 Class 1 Device Recall Customedsee related information
Date Initiated by FirmMay 20, 2014
Date PostedAugust 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-2415-2014
Recall Event ID 68536
Product Classification Orthopedic tray - Product Code OJH
ProductHIP ORTHOPEDIA PACK - (1) BLADE SURGICAL #15 STAINLES /STEEL (1) GOWN XL SMS IMPERVIOUS REINFORCED MMI Ill LIF (1) TABLE COVER REINFORCED 50" X 90" LIF (1) MAYO STAND COVER REINFORCED LIF (4) DRAPE UTILITY WIT APE LIF (4) TOWEL ABSORBENT 15" X 20" LIF (1) YANKAUER SUCTION TUBE WITHOUT VENT LIF (2) DRAPE SHEET 41" X 58" SMS LIF (1) NEEDLE & BLADE COUNTER 20C FOAM/MAG LIF (10) GAUZE 4" X 4" 4PLY (2) UTILITY BOWL 16oz LIF (1) CAUTERY TIP POLISHER LIF (1) SUTURE BAG FLORAL (1) BLADE SURGICAL#10 STAILESS STEEL (1) SURGICAL DURAPREP SOLUTION 26m! LIF (1) SPECIMEN CONTAINER 4oz. WITH LID & LABEL (10)GAUZE SPONGE 4" X 4" 16PLY XRD LIF (5) LAP SPONGE PRE WASH 18" X 18" XRD LIF (1) CAUTERY PENCIL ROCKER SWITCH LIF (1) TUBE SUCTION CONNECT. 3/16" X 12' LIF (1) DRAPE INCISE ANTIMICROBIAL 23" X 33" (2) SHEET DRAPE 70" X 100" LIF (1) DRAPE TOP WITH ADHESIVE 108" X 50" STD SMS LIF (1) SHEET SPLIT WITH ADHESIVE 108" X 77" STD SMS L/F EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Code Information Product code 900-1835, 3 lots: 113109763 131210771 140412619 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information ContactRosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce3 lots; 36 units (multiple lots per unit)
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-