| | Class 2 Device Recall Siemens |  |
| Date Initiated by Firm | April 08, 2014 |
| Date Posted | January 06, 2015 |
| Recall Status1 |
Terminated 3 on December 23, 2016 |
| Recall Number | Z-0914-2015 |
| Recall Event ID |
69992 |
| 510(K)Number | K952109 K982387 K992731 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | e.cam -emission computed tomography system
Used to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. |
| Code Information |
1) E.CAM DCR Duet Head, A (e.soft); Serial Numbers: 7090 , 7501 , 7502 , 8001 , 8002 , 8003 , 8004 , 8501 , 8502 , 8506 , 8507 ; 2) E.CAM DCR Gantry Dual Head, Bl Serial Numbers: 3097 , 8502 , 8507 , 8508 , 8509 , 8510 , 8511 , 8514 , 8516 , 8521 , 8522 , 9821 , 71101 ; 3) E.CAM DCR Gantry Single Head; Serial Numbers: 5045 , 7501 , 7502 , 7503 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7511 , 7512 , 8001 , 8002 , 8003 , 8004 , 8005 , 8006 , 8007 , 8008 , 8009 , 8010 , 8011 , 8012 , 8013 , 8014 , 8015 , 8016 , 8017 , 8018 , 8019 , 8020 , 8021 , 8022 , 8023 , 8024 , 8025 , 8026 , 8027 , 8028 , 8029 , 8030 , 8031 , 8032 , 8033 , 8501 , 8502 , 8503 , 8504 , 8505 , 8506 , 8507 , 8508 , 8509 , 8510 , 8511 , 8512 , 8513 , 8514 , 8516 , 8517 , 8518 , 8519 , 8520 , 8521 , 8522 , 8523 , 8524 , 8525 , 8526 , 8527 , 8528 , 8530 , 8532 , 8533 , 8534 , 8535 , 8536 , 8537 , 8538 , 8539 , 8540 , 8541 , 8542 , 8543 , 8544 , 8545 , 8546 , 8547 , 8548 , 8549 , 8550 , 8551 , 8552 , 8553 , 8554 , 8555 , 8556 ; 4) E.CAM DCR Gantry Single Head, B; Serial Numbers: 2011 , 5504 , 7510 , 8501 , 8502 , 8503 , 8505 , 8507 , 8508 , 8509 , 8511 , 8512 , 8513 , 8514 , 8516 , 8517 , 8518 , 8519 , 8520 , 8521 , 8501* ; 5) E.CAM Duet Head (e.soft); Serial Numbers: 2 , 3 , 6001 , 6003 , 6005 , 6006 , 6007 , 6009 , 6010 , 7001 , 7004 , 7005 , 7006 , 7502 , 7503 , 7505 , 7507 , 7508 ; 6) E.CAM extended Gantry dual head v3; Serial Numbers: 1001 , 5039 , 5161 , 5501 , 5504 , 5506 , 5507 , 5508 , 5509 , 5510 , 5511 , 5512 , 5513 , 5514 , 5515 , 5516 , 5517 , 5519 , 5520 , 5521 , 5522 , 5524 , 5525 , 5526 , 5528 , 5529 , 5530 , 5532 , 5533 , 5534 , 5535 , 5536 , 5538 , 5539 , 5540 , 5541 , 5542 , 5543 , 5544 , 5545 , 5546 , 5547 , 5548 , 5549 , 5550 , 5553 , 5554 , 5555 , 5556 , 5557 , 5558 , 5560 , 5561 , 5563 , 5564 , 5565 , 5566 , 5567 , 5568 , 5570 , 5571 , 5572 , 5576 , 5578 , 5579 , 5580 , 5582 , 5583 , 5584 , 5585 , 5586 , 5587 , 5588 , 5589 , 5590 , 5591 , 5592 , 5593 , 5594 , 5595 , 5596 , 5597 , 5598 , 5599 , 5600 , 5601 , 5602 , 5603 , 5604 , 5605 , 5606 , 5607 , 5608 , 5609 , 5610 , 5611 , 5612 , 5613 , 5614 , 5615 , 5616 , 5617 , 5618 , 5619 , 5620 , 5621 , 5622 , 5623 , 5624 , 5625 , 5626 , 5628 , 5629 , 5631 , 5632 , 5633 , 5634 , 5635 , 5636 , 5637 , 5638 , 5639 , 5640 , 5641 , 5642 , 5643 , 5644 , 5645 , 5646 , 5647 , 5649 , 5650 , 5652 , 5655 , 5658 , 7002 , 7003 , 7004 , 7005 , 7006 , 7007 , 7008 , 7009 , 7010 , 7011 , 7012 , 7013 , 7014 , 7016 , 7017 , 7018 , 7019 , 7020 , 7021 , 7022 , 7023 , 7024 , 7025 , 7026 , 7027 , 7028 , 7029 , 7030 , 7031 , 7032 , 7033 , 7034 , 7036 , 7037 , 7038 , 7039 , 7040 , 7041 , 7042 , 7043 , 7044 , 7045 , 7046 , 7047 , 7048 , 7049 , 7050 , 7051 , 7052 , 7053 , 7054 , 7055 , 7056 , 7057 , 7058 , 7059 , 7060 , 7061 , 7062 , 7063 , 7065 , 7067 , 7068 , 7069 , 7070 , 7071 , 7072 , 7073 , 7074 , 7075 , 7076 , 7077 , 7078 , 7079 , 7080 , 7081 , 7082 , 7083 , 7084 , 7085 , 7086 , 7088 , 7090 , 7091 , 7092 , 7093 , 7094 , 7095 , 7096 , 7097 , 7097 , 7098 , 7099 , 7101 , 7103 , 7104 , 7105 , 7109 , 7110 , 7111 , 7112 , 7114 , 7115 , 7116 , 7117 , 7118 , 7119 , 7120 , 7121 , 7122 , 7123 , 7124 , 7125 , 7126 , 7127 , 7129 , 7130 , 7178 , 7179 , 7501 , 7502 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7511 , 7512 , 7513 , 7514 , 7515 , 7516 , 7518 , 7521 , 7522 , 7523 , 7524 , 7526 , 7528 , 7529 , 7530 , 7531 , 7532 , 7533 , 7534 , 7535 , 7536 , 7537 , 7538 , 7539 , 7540 , 7541 , 7542 , 7543 , 7544 , 7545 , 7546 , 7549 , 7550 , 7552 , 7553 , 7554 , 7556 , 7557 , 7558 , 7559 , 7561 , 7562 , 05173/5992099 , CAX271/U2B004 ; 7) E.CAM extended gantry single head; Serial Numbers: 5501 , 5502 , 5503 , 5504 , 5505 , 5506 , 5507 , 5508 , 5509 , 5510 , 5511 , 5512 , 5513 , 5514 , 5515 , 5516 , 5517 , 5518 , 5519 , 5520 , 5521 , 5522 , 5523 , 5524 , 5525 , 5526 , 5527 , 5528 , 5529 , 5626 , 7001 , 7002 , 7003 , 7004 , 7005 , 7006 , 7007 , 7008 , 7009 , 7010 , 7011 , 7012 , 7013 , 7014 , 7015 , 7016 , 7017 , 7018 , 7019 , 7020 , 7021 , 7022 , 7023 , 7024 , 7025 , 7026 , 7501 , 7502 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7511 , 7512 , 7513 , 7013-SMS , 7505X ; 8) E.CAM gantry dual head FIX; Serial Numbers: 1002 , 1003 , 1004 , 1005 , 1006 , 1007 , 1009 , 1010 , 1011 , 1012 , 1013 , 1014 , 1015 , 1016 , 1019 , 1020 , 1021 , 1022 , 1023 , 1024 , 1025 , 1026 , 1028 , 1029 , 1030 , 1031 , 1032 , 1034 , 1035 , 1036 , 1037 , 1038 , 1039 , 1040 , 1041 , 1042 , 1043 , 1045 , 1046 , 1047 , 1048 , 1049 , 1050 , 1051 , 1052 , 1054 , 1056 , 1057 , 1058 , 1059 , 1060 , 1062 , 1063 , 1064 , 1065 , 1066 , 1067 , 1068 , 1069 , 1070 , 1071 , 1073 , 1074 , 1075 , 1076 , 1077 , 1078 , 1079 , 1080 , 1082 , 1083 , 1084 , 1087 , 1088 , 1090 , 1091 , 1093 , 1094 , 1099 , 1100 , 1101 , 1102 , 1103 , 1104 , 1105 , 1107 , 1109 , 1110 , 1111 , 1112 , 1113 , 1114 , 1116 , 1117 , 1118 , 1119 , 1121 , 1134 , 1203 , 1210 , 2001 , 2002 , 2003 , 2004 , 2005 , 2006 , 2007 , 2008 , 2009 , 2010 , 2011 , 2012 , 2013 , 2014 , 2015 , 2016 , 2017 , 2018 , 2019 , 2020 , 2021 , 2022 , 2023 , 2024 , 2026 , 2027 , 2028 , 2030 , 2031 , 2032 , 2034 , 2035 , 2036 , 2037 , 2038 , 2039 , 2040 , 2041 , 2042 , 2043 , 2044 , 2045 , 2046 , 2047 , 2048 , 2050 , 2051 , 2052 , 2053 , 2054 , 2055 , 2056 , 2057 , 2058 , 2059 , 2060 , 2061 , 2062 , 2063 , 2064 , 2065 , 2066 , 2067 , 2068 , 2069 , 2070 , 2072 , 2073 , 2074 , 2075 , 2076 , 2077 , 2078 , 2079 , 2080 , 2081 , 2082 , 2083 , 2084 , 2085 , 2086 , 2089 , 2090 , 2091 , 2092 , 2093 , 2094 , 2096 , 2097 , 2098 , 2099 , 2100 , 2111 , 2114 , 2115 , 2117 , 2130 , 3001 , 3002 , 3003 , 3004 , 3005 , 3006 , 3007 , 3008 , 3009 , 3010 , 3011 , 3012 , 3013 , 3014 , 3015 , 3016 , 3017 , 3018 , 3019 , 3020 , 3021 , 3022 , 3023 , 3024 , 3025 , 3026 , 3027 , 3028 , 3029 , 3030 , 3031 , 3032 , 3033 , 3034 , 3035 , 3036 , 3037 , 3039 , 3040 , 3041 , 3043 , 3044 , 3045 , 3046 , 3047 , 3048 , 3049 , 3050 , 3051 , 3052 , 3053 , 3054 , 3055 , 3056 , 3057 , 3058 , 3059 , 3060 , 3061 , 3062 , 3063 , 3064 , 3065 , 3066 , 3067 , 3068 , 3069 , 3070 , 3071 , 3072 , 3073 , 3074 , 3075 , 3076 , 3077 , 3078 , 3079 , 3080 , 3082 , 3083 , 3084 , 3085 , 3086 , 3087 , 3088 , 3089 , 3090 , 3091 , 3092 , 3093 , 3094 , 3095 , 3096 , 3098 , 3099 , 3100 , 3101 , 3102 , 3103 , 3104 , 3105 , 3106 , 3107 , 3108 , 3109 , 3110 , 3111 , 3112 , 3114 , 3115 , 3116 , 3117 , 3118 , 3120 , 3121 , 3122 , 3123 , 3124 , 3125 , 3126 , 3127 , 3128 , 3129 , 3130 , 3131 , 3132 , 3133 , 4001 , 4002 , 4003 , 4004 , 4005 , 4006 , 4007 , 4008 , 4009 , 4010 , 4011 , 4012 , 4014 , 4015 , 4017 , 4018 , 4019 , 4020 , 4021 , 4022 , 4023 , 4024 , 4025 , 4026 , 4027 , 4028 , 4029 , 4030 , 4031 , 4032 , 4033 , 4034 , 4035 , 4036 , 4037 , 4038 , 4039 , 4040 , 4041 , 4042 , 4043 , 4044 , 4045 , 4046 , 4048 , 4049 , 4050 , 4052 , 4053 , 4054 , 4055 , 4056 , 4057 , 4058 , 4059 , 4060 , 4061 , 4062 , 4063 , 4064 , 4065 , 4066 , 4067 , 4068 , 4069 , 4070 , 4071 , 4075 , 4076 , 4077 , 4078 , 4090 , 4111 , 7501 , 01009R , 1073DEL , 2065XX , 3070// , 4049XXX , CAR-557 ; 9) E.CAM gantry dual head v2.5; Serial Numbers: 1006 , 2002 , 4079 , 4080 , 4081 , 4083 , 4084 , 4085 , 4086 , 4087 , 4088 , 4089 , 4091 , 4092 , 4093 , 4094 , 4095 , 4096 , 4097 , 4098 , 4099 , 4100 , 4101 , 4102 , 4103 , 4104 , 4105 , 4106 , 4110 , 4112 , 4113 , 4117 , 4118 , 4119 , 4120 , 4123 , 4124 , 4126 , 4127 , 4129 , 4131 , 4132 , 4133 , 4134 , 4137 , 4138 , 4139 , 4141 , 4142 , 4143 , 4144 , 4145 , 4146 , 4148 , 4149 , 4150 , 4151 , 4152 , 4153 , 4154 , 4155 , 4156 , 4158 , 4159 , 4160 , 4162 , 4163 , 4164 , 4165 , 4166 , 4167 , 4168 , 4169 , 4170 , 4171 , 4172 , 4173 , 4174 , 4176 , 4177 , 4178 , 4179 , 4180 , 4181 , 4182 , 4183 , 4184 , 4185 , 4187 , 4188 , 4189 , 4190 , 4192 , 4194 , 4195 , 4196 , 4197 , 4198 , 4199 , 4202 , 4205 , 4206 , 4208 , 4209 , 4214 , 4217 , 4218 , 4219 , 4220 , 4221 , 4222 , 4223 , 4224 , 4225 , 4226 , 5001 , 5002 , 5003 , 5004 , 5005 , 5006 , 5007 , 5008 , 5009 , 5010 , 5011 , 5012 , 5013 , 5014 , 5015 , 5016 , 5017 , 5018 , 5019 , 5020 , 5021 , 5022 , 5023 , 5024 , 5025 , 5026 , 5027 , 5028 , 5029 , 5030 , 5031 , 5032 , 5034 , 5036 , 5037 , 5038 , 5039 , 5040 , 5041 , 5042 , 5043 , 5044 , 5045 , 5046 , 5047 , 5048 , 5049 , 5050 , 5051 , 5052 , 5053 , 5054 , 5055 , 5056 , 5057 , 5058 , 5059 , 5060 , 5061 , 5062 , 5063 , 5064 , 5065 , 5066 , 5067 , 5068 , 5069 , 5070 , 5071 , 5103 , 5106 , 5107 , 5108 , 5109 , 5114 , 5123 , 5164 , 6000 , 6001 , 6648 , 6654 , 7503 , 4084A , 4123DEL , 5027A ; 10) E.CAM gantry dual head v3.0lnt; Serial Numbers: 5104 , 5107 , 5109 , 5110 , 5111 , 5112 , 5113 , 5115 , 5116 , 5119 , 5120 , 5121 , 5122 , 5123 , 5124 , 5125 , 5126 , 5127 , 5128 , 5129 , 5131 , 5132 , 5133 , 5134 , 5135 , 5136 , 5137 , 5138 , 5139 , 5140 , 5141 , 5142 , 5143 , 5144 , 5146 , 5147 , 5148 , 5149 , 5150 , 5151 , 5152 , 5153 , 5154 , 5155 , 5157 , 5160 , 5162 , 5163 , 5165 , 5166 , 5167 , 5168 , 5169 , 5170 , 5171 , 5172 , 5173 , 5174 , 5211 , 5502 , 5578 , 6510 ; 11) E.CAM Gantry Dual Head v3.7; Serial Numbers: 1 , 4001 , 5652 , 6501 , 7057 , 7501 , 7502 , 7503 , 7504 , 7505 , 7506 , 7507 , 7508 , 7509 , 7510 , 7512 , 7513 , 7514 , 7515 , 7516 , 7517 , 7518 , 7520 , 7521 , 7524 , 7525 , 7526 , 7527 , 7528 , 7529 , 7530 , 7531 , 7532 , 7533 , 7534 , 7535 , 7536 , 7537 , 7538 , 7539 , 7540 , 7541 , 7543 , 7544 , 7545 , 7546 , 7547 , 7548 , 7549 , 7550 , 7551 , 7552 , 7554 , 7555 , 7556 , 7557 , 7558 , 7560 , 7561 , 7562 , 7564 , 7565 , 7566 , 7567 , 7568 , 7569 , 7570 , 7571 , 7572 , 7573 , 7574 , 7575 , 7577 , 7578 , 7579 , 7580 , 7581 , 7582 , 7583 , 7584 , 7585 , 7586 , 7588 , 7590 , 7591 , 7593 , 7594 , 7595 , 7596 , 7597 , 7598 , 7599 , 7601 , 7604 , 7605 , 7606 , 7607 , 7608 , 7609 , 7612 , 7613 , 7614 , 7615 , 7616 , 7617 , 7618 , 7619 , 7621 , 7622 , 7623 , 7624 , 7625 , 7627 , 7628 , 7629 , 7630 , 7631 , 7632 , 7633 , 7635 , 7636 , 7637 , 7638 , 7861 , 7889 , 8001 , 8002 , 8003 , 8004 , 8005 , 8006 , 8007 , 8008 , 8009 , 8010 , 8011 , 8012 , 8013 , 8014 , 8015 , 8016 , 8017 , 8018 , 8019 , 8020 , 8021 , 8022 , 8023 , 8024 , 8025 , 8026 , 8027 , 8028 , 8029 , 8030 , 8031 , 8032 , 8033 , 8034 , 8035 , 8036 , 8037 , 8038 , 8039 , 8042 , 8043 , 8044 , 8045 , 8046 , 8047 , 8048 , 8049 , 8050 , 8051 , 8052 , 8053 , 8054 , 8055 , 8056 , 8057 , 8058 , 8059 , 8061 , 8064 , 8065 , 8066 , 8067 , 8068 , 8069 , 8070 , 8071 , 8072 , 8073 , 8074 , 8075 , 8076 , 8077 , 8078 , 8079 , 8080 , 8081 , 8082 , 8083 , 8084 , 8085 , 8086 , 8087 , 8088 , 8089 , 8090 , 8092 , 8093 , 8094 , 8095 , 8096 , 8097 , 8098 , 8099 , 8100 , 8101 , 8102 , 8103 , 8105 , 8107 , 8108 , 8109 , 8110 , 8111 , 8112 , 8113 , 8115 , 8116 , 8117 , 8118 , 8119 , 8120 , 8121 , 8122 , 8123 , 8124 , 8125 , 8126 , 8127 , 8128 , 8130 , 8131 , 8132 , 8135 , 8136 , 8137 , 8139 , 8141 , 8142 , 8143 , 8144 , 8145 , 8146 , 8147 , 8501 , 8502 , 8503 , 8504 , 8505 , 8506 , 8507 , 8508 , 8509 , 8510 , 8511 , 8512 , 8513 , 8514 , 8515 , 8516 , 8517 , 8518 , 8519 , 8521 , 8522 , 8523 , 8524 , 8525 , 8526 , 8527 , 8528 , 8529 , 8530 , 8531 , 8532 , 8533 , 8535 , 8536 , 8537 , 8538 , 8540 , 8541 , 8542 , 8543 , 8544 , 8545 , 8546 , 8547 , 8548 , 8549 , 8550 , 8551 , 8552 , 8553 , 8554 , 8555 , 8556 , 8557 , 8558 , 8560 , 8561 , 8562 , 8563 , 8564 , 8565 , 8566 , 8567 , 8568 , 8569 , 8570 , 8571 , 8572 , 8573 , 8574 , 8575 , 8576 , 8577 , 8578 , 8579 , 8580 , 8581 , 8582 , 8583 , 8584 , 8585 , 8586 , 8587 , 8588 , 8589 , 8590 , 8591 , 8592 , 8593 , 8594 , 8595 , 8596 , 8597 , 8598 , 8599 , 8600 , 8601 , 8602 , 8603 , 8604 , 8605 , 8606 , 8608 , 8609 , 8610 , 8611 , 8612 , 8613 , 8614 , 8615 , 8616 , 8617 , 8618 , 8619 , 8620 , 8621 , 8622 , 8623 , 8624 , 8625 , 8626 , 8627 , 8628 , 8629 , 8632 , 8633 , 8634 , 8635 , 8636 , 8637 , 8638 , 8639 , 8640 , 8642 , 8643 , 8644 , 8645 , 8646 , 8647 , 8648 , 8649 , 8650 , 8651 , 8652 , 8653 , 8654 , 8655 , 8656 , 8658 , 8659 , 8660 , 8661 , 8662 , 8663 , 8665 , 8667 , 8668 , 8669 , 8670 , 8671 , 8672 , 8673 , 8674 , 8676 , 8677 , 8678 , 8679 , 8680 , 8681 , 8682 , 8683 , 8684 , 8685 , 8686 , 8687 , 8688 , 8689 , 8690 , 8691 , 8692 , 8693 , 8694 , 8695 , 8696 , 8697 , 8698 , 8699 , 8700 , 8701 , 8702 , 8703 , 8704 , 8705 , 8706 , 8707 , 8708 , 8709 , 8710 , 8711 , 8712 , 8713 , 8714 , 8715 , 8716 , 8717 , 8718 , 8719 , 8720 , 8721 , 8722 , 8723 , 8724 , 8725 , 8727 , 8728 , 8729 , 8730 , 8731 , 8732 , 8733 , 8734 , 8735 , 8737 , 8738 , 8739 , 8742 , 8743 , 8744 , 8745 , 8746 , 8748 , 8749 , 8750 , 8751 , 8753 , 8755 , 8756 , 8757 , 8758 , 8759 , 8760 , 8763 , 8764 , 8765 , 8766 , 8767 , 8768 , 8769 , 8770 , 8774 , 8775 , 8776 , 8777 , 8778 , 8779 , 8780 , 8781 , 8784 , 8785 , 8787 , 8788 , 8789 , 8792 , 8794 , 8795 , 8796 , 8797 , 8798 , 8799 , 8800 , 8801 , 8802 , 8803 , 8804 , 8805 , 8806 , 8807 , 8808 , 8809 , 8810 , 8811 , 8812 , 8813 , 8814 , 8815 , 8816 , 8817 , 8818 , 8820 , 8821 , 8823 , 8824 , 8825 , 8826 , 8827 , 8828 , 8831 , 8832 , 8833 , 8834 , 8836 , 8840 , 8843 , 8844 , 8847 , 44001 , 71101 , 71102 , 72102 , 4366618 , 5224741 , 100100018 , 08501/7823920 , 7618* , 8794DEL ; 12) E.CAM gantry dual head variable; Serial Numbers: 9 , 13 , 1002 , 1004 , 1005 , 1006 , 1007 , 1008 , 1009 , 1010 , 1011 , 1012 , 1013 , 1014 , 1015 , 1016 , 1017 , 1018 , 1019 , 1020 , 1021 , 1022 , 1023 , 1024 , 1025 , 1026 , 1028 , 1029 , 1030 , 1031 , 1032 , 1033 , 1034 , 1035 , 1036 , 1037 , 1038 , 1039 , 1041 , 1042 , 1044 , 1045 , 1046 , 1047 , 1048 , 1051 , 1053 , 1054 , 1055 , 1056 , 1057 , 1058 , 1059 , 1060 , 1061 , 1062 , 1063 , 1065 , 1066 , 1067 , 1068 , 1069 , 1070 , 1071 , 1072 , 1073 , 1074 , 1075 , 1076 , 1077 , 1078 , 1079 , 1081 , 1082 , 1084 , 1085 , 1086 , 1088 , 1089 , 1090 , 1091 , 1092 , 1093 , 1094 , 1095 , 1096 , 1098 , 1099 , 1100 , 1101 , 1102 , 1103 , 1105 , 1106 , 1107 , 1108 , 1111 , 1112 , 1113 , 1114 , 1115 , 1116 , 1117 , 1118 , 1119 , 1120 , 1121 , 1122 , 1123 , 1124 , 1125 , 1126 , 1127 , 1128 , 1129 , 1130 , 1132 , 1133 , 1134 , 1135 , 1135 , 1136 , 1137 , 1140 , 1141 , 1142 , 1143 , 1144 , 1145 , 1146 , 1147 , 1148 , 1149 , 1152 , 1153 , 1154 , 1155 , 1156 , 1157 , 1158 , 1159 , 1160 , 1161 , 1162 , 1163 , 1164 , 1165 , 1166 , 1167 , 1169 , 1170 , 1171 , 1172 , 1176 , 1180 , 1181 , 1182 , 1183 , 1184 , 1185 , 1186 , 1188 , 1189 , 1190 , 1192 , 1193 , 1195 , 1196 , 1197 , 1198 , 1199 , 1200 , 1201 , 1202 , 1203 , 1204 , 1205 , 1206 , 1207 , 1208 , 1209 , 1212 , 2050 , 2089 , 2100 , 3005 , 3066 , 3111 , 3125 , 4003 , 4052 , 4069 , 4081 , 4117 , 4166 , 5009 , 5090 , 5145 , 7030 , 7582 , 8149 , 8462 , 8635 , 9601 , 10351 , 10352 , 31827 , 49641 , 86439 , NU6457-01 ; 13) e.cam Gantry Foresight Sign_Ser 1Hd; Serial Numbers: 1002 , 1003 , 1004 , 1005 , 1006 , 1007 , 1008 , 1009 , 1010 , 1011 , 1012 , 1013 , 1014 , 1015 , 1016 , 1017 , 1018 , 1019 , 1020 , 1021 , 1022 , 1023 , 1024 , 1025 , 1026 , 1027 , 1028 , 1029 , 1030 , 1031 , 1032 , 1033 , 1034 , 1035 , 1036 , 1037 , 1038 , 1039 , 1040 , 1041 , 1042 , 1043 , 1044 , 1045 , 1046 , 1047 , 1048 , 1049 , 1050 , 1051 , 1052 , 1053 , 1054 , 1055 , 1056 , 1057 , 1058 , 1059 , 1060 , 1061 , 1062 , 1063 , 1064 , 1065 , 1066 , 1067 , 1068 , 1069 , 1070 , 1071 , 1072 , 1073 , 1074 , 1075 , 1076 , 1077 , 1078 , 1079 , 1080 , 1081 , 1082 , 1083 , 1084 , 1085 , 11128 , 11144 , 70001 , 70002 , 70003 ; 14) E.CAM gantry single head; Serial Numbers: 1 , 2 , 4 , 5 , 1003 , 1004 , 1005 , 1006 , 1007 , 1008 , 1009 , 1010 , 1011 , 1012 , 1013 , 1014 , 1015 , 1016 , 1017 , 1018 , 1019 , 1020 , 1021 , 1022 , 1065 , 2001 , 2002 , 2003 , 2004 , 2005 , 2006 , 2007 , 2008 , 2009 , 2010 , 2012 , 2013 , 2014 , 2015 , 2016 , 2017 , 2018 , 2019 , 2020 , 2021 , 2022 , 2023 , 2024 , 2025 , 2026 , 2027 , 2028 , 3001 , 3002 , 3003 , 3004 , 3005 , 3006 , 3007 , 3008 , 3009 , 3010 , 3011 , 3012 , 3013 , 3014 , 3015 , 3016 , 3017 , 3018 , 3020 , 3021 , 3022 , 3023 , 3024 , 3025 , 3026 , 3027 , 3028 , 3029 , 3030 , 3031 , 3032 , 3034 , 4001 , 4002 , 4003 , 4004 , 4005 , 4006 , 4007 , 4008 , 4009 , 4010 , 4011 , 4012 , 4013 , 4014 , 4015 , 4016 , 4219 , 6104 , 6303 , 7912 , 9965 , 49518 , 96302 , 3024// , 3030// , 6134R ; 15) E.CAM gantry single head v3.0; Serial Numbers: 1001 , 4017 , 4018 , 4019 , 4020 , 4022 , 4023 , 4024 , 4025 , 4027 , 4029 , 4030 , 4031 , 4032 , 4033 , 4034 , 4035 , 4036 , 4037 , 4040 , 4041 , 4042 , 4043 , 4044 , 4045 , 4046 , 4047 , 4047 , 4048 , 4049 , 4050 , 4090 , 4091 , 5001 , 5002 , 5003 , 5004 , 5005 , 5006 , 5008 , 5009 , 5010 , 5011 , 5012 , 5013 , 5014 , 5015 , 5016 , 5017 , 5018 , 5019 , 5020 , 5021 , 5022 , 5023 , 5024 , 5025 , 5026 , 5027 , 5028 , 5029 , 5030 , 5031 , 5032 , 5033 , 5034 , 5035 , 5036 , 5037 , 5038 , 5039 , 5040 , 5041 , 5042 , 5043 , 701001 ; 16) E.Cam Sngl Special Configuration; Serial Number: 9995; 17) Prvn Excellence e.cam Cardiac; Serial Numbers: 556 , 572 , RC492 , RC493 , RC524 , RC538 , RC540 , RC541 , RC545 , RC546 ; 18) Prvn Excellence e.cam Fixed 180; Serial Numbers: 551 , 574 , 584 , 591 , RC521 , RC536 ; 19) Prvn Excellence e.cam Variablel Serial Numbers: 510 , 529 , 530 , 557 , 559 , 563 , 565 , 573 , 575 , 580 , 582 , 587 , 594 , 7127 , 9325 , RC495 , RC498 , RC543 ; 20) Signature E.CAM Dual Camera, A; Serial Numbers: 8191 , 8501 , 8502 , 8503 , 8506 , 8507 , 8508 , 8509 , 8510 , 8511 , 8512 , 8513 , 8514 , 8515 , 8516 , 8517 , 8518 , 8519 , 8520 , 8521 , 8522 , 8523 , 8524 , 8525 , 8526 , 8527 , 8529 , 8530 , 8531 , 8532 , 8533 , 8534 , 8535 , 8536 , 8537 , 8538 , 8539 , 8540 , 8541 , 8542 , 8543 , 8544 , 8545 , 8546 , 8547 , 8548 , 8549 , 8550 , 8551 , 8552 , 8553 , 8554 , 8555 , 8556 , 8557 , 8558 , 8560 , 8561 , 8562 , 8563 , 8564 , 8565 , 8566 , 8567 , 8568 , 8569 , 8570 , 8571 , 8572 , 8573 , 8574 , 8575 , 8576 , 8577 , 8578 , 8579 , 8580 , 8581 , 8582 , 8583 , 9001 , 9003 , 9004 , 9005 , 9006 , 9007 , 9008 , 9009 , 9010 , 9011 , 9012 , 9013 , 9014 , 9015 , 9016 , 9017 , 9018 , 9019 , 9020 , 9021 , 9022 , 9023 , 9024 , 9026 , 9027 , 9028 , 9029 , 9030 , 9031 , 9032 , 9033 , 9034 , 9035 , 9036 , 9037 , 9038 , 9039 , 9040 , 9041 , 9042 , 9043 , 9044 ; 21) Signature E.CAM Dual Camera, B; Serial Numbers: 1013 , 1278 , 5382 , 5389 , 5517 , 5764 , 6481 , 7778 , 7853 , 7883 , 7990 , 8484 , 8490 , 9001 , 9002 , 9003 , 9004 , 9005 , 9006 , 9007 , 9008 , 9009 , 9010 , 9011 , 9012 , 9013 , 9014 , 9015 , 9016 , 9017 , 9018 , 9019 , 9020 , 9021 , 9022 , 9023 , 9024 , 9025 , 9027 , 9028 , 9029 , 9030 , 9031 , 9033 , 9034 , 9035 , 9036 , 9038 , 9039 , 9040 , 9041 , 9042 , 9043 , 9044 , 9045 , 9046 , 9047 , 9048 , 9049 , 9050 , 9051 , 9052 , 9053 , 9054 , 9055 , 9056 , 9057 , 9058 , 9059 , 9060 , 9061 , 9062 , 9063 , 9064 , 9065 , 9066 , 9067 , 9068 , 9069 , 9070 , 9071 , 9072 , 9073 , 9074 , 9075 , 9076 , 9077 , 9078 , 9079 , 9080 , 9081 , 9082 , 9083 , 9084 , 9085 , 9086 , 9087 , 9088 , 9089 , 9090 , 9091 , 9092 , 9093 , 9094 , 9095 , 9096 , 9097 , 9098 , 9099 , 9100 , 9101 , 9102 , 9103 , 9104 , 9105 , 9106 , 9107 , 9108 , 9109 , 9110 , 9111 , 9112 , 9113 , 9114 , 9115 , 9116 , 9117 , 9118 , 9119 , 9122 , 9123 , 9124 , 9125 , 9126 , 9127 , 9128 , 9129 , 9130 , 9131 , 9132 , 9133 , 9134 , 9135 , 9136 , 9137 , 9138 , 9139 , 9141 , 9142 , 9143 , 9144 , 9145 , 9146 , 9147 , 9148 , 9149 , 9150 , 9151 , 9152 , 9153 , 9154 , 9155 , 9156 , 9157 , 9158 , 9159 , 9160 , 9161 , 9162 , 9163 , 9164 , 9165 , 9166 , 9167 , 9168 , 9169 , 9170 , 9171 , 9172 , 9173 , 9174 , 9178 , 9179 , 9180 , 9181 , 9182 , 9183 , 9184 , 9185 , 9186 , 9187 , 9188 , 9189 , 9190 , 9191 , 9192 , 9193 , 9194 , 9195 , 9197 , 9198 , 9199 , 9200 , 9201 , 9202 , 9203 , 9204 , 9205 , 9206 , 9207 , 9208 , 9209 , 9210 , 9211 , 9212 , 9213 , 9214 , 9215 , 9216 , 9217 , 9219 , 9220 , 9221 , 9222 , 9223 , 9224 , 9225 , 9226 , 9228 , 9229 , 9230 , 9231 , 9232 , 9233 , 9234 , 9235 , 9236 , 9237 , 9238 , 9239 , 9240 , 9241 , 9242 , 9243 , 9244 , 9245 , 9246 , 9247 , 9248 , 9249 , 9250 , 9251 , 9252 , 9253 , 9254 , 9255 , 9256 , 9257 , 9259 , 9261 , 9262 , 9263 , 9264 , 9265 , 9266 , 9267 , 9268 , 9270 , 9273 , 9274 , 9275 , 9276 , 9277 , 9278 , 9280 , 9281 , 9282 , 9283 , 9284 , 9287 , 9288 , 9289 , 9290 , 9291 , 9292 , 9293 , 9294 , 9295 , 9296 , 9297 , 9298 , 9299 , 9300 , 9301 , 9302 , 9303 , 9305 , 9306 , 9307 , 9308 , 9309 , 9310 , 9311 , 9313 , 9724 , 10000 , 10001 , 10002 , 10004 , 10005 , 10006 , 10007 , 10008 , 10009 , 10010 , 10011 , 10012 , 10013 , 10014 , 10015 , 10016 , 10017 , 10018 , 10019 , 10020 , 10021 , 10022 , 10023 , 10024 , 10025 , 10026 , 10027 , 10028 , 10029 , 10030 , 10031 , 10032 , 10033 , 10034 , 10035 , 10036 , 10037 , 10038 , 10039 , 10040 , 10041 , 10042 , 10043 , 10044 , 10045 , 10046 , 10047 , 10048 , 10049 , 10050 , 10051 , 10052 , 10053 , 10054 , 10055 , 10056 , 10057 , 10058 , 10059 , 10060 , 10061 , 10062 , 10063 , 10064 , 10065 , 10067 , 10068 , 10069 , 10070 , 10071 , 10072 , 10073 , 10074 , 10075 , 10077 , 10078 , 10079 , 10080 , 10082 , 10083 , 10084 , 10085 , 10086 , 10087 , 10088 , 10089 , 10090 , 10091 , 10092 , 10093 , 10094 , 10095 , 10096 , 10097 , 10098 , 10099 , 10100 , 10101 , 10102 , 10103 , 10104 , 10105 , 10106 , 10107 , 10108 , 10109 , 10110 , 10111 , 10112 , 10113 , 10114 , 10115 , 10116 , 10117 , 10118 , 10119 .
10120 , 10121 , 10122 , 10123 , 10124 , 10125 , 10126 , 10127 , 10128 , 10129 , 10130 , 10131 , 10133 , 10134 , 10135 , 10136 , 10137 , 10140 , 10141 , 10142 , 10143 , 10144 , 10145 , 10146 , 10147 , 10148 , 10149 , 10150 , 10151 , 10152 , 10154 , 10155 , 10156 , 10157 , 10158 , 10159 , 10160 , 10161 , 10162 , 10163 , 10164 , 10165 , 10168 , 10169 , 10170 , 10171 , 10172 , 10173 , 10174 , 10175 , 10177 , 10178 , 10179 , 10180 , 10181 , 10183 , 10184 , 10185 , 10186 , 10187 , 10188 , 10190 , 10191 , 10192 , 10193 , 10194 , 10195 , 10196 , 10197 , 10198 , 10199 , 10200 , 10202 , 10203 , 10204 , 10206 , 10207 , 10208 , 10209 , 10210 , 10211 , 10212 , 10214 , 10215 , 10216 , 10217 , 10218 , 10219 , 10220 , 10221 , 10222 , 10223 , 10224 , 10225 , 10226 , 10227 , 10228 , 10229 , 10230 , 10231 , 10232 , 10233 , 10234 , 10235 , 10236 , 10237 , 10238 , 10239 , 10240 , 10241 , 10242 , 10243 , 10244 , 10245 , 10247 , 10248 , 10249 , 10250 , 10251 , 10252 , 10253 , 10254 , 10255 , 10256 , 10257 , 10258 , 10260 , 10263 , 10264 , 10266 , 10267 , 10268 , 10270 , 10271 , 10272 , 10273 , 10274 , 10275 , 10276 , 10277 , 10278 , 10279 , 10280 , 10281 , 10282 , 10283 , 10284 , 10286 , 10287 , 10288 , 10289 , 10290 , 10291 , 10292 , 10293 , 10294 , 10296 , 10297 , 10299 , 10300 , 10301 , 10302 , 10303 , 10304 , 10305 , 10306 , 10307 , 10309 , 10310 , 10311 , 10312 , 10313 , 10314 , 10315 , 10316 , 10318 , 10319 , 10320 , 10321 , 10322 , 10323 , 10324 , 10325 , 10326 , 10327 , 10328 , 10330 , 10331 , 10333 , 10334 , 10335 , 10336 , 10337 , 10338 , 10339 , 10340 , 10341 , 10342 , 10343 , 10344 , 10345 , 10346 , 10347 , 10348 , 10349 , 10350 , 10352 , 10353 , 10354 , 10355 , 10356 , 10357 , 10358 , 10359 , 10360 , 10361 , 10362 , 11003 , 11006 , 11007 , 11008 , 11009 , 11010 , 11012 , 11013 , 11014 , 11015 , 11016 , 11017 , 11018 , 11019 , 11020 , 11021 , 11022 , 11023 , 11024 , 11025 , 11026 , 11027 , 11029 , 11030 , 11031 , 11032 , 11033 , 11034 , 11035 , 11038 , 11040 , 11041 , 11042 , 11043 , 11044 , 11045 , 11046 , 11048 , 11049 , 11050 , 11051 , 11052 , 11053 , 11054 , 11055 , 11056 , 11057 , 11059 , 11060 , 11061 , 11062 , 11063 , 11064 , 11065 , 11066 , 11067 , 11068 , 11069 , 11070 , 11071 , 11072 , 11073 , 11074 , 11075 , 11076 , 11077 , 11078 , 11079 , 11081 , 11082 , 11083 , 11084 , 11085 , 11086 , 11087 , 11088 , 11089 , 11090 , 11091 , 11092 , 11093 , 11094 , 11095 , 11096 , 11097 , 11098 , 11099 , 11100 , 11101 , 11102 , 11103 , 11104 , 11105 , 11106 , 11107 , 11108 , 11109 , 11110 , 11111 , 11112 , 11113 , 11114 , 11115 , 11116 , 11117 , 11118 , 11119 , 11120 , 11121 , 11122 , 11123 , 11124 , 11125 , 11126 , 11127 , 11128 , 11129 , 11130 , 11131 , 11132 , 11133 , 11134 , 11135 , 11136 , 11137 , 11139 , 11140 , 11141 , 11142 , 11143 , 11144 , 11145 , 11147 , 11148 , 11149 , 11150 , 11151 , 11152 , 11153 , 11154 , 11155 , 11156 , 11157 , 11158 , 11159 , 11160 , 11162 , 11163 , 11164 , 11165 , 11166 , 11167 , 11168 , 11169 , 11170 , 11171 , 11172 , 11173 , 11174 , 11175 , 11176 , 11177 , 11178 , 11179 , 11181 , 11182 , 11184 , 11185 , 11186 , 11188 , 11189 , 11190 , 11192 , 11194 , 11195 , 11196 , 11197 , 11198 , 11199 , 11200 , 11201 , 11202 , 11203 , 11204 , 11205 , 11206 , 11207 , 11208 , 11210 , 11211 , 11212 , 11213 , 11214 , 11215 , 11216 , 11217 , 11219 , 11220 , 11221 , 11222 , 11223 , 11224 , 11225 , 11226 , 11227 , 11228 , 11229 , 11230 , 11231 , 11232 , 11233 , 11234 , 11235 , 11236 , 11237 , 11238 , 11239 , 11241 , 11243 , 11244 , 11245 , 11246 , 11247 , 11248 , 11249 , 11250 , 11252 , 11253 , 11254 , 11255 , 11256 , 11258 , 11259 , 11260 , 11261 , 11262 , 11263 , 11264 , 11265 , 11266 , 11267 , 11268 , 11269 , 11270 , 11271 , 11272 , 11273 , 11274 , 11275 , 11276 , 11277 , 11278 , 11279 , 11281 , 11282 , 11283 , 11284 , 11285 , 11286 , 11287 , 11288 , 11289 , 11290 , 11291 , 11292 , 11293 , 11294 , 11295 , 11296 , 11298 , 11299 , 11300 , 11301 , 11303 , 11306 , 11307 , 11308 , 11309 , 11310 , 11311 , 11312 , 11313 , 11314 , 11315 , 11316 , 11317 , 11318 , 11319 , 11320 , 11321 , 11323 , 11324 , 11325 , 11326 , 11327 , 11328 , 11329 , 11330 , 11331 , 11332 , 11333 , 11334 , 11335 , 11336 , 11337 , 11338 , 11339 , 11340 , 11341 , 11342 , 11344 , 11345 , 11347 , 11348 , 11350 , 11355 , 11356 , 11357 , 11358 , 11359 , 11361 , 11362 , 11363 , 11365 , 11366 , 11368 , 11369 , 11370 , 11371 , 11372 , 11373 , 11374 , 11375 , 11376 , 11377 , 11378 , 11380 , 11381 , 11382 , 11383 , 11384 , 11386 , 11387 , 11388 , 11389 , 11390 , 11391 , 11392 , 11395 , 11396 , 11397 , 11399 , 11401 , 11402 , 11404 , 11405 , 11406 , 11407 , 11408 , 11409 , 11410 , 11411 , 11412 , 11414 , 11418 , 11419 , 11420 , 11422 , 11423 , 11424 , 11426 , 11427 , 11428 , 11429 , 11430 , 11431 , 11432 , 11434 , 11435 , 11436 , 11437 , 11438 , 11439 , 11440 , 11441 , 11442 , 11443 , 11445 , 11446 , 11447 , 11448 , 11449 , 11451 , 11452 , 11453 , 11454 , 11455 , 11456 , 11457 , 11458 , 11459 , 11460 , 11462 , 11463 , 11464 , 11465 , 11466 , 11467 , 11468 , 11469 , 11470 , 11471 , 11472 , 11473 , 11474 , 11475 , 11476 , 11477 , 11478 , 11479 , 11480 , 11481 , 11482 , 11483 , 11484 , 11486 , 11487 , 11488 , 11489 , 11490 , 11495 , 11496 , 11497 , 11498 , 11499 , 11500 , 11501 , 11502 , 11503 , 11504 , 11505 , 11506 , 11508 , 11509 , 11510 , 11511 , 11512 , 11513 , 11514 , 11515 , 11516 , 11517 , 11518 , 11519 , 11520 , 11521 , 11522 , 11523 , 11524 , 11525 , 11526 , 11527 , 11528 , 11529 , 11531 , 11532 , 11534 , 11535 , 11536 , 11537 , 11538 , 11540 , 11541 , 11542 , 11543 , 11544 , 11545 , 11546 , 11547 , 11548 , 11549 , 11550 , 11551 , 11552 , 11554 , 11555 , 11556 , 11558 , 11559 , 11562 , 11563 , 11565 , 11566 , 11567 , 11568 , 11569 , 11570 , 11571 , 11572 , 11573 , 11574 , 11575 , 11577 , 11578 , 11579 , 11580 , 11581 , 11582 , 11583 , 11584 , 11585 , 11586 , 11587 , 11588 , 11589 , 11590 , 11591 , 11592 , 11593 , 11594 , 11595 , 11596 , 11597 , 11598 , 11599 , 11600 , 11601 , 11602 , 11603 , 11604 , 11605 , 11606 , 11607 , 11608 , 11609 , 11610 , 11611 , 11612 , 11614 , 11615 , 11616 , 11617 , 11618 , 11619 , 11620 , 11621 , 11622 , 11623 , 11624 , 11625 , 11626 , 11627 , 11628 , 11630 , 11631 , 11632 , 11633 , 11634 , 11636 , 11637 , 11638 , 11640 , 11641 , 11642 , 11643 , 11644 , 11645 , 11646 , 11647 , 11648 , 11649 , 11650 , 11651 , 11652 , 11653 , 11654 , 11655 , 11656 , 11657 , 11658 , 11659 , 11660 , 11661 , 11662 , 11663 , 11664 , 11665 , 11666 , 11668 , 11669 , 11670 , 11671 , 11673 , 11674 , 11677 , 11678 , 11679 , 11680 , 11681 , 11682 , 11683 , 11688 , 11689 , 11690 , 11691 , 11692 , 11693 , 11694 , 11695 , 11696 , 11697 , 11698 , 11701 , 11702 , 11703 , 11706 , 11708 , 11709 , 11710 , 11711 , 11712 , 11714 , 11715 , 11716 , 11717 , 11723 , 11726 , 11728 , 11729 , 11730 , 11731 , 11732 , 11734 , 11735 , 11736 , 11738 , 11739 , 11740 , 11741 , 11742 , 11743 , 11744 , 11746 , 11748 , 11749 , 11751 , 11752 , 11753 , 11755 , 11756 , 11757 , 11758 , 11759 , 11760 , 11761 , 11762 , 11763 , 11764 , 11765 , 11766 , 11767 , 11768 , 11769 , 11770 , 11771 , 11772 , 11773 , 11774 , 11775 , 11776 , 11777 , 11778 , 11779 , 11780 , 11781 , 11782 , 11783 , 11784 , 11785 , 11786 , 11787 , 11788 , 11789 , 11790 , 11791 , 11792 , 11793 , 11794 , 12345 , 115930 , ECGS1 , 8501 , 8502 , 8503 , 6324 , 6683 , 9001 , 9002 , 9004 , 9005 , 9006 , 9010 , 9011 , 9012 , 9013 , 9014 , 9015 , 9016 , 9017 , 9020 , 9021 , 9022 , 9023 , 9025 , 9026 , 9028 , 9029 , 9031 , 9304 ; 22) Signature E.CAM Single Camera, A; Serial Numbers: 8501 , 8502 , 8503 , 8504 , 8505 , 8506 , 8507 , 8508 , 8509 , 8510 , 8511 ; 23) Signature E.CAM Single Camera, B; Serial Numbers: 4626 , 6116 , 8502 , 8503 , 8504 , 8505 , 8506 , 8507 , 8509 , 9001 , 9002 , 9003 , 9004 , 9005 , 9006 , 9007 , 9008 , 9009 , 9010 , 9011 , 9012 , 9013 , 9014 , 9015 , 9016 , 9017 , 9018 , 9019 , 9020 , 9021 , 9022 , 9023 , 9024 , 9025 , 9026 , 9027 , 9028 , 9029 , 9030 , 9031 , 9032 , 9033 , 9034 , 9035 , 9036 , 9037 , 9038 , 9039 , 9040 , 9041 , 9042 , 9043 , 9044 , 9045 , 9658 , 10000 , 10001 , 10002 , 10003 , 10005 , 10006 , 10007 , 10008 , 10009 , 10010 , 10011 , 10012 , 10013 , 10014 , 10015 , 10016 , 10017 , 10018 , 10019 , 10020 , 10022 , 10023 , 10024 , 10025 , 10026 , 10027 , 10028 , 10029 , 10030 , 10031 , 10032 , 10033 , 10034 , 10035 , 10036 , 10037 , 10038 , 10039 , 10040 , 10041 , 10042 , 10043 , 10044 , 10045 , 10046 , 10047 , 10048 , 10049 , 10050 , 10051 , 10052 , 10053 , 10054 , 10055 , 10056 , 10057 , 10058 , 10059 , 10060 , 10061 , 10062 , 10063 , 10064 , 11001 , 11002 , 11003 , 11004 , 11005 , 11006 , 11007 , 11008 , 11009 , 11010 , 11011 , 11012 , 11013 , 11014 , 11015 , 11016 , 11017 , 11018 , 11019 , 11020 , 11021 , 11022 , 11023 , 11024 , 11025 , 11026 , 11027 , 11028 , 11029 , 11030 , 11031 , 11032 , 11033 , 11034 , 11035 , 11036 , 11037 , 11038 , 11039 , 11040 , 11041 , 11042 , 11043 , 11044 , 11045 , 11046 , 11047 , 11048 , 11049 , 11050 , 11051 , 11052 , 11053 , 11054 , 11055 , 11056 , 11057 , 11058 , 11059 , 11060 , 11061 , 11063 , 11064 , 11065 , 11066 , 11067 , 11068 , 11069 , 11070 , 11071 , 11072 , 11073 , 11074 , 11075 , 11076 , 11077 , 11078 , 11079 , 11080 , 11081 , 11082 , 11083 , 11084 , 11085 , 11086 , 11087 , 11088 , 11089 , 11090 , 11091 , 11092 , 11093 , 11094 , 11095 , 11096 , 11097 , 11098 , 11099 , 11100 , 11101 , 11102 , 11103 , 11104 , 11105 , 11106 , 11107 , 11108 , 11109 , 11110 , 11111 , 11112 , 11113 , 11114 , 11115 , 11116 , 11117 , 11118 , 11119 , 11120 , 11121 , 11122 , 11123 , 11124 , 11126 , 11128 , 11130 , 11132 , 11133 , 11134 , 11136 , 11137 , 11138 , 11139 , 11140 , 11141 , 11142 , 11143 , 11144 , 11145 , 11146 , T1039 ; 24) Signature Foresight Ecam dualcamera; Serial Numbers: 1001 , 1003 , 1004 , 1005 , 1006 , 1007 , 1008 , 1009 , 1010 , 1011 , 1013 , 1014 , 1015 , 1016 , 1017 , 1018 , 1019 , 1020 , 1021 , 1022 , 1023 , 1024 , 1025 , 1026 , 1027 , 1028 , 1029 , 1030 , 1031 , 1032 , 1033 , 1034 , 1035 , 1036 , 1037 , 1038 , 1039 , 1040 , 1041 , 1042 , 1043 , 1044 , 1045 , 1047 , 1048 , 1049 , 1050 , 1051 , 1052 , 1053 , 1054 , 1055 , 1056 , 1057 , 1058 , 1059 , 1060 , 1061 , 1062 , 1063 , 1064 , 1065 , 1066 , 1067 , 1068 , 1069 , 1070 , 1071 , 1072 , 1073 , 1074 , 1075 , 1076 , 1077 , 1078 , 1079 , 1080 , 1081 , 1082 , 1083 , 1084 , 1085 , 1086 , 1087 , 1088 , 1089 , 1090 , 1091 , 1092 , 1094 , 1095 , 1096 , 1097 , 1098 , 1099 , 1100 , 1101 , 1102 , 1103 , 1104 , 1105 , 1106 , 1107 , 1108 , 1109 , 1110 , 1111 , 1112 , 1113 , 1114 , 1115 , 1116 , 1117 , 1118 , 1119 , 1120 , 1121 , 1122 , 1123 , 1124 , 1125 , 1126 , 1127 , 1128 , 1129 , 1130 , 1131 , 1132 , 1133 , 1134 , 1135 , 1136 , 1137 , 1138 , 1139 , 1140 , 1141 , 1142 , 1143 , 1144 , 1145 , 1146 , 1147 , 1148 , 1149 , 1150 , 1151 , 1152 , 1153 , 1154 , 1155 , 1156 , 1157 , 1158 , 1159 , 1160 , 1161 , 1162 , 1163 , 1164 , 1165 , 1166 , 1167 , 1168 , 1169 , 1170 , 1171 , 1172 , 1173 , 1174 , 1175 , 1177 , 1178 , 1179 , 1180 , 1181 , 1182 , 1183 , 1184 , 1185 , 1186 , 1187 , 1188 , 1189 , 1190 , 1191 , 1192 , 1193 , 1194 , 1195 , 1196 , 1198 , 1199 , 1200 , 1201 , 1202 , 1203 , 1204 , 1205 , 1206 , 1207 , 1208 , 1209 , 1210 , 1211 , 1212 , 1213 , 1214 , 1215 , 1216 , 1217 , 1218 , 1219 , 1220 , 1221 , 1222 , 1223 , 1224 , 1225 , 1226 , 1227 , 1228 , 1229 , 1230 , 1231 , 1232 , 1233 , 1234 , 1235 , 1236 , 1237 , 1238 , 1239 , 1240 , 1241 , 1242 , 1243 , 1244 , 1245 , 1246 , 1247 , 1248 , 1249 , 1250 , 1251 , 1252 , 1253 , 1254 , 1255 , 1256 , 1257 , 1258 , 1259 , 1261 , 1263 , 1264 , 1265 , 1266 , 1267 , 1268 , 1269 , 1270 , 1271 , 1272 , 1273 , 1274 , 1275 , 1276 , 1277 , 1278 , 1279 , 1280 , 1281 , 1282 , 1283 , 1284 , 1285 , 1286 , 1287 , 1288 , 1289 , 1290 , 1291 , 1292 , 1293 , 1294 , 1295 , 1296 , 1297 , 1298 , 1299 , 1300 , 1301 , 1302 , 1303 , 1304 , 1305 , 1306 , 1307 , 1308 , 1309 , 1310 , 1311 , 1312 , 1313 , 1314 , 1315 , 1316 , 1317 , 1318 , 1319 , 1320 , 1321 , 1322 , 1323 , 1324 , 1325 , 1326 , 1327 , 1328 , 1329 , 1330 , 1331 , 1332 , 1333 , 1334 , 1335 , 1336 , 1337 , 1338 , 1339 , 1340 , 1341 , 1342 , 1343 , 1344 , 1345 , 1346 , 1347 , 1348 , 1349 , 1350 , 1351 , 1352 , 1353 , 1354 , 1355 , 1356 , 1357 , 1358 , 1359 , 1360 , 1361 , 1362 , 1363 , 1364 , 1365 , 1366 , 1367 , 1368 , 1369 , 1370 , 1371 , 1372 , 1373 , 1374 , 1375 , 2107 , 7197 , 10016 , 10319 , 70001 , 70002 , 70003 , 70004 , 70005 , 70006 , 70007 , 70008 , 70009 , 70010 , 70011 , 70012 , 70013 , 70014 , 70015 , 70017 , 70018 , 70019 , 70020 , 70021 , 70022 ; 25) SRP e.cam Single Detector Camera; Serial Numbers: 542 , 548 , 550 , 558 , 564 , 579 , 581 , 585 , 588 , 589 , 590 , 592 , 593 , 595 , 604 , 1003 , 9164 , RC494 , RC518 , RC528 , RC537 , RC539 , RC549 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc. 2501 Barrington Rd Hoffman Estates IL 60195-2061
|
| For Additional Information Contact | 800-888-7436 |
Manufacturer Reason for Recall | There is a potential for patient finger injury on the e.cam systems. |
FDA Determined Cause 2 | Use error |
| Action | The firm, Siemens, sent a "Customer Safety Advisory Notice" letter dated April 8, 2014 to all direct accounts (customers). The letter described the product, problem and actions to be taken. The letter instructed the customers to follow the instructions in the e.cam operator's manual and the e.cam Patient Bed Safety Addendum of the operator's manual; and to always monitor the patient at all times while they are on the system during initiation of any system movement. The letter further requested that customers forward the letter to the new owner of the equipment in the event that the system is sold or transferred.
Customers with questions about this letter can contact Siemens Medical Solutions USA at 800-888-7463 (Monday - Friday, 7:00 AM - 8:00 PM EST). |
| Quantity in Commerce | 4,462 systems |
| Distribution | Worldwide Distribution: US (nationwide) including states of: AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WY and Puerto Rico; and countries of: Algeria , Angola , Argentina , Australia , Austria , Azerbaijan , Bahrain , Bangladesh , Belarus , Belgium , Bolivia , Bosnia Herzeg. , Brazil , Bulgaria , Cameroon , Canada , Chile , Colombia , Costa Rica , Croatia , Cuba , Czech Republic , Dem. Rep. Congo , Denmark , Dominican Rep. , Ecuador , Egypt , El Salvador , Finland , France , Georgia , Germany , Ghana , Greece , Guam , Honduras , Hong Kong , India , Indonesia , Iran , Iraq , Ireland , Israel , Italy , Jamaica , Japan , Jordan , Kazakhstan , Kosovo , Kuwait , Latvia , Lebanon , Luxembourg , Macedonia , Malaysia , Mauritius , Mexico , Morocco , Myanmar , Nepal , Netherlands , New Zealand , Niger , Nigeria , Norway , Oman , P.R. China , Pakistan , Palestinian Ter , Peru , Philippines , Poland , Portugal , Qatar , Rep. of Yemen , Republic Korea , Romania , Russian Fed. , Saudi Arabia , Serbia , Singapore , Slovakia , Slovenia , South Africa , Spain , Sri Lanka , Sweden , Switzerland , Syria , Taiwan , Thailand , Trinidad,Tobago , Tunisia , Turkey , Uganda , Ukraine , United Arab Emirates , United Kingdom , Uruguay , Uzbekistan , and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|