|
Class 2 Device Recall CDI System 100/101 |
 |
Date Initiated by Firm |
July 26, 2012 |
Date Posted |
August 21, 2012 |
Recall Status1 |
Terminated 3 on June 27, 2014 |
Recall Number |
Z-2232-2012 |
Recall Event ID |
62684 |
510(K)Number |
K022947
|
Product Classification |
Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
|
Product |
Interface module for CDl System 100/101
Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the system. The module performs the following functions: Communicates primary pump flow data to the CDl 500 Receives blood parameter data from the CDl 500 |
Code Information |
catalog number: 802558 and serial numbers:00008-00013, and 00015-00241. |
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor MI 48103-9586
|
For Additional Information Contact |
734-741-6173
|
Manufacturer Reason for Recall |
Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu
|
FDA Determined Cause 2 |
Device Design |
Action |
Terumo sent an URGENT MEDICAL DEVICE Recall letter dated July 24, 2012 to all affected customers. The letter identified the affected product, reason for correction, potential hazard and actions to be taken. The letter stated that Terumo CVS will replace all affected modules. Customers were informed that a Terumo VS Representative will contact the user to schedule a service call for the replacement upon receipt of the Response Form attached. For questions
contact Terumo CVS Customer Service at 800-521-2818. |
Quantity in Commerce |
232 units |
Distribution |
Worldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, and WV and the countries of:
AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Germany, Hong Kong, Indonesia, Japan, Jordan, Malaysia, Singapore, South Africa, South Korea, Taiwan, Thailand, UNITED ARAB EMIRATES (UAE), and Vietnam |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DTQ and Original Applicant = TERUMO CARDIOVASCULAR SYSTEMS CORP.
|
|
|
|