| | Class 2 Device Recall LifeStar ACT III Platinum Sensor |  |
| Date Initiated by Firm | September 26, 2011 |
| Date Posted | November 17, 2011 |
| Recall Status1 |
Terminated 3 on May 15, 2012 |
| Recall Number | Z-0207-2012 |
| Recall Event ID |
59990 |
| 510(K)Number | K081257 K110499 |
| Product Classification |
Transmitters and receivers, electrocardiograph, telephone - Product Code DXH
|
| Product | LifeStar ACT III Platinum Sensor Ambulatory Cardiac Telemetry; an automatically activated system with a high performance triple channel EKG, office hookup and auto send/auto detect that requires no patient intervention to either capture or transmit an arrhythmia when it occurs; Manufactured by Card Guard, 2 Pekeris St, Rehovot, 76100, Israel; Model FG-00084
LifeStar ACT provides critical ECG information by capturing the onset and termination/escape of an arrhythmia to assist in correctly identifying and treating the patient. Patients are monitored for up to 30 days and all of the data is available at the request of the physician during the service period and for 7 days following the end of service. The ACT III will provide more sensitive data for initial or early detection of arrhythmia in patients that have minimal or atypical symptoms. |
| Code Information |
Model FG-00084, all serial numbers with firmware REV 10.4 |
| FEI Number |
1000127138
|
Recalling Firm/ Manufacturer |
Lifewatch Services Inc
10255 W Higgins Rd
O'Hare International Center II
Rosemont
IL
60018-5606
|
| For Additional Information Contact | 847-459-7799 |
Manufacturer Reason for Recall | LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | The firm, LifeWatch, sent an " Urgent Device Recall" letter dated September 27, 2011 to the affected customers. The letter describes the product, problem and action to be taken. The customers were instructed to immediately refrain from providing these devices to their customers; stop using the affected sensors, return their entire kit in the pre-paid mailer, and use the new device kit included with the recall letter. In addition, the letter informed the users that the replacement device has upgraded software and firmware that allows for detection of shorts in the cable, and if detected, cuts off current flow to the patient to prevent minor shocks and burns.
If you have any questions, please call (847) 813-4258. |
| Quantity in Commerce | 442 sensors |
| Distribution | Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|