• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopessee related information
Date Initiated by FirmApril 26, 2010
Date PostedAugust 22, 2012
Recall Status1 Terminated 3 on August 22, 2012
Recall NumberZ-2241-2012
Recall Event ID 62729
510(K)NumberK021049 K022069 K024012 K041932 
Product Classification Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
ProductGE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5. This product is used for image intensified fluoroscopic x-rays for interventional radiologic procedures.
Code Information Serial numbers: 69-0128, 82-0157, 82-0158, 82-0185, 82-0191, 82-0210, 82-0215, 82-0223, 82-0229, 82-0266, 82-0324, 82-0328, 82-0464, 82-0477, 82-0484, 82-0505-RC, 82-0507, 82-0542, 82-0544, 82-0546, 82-0551, 82-0604, 82-0648, 82-0673, 82-0680, 82-0714, 82-0733, 82-0779, 82-0858, 82-0869, 82-0884, 82-0906, 82-0918, 82-0965, 82-1004, 82-1009, 82-1057-RC, 82-1070, 82-1139, 82-1191, 82-1194, 82-1225, 82-1247, 82-1303, 82-1405, 82-1434, 82-1458, 82-1480, 82-1534, 82-1547-RC, 82-1564, 82-1585, 82-1588, 82-1601, 82-1604, 82-1621, 82-1694, 82-1695, 82-1699, 82-1797, 82-1798, 82-1842, 82-1850, 82-1921, 82-1959, 82-2015, 82-2120, 82-2192, 82-2245, 82-2247, 82-2274, 82-2277, 82-2306, 82-2386, 82-2487, 82-2661, 82-2662, 82-2716, 82-2721, 82-2741, 82-2745, 82-2797, 82-2846, 82-2905, 82-2965, 82-3102, 82-3107, 82-3122, 82-3135, 82-3151, 82-3351, 82-3366, 82-3403, 82-3431, 82-3492, 82-3498, 82-3563, 82-3570, 82-3614, 82-3634, 82-3762, 82-3815, 82-3819, 82-3824, 82-3839, 82-3851, 82-3882, 82-3884, 82-3887, 82-3888, 82-3890, 82-3894, 82-3895, 82-3896, 82-3898, 82-3899, 82-3901, 82-3903, 82-3904, 82-3910, 82-3911, 82-3912, 82-3913, 82-3914, 82-3915, 82-3917, 82-3919, 82-3920, 82-3921, 82-3922, 82-3923, 82-3925, 82-3928, 82-3929, 82-3930, 82-3931, 82-3933, 82-3934, 82-3935, 82-3936, 82-3937, 82-3938, 82-3939, 82-3941, 82-3942, 82-3943, 82-3944, 82-7011-MH, 82-7024-MH, 82-7154-CMH, 82-7189-MH, 82-7190-MH, 82-7191-MH, 82-7197-MH, 82-7200-MH, 82-7206-MH, 89-0100, 89-0128, 89-0135-RC, 89-0210, 89-0248, 89-0315, 89-0362, 89-0385, 89-0401, 89-0416, 89-0457, 89-0464, 89-0467, 89-0505, 89-0506, 89-0537, 89-0556, 89-0559, 89-0577, 89-0612, 89-0643, 89-0660, 89-0662, 89-0729, 89-0738, 89-0780, 89-0781, 89-0832, 89-0853, 89-0868, 89-0876, 89-0882, 89-0922, 89-0954-RC, 89-0963, 89-1040, 89-1071, 89-1172, 89-1195, 89-1228, 89-1245, 89-1248, 89-1255, 89-1262, 89-1268, 89-1271, 89-1274, 89-1286, 89-1295, 89-1308, 89-1343, 89-1373, 89-1400-RC, 89-1408, 89-1426, 89-1494, 89-1531, 89-1576, 89-1598, 89-1606, 89-1612, 89-1640, 89-1663, 89-1673, 89-1721, 89-1729, 89-1739, 89-1749, 89-1868, 89-2121, 89-2259, 89-2264, 89-2275, 89-2309, 89-2315, 89-2350, 89-2354, 89-2456, 89-2464, 89-2500, 89-2680, 89-2866, 89-3211, 89-3294, 89-3312, 89-3349, 89-3352, 89-3355, 89-3377, 89-3420, 89-3433, 89-3461, 89-3479, 89-3483, 89-3488, 89-3490, 89-3494, 89-3495, 89-3499, 89-3501, 89-3504, 89-3508, 89-3509, 89-3511, 89-3512, 89-3513, 89-3514, 89-3515, 89-3519, 89-3520, 89-3522, 89-3524, 89-3527, 89-3529, 89-3530, 89-3533, 89-3537, 89-3538, 89-3539, 89-3543, 89-3544, 89-3545, 89-3546, 89-3547, 89-3548, 89-3549, 89-3551, 89-3552, 89-3553, 89-3554, 89-3555, 89-3556, 89-3557, 89-3558, 89-3559, 89-3560, 89-3561, 89-3562, 89-3563, 89-3564, 89-3565, 89-3567, 89-3568, 89-3569, 89-3570, 89-3571, 89-3572, 89-3573, 89-3574, 89-3578, 89-3579, 8S-0027, 8S-0076-RC, 8S-0086, 8S-0121-C, 8S-0158, 8S-0165, 8S-0245, 8S-0291, 8S-0301-C, 8S-0315, 8S-0413-C, 8S-0494, 8S-0615-C, 8S-0642, 8S-0691, 8S-0707, 8S-0711, 8S-0735, 8S-0747, 8S-0776, 8S-0840, 8S-0844-RC, 8S-0858, 8S-0867, 8S-0872, 8S-0921, 8S-0974, 8S-1013-C, 8S-1046, 8S-1056, 8S-1058, 8S-1065-C, 8S-1174, 8S-1203, 8S-1224-C-RC, 8S-1285-RC, 8S-1360, 8S-1445, 8S-1644, 8S-1856, 8S-1930-N, 8S-1954, 8S-1959, 8S-1961, 8S-2052, 8S-2053, 8S-2063, 8S-2063, 8S-2113, 8S-2145, 8S-2164, 8S-2170, 8S-2207, 8S-2256, 8S-2329, 8S-2439, 8S-2451, 8S-2452, 8S-2452, 8S-2462, 8S-2532, 8S-2622, 8S-2865, 8S-2956, 8S-2985, 8S-3002, 8S-3108, 8S-3111, 8S-3112, 8S-3130, 8S-3179, 8S-3183, 8S-3197, 8S-3205, 8S-3207, 8S-3211-CH, 8S-3216, 8S-3222, 8s-3231, 8S-3233, 8S-3243, 8S-3245, 8S-3246, 8S-3247, 8S-3248, 8S-3249, 8S-3253, 8S-3254, 8S-3255, 8S-3256-H, 8S-3257, 8S-3258-CH, 8S-3259, 8S-3260, 8S-3261, 8S-3262, 8S-3263, 8S-3264, 8S-3265, 8S-3266, 8S-3267, 8S-3268, 8S-3269-H, 8S-3270-H, 8S-3271-H, 8S-3272-H, 8S-3273-CH, 8S-3274-CH, 8S-3275, 8S-3276, 8S-3278, 8S-3279, 8S-3280, 8S-3281, 8S-3282, 8S-3283, 8S-3285, 8S-3286, 8S-3287, 8S-3288-H, 8S-3289-H, 8S-3291, 8S-3292, 8S-3296, 8S-3299, 8S-3301-H, 8S-3302, 8S-3305, 8S-3306, 8S-3307, 8S-3308, 8S-3309, 8S-3310-H, 8S-3311, 8S-3315, 8S-7094-CMH, 8S-7228-CMH, 8S-7229-CMH, 8S-7230-CMH, 8S-7231-MH, 8S-7232-MH, 8S-7233-MH, E2-0001, E2-0015, E2-0026, E2-0028, E2-0034, E2-0078, E2-0084, E2-0085, E2-0108, E2-0110, E2-0111, E2-0117, E2-0120, E2-0122, E2-0130, E2-0142, E2-0143, E2-0148, E2-0151, E2-0152, E2-0153, E2-0154, E2-0155, E2-0156, E2-0157, E2-0158, E2-0159, E2-0160, E2-0161, E2-0162, E2-0163, E2-0164, E2-0165, E2-0167, E2-0168, E2-0169, E2-0171, E2-0172, E2-0173, E2-0174, E2-0175, E2-0176, E2-0177, E2-0178, E2-0179, E2-0180, E2-0181, E2-0182, E2-0183, E2-0184, E2-0185, E2-0186, E2-0187, E2-0188, E2-0189, E2-0190, E2-0191, E2-0192, E2-0193, E2-0194, E2-0195, E2-0196, E2-0197, E2-0198, E2-0199, E2-0200, E2-0201, E2-0202, E2-0203, E2-0204, E2-0205, E2-0206, E2-0207, E2-0208, E2-0209, E2-0210, E2-0211, E2-0212, E2-0213, E2-0215, E2-0216, E2-0217, E2-0222, E2-0224, E2-0225, E2-0230, E2-0231, E2-0232, E2-0233, E2-0234, E2-0235-C, E2-0237, E2-0238, E2-0239, E2-0240, E2-0241, E2-0242, E2-0243, E2-0244, E2-0245, E2-0246, E2-0247, E2-0248, E2-0250, E2-0251, E2-0252, E2-0253, E2-0254, E2-0255, E2-0256, E2-0257, E2-0258, E2-0259, E2-0260, E2-0261, E2-0262, E2-0263, E2-0264, E2-0265, E2-0266, E2-0267, E2-0268, E2-0269, E2-0270, E2-0271, E2-0272, E2-0273, E2-0274, E2-0275, E2-7014-MH, E2-7020-MH, E2-7048-MH, E2-7050-MH, E2-7051-MH, E2-7057-MH, E2-7058-MH, E2-7059-MH, E2-7060-MH, E2-7061-MH, E2-7062-MH, E2-7063-MH, E2-7064-MH, E2-7065-MH, E2-7066-MH, E2-7067-MH, E2-7068-MH, E2-7069-CMH, E2-7070-MH, E2-7071-MH, E2-7072-MH, E2-7074-MH, E2-7075-MH, E2-7076-MH, E2-7077-CMH, E2-7078-MH, E2-7079-MH, E9-0008, E9-0025, E9-0031, E9-0035-G, E9-0037, E9-0050, E9-0083, E9-0085, E9-0086, E9-0088, E9-0089-G, E9-0090, E9-0091, E9-0092, E9-0093, E9-0094, E9-0095, E9-0098, E9-0099, E9-0100, E9-0101, E9-0102, E9-0103, E9-0108, E9-0109, E9-0110, E9-0111, E9-0112, E9-0113, E9-0114, E9-0115, E9-0116, E9-0118, E9-0119, E9-0120, E9-0121, E9-0122, E9-0123, E9-0124, E9-0125, E9-0126, E9-0127, E9-0128, E9-0129, E9-0129, E9-0130, E9-0139, E9-0141, E9-0142, E9-0143, E9-0144, E9-0145, E9-0146, E9-0147, E9-0148, E9-0149, E9-0150, E9-0151, E9-0152, E9-0153, E9-0155, E9-0157, E9-0158, E9-0159, E9-0162, E9-0163, E9-0164, E9-0165, E9-0166, E9-0169, E9-0170, E9-0171, E9-0172, E9-0173, E9-0177, E9-0178, E9-0179, E9-0180, E9-0181, E9-0183, E9-0184, E9-0185, E9-0188, E9-0249, E9-0253, ES-0005, ES-0006, ES-0007-C, ES-0022-CH, ES-0029-CH, ES-0034-G, ES-0056, ES-0069, ES-0071, ES-0072, ES-0076, ES-0078, ES-0084-G, ES-0086-G, ES-0087, ES-0088, ES-0089, ES-0090, ES-0091-H, ES-0092-CH, ES-0096, ES-0097, ES-0098, ES-0099-CH, ES-0101-CH, ES-0102-CH, ES-0103, ES-0104, ES-0105, ES-0106, ES-0107, ES-0108, ES-0109, ES-0110, ES-0111, ES-0112, ES-0113, ES-0114, ES-0115, ES-0116, ES-0117, ES-0118, ES-0119, ES-0120, ES-0121, ES-0122, ES-0123, ES-0124, ES-0125, ES-0126, ES-0127, ES-0128, ES-0129, ES-0130, ES-0131-H, ES-0132, ES-0133, ES-0134, ES-0135, ES-0136, ES-0138, ES-0139, ES-0140, ES-0141, ES-0142, ES-0147, ES-0152, ES-0155-H, ES-0156, ES-0158, ES-0159, ES-0160, ES-0163-G, ES-0169-CH, ES-0170, ES-0171, ES-0172, ES-0176, ES-0178, ES-0180, ES-0181, ES-0182, ES-0184-CH, ES-0185-H, ES-0186-H, ES-0187-H, ES-0190-CH, ES-0191-H, ES-0192-H, ES-7010-CMH, ES-7015-MH, ES-7018-MH, ES-7021-MH, ES-7028-CMH, ES-7029-MH, ES-7030-CMH, ES-7031-CMH, ES-7033-CMH, ES-7034-CMH, ES-7035-CMH, ES-7036-CMH, ES-7037-CMH, ES-7039-MH.
FEI Number 1720753
Recalling Firm/
Manufacturer
GE OEC Medical Systems, Inc
384 Wright Brothers Dr
Salt Lake City UT 84116-2862
For Additional Information ContactMs. Deena Pease
801-536-4952
Manufacturer Reason
for Recall
GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
FDA Determined
Cause 2
Component change control
ActionNo customer communication took place. On April 26, 2010, GE began opening service cases for field engineers to begin replacing the snubber boards. This action has been completed as of October 29, 2010.
Quantity in Commerce845 units
DistributionWorldwide and Nationwide (USA) distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = OXO
510(K)s with Product Code = OXO
-
-