Date Initiated by Firm | November 21, 2014 |
Date Posted | December 19, 2014 |
Recall Status1 |
Terminated 3 on February 20, 2015 |
Recall Number | Z-0834-2015 |
Recall Event ID |
69845 |
510(K)Number | K101077 |
Product Classification |
Laparoscope, general & plastic surgery - Product Code GCJ
|
Product | Fluorescence Imaging Procedure Kit, designed
for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments. |
Code Information |
PN 950156-01 |
Recalling Firm/ Manufacturer |
Intuitive Surgical, Inc. 1266 Kifer Rd Bldg 100 Sunnyvale CA 94086-5304
|
For Additional Information Contact | Mark Johnson 408-523-2100 |
Manufacturer Reason for Recall | The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence. |
FDA Determined Cause 2 | Labeling design |
Action | Urgent Medical Device Correction Letters were sent to customers on November 24, 2014. |
Quantity in Commerce | 1532 total both devices |
Distribution | Nationwide Distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = GCJ
|