| | Class 2 Device Recall Cryopette |  |
| Date Initiated by Firm | November 28, 2011 |
| Date Posted | February 23, 2012 |
| Recall Status1 |
Terminated 3 on February 04, 2013 |
| Recall Number | Z-1094-2012 |
| Recall Event ID |
60824 |
| 510(K)Number | K100596 |
| Product Classification |
Labware, assisted reproduction - Product Code MQK
|
| Product | ORIGIO MIDATLANTIC DEVICES CRYOPETTE (r) Rx only Single use only CRY-PETTE-CL LOT 123456 2013-10-10 Exp. date *** MADE IN THE USA STERILE R ***Origio, Inc. 3400 Hunter's Way, Charlottesville, VA 22911 www.origio.com
The Cryopette (r) is a cryopreservative storage device that is intended for use in vitrification procedures to contain and maintain human blastocyst stage embryos. |
| Code Information |
Lot numbers/Expiration: 27109A exp. 9/2011; 29909A, exp. 10/2011; 06210A, exp. 3/2012; 06210B, exp. 3/3/2012; 16210A, exp. 6/2012; 20810A, exp. 7/2012; 22410A, exp. 8/2012; 25610A, exp. 9/2012; 32010A, exp. 11/2012; 32010B, exp. 11/2012; 35510A, exp. 12/2012; 03210A, exp. 2/2012; 05411A, exp. 2/2013; 09811C, exp. 4/2013; 09811A, exp. 4/2013; 09811B, exp. 4/2013; 14611A, exp. 5/2013; 14611B, exp. 2013; 020111, exp. 8/1/13; 020111, exp. 8/8/13; 020111, exp. 8/15/13; 020111, exp. 9/5/13; 020111, exp. 9/12/13; 020111, exp. 9/26/13 |
| FEI Number |
1216677
|
Recalling Firm/ Manufacturer |
ORIGIO HUMAGEN PIPETS INC.
2400 Hunters Way
Charlottesville
VA
22911-7930
|
| For Additional Information Contact | Cindy Showalter 434-979-4000 |
Manufacturer Reason for Recall | Cryopettes were rupturing during the warming process following vitrification. |
FDA Determined Cause 2 | Device Design |
| Action | ORIGIO sent a Field Safety Corrective Action letter dated November 21, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
ORIGIO feels it is important to eliminate this risk completely. As soon as we are confident that this danger of rupture has been eliminated, we will make Cryopettes available again. Please use the attached UPS call tag to return any unused Cryopettes to ORIGIO Inc. Upon receipt, we will issue a credit to your account.
We sincerely apologize for the problems this issue has caused in your laboratories. For further questions please call (434) 979-4000. |
| Quantity in Commerce | 73850 Cryopettes |
| Distribution | Worldwide Distribution - U.S. included states of AL, AR, DE, PR, NY, MA, NH, VT, IL, CT, NJ, PA, VA, WV, NV, NC, SC, GA, FL, OH, MD, MS, IN, MI, PR, WI, MN, SD, MT, MO, TX, CO, UT, AZ, CA, HI, OR, AK, WA and in countries of Turkey, Saudi Arabia, Kuwait, Jordon, So. Korea, Taiwan, Brazil, So. Africa, Oman, Kazakhstan, Honduras, Slovenia, Serbia, Poland, Czech Republic, Romania, Croatia, Ukraine, Netherlands, Bulgaria, Hungary, Russia, Lebanon, Singapore, Malaysia, Vietnam, China, Thailand, Nigeria, Japan, Spain, Greece, Canada, Denmark, France, United Kingdom, Italy, Germany, and Australia, |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = MQK
|
|
|
|