manufacturer Reason for Recall |
Medical device bar code reader does not function as intended and could result in associating clinical test results to wrong patient.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated May 2011, to all affected customers. The letter identified the products, the problem, and the action to be taken by the customers. The letter advised customers to discontinue use of the Sample Preparation Reagent tubes in the referenced kits and discard any remaining products. Customers were requested to complete an attached reply form to acknowledge their receipt of the notification and fax it to (410) 316-4258. For further assistance regarding replacements contact BD Customer Service Dept. 1-800-675-0908. For all other inquiries, contact BD Technical Services Dept. 1-800-638-8663. |
Quantity in Commerce |
61 kits |
Distribution |
Worldwide Distribution-USA (nationwide) including the states of CA, KY, MD, MI, NC, NY, TX, and WI and the country Belgium. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|