| | Class 1 Device Recall AVEA ventilator |  |
| Date Initiated by Firm | April 03, 2015 |
| Date Posted | May 26, 2015 |
| Recall Status1 |
Terminated 3 on July 23, 2018 |
| Recall Number | Z-1609-2015 |
| Recall Event ID |
70920 |
| 510(K)Number | K103211 |
| Product Classification |
Ventilator, continuous, facility use - Product Code CBK
|
| Product | CareFusion AVEA ventilator, all models.
Product Usage:
A continuous ventilator (respirator), intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. |
| Code Information |
Serial No. AKV01874 AGV01443 AEV02197 AKV02580 AEV02375 BBV01304 BBV01303 AJV02510 AJV02502 AGV02650 AFV01802 AFV01792 AFV01400 AFV01361 AEV01622 AEV01608 AKV01915 BBV03730 BBV03557 BBV03556 BBV03554 BBV03553 BBV 3558 BBV01406 BBV01396 AHV01880 AGV02205 BBV01432 AFV01657 AFV01655 AFV01647 BAV05234 AHV03069 AGV02249 BBV03797 BAV06514 BDV01011 BDV01000 BCV01938 BAV04011 BAV04010 BAV04009 BAV04005 BAV04004 BAV04003 BAV04002 BAV01313 AKV04053 AJV02289 BAV04021 AEV02264 AEV02225 AEV02205 ADV06091 AKV02569 AHV01206 AKV04351 AKV04349 AKV02550 AGV01385 AHV03128 ADV04116 BBV06059 BBV06057 BAV05153 AFV01492 BBV01620 BDV01371 BAV06447 BAV06446 BAV06445 AEV01724 AEV01712 BAV06463 AEV01694 AJV02446 AGV02139 BAV04188 BAV04187 AKV01153 AKV01147 AEV02720 AEV02717 AEV02713 BAV04191 BAV04993 AHV02129 ADV06297 ADV01637 ADV01592 AFV01922 AFV02277 AHV01274 BAV05235 BCV02186 BCV01822 BBV06061 BDV01249 BDV01242 ADV03032 AJV01685 AHV01967 AHV01722 AHV01701 AKV02230 AKV02226 AKV02224 AKV02219 AKV02217 AKV02215 AKV02214 BEV01326 BEV01325 BEV01324 BEV01306 BEV01205 BDV02001 BDV01994 BDV01810 AKV01279 AKV04320 BAV06124 BBV03028 BBV03027 BBV01483 BBV01019 BBV01018 BBV01016 BAV01015 BAV01020 BBV01010 AKV01029 AKV01012 AKV01007 AKV01000 BCV02865 BAV01088 BAV01026 BAV01018 AKV01018 AKV01017 AHV03287 AHV03061 AFV02243 AFV02230 ADV01638 AGV01808 BFY01003 AGV02074 AFV01176 AKV04033 AEV01588 BAV01089 AKV01009 AKV01005 AFV02205 BBV01005 ACV01718 AKV01028 BFV01053 BEV02183 BAV01099 ADV06397 ADV03003 BAV04113 BAV04111 BEV02185 BEV02184 BDV01816 ajv01601 AGV01333 AGV01278 AFV01962 AHV01326 AJV02876 AEV02362 AEV02359 AEV02340 AFV01616 AFV01744 AFV01661 AFV01633 AFV01632 AFV01631 AFV01629 AFV01625 AFV01545 AFV01481 AFV01449 BEV01356 BEV01352 BDV02093 BCV01220 BCV01206 AGV02366 BBV06021 BBV06016 BBV06007 BBV06006 AEV02120 AKV01274 AKV01240 BCV01810 BCV01619 BCV01176 BCV01072 BCV01059 BCV01057 BBV03951 BBV03598 BBV03570 BAV04368 AKV01068 AKV01047 BAV04362 BDV01515 BDV01480 BEV01663 BEV01648 BCV02271 BCV02270 AJV02640 AHV01549 AEV02727 BBV01207 BBV01206 AKV01466 AKV01442 AKV01367 AKV01046 AKV01003 AJV02681 BAV04017 BAV04016 BAV04015 BAV04014 BAV01306 AKV04488 AGV02552 AGV02441 AGV02432 AJV01532 AJV01476 AJV01424 BBV03552 AJV02441 AJV02950 AJV02504 AJV02467 AJV02432 AJV02427 AJV04226 AJV04195 AJV04219 AJV04225 AJV04192 AJV04228 AJV04232 AJV04223 AJV04229 AJV04201 AJV04205 AJV04203 AJV04202 AJV04221 AJV04189 AJV04176 AJV04046 AJV04233 AJV04224 AKV04118 AHV01285 AJV04196 BEV01544 BEV01543 BEV01542 BEV01541 BEV01540 BEV01539 BEV01534 BEV01531 BEV01522 BEV01519 BEV01515 AJV01591 AKV02755 ADV06008 BDV02092 BDV02087 BDV02022 BDV02021 BDV02002 BDV01980 BDV01668 AFV02626 AKV04119 AFV01567 BCV02763 BCV02757 AFV02102 BBV01459 BDV01474 BDV01461 BCV01732 BCV01729 BCV01722 BAV04354 ADV01559 BBV01300 BBV01292 BBV01278 BDV01818 BDV01646 BBV01017 BEV02141 BEV02137 BEV02136 BEV02135 BEV02134 BEV02133 BEV02131 BEV02127 BEV02123 BEV02113 BDV02011 BDV02006 BDV01995 BDV01982 BDV01971 AKV04324 AJV01076 AKV04325 BCV02209 BCV02208 BCV02205 AJV02216 AJV02173 AFV02439 AFV02368 AFV02299 AFV02281 AFV02183 AFV02182 AFV02170 AFV02129 AFV02000 AFV01688 AFV01668 AFV01296 AGV01584 AGV01527 BAV01218 BAV01217 BAV06470 BEV01359 BCV02187 BCV01831 BCV01830 AJV01531 AJV01524 AHV01705 AGV01413 AFV02402 AFV01572 AFV01546 AFV01250 AFV01247 AFV01165 AFV01106 AEV02698 AGV01762 BDV01385 BDV01375 AHV01434 BEV01140 BEV01135 BCV01708 BCV01705 BCV01704 BCV01703 BCV01702 BCV01701 BBV01480 BBV01472 BBV01471 BBV01470 BBV01468 BBV01465 BBV01462 BAV05231 BAV05230 BAV05229 BAV05228 BEV01131 AJV02530 AJV02529 AJV02505 AGV02710 AKV04061 AKV04060 AKV01478 AHV01745 ADV01758 AGV01287 AKV01475 AKV01022 AJV03054 AJV03053 AJV01139 AHV03369 AHV01916 AGV02297 AFV02424 ADV04002 ADV02151 AGV01352 AKV04114 AKV04116 BDV01830 BDV01745 BAV05103 BAV05102 BAV05101 AKV04163 AKV04129 AKV01926 AKV01906 AKV04104 AKV01402 BAV05001 BAV05002 BBV01176 AJV02144 AJV02130 AJV02121 AHV01778 AHV01748 BAV01041 BCV01785 BCV01734 BBV03213 BBV03206 BBV03112 AKV01964 AKV01959 AKV01957 AGV01497 BDV01264 BFV01058 BEV01865 BDV01285 BEV01235 BEV01233 BEV01226 BEV01224 BEV01217 BEV01208 BEV01202 BDV01974 BDV01970 BDV01357 BDV01349 BDV01347 BDV01338 BDV01336 BDV01333 BDV01318 BDV01317 BDV01181 BDV01175 BDV01104 BCV02335 BAV05008 BAV05007 BAV05006 BAV05005 BAV05004 BAV05003 BAV05000 BAV04999 AGV02364 BBV01409 BBV01380 BBV01379 BBV01357 BBV01197 BBV01181 BBV01173 BBV01167 BBV01164 BAV04256 BAV04254 BCV02306 BCV02304 BCV02303 AEV02657 AJV01699 AGV01964 BBV03145 AHV01184 AGV01160 BEV01264 BEV01227 BEV01028 BEV01017 AGV01986 AGV01259 AGV01255 AGV01242 AGV01241 AGV01156 AGV01138 AGV01037 AFV02619 AFV02616 BBV01213 BBV01203 BBV01201 BBV01000 AHV01575 AHV01591 AGV02223 AGV01237 AGV01222 AGV01198 BCV02262 BCV02261 BCV02260 BCV02259 BCV02258 BCV02257 BCV02256 BCV02254 BCV02253 BCV02252 BCV02249 BCV02248 BCV02247 BCV02246 BCV02245 BCV02244 BCV02243 BCV02241 BCV02239 BCV02238 BCV02237 BCV02236 BCV02235 BCV02234 BCV02231 BCV02222 BCV02220 BBV03803 BBV03796 AFV01043 BAV01150 AJV02703 AJV01645 AHV01370 AGV01811 AFV01503 AEV02548 AEV02551 BDV01601 BCV02633 BBV03599 AGV02040 AGV01257 BCV02240 AKV02765 AKV02763 AKV02762 AKV02761 AKV02760 AKV02759 AKV02758 AKV02757 AKV02756 AKV02225 AKV01893 AJV02967 AHV01636 AHV01621 AHV01585 AHV01526 AHV01648 AGV02704 BEV01182 BEV01181 BEV01180 BEV01179 BEV01178 BEV01176 BEV01175 BEV01173 BEV01172 BEV01169 BEV01168 BEV01150 BDV01308 BDV01307 BDV01306 BDV01303 BDV01302 BDV01301 BDV01147 BCV02338 BCV02266 BCV02265 BCV02264 BCV02263 BCV01748 BAV04434 BAV04432 BAV04431 BAV04430 BAV04429 BAV04428 BAV04427 AKV04029 AKV04028 AKV04026 AKV04025 AKV04024 AKV04022 AKV04020 AKV04019 AKV04018 AKV04017 AKV04016 AKV04015 AKV04014 AKV04013 AKV04010 AKV04009 AKV04008 AKV04007 AKV04006 AKV04005 AKV04003 AHV01620 AHV01615 AGV02580 AGV02556 AGV02545 AGV02481 AGV01940 BAV04433 AGV02590 AKV04023 AKV04027 AKV04021 AKV04001 AKV04011 AKV04004 BDV01645 BDV01559 BAV04772 BAV04768 BFV01055 BFV01054 BFV01051 BBV01223 BAV06784 BAV06780 BAV06777 BAV06796 BAV06782 BAV06783 BAV06781 BAV06778 BAV06775 BBV03005 BBV03000 BAV04489 BAV04488 BAV04487 BAV04486 BAV04485 BAV04481 AJV02381 BCV01312 BCV01262 BDV01684 BCV01738 BCV02176 BCV02174 AKV04316 AKV04315 AKV04314 AKV02719 AKV02718 AKV02201 AKV02175 AKV01616 AEV01908 BCY01036 BBV06068 AJV01609 AHV01343 AGV01327 AEV02473 AEV01808 AGV01318 AEV02648 AJV02165 AJV01665 AFV01117 AEV02712 AEV02700 AEV02581 AEV02380 BDV01263 BDV01259 AKV04120 AGV02357 AEV01520 AGV01855 AGV01852 AGV01851 AGV01850 AGV01849 ADV06048 ADV05035 ADV05033 ADV05031 AGV01857 BCV02063 AKV04318 AKV04317 BCV01930 BCV01924 BCV01920 BCV01917 BCV01902 BCV01895 BCV01894 BCV01888 BCV01886 BCV01885 BCV01784 BCV01782 BCV01775 BCV01770 BCV01766 BCV01762 BCV01760 BCV01758 BCV01757 BCV01756 BCV01754 BCV01751 BCV01683 BCV01663 BCV01657 AKV01619 AKV01606 AKV01604 AKV01602 AKV01500 AJV02688 AJV02473 AJV02425 BAV04112 BAV04094 BCV02558 AJV01073 AJV01072 AJV01067 AJV01065 AHV03215 AJV01625 BDV01981 BDV01848 BDV01842 BDV01841 BDV01836 BDV01824 BDV01822 BDV01820 BDV01815 BDV01748 BCV01771 BCV01763 BCV01710 BCV01699 BCV01696 BCV01689 BAV04770 AKV04460 AKV04458 BAV04777 BAV04771 BAV04769 BAV04776 AKV04459 AKV04462 AKV04461 AKV04457 BAV04792 BAV04790 BAV04789 BAV04786 AKV02753 AKV02752 AKV02581 AKV02751 BDV01029 BCV02805 AKV01614 AKV01591 BEV01304 BEV01195 BEV01193 BEV01192 BEV01191 BEV01190 BEV01188 BEV01186 BEV01185 BEV01184 BEV01183 BEV01171 BEV02119 BEV02116 BEV02114 BEV02110 BEV02099 BEV02090 BEV01879 BEV01857 BEV01290 BEV01288 BEV01287 BEV01285 BEV01282 BEV01281 BEV01280 BEV01278 BEV01277 BEV01275 BEV01274 BEV01273 BEV01272 BEV01271 BEV01270 BEV01269 BEV01268 BEV01260 BEV01259 BEV01258 BEV01257 BEV01256 BEV01255 BEV01245 BEV01241 BEV01240 BEV01222 BBV06026 BBV06025 AFV02157 BAV04061 AJV02390 AJV01101 AGV01322 AGV01056 AFV02070 AFV01653 AEV01780 AEV01726 ADV06195 ADV02030 ADV01587 AEV01701 AJV01633 BAV04063 AHV01492 AHV01477 AFV01372 ADV01052 AJV02992 BDV01297 BDV01295 BDV01294 BDV01292 BDV01287 BDV01280 BDV01279 BDV01276 BDV01274 AKV04494 AKV04483 AKV04481 AKV04480 AKV04479 AHV01409 AKV04472 AJV02713 AKV02717 ADV06065 AHV01976 BEV01358 BEV01357 AKV04058 AKV04055 AKV01767 AKV01760 AKV01757 AKV01756 AKV01753 AKV01752 AKV01751 AKV01750 AKV01716 AKV01715 AKV01708 AKV01706 AKV01758 AGV02163 AGV01212 AGV01209 AGV01201 AGV01203 AFV02211 AFV01582 AFV01580 AFV01352 AKV04456 AKV04455 AHV01488 agv01509 AKV04454 AKV01697 AGV01493 AGV01104 BCV02331 BCV02327 BCV02320 AKV04117 AKV04115 BEV01321 BEV01318 BEV01315 BEV01314 BEV01311 BEV01310 BEV01309 BEV01305 BEV01303 BEV01301 BEV01300 BEV01299 BEV01298 BEV01296 BEV01295 BEV01294 BEV01293 BEV01292 BEV01291 BEV01284 BCV02747 BCV02681 BCV02678 BCV02677 BCV02669 BCV02529 BCV02519 BCV02328 BCV02321 BCV01678 AGV02025 AFV01454 AFV01419 AFV01336 BAV04700 AKV01165 AJV02961 AHV03116 AFV02187 ADV06096 ADV01808 ADV01689 BCV02861 BEV01267 BEV01263 BEV01238 BEV01232 BEV01230 BEV01228 BEV01225 BEV01221 BEV01220 BEV01219 BEV01218 BEV01215 BEV01214 BEV01213 BEV01210 BEV01209 BEV01207 BEV01206 BEV01201 BEV01200 BBV03128 BBV03125 BBV03124 AHV01899 BEV01470 BDV01991 BAV01281 AKV01587 AKV01585 AKV01582 ADV02050 BDV01804 BCV01903 BCV01899 BCV01897 BCV01753 BBV06044 BAV04797 BAV04796 AHV01917 ADV06365 BBV06035 ADV01872 ADV01696 ADV01025 AKV01438 AEV02580 ADV01575 BCV01315 BBV01601 BAV06476 BAV04732 AJV02146 AJV02135 AJV02131 AJV02116 AJV02065 AJV02037 AJV02117 BFV01007 BFV01006 BFV01004 BFV01001 BFV01000 BEV01978 AHV02249 AHV02202 BDV01863 BDV01847 BDV01845 BDV01844 BDV01838 BDV01837 BDV01806 BDV01801 BDV01799 BDV01585 BCV01390 BCV01388 AHV01466 AFV01557 AFV01554 AFV01482 AFV01383 AFV01280 AEV02172 AFV01462 AKV01302 AKV01273 AKV01269 AKV01268 AKV01267 AKV01266 AKV01264 AKV01261 AKV01260 AKV01258 AKV01255 AKV01252 AKV01250 AGV02679 AGV02677 AGV02559 BBV03127 AKV01783 AKV01778 ADV01031 BAV06457 AKV04276 BAV06459 AKV01855 AEV01621 BBV06099 BBV06069 BBV06043 BEV01466 BEV01329 BEV01328 BAV05138 AEV01528 BCV01139 BCV01137 BCV01135 BCV01134 BCV01133 BCV01132 BCV01131 BCV01130 BCV01129 BCV01128 BCV01127 BCV01117 BBV06117 BBV06113 BBV06095 AHV01637 AKV04348 AHV03316 AHV01926 AEV01577 AEV01638 AEV02667 BBV01421 BBV01408 BBV01400 BBV01393 BBV01381 BBV01163 AKV04453 AKV04450 AKV04449 AKV04446 AKV04445 AKV04443 AKV04442 AKV04441 AKV04440 AKV04439 AKV04438 AKV04437 AKV04434 AKV04432 AKV04431 AKV04430 AKV04429 AJV01545 AJV01519 AJV01492 AJV01483 AGV02078 AGV02000 AFV01003 AEV02693 AEV02670 AEV02644 AEV02513 AEV02505 AEV02477 AEV02472 AEV02432 AEV02423 AEV02422 AEV02419 AEV02656 ADV06102 ADV06081 AJV01536 ajv01521 AJV01533 ADV01570 AEV02613 AKV04436 AKV04452 AKV04451 AKV04444 AHV02082 AGV02700 AJV02631 AJV02629 AJV02633 BCV01399 BCV01398 BCV01396 BCV01393 BCV01392 BCV01389 BCV01967 BCV01884 AJVO2812 AHV02047 AGV01206 AFV01017 BEV01212 BEV01203 AEV01542 ADV06284 AKV04127 AJV02379 AEV01549 AEV01545 AEV01509 AEV01503 AEV01002 ADV06398 ADV01706 ADV01531 AHV01927 BDV01051 BCV02792 BCV02790 BCV02760 BBV03260 BBV03070 AGV02488 BBV01159 BBV01168 BBV06104 BAV05149 ADV05026 BDV01563 BDV01290 BDV01639 BDV01628 BDV01582 BDV01577 BDV01574 BDV01568 BBV03690 BDV01558 AGV02330 BCV02881 BCV02423 AFV01463 AFV01425 AKV01807 AKV01795 AKV01720 AGV01462 BCV01530 BCV01529 BCV01947 BCV01946 BCV01945 BCV01944 BCV01943 BCV01942 BCV01941 BCV01940 BCV01939 BCV01937 BCV01936 BCV01925 BCV01922 BCV01921 BCV01919 BCV01915 BCV01914 BCV01913 BCV01912 BCV01911 BCV01910 BCV01909 BCV01908 BCV01907 BCV01906 BCV01905 BCV01904 BCV01898 BCV01893 BCV01891 BCV01890 BCV01887 BCV01879 BCV01783 BCV01755 BCV01666 BCV01535 BCV01531 BCV01892 BAV06757 BAV06691 BAV06686 BAV06685 BAV06683 BAV06680 BAV06679 BAV06677 BAV06675 BAV06751 BAV06692 BAV06688 BAV06830 BAV06672 BAV06684 BAV06676 BAV06673 BAV06690 BAV06674 BAV06678 BAV06762 BAV06760 BAV06756 BAV06755 BAV06752 BAV06750 BAV06748 BAV06695 BAV06694 BAV06689 BAV06577 BAV06758 BAV06746 BAV06747 BAV06753 BAV06681 AKV02754 BCV01995 AHV01903 ADV03053 AGV01662 AFV01496 AFV01485 AEV02110 AEV02106 BAV04309 BAV04299 BAV04277 BAV04259 BAV04236 BAV04186 ADV03045 AKV04081 AKV04071 AKV04069 AKV04068 AKV04066 AHV01253 AHV01196 AHV01149 ADV06302 ADV06205 ADV06034 ADV02071 AKV04067 AKV04070 AJV02623 BAV04706 AHV01486 AFV02609 AFV02610 BBV03010 AFV01191 AEV01974 AEV01777 ADV01692 AFV02006 AKV02507 BCV02706 BCV02687 BCV02682 BCV02671 AKV04344 AKV04343 AKV04342 AKV04341 AKV04340 AKV04338 AKV04327 AKV04326 AKV04339 AJV02892 AJV01695 AJV02201 BEV01244 AJV01610 AEV02390 BEV01137 BEV01127 BEV01126 BEV01124 BEV01123 BEV01118 BEV01117 BEV01116 BEV01111 BCV02354 BCV02346 BCV02340 AGV02721 AGV01009 AGV01004 AEV02594 ADV02004 ADV02021 AFV02576 BAV04255 BAV04252 BBV06033 BBV06003 BAV06741 AJV01275 AHV02084 AJV04035 AGV01074 AFV02577 AFV02319 BAV05164 BAV01024 BDV01421 BDV01403 BDV01401 BBV01586 BBV01572 BAV04049 BCV02020 BCV02019 BCV02018 BCV02017 BCV02016 BCV02014 BCV02013 BCV02012 BCV02011 BCV02010 BCV02009 BCV02008 BCV02007 BCV02006 BCV02005 BCV02004 BAV05033 BAV05031 BAV04997 BAV04984 BBV01347 BBV06019 BBV06009 BBV06008 AGV01165 AFV02624 AFV02615 AFV02595 AFV02155 AGV01146 AFV02574 AFV02572 AFV02540 AFV02529 AFV02490 AFV02605 BDV01589 BDV01583 BDV01572 BDV01571 BDV01564 BDV01553 BDV01551 BDV01289 BDV01273 BDV01260 BCV02719 AJV02150 AJV02132 AJV02026 AJV02021 AJV02013 AJV02012 AJV02011 AJV01711 AJV02154 AKV01476 BDV01986 BDV01811 BDV01809 BEV01071 BBV03561 AGV02553 AJV02483 AJV02480 BCV01308 BCY01026 BCY01025 BEV01813 AJV01806 AJV01687 BCV02495 BCV02491 BCV02329 BCV02315 BCV02312 BCV01318 BCV01317 BCV01043 BBV3596 BBV03595 AFV02283 AEV01046 ADV06282 AGV01553 AEV02745 AEV01727 ADV01538 ADV01535 AGV01338 AGV01312 BAV06268 AGV01721 ADV05015 AFV01327 BCV01542 BCV01540 BCV01537 BCV01536 BCV01523 AKV02737 AKV02736 BCV01255 BEV01469 BAV04245 AKV01648 BAV04069 BCV01313 AHV01633 AHV01617 AKV01451 AKV01280 AKV01167 AKV01452 AKV04131 AKV04130 AKV04128 AKV04063 AKV04062 AKV04059 AKV04057 AKV04052 AKV04051 AKV04050 AKV04049 AKV04048 AKV04047 AKV04046 AKV04044 AKV04043 AKV04042 AKV04039 AKV04038 AKV04037 AKV04036 AKV04035 AKV04034 AKV04032 AKV04031 AKV04030 AKV01775 AHV01091 AJV02743 BAV01154 BAV01158 BAV01156 BAV01164 AJV04113 AFV01717 AEV02043 AEV01997 AEV01986 AEV01979 AEV01968 BCV01210 BCV01204 BCV01200 BCV01194 BCV01193 BCV01192 BCV01184 BCV01180 BCV01174 AGV02698 AGV01882 AGV01798 AFV01386 AEV02080 AEV02079 AFV01375 AGV02522 BBV03061 AJV02315 BBV03062 BEV01343 BEV01342 BEV01341 BEV01337 BEV01334 BEV01818 BEV01812 BAV06650 BAV06649 BAV06646 BAV06652 AKV01736 AKV01723 BAV04747 BAV04745 BAV04751 BAV04750 BAV04755 BAV04748 AHV01867 BAV04805 BAV04806 BAV04816 BAV04813 BAV04810 BAV04804 AKV01737 AKV01840 BAV04817 BEV01327 BEV01262 AKV01462 BDV01860 BDV01851 BDV01840 BCV02283 BCV02282 BAV04366 BAV04360 BAV04359 BAV04369 BAV04365 AEV01711 ADV03111 ADV01585 ADV03042 ADV01899 BAV06024 BAV06018 BAV06012 BAV05295 BAV05262 BAV05257 BAV05246 BAV05244 BAV06014 BAV06022 BAV06007 BAV06025 AFV02585 ADV02092 AJV01640 BBV06100 BAV06017 BAV05249 BAV05245 AKV04499 AKV04123 ADV06308 AJV02468 AFV02359 AEV02222 AEV02207 AEV02182 ADV06238 BEV02138 BEV02130 BEV02129 BEV02128 BEV02126 BEV02122 BEV02121 BEV02118 BEV02115 BEV02094 BEV01506 BEV01505 BEV01504 BEV01503 BEV01501 BEV01500 BEV01499 BEV01498 BEV01497 BEV01495 BEV01493 BEV01492 BEV01491 BEV01490 BEV01489 BEV01488 BEV01487 BEV01486 BEV01485 BEV01484 BEV01483 BEV01482 BEV01481 BEV01480 BEV01479 BEV01478 BEV01476 BEV01474 BEV01473 BEV01354 AJV02428 AGV02405 AKV01694 BAV06019 BAV06011 BAV06010 BAV06008 BAV05258 BAV05255 BAV05254 BAV05248 BAV05260 BAV06015 BAV05241 BAV06184 AKV01693 AKV01692 AKV01690 AKV01688 AKV01686 AKV01683 AKV01669 AKV01661 AKV01660 AGV02015 AKV04500 AJV02782 AJV02780 AJV02775 AJV02771 AJV02744 BDV01316 BCV01397 BBV03241 BBV03239 BBV03238 BBV03237 AEV01829 BAV06518 BAV05177 BCV01391 ADV04115 BAV01139 BEV01964 BDV01862 AEV02170 AEV02169 AEV02148 AEV02051 AEV01866 AJV02948 BDV01374 BBV01612 BAV06800 BAV06797 BAV06794 BAV06793 BAV06788 BAV06787 BAV06786 BAV06785 BAV06776 BAV06095 BAV06094 AJV04187 AHV03024 AEV02229 AEV02158 AEV02147 AEV02126 AEV02095 AEV01675 ADV06223 ADV06140 ADV02130 ADV02081 AEV02160 BDV01454 BBV03122 BCV02855 BCV02606 BCV02345 BCV02268 BCV02267 AJV02829 AKV01224 AKV01180 ADV02134 BDV01342 BDV01331 BDV01319 BDV01304 BDV01173 BDV01471 BDV01465 BDV01358 BDV01355 BDV01354 BDV01350 BDV01348 BDV01321 BAV06795 BAV06792 BAV06790 BAV06789 BAV06779 BAV06006 BAV05298 AEV02121 BAV06806 BAV06799 BAV05300 BEV01521 BEV01516 BEV01194 BEV01189 BEV01187 BEV01115 BEV01110 BDV01757 BDV01755 BDV01537 BDV01536 BDV01532 BDV01527 BDV01458 AJV02827 BBV01614 BAV06456 BAV04059 BAV04056 AKV01925 AKV01923 AKV01912 AKV01910 AKV01909 AKV01908 AKV01904 BEV02203 BEV02202 BEV02201 BEV02200 BEV02199 BEV02198 BEV02196 BEV02195 BEV02194 BEV02192 BEV02190 AJV01182 AJV02598 AHV03106 AHV03104 AFV02030 BDV01363 AKV01443 AKV01434 AKV01417 AJV02872 ADV01827 BAV01170 BAV01167 BAV01127 BDV01507 BDV01499 BDV01493 BDV01492 BDV01486 BDV01484 BDV01483 BDV01475 BDV01473 BDV01468 BDV01282 BCV02703 BCV02702 BCV02679 BAV01168 AJV02600 AEV02704 BBV06103 AJV01106 BCV01138 BCV01250 BAV05171 BAV04333 AKV04125 BAV04339 AJV02585 AJV02583 AJV02581 AJV02548 AJV01107 AHV01151 AHV01145 AJV04209 BBV03549 BBV03548 BBV03547 BBV03546 BBV03545 BBV03543 BBV03542 BBV03535 BBV03514 BBV03513 BBV03512 BBV03511 BBV03409 BBV03408 BBV03407 BBV03406 BBV03405 BBV03404 BBV03401 BBV03400 BBV03399 BBV03397 BBV03396 BBV03395 BBV03392 BBV03391 BBV03390 BBV03387 BBV03386 BBV03383 BBV03382 BBV03381 BBV03380 BBV03249 BBV03248 BBV03247 BDV02016 BDV02015 BDV02014 BDV02003 BDV01997 BDV01985 BDV01979 BDV01972 BDV01875 BDV01813 BBV03240 BBV03130 ADV01645 BCV02879 BCV02876 BCV02859 BAV06693 BAV06464 AKV04484 AKV04482 BAV06772 BAV06771 BAV06040 BAV05159 BAV05158 BAV05157 BAV05154 BAV05152 BAV05151 BAV05148 BAV05146 BAV05145 BAV05144 BAV05143 BAV05142 BAV05141 BAV05140 BAV05137 BAV05136 BAV05135 BAV05134 BAV05133 BAV05132 BAV05130 AKV02195 AKV02194 AKV02178 AKV02169 AKV02135 AHV03331 AHV03327 AHV03302 BAV05147 BBV01007 AJV01502 BAV04152 BAV04151 BAV04150 BAV04144 BAV04140 BAV04131 BAV04128 BBV03387 BAV04838 BAV04811 BAV04145 AHV02016 AEV01765 AEV01753 ADV04100 ADV04008 BAV06763 BAV06761 BAV06759 ADV06327 ADV06309 ADV06190 BAV06773 AGV02291 AEV01703 BBV01026 BBV01024 AKV01780 AKV01373 AKV01385 BDV01610 BDV01597 BDV01505 BDV01497 BDV01298 BDV01271 BDV01270 BDV01266 BDV01265 BDV01250 BCV02707 BCV02680 BCV02662 BCV02650 BCV01261 BCV01260 BCV01259 BCV01258 BCV01254 BCV01252 BCV01251 BCV01149 BCV01148 BCV01147 BCV01146 BCV01145 BCV01144 BCV01143 BCV01142 BCV01141 BCV01140 BCV01136 BCV01111 BBV06111 BBV06074 BBV01059 BBV01058 BBV01057 BBV01055 BBV01054 BBV01003 BBV01056 BAV05168 BAV05166 BAV05165 BAV04340 BAV04338 BAV04337 BAV04335 BAV04275 AJV02490 AJV02486 AHV01166 AGV02304 AGV02226 AGV02193 AGV02081 AGV02217 AEV02266 AEV02214 AEV01831 BBV01053 BAV04721 AEV01822 AHV02140 BDV01356 BDV01345 BDV01312 BEV01464 BAV06197 AJV02391 AJV01693 AJV01690 AJV01569 AJV01207 AKV04346 BCV01709 BCV01621 BCV01618 BCV01615 BCV01610 BCV01604 AKV02547 BAV05100 BAV04825 BAV04798 AKV01474 AKV01473 AKV01467 BAV04814 AKV01455 AKV01454 AKV01458 AKV01427 AKV01318 AKV01314 AKV01967 AKV01460 AKV01461 AKV01464 AKV01414 AKV01457 AKV01344 AKV01292 BCV02870 BCV02868 BCV02866 BCV02864 BCV02857 BDV01777 BDV01764 BDV01533 BDV01530 AKV02721 AKV02720 AKV02706 AKV02705 ADV01522 AKV01300 BDV01977 BDV01814 BEV01109 BEV01106 BEV01105 BEV01102 BEV01100 BDV01861 BDV01858 BDV01857 BDV01854 BDV01513 BDV01511 BDV01496 BDV01464 BDV01455 BDV01389 BDV01386 BDV01352 BDV01346 BDV01332 BDV01498 BAV01108 BAV01109 BAV01106 BAV01107 BBV06116 BBV06119 AHV01395 AHV01116 AHV01018 AHV01002 BAV06475 ADV03113 BDV01414 BDV01404 BBV03138 AKV01930 AGV01809 BBV06142 BBV06141 BBV06140 BBV06139 BBV06137 BBV06136 BBV06135 BBV06134 BBV06133 BBV06122 BBV06121 BBV06118 BBV06115 BBV06114 BBV06110 BBV06087 BBV01049 BBV01022 BBV01021 BAV06728 BAV06715 BAV06714 BAV06711 BAV06707 BAV06705 BAV06704 BAV06699 BAV06698 BAV06697 BAV06696 AHV01482 AHV01404 AHV01394 AHV01357 AFV01523 AFV01002 BAV06710 ADV06406 ADV04031 ADV03055 ADV05003 BBV06138 BAV06701 AJV02501 AJV02179 AJV02164 AJV02159 BCV01690 BCV01626 BBV03591 BAV05119 BDV01654 BDV01652 BDV01651 BDV01447 BDV01441 BDV01439 AKV01489 AEV02030 AEV01948 AEV01714 AEV01682 AEV01610 AEV02187 BDV01988 BDV01596 AKV02597 BEV02167 BEV02164 BEV02158 BEV02156 BEV02152 BEV02151 BEV02002 BEV01496 BEV01351 BEV01350 BDV01216 BDV01213 BDV01204 BCV1009 BCV02251 BBV03246 BBV03245 BBV03146 BBV03123 BBV03001 BBV01635 BBV01634 BBV01630 BAV05233 BAV04812 AKV04074 AKV04073 AKV01670 AKV01668 AKV01662 AKV01295 AJV01742 AJV01135 AJV01104 AJV01064 AHV01308 AGV01618 AGV01535 AGV01532 AGV01464 AGV01410 BBV03002 BAV06429 AGV01569 AGV01463 BDV01555 BDV01615 BDV01595 BDV01584 BDV01323 BDV01322 BCV02630 BBV03544 BBV03534 BBV03533 BBV03532 BBV03531 BAV04593 AHV01910 AFV02473 AFV02332 AHV03304 BAV04592 BCV02305 BAV01115 AGV01871 AKV01931 AJV02769 AJV02735 AJV02734 AHV03158 AGV01968 AGV01873 AGV01870 AFV01499 AFV01471 AFV01468 AFV01467 AFV01441 AFV01430 AFV01395 ADV02154 AFV01435 AJV01700 AEV02747 AEV02617 AJV01248 BDV02009 BDV02008 BDV02000 BDV01999 AJV01196 AJV01222 AHV01060 AGV02662 AJV01197 aev01792 BEV01250 BCV01003 BCV01000 BAV06660 AKV01525 AJV02915 AJV01151 AEV01536 AJV02815 BCV01001 BBV03835 BBV01205 AGV01526 AGV01524 AKV01698 AKV01710 AKV01709 AKV01699 BDV01642 AEV01586 BEV02140 BBV06106 AJV02376 AGV02282 AFV01217 aev01921 BAV05098 BAV05094 BAV05099 AKV01530 AGV02089 ADV06381 AFV02500 ADV01553 BEV01764 BEV01760 AJV03041 AJV03026 AJV03007 AJV02989 AJV02907 AJV02895 AJV02666 AGV01846 AEV01764 BCV02163 BCV02161 BCV02843 BCV02842 BBV03129 BBV03108 AKV01223 AKV01174 AKV01533 AKV01528 AKV01527 AKV01523 AKV01522 AKV01520 BBV01280 AGV02564 AFV02061 AEV02092 AEV01619 AEV01617 BBV01204 BEV01266 BEV01265 BEV01261 BEV01253 BEV01252 BEV01249 BEV01248 BEV01247 BEV01246 BEV01243 BEV01239 BEV01237 BEV01236 BEV01234 BEV01229 BEV01223 BEV01216 BEV01204 BDV01768 BDV01706 BDV01641 BDV01638 BDV01637 BDV01636 BDV01634 BDV01630 BDV01617 BDV01580 BCV02025 BCV02024 BCV02023 BCV02001 BCV02000 BCV01998 BCV01270 BCV01268 BCV01267 BBV01011 BAV06048 BAV06433 BAV06426 BAV06052 BAV06051 BAV06049 AKV01771 AKV01743 AKV01218 AKV01217 AKV01216 AKV01214 AKV01194 AKV01163 AKV01158 AKV01149
AKV01144 AHV03310 AHV03213 AKV01215 BAV06428 BEV01165 BEV01162 BEV01160 BEV01145 BEV01119 BCV02608 BCV02607 BCV02604 BCV02598 BCV02590 BCV02567 BBV03712 BBV03711 BBV03710 BBV03659 BBV03658 BBV03657 BBV03656 BBV03655 BBV03653 BBV03652 BBV03649 BBV03648 BBV03647 BBV03646 BBV03645 BBV03644 BBV03643 BBV03642 BBV03641 BBV03640 BBV03639 BBV03638 BBV03637 BBV03636 BBV03635 BBV03634 BBV03633 BBV03632 BBV03631 BBV03630 BBV03629 BBV03628 BBV03627 BBV03555 BBV03366 BBV03361 BBV03357 BBV03356 BBV03354 BBV03353 BBV03347 BBV01271 BBV01270 BBV01269 BBV01268 BBV01267 BBV01266 BBV01265 BBV01264 BBV01263 BBV01262 BAV04803 BAV04802 BAV04801 BAV04799 BAV04626 BAV04625 BAV04624 BAV04623 BAV04621 BAV04620 BAV04619 AKV01944 AKV01920 AKV01362 AKV01358 AKV01355 AKV01351 AKV01348 AKV01345 AKV01336 AKV01334 AKV01333 AKV01329 AKV01328 AKV01326 AKV01325 AKV01324 AKV01323 AKV01313 AKV01312 AKV01311 AKV01310 AKV01309 AKV01308 AKV01307 AKV01306 AKV01305 AKV01304 AKV01303 AKV01272 AKV01257 AJV02342 AGV01024 AJV02325 BAV04629 AKV01327 BAV04800 AJV02326 BBV03654 BAV04618 BAV04617 AKV01347 AKV01350 AKV01339 AKV01338 AKV01337 BBV06102 AJV01605 BDV01644 BAV01268 AJV01692 BDV02012 BDV02007 BDV01983 BDV01978 BDV01976 BDV01649 AHV01161 AFV02497 AFV01981 AFV01907 AFV01893 AFV01870 AFV01835 AFV01819 BDV01812 BBV01276 BCV02193 BCV02185 BCV01834 BCV01833 BCV01265 BCV01264 AKV01789 AKV01787 AKV01786 AKV01784 AKV01782 AKV01779 AHV02014 AGV01812 AGV01782 AGV01764 AGV01757 AGV01780 BAV06512 AKV04312 BEV01807 BEV01802 BCV01513 BAV04224 BAV04051 BAV04050 BAV04052 BAV04053 ADV05172 AHV01082 AHV01013 AHV01009 AGV02717 AJV01674 AJV01656 AJV01682 BCV01849 BAV06513 BAV06511 AKV04313 AFV02053 AJV02434 BCV01517 AFV01267 AFV01284 AFV01275 AFV01271 AFV01262 AFV01241 AEV02351 AHV01296 AHV03336 AHV01123 AHV01061 AHV01006 AGV02708 BAV04268 AHV03271 BAV05170 BAV05163 AKV01079 AKV01062 AJV01325 AFV02530 AFV02083 AEV02332 AEV02288 AFV02329 BDV01296 BDV01272 BDV01269 BDV01268 BDV01267 AFV02260 BAV04702 BAV04467 AJV02957 AJV02880 AJV01038 AFV02416 AFV02264 AFV02201 AEV02387 BEV01477 BEV01465 BEV01254 BEV01251 BEV01242 BEV01231 BEV01655 BEV01652 BEV01650 BEV01649 BEV01645 BEV01177 BDV01850 BDV01788 BDV01360 BDV01343 BCV01583 BCV01582 BCV01581 BCV01580 BCV01579 BCV01577 BCV01574 BCV01566 BCV01506 BCV01395 BCV01386 BAV04783 AKV01792 AKV01790 AKV01733 AKV01729 AJV04075 AJV04071 AJV04065 AHV01036 AGV02400 AGV02324 BAV06534 BAV06626 BAV06535 BAV06769 BAV06765 BAV06768 BAV06767 AKV01419 AKV01343 AKV01331 AKV01330 BBV03216 BBV03215 BEV01752 BEV01747 BEV01679 BEV01340 BEV01339 BEV01338 BEV01336 BEV01335 BEV01332 BEV01331 BEV01330 BEV01302 BEV01297 AFV02357 BAV06042 AKV04322 AGV01020 AKV04323 AKV04321 BCV02309 BCV02308 BCV02295 BCV02292 BCV02286 BCV01314 BCV01311 BCV01310 BCV01302 BCV01049 BCV01032 BCV01026 BCV01018 BCV01017 BBV03811 BBV03707 BBV03705 BBV03702 BBV03700 BBV03502 BBV03501 BBV03500 BBV03491 BBV03490 BBV03489 BBV03488 BBV03487 BBV03482 BBV03473 AKV04497 AKV04496 AKV04495 AKV04493 AKV01973 AJV01621 AJV01618 AHV03262 AHV03248 AHV03245 BCV01030 BAV06659 AEV01597 BAV06658 BAV06656 BAV06653 BAV06655 BAV06754 BCV01781 BCV01780 BCV01779 BCV01764 BCV01778 BCV01949 BCV01948 BBV01513 BBV01512 BBV01488 BBV01485 BBV01327 BBV01509 AKV01058 AHV02064 AHV02040 AHV01849 AHV01837 BEV01502 BDV01834 BDV01829 BDV01828 BDV01794 BCV02609 BCV02602 BCV02592 BBV03236 BBV03234 BBV03233 BBV03230 BCV02391 BCV02376 BCV02367 BCV02362 BCV01744 BCV01695 BCV01687 BCV01686 BCV01643 BCV01629 BCV01082 BDV01643 AJV02374 AJV01115 BDV02004 BBV06042 BBV06058 BAV04336 BAV04315 BAV04314 BAV04311 BAV04307 BAV04303 BAV04298 BAV04297 BAV04296 BAV04284 BDV01989 BDV01973 BDV01817 BCV02716 BCV02688 BCV02686 BCV01698 BCV01697 BCV01688 BBV03594 BBV03593 BBV03592 BBV03141 BBV01610 BAV06766 AKV02227 AKV01086 AJV04222 AHV02124 AHV01137 AGV02568 AGV02530 AGV02210 AFV02532 AFV02298 AFV02120 AFV01343 AFV01289 AFV01209 AEV02708 AEV02339 AEV01569 AEV01551 ADV01876 BBV03388 BBV03385 BAV04622 BAV04614 BAV04613 BAV04612 BAV04611 BAV04610 BAV04609 BAV04608 AKV01731 AKV01730 AKV01727 AKV01721 AKV01691 BAV04615 AJV01548 AJV01542 BEV01766 BCV02844 BCV02841 BCV02840 BCV02632 BCV02631 BCV02360 BCV02233 BCV02228 BCV02226 BCV02225 BCV02223 BCV02221 BCV02214 BCV02207 BBV03720 BBV03718 BBV03717 BBV03716 BBV03715 BBV03144 BBV03133 BBV03132 BBV03114 BBV03113 BBV03006 BBV01629 BBV01628 BBV01625 BBV01623 BBV01621 BAV06246 BAV06236 BAV06183 AGV01752 AGV01751 AFV02561 AFV02008 AFV02004 AFV01761 ADV06146 ADV06066 ADV04057 BAV06177 BAV06176 BAV06185 BAV06241 BAV06238 BAV06237 BAV01175 BBV01616 BBV01615 BBV01611 BBV01609 BBV01603 BBV01546 BBV01332 BBV01273 AHV02098 BBV01613 BAV04787 BAV04782 BAV04778 BAV04773 AFV01933 BAV04774 AHV02118 BCV02347 BBV03692 BAV04455 BAV04424 BAV04414 BAV04413 BAV04401 BAV04394 AJV02103 BEV01463 BBV01605 BBV01604 BEV01089 BEV01067 BCV01377 BCV01375 BCV01373 BCV01363 AEV01914 AKV04413 AJV01698 AHV02093 AHV01401 AHV02121 BBV01046 AKV01463 AJV01683 AEV02364 AEV02367 AEV02321 BCV02696 BCV02684 BBV03218 BBV03147 BBV03020 BBV03018 BBV01608 BBV01607 BAV06770 BAV04763 BAV04060 BAV01112 BAV01050 BAV01049 AKV01650 AKV01649 AJV04199 AJV03071 AJV03017 AJV02994 AFV02308 AEV02446 ADV06252 ADV06181 ADV04058 BAV04761 BAV04749 BAV01110 BAV01114 BAV04754 AGV01743 AEV01775 AEV02570 AGV02515 AGV02106 AGV02075 AEV02746 BEV01008 BDV02049 BDV02048 BDV01664 BDV01662 BDV01442 BAV06642 AKV01997 AKV01989 AKV01988 AJV01214 AHV03378 AFV01150 AGV01929 BDV01831 BDV01786 BDV01776 BDV01743 BDV01741 BDV01738 BDV01722 BDV01717 BDV01713 BDV01710 AFV01272 BCV01829 BCV01828 BEV02106 BEV01875 BEV01860 BEV01856 AKV01453 AKV01277 AKV01241 AKV01210 AKV01190 AKV01188 AKV01187 AKV01171 AJV04082 AJV02396 AJV02290 AFV02210 BDV01197 AHV01888 BAV04124 BAV04118 BAV04070 BAV01044 BAV01039 BFV01027 BFV01026 BFV01025 BFV01023 BFV01020 BFV01019 BFV01013 BFV01003 AHV03355 AGV02222 ADV06041 BDV02010 BDV02005 BDV01998 BDV01996 BDV01993 BDV01992 BDV01990 BDV01987 BDV01984 ADV06314 ADV06199 BBV03590 AFV02434 AEV02549 BDV02082 ADV01765 AEV01796 AEV01761 BBV01275 AKV02739 BEV01494 BEV01475 BEV01355 BEV01353 BBV03140 BAV04367 BAV04364 BAV04363 BAV04361 BAV04756 AJV02983 AJV02766 AHV02111 AHV02094 AFV02536 AFV02427 AFV02380 AFV02363 AFV02117 AEV01564 AEV01541 ADV02138 AGV02152 AEV01038 ADV06070 ADV01846 ADV01844 BAV05112 BBV01060 AJV02941 BCV02307 AJV02840 AJV01409 AGV01183 AFV02588 AFV02580 AFV02564 BBV01631 BAV08013 AKV02276 AEV02500 AGV01667 AFV01211 AEV02480 BCV02849 BCV02846 BCV02845 AFV01302 BDV01746 BDV01740 BBV01633 BBV01627 BBV01624 BBV01622 AGV01533 AGV01487 AGV01445 AGV01300 BDV01771 BDV01767 BDV01763 BDV01341 AJV01644 BAV01121 AKV01977 BFV01005 BAV06837 BAV06824 BAV05215 BDV01310 AKV01876 AFV02014 BEV01615 AFV01767 AKV01871 BDV01592 BBV06081 AJV01772 AKV01971 AKV01969 AJV01313 AHV03288 AGV02309 AGV02302 BCV02717 AKV02544 AKV02543 AKV02257 AGV01860 AKV02542 AKV02244 BDV01647 BAV04419 BBV06108 BBV01027 BBV01020 AHV01432 BAV04423 BAV04422 BAV04421 BAV04420 BAV04417 BAV04416 BAV04345 BAV04344 BAV04343 BAV04342 AJV01775 AJV01631 AGV02377 AKV04065 AKV04064 AKV04002 AKV04000 aev01685 AKV02590 BAV04694 BAV04692 BCV01374 BCV01356 AEV02753 BCV02614 BCV01723 BAV06106 BAV06105 BAV06093 BAV06084 BAV06083 BAV06082 BAV06080 BAV06033 BAV01161 AJV02288 AHV01271 AHV01224 ADV01048 BBV01235 BAV01148 BBV01475 BBV01473 BAV01128 BAV01119 BEV01735 BAV04606 BAV04718 BBV03681 AJV02594 AJV02592 AJV02587 AJV02591 AJV02595 AJV02067 BCV01661 BBV01236 BAV06104 BAV06103 BAV06092 BAV06087 BAV06085 BAV06081 BAV06079 BAV06077 BAV06037 AKV04489 AKV04414 AKV04412 AKV04411 AJV04191 AJV03045 AJV02991 AJV02287 AHV01297 AGV02397 AEV02186 AEV01664 BCV01272 BCV01664 BAV04951 BDV01500 AKV04112 AGV02391 ADV03007 BBV06029 BAV04206 BBV06034 BBV06032 BBV06031 BBV06028 BBV06027 BBV06022 BBV06020 BBV06018 BBV06017 BBV06014 BBV06013 BBV06005 BEV01756 BEV01755 AJV02168 BCV01767 BCV01750 BCV01749 BCV01743 BCV01742 BCV01741 BCV01740 BCV01739 BCV01737 BCV01736 BCV01735 BCV01684 BCV01682 BCV01681 BCV01680 BCV01679 BCV01677 BCV01676 BCV01675 BCV01674 BCV01673 BCV01672 BCV01671 BCV01670 BCV01669 BCV01668 BCV01667 BCV01662 BCV01660 BCV01659 BCV01658 BCV01656 BCV01655 BCV01654 BCV01653 BCV01652 BCV01651 BCV01650 BCV01649 BCV01648 BCV01647 BCV01646 BCV01645 BCV01644 BCV01641 BCV01640 BCV01286 BCV01285 BCV01284 BCV01283 BCV01282 BCV01281 BCV01279 BCV01277 BCV01276 BCV01271 BCV01745 BAV05110 AJV02767 AJV02764 AJV02759 AJV02757 AJV02755 AJV02753 AJV02752 AJV02749 AJV02717 BBV06010 AEV02557 ADV06362 AGV01460 AFV01202 AFV01185 AFV01184 AFV01179 AFV01177 AFV01169 AFV01013 AEV01527 ADV06393 ADV06358 ADV06338 ADV06336 ADV06332 ADV06287 ADV06278 ADV06072 ADV06033 ADV01635 BCV01700 BAV06100 AJV01696 International Serial No. BAV01034 BAV01042 BBV06009 BCY01010 BDY01008 BDY01009 BDY01032 BDY01105 BDY01107 BDY01108 BDY01712 BDY01716 BBY01208 BCY01141 BCY01144 BCY01241 BCY01474 BCY01479 AGV02512 AGV01694 AKV01152 AKV01159 BAV06458 BBV03008 BBV03009 BBV03015 BBV03019 BEV02096 AKV01316 AKV01322 AJV01036 AKV01239 AKV01242 AKV01651 AKV01652 AKV01653 AKV01655 AKV01672 AKV01673 AKV01674 AKV01675 AKV01682 AKV01684 AKV01685 AKV01822 AKV01837 AKV01842 AKV01843 AKV01844 AKV01847 AKV01848 AKV01851 AKV01856 AKV01857 AKV01859 AKV01860 AKV01861 AKV01862 AKV01863 AKV01865 AKV01866 AKV01870 AKV01873 AKV01892 AKV01895 AKV01927 AKV01928 AKV01929 AKV01935 AKV01972 AKV01974 AKV02722 AKV02723 AKV02724 AKV02725 AKV02726 AKV02727 AKV02728 AKV02729 AKV02730 AKV02731 AKV02732 AKV02733 BAV01031 BAV04091 BAV04099 BAV05089 BAV05129 BAV05176 BAV05182 BAV06000 BAV06001 BAV06002 BAV06003 BAV06004 BAV06060 BAV06061 BAV06063 BAV06064 BAV06066 BAV06067 BAV06068 BAV06521 BAV06522 BAV06523 BAV06524 BAV06525 BAV06526 BAV06527 BAV06528 BAV06529 BAV06530 BAV06531 BAV06533 BBV01108 BBV01109 BBV01110 BBV01111 BBV01112 BBV01113 BBV01114 BBV01587 BBV01590 BBV01591 BBV01594 BBV01596 BBV03341 BBV03370 BBV03377 BBV03410 BBV03412 BBV03422 BBV03430 BBV03433 BBV03434 BBV03435 BBV03436 BBV03437 BBV03438 BBV03439 BBV03440 BBV03441 BBV03442 BBV03443 BBV03444 BBV03445 BBV03446 BBV03447 BBV03448 BBV03449 BBV03450 BBV03451 BBV03452 BBV03453 BBV03454 BBV03455 BBV03456 BBV03457 BBV03458 BBV03459 BBV03460 BBV03461 BBV03462 BBV03465 BBV03466 BBV03467 BBV03468 BBV03469 BBV03470 BBV03471 BBV03472 BBV03474 BBV03475 BBV03476 BBV03477 BBV03478 BBV03481 BBV03585 BCV01046 BCV01300 BCV01305 BCV01307 AKV01841 AKV01845 AKV01902 AKV01918 AKV01921 AKV01922 AKV01991 BAV04058 BAV04062 BAV04065 BAV04068 BAV04072 BAV04073 BAV04074 BAV04075 BAV04076 BAV04077 BAV04078 BAV04079 BAV04081 BAV04082 BAV04083 BAV04084 BAV04085 BAV04086 BAV04087 BAV04088 BAV04089 BAV04090 BAV04097 BAV04100 BAV04101 BAV04102 BAV04103 BAV04104 BAV04105 BAV04106 BAV04107 BAV04108 BAV04126 BAV04130 BAV04132 BAV05282 BAV05285 BAV05286 BAV05292 BAV05297 BAV06343 BAV06363 BAV06581 BAV06590 BCV01207 BCV01208 BCV01209 BCV01211 BCV01212 BCV01213 BCV01214 BCV01216 BCV01218 BCV01221 BCV01225 BCV01228 BCV01588 BCV01589 BCV01591 BCV01592 BCV01594 BCV01598 BCV01599 BCV01600 BCV01602 BCV01605 BCV01611 BCV01612 BCV01633 BCV01685 BCV01691 BCV01692 BCV01693 BCV01694 BCV01707 BCV01791 BCV01793 BCV01794 BCV01795 BCV01796 BCV01798 BCV01799 BCV01802 BCV01803 BCV01805 BCV01806 BCV01809 BCV02569 BCV02596 BCV02599 BCV02601 BCV02620 BCV02621 BCV02622 BCV02623 BCV02624 BCV02625 BCV02626 BCV02628 BCV02629 BCV02634 BDV01478 BDV01765 BDV01769 BDV01778 BEV01196 BEV01518 BEV01520 BEV01526 BEV01535 BEV01616 BEV01637 BEV01643 BEV01646 BEV01651 BEV01653 BEV01654 BEV01657 BEV01658 BEV01659 BEV01660 BEV01661 BEV01662 BEV01664 BEV01665 BEV01666 BEV01667 BEV01668 BEV01669 BEV01670 BEV01671 BEV01672 BEV01673 BEV01674 BEV01676 BEV01677 BEV01927 BEV02004 BEV02005 BEV02009 BBV01101 BBV01102 BBV01103 BBV03050 BBV03052 BBV03053 BBV03054 BCY01052 BCY01057 BCY01058 BCY01059 BCY01060 BCY01061 BCY01082 BCY01086 BCY01088 BCY01090 BCY01091 BDY01100 BDY01101 BDY01102 BDY01103 BDY01104 BDY01168 BEY01399 BEY01400 BAV01116 BCY01441 BCY01448 BEY01207 BEY01208 ADV02052 ADV01877 AHV01993 AHV01415 AHV02049 AJV04018 AKV01484 AKV01485 AKV01501 AKV01502 AKV01505 AKV01506 AKV01511 AKV01514 AKV01515 AKV02636 AKV02637 AKV02638 AKV02639 AKV02641 AKV02642 AKV02644 AKV02645 BAV01117 BAV01120 BAV01198 BAV01199 BAV04218 BAV04234 BAV04237 BAV04238 BAV04240 BAV04241 BAV04752 BAV04757 BAV04762 BAV04765 BAV06732 BAV06734 BAV06735 BAV06736 BAV06737 BAV06738 BAV06739 BAV06743 BBV01001 BBV01002 BBV01044 BBV01045 BBV01050 BBV01051 BBV01052 BBV03013 BBV03014 BBV06073 BBV06075 BBV06076 BBV06077 BBV06078 BBV06088 BBV06096 BBV06097 BBV06098 BBV06105 BBV06146 BCV01100 BCV01107 BCV01108 BCV01110 BCV01113 BCV01114 BCV01115 BCV01116 BCV01266 BCV01291 BCV01292 BCV01639 BCV01746 BCV01923 BCV01926 BCV01927 BCV01928 BCV01929 BCV01931 BCV01933 BCV01934 BCV01985 BCV01986 BCV01988 BCV01989 BCV01992 BCV01994 BCV02002 BCV02015 BCV02022 BCV02027 BCV02028 BCV02029 BCV02030 BCV02031 BCV02032 BCV02034 BCV02035 BCV02036 BCV02037 BCV02038 BCV02039 BCV02040 BCV02041 BCV02042 BCV02043 BCV02044 BCV02045 BCV02046 BCV02047 BCV02048 BCV02049 BCV02050 BCV02051 BCV02052 BCV02053 BCV02054 BCV02055 BCV02056 BCV02058 BCV02059 BCV02061 BCV02062 BCV02064 BCV02065 BCV02066 BCV02067 BCV02068 BCV02069 BCV02070 BCV02071 BCV02072 BCV02073 BCV02074 BCV02075 BCV02076 BCV02077 BCV02082 BCV02083 BCV02635 BCV02636 BCV02637 BCV02639 BCV02640 BCV02644 BCV02646 BCV02648 BCV02649 BCV02654 BCV02655 BCV02657 BCV02658 BCV02660 BCV02661 BCV02663 BCV02664 BCV02665 BCV02666 BCV02667 BCV02668 BCV02670 BCV02672 BCV02673 BCV02674 BCV02675 BDV01819 BEV01211 BAV06038 BAV06043 BBV01005 BBV01047 BBV01048 BAV04266 BAV06866 BAV06871 BAV06872 BAV06873 BAV06874 BAV06877 BAV06878 BAV06880 BAV06881 BAV06882 BAV06884 BAV06885 BAV06921 BCV01522 BCV01525 BCV02287 BCV02319 BCV02497 BCV02500 BCV02502 BCV02506 BCV02507 BCV02515 BCV02517 BCV02520 BCV02521 BCV02522 BCV02523 BCV02525 AHV03376 AJV04119 AKV01271 AKV01346 AKV01357 AKV01369 AKV01374 AKV01379 AKV01382 AKV01399 AKV01409 AKV01468 AKV01622 AKV01625 AKV01626 AKV01627 AKV01628 AKV01629 AKV01630 AKV01633 AKV01634 AKV01635 AKV01636 AKV01637 AKV01638 AKV01639 AKV01641 AKV01642 AKV01643 AKV01644 AKV02237 AKV02275 AKV02314 AKV02603 AKV02604 AKV02605 AKV02646 AKV02647 AKV02648 AKV02649 AKV02652 AKV02653 AKV02654 AKV02655 AKV02664 AKV02666 AKV02667 AKV02668 AKV02669 AKV02670 AKV02671 AKV02672 AKV02673 AKV02674 AKV02675 AKV02676 AKV02677 AKV02678 AKV02679 AKV04090 AKV04091 AKV04092 AKV04093 AKV04095 AKV04097 AKV04098 AKV04099 AKV04100 BAV01149 BAV01204 BAV01205 BAV01206 BAV01207 BAV01208 BAV01209 BAV01210 BAV01212 BAV01262 BAV04080 BAV04127 BAV04129 BAV04137 BAV04139 BAV04142 BAV04143 BAV04147 BAV04148 BAV04158 BAV04159 BAV04161 BAV04162 BAV04163 BAV04164 BAV04165 BAV04166 BAV04175 BAV04176 BAV04178 BAV04179 BAV04180 BAV04181 BAV04182 BAV04183 BAV04184 BAV04185 BAV04197 BAV04198 BAV04199 BAV04200 BAV04201 BAV04202 BAV04203 BAV04204 BAV04205 BAV04207 BAV04208 BAV04209 BAV04210 BAV04310 BAV04596 BAV04597 BAV04599 BAV04600 BAV04601 BAV04602 BAV04603 BAV04604 BAV04605 BAV04739 BAV06026 BAV06028 BAV06029 BAV06030 BAV06031 BAV06034 BAV06035 BAV06076 BAV06078 BAV06091 BAV06835 BAV06838 BAV06839 BAV06840 BAV06842 BAV06843 BAV06844 BAV06846 BAV06847 BAV06848 BAV06849 BAV06850 BAV06851 BAV06852 BAV06853 BAV06854 BAV06948 BAV06949 BAV06950 BAV06951 BAV06952 BAV06953 BBV01277 BBV01284 BBV01290 BBV01295 BBV01296 BBV01302 BBV01305 BBV01306 BBV01307 BBV01308 BBV01310 BBV01311 BBV01320 BBV01346 BBV01526 BBV01532 BBV01533 BBV01535 BBV01536 BBV01537 BBV01538 BBV01539 BBV01540 BBV01542 BBV01544 BBV01545 BBV01547 BBV01548 BBV03109 BBV03126 BBV03134 BBV03137 BBV03142 BBV03200 BBV03203 BBV03212 BBV03214 BBV03219 BBV03222 BBV03223 BBV03224 BBV03225 BBV03242 BBV03536 BBV03625 BBV03626 BBV03722 BBV03723 BBV03724 BBV03725 BBV03726 BBV03727 BBV03728 BBV03729 BBV03734 BBV03735 BBV03736 BBV03737 BBV03738 BBV03741 BBV03743 BBV03747 BBV03798 BCV01175 BCV01177 BCV01178 BCV01179 BCV01181 BCV01182 BCV01183 BCV01187 BCV01188 BCV01189 BCV01195 BCV01197 BCV01198 BCV01199 BCV01205 BCV01451 BCV01491 BCV01501 BCV01503 BCV01509 BCV01560 BCV01562 BCV01563 BCV01565 BCV01567 BCV01568 BCV01569 BCV01571 BCV01572 BCV01573 BCV01575 BCV01576 BCV01586 BCV01603 BCV01607 BCV01608 BCV01613 BCV01614 BCV01616 BCV01617 BCV01622 BCV01624 BCV01625 BCV01711 BCV01712 BCV01713 BCV01714 BCV01715 BCV01716 BCV01717 BCV01718 BCV01719 BCV01720 BCV01721 BCV01724 BCV01725 BCV01726 BCV01727 BCV01728 BCV01730 BCV01731 BCV01801 BCV01812 BCV01814 BCV01815 BCV01821 BCV02198 BCV02199 BCV02206 BCV02272 BCV02273 BCV02275 BCV02276 BCV02277 BCV02278 BCV02279 BCV02280 BCV02284 BCV02348 BCV02350 BCV02351 BCV02352 BCV02353 BCV02355 BCV02356 BCV02357 BCV02358 BCV02359 BCV02363 BCV02364 BCV02365 BCV02366 BCV02369 BCV02370 BCV02372 BCV02373 BCV02374 BCV02375 BCV02378 BCV02379 BCV02380 BCV02381 BCV02382 BCV02383 BCV02385 BCV02386 BCV02387 BCV02388 BCV02389 BCV02390 BCV02393 BCV02394 BCV02395 BCV02396 BCV02397 BCV02398 BCV02399 BCV02400 BCV02401 BCV02402 BCV02403 BCV02404 BCV02405 BCV02406 BCV02407 BCV02409 BCV02410 BCV02411 BCV02412 BCV02413 BCV02414 BCV02415 BCV02416 BCV02417 BCV02418 BCV02419 BCV02420 BCV02421 BCV02422 BCV02425 BCV02426 BCV02427 BCV02430 BCV02431 BCV02432 BCV02433 BCV02435 BCV02437 BCV02438 BCV02440 BCV02441 BCV02442 BCV02443 BCV02444 BCV02445 BCV02446 BCV02447 BCV02448 BCV02449 BCV02450 BCV02451 BCV02452 BCV02453 BCV02454 BCV02455 BCV02456 BCV02457 BCV02458 BCV02459 BCV02460 BCV02461 BCV02462 BCV02463 BCV02464 BCV02465 BCV02466 BCV02467 BCV02468 BCV02469 BCV02470 BCV02471 BCV02472 BCV02473 BCV02474 BCV02475 BCV02476 BCV02477 BCV02478 BCV02479 BCV02480 BCV02481 BCV02482 BCV02483 BCV02484 BCV02535 BCV02536 BCV02537 BCV02538 BCV02539 BCV02540 BCV02541 BCV02542 BCV02543 BCV02544 BCV02545 BCV02546 BCV02547 BCV02548 BCV02549 BCV02550 BCV02551 BCV02552 BCV02553 BCV02554 BCV02557 BCV02559 BCV02561 BCV02562 BCV02563 BCV02565 BCV02566 BCV02570 BCV02573 BCV02574 BCV02575 BCV02576 BCV02577 BCV02578 BCV02580 BCV02581 BCV02582 BCV02583 BCV02584 BCV02585 BCV02586 BCV02610 BCV02611 BCV02616 BCV02617 BCV02618 BCV02619 BCV02627 BCV02851 BCV02854 BDV01739 BDV01742 BDV01749 BDV01779 BDV01783 BDV01784 BDV01785 BDV01787 BDV01789 BDV01790 BDV01791 BDV01795 BDV01797 BDV01798 BDV01821 BDV01825 BDV01833 BEV01107 BEV01112 AKV01015 BBV01003 BBV01004 BBV01005 AGV01623 AHV02214 AFV01678 BBV01028 BBV01029 BBV01030 BBV01031 BBV01032 BBV01033 BBV01034 BBV01035 BBV01036 BBV01037 BBV01038 BBV01039 BBV01040 BBV01041 BBV01042 BBV01043 BBV06023 BBV06037 BEV01507 BEV01508 BEV01830 BEV01834 AJV02274 AKV02656 AKV02657 AKV02658 AKV02659 AKV02660 BAV01228 BAV01229 BAV01230 BAV01231 BAV01232 BAV01233 BAV01234 BAV01235 BAV01236 BAV01237 BAV01238 BAV01239 BAV01240 BAV01242 BAV01243 BAV01244 BAV01245 BAV01246 BAV01247 BAV01248 BAV01249 BAV01250 BAV01251 BAV01252 BAV01253 BAV01254 BAV01255 BAV01257 BAV01258 BAV01259 BAV01260 BAV01261 BAV04214 BAV04215 BAV04216 BAV04219 BAV04220 BAV04221 BAV04222 BAV04223 BAV04227 BAV04228 BAV04229 BAV04247 BAV04248 BAV04251 BAV04257 BAV04648 BAV04650 BAV04651 BAV04652 BAV04653 BAV04654 BAV04655 BAV04656 BAV04657 BAV04658 BAV04659 BAV04660 BAV04661 BAV04662 BAV04667 BAV04675 BAV06662 BAV06664 BAV06665 BAV06856 BAV06857 BAV06858 BBV01115 BBV01116 BBV01117 BBV01119 BBV01120 BBV01121 BBV01122 BBV01123 BBV01124 BBV01125 BBV01126 BBV01127 BBV01128 BBV01129 BBV01130 BBV01131 BBV01132 BBV01551 BBV01552 BBV01553 BBV01554 BBV01555 BBV01556 BBV01557 BBV01558 BBV01559 BBV01560 BCV02131 BCV02150 BCV02157 BCV02164 BCV02165 BCV02166 BCV02167 BCV02168 BCV02169 BCV02170 BCV02171 BCV02172 BCV02178 BCV02179 BCV02180 BAV04214 BBV01246 BDV01859 BAV01004 ADV02103 AJV04055 AKV01503 AKV01532 AKV01599 AKV01600 AKV01603 AKV01607 AKV01608 AKV01617 AKV01618 AKV01621 AKV01645 AKV01646 AKV01647 AKV01676 AKV01677 AKV01678 AKV01679 AKV01680 AKV01681 AKV01700 AKV01701 AKV01744 AKV02074 AKV02080 AKV02082 AKV02085 AKV02088 AKV02089 AKV02740 AKV02741 AKV02742 AKV02744 AKV02745 AKV02746 AKV02747 AKV02748 AKV02749 AKV02750 AKV02766 AKV02767 AKV02768 AKV02769 AKV02770 AKV02771 AKV02774 AKV02775 AKV02776 AKV02777 AKV02778 AKV02779 AKV02780 AKV02781 AKV02782 AKV02783 AKV02784 AKV02785 AKV02786 AKV02787 AKV02788 AKV02789 AKV02790 AKV02791 AKV02792 AKV02793 AKV02794 AKV02795 AKV02797 AKV02798 AKV02799 AKV02800 AKV02802 AKV02803 AKV02804 AKV02805 AKV02812 AKV02813 AKV02815 AKV02816 AKV02817 AKV02818 AKV02819 AKV02820 AKV02821 AKV02822 AKV02823 AKV02824 AKV02825 AKV02826 AKV02827 AKV02828 AKV02829 AKV02830 AKV02831 AKV02832 AKV02833 AKV02834 AKV02835 AKV02836 AKV02837 AKV02838 AKV02839 AKV02840 AKV02841 AKV02842 AKV02843 AKV02844 AKV02845 AKV02846 AKV02847 AKV02849 AKV02850 AKV02851 AKV02852 AKV02853 AKV02854 AKV02855 AKV02856 AKV02857 AKV02858 AKV02859 BAV01035 BAV01036 BAV01038 BAV01046 BAV01047 BAV01048 BAV01053 BAV01056 BAV04760 BAV06439 BDV01566 BDV01567 BDV01570 BDV01573 BDV01576 BDV01579 BDV01586 BDV01590 BDV01591 BDV01594 BDV01598 BDV01599 BDV01602 BDV01603 BDV01604 BDV01606 BDV01607 BDV01608 BDV01609 BDV01611 BDV01612 BDV01614 BDV01616 BVD01560 BAV01091 BDY01022 BDY01028 BDY01030 BDY01031 BDY01034 BAV06044 BCY01589 BCY01594 BCY01596 BVY01598 ADV01067 AFV02111 AFV02212 AFV01224 AFV02033 AKV01477 AKV02508 AKV02509 AKV02510 AKV02511 AKV02513 AKV02514 AKV02516 AKV02517 AKV02518 AKV02519 AKV02520 AKV02523 AKV02524 AKV04486 AKV04491 BAV01034 BAV04055 BBV01013 BBV01014 BCV02701 BDV01220 BDV01221 BDV01222 BDV01223 BDV01224 BDV01225 BDV01229 BDV01234 BDV01237 BDV01618 BDV01622 BDV01624 BDV01626 BDV01627 BDV01629 BDV01631 BDV01632 BDV01633 BDV01635 AKV04358 BAV01016 BAV01017 BAV01025 BAV04544 BAV04780 BAV04788 BBV01155 BBV01550 BBV01584 BBV01585 BBV01588 BBV01593 BBV01595 BBV01598 BAV06774 BBV03713 BCV01360 BCV01361 BCV01706 BCV02232 BCV02242 BCV02255 BEY01009 BEY01011 BAV01094 BCY01003 BCY01004 BCY01013 BCY01014 BCY01015 BCY01016 BCY01017 BCY01027 BCY01037 BBV01203 BCY01431 BDY01693 BCV01866 BCV01870 BCV01871 BCV01872 BCV01873 BCV01876 BCV01877 BCV01878 BCV01880 BCV01881 BCV01882 BCV01883 BEV01113 BEV01114 BEV01125 BEV01136 BEV01142 BAV04815 BBV01640 BCV01792 BCV02371 BCV02384 BCV1595 BEV01197 BEV01510 BEV01511 BEV01512 BEV01538 AKV02801 BAV05160 BAV06442 BAV06444 BAV06450 BAV06461 BCV02033 BCV02084 BCV02638 BCV02641 BCV02642 BCV02643 BCV02645 BCV02647 BDV01620 BDV01621 BDV01625 BAV04253 BAV04261 BAV04262 BAV04271 BBV03355 BBV03358 BBV03359 BBV03360 BBV03362 BBV03363 BBV03364 BBV03367 BBV03371 BBV03373 BBV03376 BBV03378 BBV03379 BBV03416 BBV03417 BBV03428 BBV03431 BBV03432 BCV01975 BCV02090 BCV02780 BDV01017 BDV01032 BDV01044 BDV01048 BDV01062 BDV01074 BDV01075 BDV01079 BDV01084 BDV01400 BDV01430 BDV01433 BVD01037 AKV01994 BBV03776 BDV01546 BDV01228 BEV01320 BDV01435 BCV02269 BCV02862 BDV01106 BDV01122 BDV01128 BDV01320 BDV01339 BDV01340 BDV01364 BDV01366 BDV01369 BDV01373 BDV01377 BDV01379 BDV01382 BDV01388 BDV01392 BDV01397 BDV01398 BDV01451 BDV01466 BDV01467 BDV01472 BDV01476 BDV01485 BDV01487 BDV01489 BDV01494 BDV01506 BDV01510 BDV01516 BDV01520 BDV01539 BDV01542 BDV01545 BDV01548 BDV01549 BDV01751 BDV01752 BDV01756 BDV01761 BDV01793 BDV01796 BDV01849 BDV01853 BDV01856 BDV01864 BDV01865 BEV01121 BEV01130 BEV01148 BEV01152 BEV01154 BEV01163 BEV01164 BEV01167 BEV01737 BEV01748 BEV01750 BEV01753 BEV01754 BAV01095 AGV02148 BAV01126 BAV01130 BAV01131 BCY01586 ADV01501 ADV01518 ADV06412 AFV02024 ADV06269 ADV06369 ADV06385 ADV06392 AEV02200 AFV01074 AGV01086 AGV01325 AGV01651 AGV01687 AGV02273 AHV01040 AHV01043 AKV01201 AKV01294 AKV01315 AKV04079 AKV04121 BAV04093 BAV04109 AKV01172 AKV01186 AKV01191 AKV01208 AKV01213 AKV01236 AKV01238 AKV01275 AKV01282 AKV01524 AKV02587 AKV04448 BCV01020 BCV01021 BCV01022 BCV01965 BDV01024 BDV01061 BDV01663 BEV01444 BEV01448 BDV01334 BDV01457 BDV01463 BCY01300 BAV01078 BEY01315 BEY01320 BEY01321 BEY01322 BBV03113 AJV02481 BAV04110 AKV01199 AKV01745 AKV02147 AKV02184 AKV02187 AKV02484 AKV02485 AKV02515 BAV01132 BAV01151 BAV01152 BAV01153 BAV01171 BAV01172 BAV04057 BDV01800 BAV04676 BAV04678 BAV04680 BAV04681 BAV04337 BBV01272 BAV04133 BAV04149 BAV01022 BAV01028 BAV01035 BAV01036 BAV01037 BAV01039 BBV01007 AKV01016 BAV01030 BAV01044 BBV03009 BBV03010 BBV06025 BCY01000 BCY01001 BBV03150 BAV01072 BAV01073 BAV01074 BBV03156 AJV01169 AKV02125 BAV01163 BAV01174 AKV01611 AKV01702 AKV01703 AKV01704 AKV01705 AKV01932 AKV01970 AKV01975 AKV04113 BAV06637 BAV06639 BAV06641 BCV02534 BCV02746 BCV02764 BCV02771 BCV02783 BCV02786 BCV02799 BEV01030 BEV01031 BCY01067 BCY01068 BCY01069 BEV01656 BEV01730 BEV01731 BEV01732 BEV01733 BEV01734 BEV01736 BEV01738 BEV01739 BEV01740 BEV01741 AKV02807 AKV02808 AKV02809 AKV02810 BAV01030 AJV01276 AJV02514 BAV04134 BBV01238 BBV01239 BBV01240 BBV01241 BBV01242 BBV01243 BBV01244 BBV01245 BBV01247 BBV01248 BBV01249 BBV01250 BBV01251 BBV01253 BBV01255 BBV01256 BBV01257 BBV01259 BBV01260 BBV01261 BCV01498 BCV01499 BCV01500 BCV01502 BCV01504 BCV01505 BCV01507 BCV01508 BCV01561 BCV01564 BCV01570 BCV02212 BCV02224 BCV02227 BCV02229 BCV02560 BCV02571 BCV02600 BDV01359 BDV01362 BDV01381 BDV01383 BEV01132 BEV01155 BEV01161 BEV01166 BEV01627 BEV01635 BEV01636 BEV01640 BCY01011 BCY01181 BCY01449 BCY01465 BDY01610 BEY01071 BEY01098 BAV06427 BAV06434 BAV06438 BAV06440 BAV06465 BBV06120 BBV06143 BBV06144 BBV06145 BBV06148 BCV01999 BBV01606 AKV01183 AKV01225 AKV01230 AKV01232 AKV01243 AKV01244 AKV01246 AKV01248 AKV01249 AKV01251 AKV01253 AKV01256 AKV01262 AKV01263 AKV01332 AKV01356 AKV01361 AKV01363 AKV01364 AKV01365 AKV01366 AKV01368 AKV01371 AKV01376 AKV01420 AKV01423 AKV01428 AKV01430 AKV01432 AKV01433 AKV01436 AKV01440 AKV01446 AKV01449 AKV01470 AKV01471 AKV01601 AKV01605 BAV01286 BAV01287 BAV01288 BAV01289 BAV01290 BAV01291 BAV01292 BAV01294 BAV01297 BAV01298 BAV01299 BAV01300 BAV01302 BAV01303 BAV01304 BAV01307 BAV04013 BAV04019 BAV04020 BAV04663 BAV04666 BAV04668 BAV04688 BAV04689 BAV04691 BAV04693 BAV04696 BAV04714 BAV04719 BAV04720 BAV04723 BAV04746 BAV04753 BAV04758 BAV04759 BAV04764 BAV04766 BAV04767 BAV06592 BAV06593 BAV06594 BAV06595 BAV06596 BAV06598 BAV06599 BAV06600 BAV06602 BAV06603 BAV06604 BAV06605 BAV06606 BAV06608 BAV06609 BAV06610 BAV06611 BAV06612 BAV06613 BAV06614 BAV06615 BAV06616 BAV06618 BAV06619 BAV06620 BAV06621 BBV01328 BBV01334 BBV01336 BBV01343 BBV01452 BBV01453 BBV01454 BBV01455 BBV01456 BBV01457 BBV01458 BBV01461 BBV01463 BBV01464 BBV01466 BBV01467 BBV01469 BBV01474 BBV01476 BBV01477 BBV01478 BBV01479 BBV01484 BBV03804 BBV03805 BBV03807 BBV03808 BBV03809 BBV03920 BBV03921 BBV03922 BBV03923 BBV03924 BBV03925 BBV03926 BBV03927 BBV03928 BBV03929 BBV03930 BBV03931 BBV03932 BBV03933 BBV03934 BBV03935 BBV03936 BBV03937 BBV03938 BBV03939 BAV01101 BDY01006 AEV01010 AKV02773 BAV06663 BAV06859 BCV02298 BCV02300 BDV01313 BDV01325 BAV01048 BDY01106 BDY01110 BDY01112 BDY01113 BDY01114 BDY01116 BDY01119 BDY01120 BDY01123 BDY01126 BDY01128 BDY01131 BDY01132 BDY01133 BDY01134 BDY01136 BDY01137 BDY01142 BDY01144 BDY01145 BDY01146 BDY01147 BDY01148 BDY01149 BDY01152 BDY01153 BDY01155 BDY01156 BDY01157 BDY01159 BDY01160 BDY01161 BDY01163 BDY01164 BDY01165 BDY01166 BDY01177 BDY01181 BDY01714 BDY01731 BDY01732 BDY01739 BDY01749 BDY01812 BCV01290 BAV04991 BAV04994 BDV01232 BDV01239 BDV01288 BEV01276 BEV01333 BEV01460 BEV01461 AKV01667 AKV01713 AKV01749 AKV01763 AKV01768 AKV01769 AKV01846 AKV02526 AKV02529 AJV02055 AKV02242 AKV02251 AKV02261 AKV02262 AKV02265 AKV02440 AKV02589 AKV04447 BAV01311 BAV01312 BAV05180 BAV05181 BAV05183 BAV06058 BAV06876 BAV06919 BAV06923 BBV01575 BCV01532 BCV01962 BCV02135 BCV02136 BCV02153 BCV02155 BCV02156 BCV02158 BCV02159 BCV02160 BCV02162 BCV02302 BCV02488 AKV02582 BAV04092 BAV04098 BAV04115 BAV04116 BAV04117 BAV04120 BAV04122 BAV04123 BAV04154 BAV04155 BAV04167 BAV04168 BAV04170 BAV04172 BAV04174 BAV04212 BAV04263 BAV04669 BAV04670 BAV04671 BAV04672 BAV04673 BAV04674 BAV04677 BAV04679 BAV04682 BAV04683 BAV04684 BAV04685 BAV04686 BAV04687 BAV04703 BAV04704 BAV04705 BAV04707 BAV04986 BAV05010 BAV05011 BAV05024 BCV01558 BCV01852 BCV01853 BCV01854 BCV01855 BCV01856 BCV01857 BCV01859 BCV01860 BCV01861 BCV01862 BCV01863 BCV01864 BCV01865 BCV01867 BCV01869 BCV01955 BCV01956 BCV01963 BCV02182 BCV02183 BCV02314 BCV02344 BCV02489 BCV02493 BCV02494 BCV02496 BCV02505 BCV02508 BCV02509 BCV02511 BCV02512 BCV02516 BCV02528 BCV02530 BCV02745 BCV02753 BCV02754 BCV02758 BCV02775 BDV01417 BDV02041 BDV02089 BEV01012 BEV01013 BEV01025 BEV01691 BEV01693 BEV01714 BEV01781 BEV01785 BEV01786 BEV01809 BEV01810 BEV01814 BEV01815 BEV01816 BEV01817 BEV01821 BEV01824 BEV01829 BEV02191 BEV02204 BEV02205 BEV02206 BEV02207 BFV01113 BFV01116 BFV01117 BFV01118 BFV01119 BFV01120 BFV01121 BFV01122 BFV01123 BFV01124 BFV01127 BFV01131 BFV01133 AKV01717 AKV01718 AKV01719 AKV01722 AKV01725 AKV01735 AKV01958 AKV01962 AKV01996 AKV02000 AKV02006 AKV02009 AKV02010 AKV02011 AKV02012 AKV02013 AKV02015 AKV02017 BAV04135 BAV04136 BBV01200 BBV01340 BBV01646 BBV03208 BBV03525 BBV03940 BCV01058 BCV01061 BCV01068 BCV01480 BCV01488 BCV01628 BCV01631 BCV01634 BCV02368 BCV01159 BAV04341 BAV04370 BAV04698 BAV04728 BAV04992 BAV04994 BAV05090 BAV05091 BAV05092 BAV05114 BAV05205 BAV05207 BAV05208 BAV05209 BAV05210 BAV05211 BAV05212 BAV05213 BAV05214 BAV05236 BAV05237 BAV05239 BAV05240 BAV05242 BAV05243 BAV05251 BAV05261 BAV05291 BAV05293 BAV06239 BAV06245 BAV06252 BAV06328 BAV06341 BAV06344 BAV06347 BBV01216 BBV01224 BBV01226 BBV01227 BBV01230 BBV01231 BBV01232 BBV01233 BBV01234 BBV01237 BBV01279 BBV01281 BBV01282 BBV01283 BBV01285 BBV01286 BBV01287 BBV01288 BBV01289 BBV01294 BBV01297 BBV01298 BBV01299 BBV01312 BBV01314 BBV01315 BBV01316 BBV01317 BBV01318 BBV01321 BBV01322 BBV01323 BBV01324 BBV01325 BBV01326 BBV01329 BBV01331 BBV01333 BBV01335 BBV01337 BBV01338 BBV01339 BBV01341 BBV01342 BBV01344 BBV01345 BBV01348 BBV01349 BBV01450 BBV01451 BBV01481 BBV01487 BBV01489 BBV01490 BBV01492 BBV01493 BBV01494 BBV01495 BBV01496 BBV01497 BBV01498 BBV01499 BBV01500 BBV01501 BBV01502 BBV01503 BBV01504 BBV01505 BBV01506 BBV01507 BBV01508 BBV01541 BBV01543 BBV01549 BBV01600 BBV01602 BBV03110 BBV03111 BBV03116 BBV03119 BBV03120 BBV03121 BBV03135 BBV03136 BBV03148 BBV03149 BBV03201 BBV03202 BBV03204 BBV03205 BBV03207 BBV03209 BBV03210 BBV03211 BBV03221 BBV03231 BBV03232 BBV03235 BBV03243 BBV03244 BBV03247 BBV03249 BBV03380 BBV03381 BBV03382 BBV03384 BBV03384 BBV03386 BBV03389 BBV03393 BBV03394 BBV03398 BBV03402 BBV03403 BBV03510 BBV03539 BBV03732 BBV03733 BBV03742 BBV03773 BBV03781 BBV03783 BBV03784 BBV04698 BCV01067 BCV01076 BCV01081 BCV01163 BCV01185 BCV01186 BCV01190 BCV01191 BCV01217 BCV01223 BCV01224 BCV01232 BCV01236 BCV01237 BCV01238 BCV01239 BCV01240 BCV01241 BCV01355 BCV01357 BCV01359 BCV01364 BCV01597 BCV01787 BCV01788 BCV01797 BCV01800 BCV02336 BDV01351 BDV01353 BDV01361 BDV01367 BDV01368 BDV01370 BDV01372 BDV01376 BDV01378 BDV01387 BDV01391 BDV01393 BDV01394 BDV01395 BDV01399 BDV01450 BDV01452 BDV01453 BDV01456 BEV01941 BEV02001 BEV02006 BEV02007 BEV02010 BEV02011 BDV01839 BEV01545 BEV01546 BEV01547 BEV01548 BEV01549 BEV01551 BEV01552 BEV01553 BEV01554 BEV01555 BEV01556 BEV01557 BEV01558 BEV01559 BEV01610 BEV01611 BEV01613 BEV01614 BEV01617 BEV01618 BEV01619 BEV01620 BEV01621 BEV01622 BEV01623 BEV01624 BEV01625 BEV01626 BEV01629 BEV01630 BEV01632 BEV01633 BEV01634 BEY01138 BDY01005 BDY01012 BDY01019 BEY01000 BCY01337 BCY01547 BDY01129 BDY01130 BDY01143 BCY01618 BCY01626 BCY01629 BDY01171 BDY01175 BDY01180 BDY01708 BDY01728 BEY01053 BEY01386 BEY01387 BEY01394 BBV03120 BBV03143 BBV03244 BAV01080 BAV01081 BCY01237 BCY01242 BEY01099 BDY01521 BDY01629 BDY01640 BDY01645 BDY01652 BDY01674 BDY01677 BDY01679 BDY01682 BDY01683 BDY01684 BDY01687 BEY01095 BEY01097 BEY01179 AFV01859 BCV01637 BEV01831 BEV01833 BEV01841 BEV01849 BAV04054 BAV04709 BAV04710 BAV04712 BAV04717 BAV05185 BAV06065 BAV06209 BAV06213 BAV06214 BAV06216 BAV06219 BAV06222 BAV06223 BAV06226 BAV06229 BAV06230 BBV03343 BBV03504 BBV03607 BBV03608 BBV03662 BBV03663 BBV03664 BBV03665 BBV03667 BBV03668 BBV03671 BBV03672 BBV03673 BBV03674 BBV03675 BBV03678 BBV03693 BBV03695 BBV03696 BBV03701 BBV03703 BBV03706 BBV03708 BBV03869 BCV01002 BCV01004 BCV01006 BCV01007 BCV01008 BCV01012 BCV01013 BCV01014 BCV01015 BCV01983 BCV02095 BCV02099 BCV02100 BCV02101 BCV02102 BCV02103 BCV02104 BCV02105 BCV02106 BCV02107 BCV02108 BCV02109 BCV02110 BCV02111 BCV02112 BCV02113 BCV02114 BCV02115 BCV02117 BCV02118 BCV02121 BCV02130 BCV03867 BDV01020 BDV01030 BDV01045 BDV01052 BDV01058 BDV01070 BDV01080 BDV01086 BDV01087 BDV01090 BDV01094 BDV01095 BDV01097 BDV01405 BDV01407 BDV01408 BDV01424 BDV01425 BDV01426 BDV01428 BDV01429 BDV01432 BDV01434 BEV01716 BEV02193 BEV02210 BEV02212 BFV01100 BCV01056 BAV01203 BAV01213 BAV01214 BAV01215 BAV01216 BAV05290 BAV06021 BAV06098 BAV06099 BAV06102 BAV06112 BBV01482 BBV03650 BBV03651 BBV03779 BBV03786 BBV03787 BBV03788 BBV03789 BBV03790 BBV03791 BBV03792 BBV03793 BCV01060 BCV01065 BCV01066 BCV01074 BCV01075 BCV01078 BCV01832 BCV02189 BCV02195 BCV02196 BCV02197 BCV02201 BCV02202 BCV02203 BCV02204 BCV02213 BCV02215 BCV02218 AKV02606 AKV02607 AKV02608 AKV02609 AKV02610 AKV02612 AKV02613 AKV02614 AKV02616 BDV01344 BDV01477 BCY01084 BCY01094 BCY01096 BCY01098 BCY01341 BCY01342 BCY01343 BCY01345 BCY01347 BCY01501 BDY01109 BDY01111 BDY01117 BDY01121 BDY01124 BDY01125 BDY01127 BDY01150 BDY01154 BDY01162 BDY01169 BDY01170 BDY01174 BDY01176 BDY01179 BDY01183 BDY01185 BDY01189 BDY01190 BDY01192 BDY01702 BDY01707 BDY01709 BDY01710 BDY01711 BDY01717 BDY01720 BDY01726 BDY01727 BDY01729 BDY01730 BDY01733 BDY01734 BDY01736 BDY01737 BDY01738 BDY01740 BDY01741 BDY01742 BDY01743 BDY01744 BDY01745 BDY01746 BDY01747 BDY01748 BDY01802 BDY01806 BDY01809 BDY01810 BEY01021 BEY01299 BEY01306 BEY01312 BEY01316 BEY01317 BEY01318 BEY01319 BEY01349 BEY01350 BEY01351 BEY01388 BEY01389 BEY01391 BEY01392 BEY01393 BEY01395 BEY01396 BEY01397 BEY01398 BEY01401 BCY01238 BCY01239 BCY01240 BCY01274 BCY01351 BCY01352 BCY01358 BCY01360 BCY01361 BCY01362 BDY01660 BEY01205 BEY01206 BEY01210 BEY01211 BEY01212 BEY01213 BEY01214 BEY01215 BEY01218 BEY01231 BEV01675 BEV01744 BEV01761 BEV01762 BEV01763 BEV01765 BEV01767 BEV01768 BEV01769 BEV01772 BEV01777 BEV01778 BEV01880 BEV01881 BEV01882 BEV01883 BEV01884 BEV01886 BEV01887 BEV01888 BEV01890 BEV01891 BEV01892 BEV01893 BEV01894 BEV01895 BEV01896 BEV01897 BEV01898 BEV01899 BEV01900 BEV01901 BEV01902 BEV01903 BEV01904 BEV01905 BEV01906 BEV01907 BEV01908 BEV01909 BEV01910 BEV01911 BEV01913 BEV01914 BEV01915 BEV01916 BEV01917 BEV01918 BEV01919 BEV01920 BEV01921 BEV01922 BEV01923 BEV01925 BEV01926 BEV01928 BEV01929 BEV01930 BEV01931 BEV01932 BEV01933 BEV01934 BEV01935 BEV01936 BEV01938 BEV01940 BEV01942 BEV01943 BEV01944 BEV01947 BEV01948 BEV01949 BEV01950 BEV01951 BEV01952 BEV01953 BEV01954 BEV01955 BEV01956 BEV01958 BEV01959 BEV01961 BEV01962 BEV01965 BEV01967 BEV01968 BEV01971 BEV01972 BEV01973 BEV01974 BEV01975 BEV01976 BEV01977 BEV01979 BEV01980 BEV01981 BEV01982 BEV01983 BEV01984 BEV01985 BEV01986 BEV01987 BEV01993 BEY01139 BEY01140 AFV02316 AKV04415 AKV04416 AKV04417 AKV04418 AKV04419 AKV04420 AKV04421 AKV02584 AKV04422 AKV04423 AKV04424 AKV04425 AKV04426 AKV04427 AKV04428 BAV04249 BAV04250 BAV01202 BCV01123 BDV01600 BDV01613 BDV01623 BDV01640 BCY01032 BAV04239 BDV01277 BDV01557 BEV01286 BEV01289 BEV01308 BEV01316 AKV01584 AKV01609 AKV02341 AKV02372 AKV02373 AKV02374 AKV02375 AKV02379 AKV02380 AKV02381 AKV02382 AKV02383 AKV02386 AKV02387 AKV02388 AKV02389 AKV02393 AKV02395 AKV02396 AKV02397 AKV02398 AKV02400 AKV02401 AKV02402 AKV02403 AKV02404 AKV02405 AKV02406 AKV02408 AKV02411 AKV02412 AKV02413 AKV02414 BAV04169 BAV04171 BAV04213 BAV04225 BAV04226 BAV04230 BAV06269 BAV06272 BAV06273 BAV06274 BCV02334 BCV02765 BDV01043 BDV01063 BDV01081 BEV01026 BCV01329 BCV01331 BCV01515 BCV01969 BCV01972 BAV04211 BEY01120 BAV01090 BAV01092 BAV01093 BBV01016 BCY01506 BCY01510 BCY01513 BCY01523 BCY01534 BCY01538 BCY01539 BCY01540 BCY01541 BCY01542 BEY01022 BEY01023 BEY01024 BEY01029 BEY01030 BEY01033 BEY01035 BEY01036 BEY01039 BEY01043 BEY01044 BEY01045 BEY01048 BEY01296 BBV03055 BBV03056 BBV03057 BBV03058 BCY01074 BCY01077 BCY01087 BCY01089 BCY01333 BCY01334 BCY01336 BEY01104 BEY01113 BEY01115 BEY01150 BEY01152 BEY01096 AGV01291 AGV01296 AGV01309 AGV01451 AGV01467 AGV02504 AGV02517 AGV02529 AFV01648 AEV01879 AEV01917 BAV01033 BCV02685 BCV02691 BCV02694 BCV02700 BCV02709 BCV02710 BCV02711 BCV02712 BCV02713 BDV01226 BDV01233 BDV01235 BDV01236 BDV01243 BDV01244 BDV01245 BDV01246 BDV01248 BDV01251 BDV01252 BDV01253 BDV01254 BDV01255 BDV01256 BDV01257 BDV01258 BDV01262 BDV01275 BDV01286 BDV01299 BDV01550 BAV01043 AJV04100 AKV02335 AKV02336 AKV02337 AKV02338 AKV02339 AKV02340 AKV04328 AKV04329 AKV04330 AKV04331 AKV04332 AKV04333 AKV04334 AKV04335 AKV04360 AKV04361 AKV04362 AKV04363 BAV01006 BAV01011 BAV01014 BAV01018 BAV01026 BAV08000 BAV08001 BAV08002 BAV08003 BAV08005 BAV08006 BAV08007 BAV08008 BAV08009 BAV08010 BAV08011 BAV08012 BAV08014 BAV08015 BAV08016 BAV08017 BAV08018 BAV08019 BAV08020 BBV08010 BBV08011 BBV08012 BBV08013 BBV08014 BBV08015 BBV08016 BBV08017 BBV08018 BBV08019 BBV09000 BCV11000 BCV11001 BDV01078 BDV01438 BDV01440 BDV01443 BDV01444 BDV01445 BDV01446 BDV01448 BDV01449 BDV01650 BDV01655 BDV01656 BDV01658 BEV01692 BEV01709 BEV01713 BEV01719 BEV01728 BEV01799 BAV01282 BAV01285 BBV08000 BBV09025 BBV09026 BBV09027 BBV09028 BCV02343 BCV02377 BCV02856 BDV01194 BDV01744 BEV01751 BFY01001 BAV01003 BAV01005 BAV01023 AGV01290 AGV01297 AGV01302 AGV02449 AGV02457 AJV01638 AJV01646 AJV01686 AJV01691 AJV02372 AJV02947 AKV01482 AKV02811 BAV01007 BAV05178 BAV05250 BBV08020 ADV01779 ADV02041 ADV02060 AKV01164 AKV01205 AKV01235 AKV01276 AKV01278 AKV01281 BCV02485 BCV02486 BEV01945 BEY01137 BFY01002 AHV01059 AGV02665 BEV01509 BEV01835 BEV01836 BEV01837 BEV01838 BEV01839 BEV01842 BEV01843 BEV01844 BEV01845 BEV01848 BEV01852 BEV01854 AKV01868 BAV04119 BAV04153 BAV04809 BAV05184 BAV06192 BAV06204 BAV06215 BAV06510 BAV06916 BAV06917 BAV06918 BAV06920 BAV06922 BBV03365 BBV03492 BBV03494 BBV03495 BBV03496 BBV03497 BBV03687 BCV01968 BCV02526 BCV02527 BCV02749 BDV01437 BEV01608 BEV01819 BBV03170 BBV03173 BBY01100 BBY01101 BCY01543 BDY01138 BDY01139 BDY01191 BDY01197 BDY01713 BFY01014 BFY01018 BDY01202 AEV02522 AFV01256 BAV01256 BAV04690 BAV04695 BAV04697 BAV04701 BAV04722 BAV04724 BAV04725 BAV04726 BAV04727 BAV04729 BAV04730 BAV04731 BAV04733 BAV04734 BAV04735 BAV04736 BAV04738 BAV04741 BAV04742 BAV04743 BAV04744 BAV05283 BAV05284 BAV05287 BAV05288 BAV05289 BAV05294 BAV05296 BAV06160 BAV06174 BAV06179 BAV06181 BAV06243 BAV06244 BAV06249 BAV06250 BAV06251 BAV06253 BAV06254 BAV06286 BAV06287 BAV06288 BAV06289 BAV06290 BAV06291 BAV06292 BAV06293 BAV06294 BAV06296 BAV06297 BAV06298 BAV06299 BAV06300 BAV06301 BAV06302 BAV06303 BAV06304 BAV06305 BAV06306 BAV06307 BAV06308 BAV06309 BAV06310 BAV06311 BAV06312 BAV06313 BAV06314 BAV06315 BAV06316 BAV06317 BAV06318 BAV06319 BAV06320 BAV06322 BAV06323 BAV06324 BAV06325 BAV06326 BAV06327 BAV06329 BAV06330 BAV06331 BAV06332 BAV06333 BAV06334 BAV06335 BAV06336 BAV06337 BAV06338 BCV01234 BDV01780 BDV01781 BDV01792 BDV01823 BDV01826 BDV01832 BDV01835 BDV01843 BDV01846 BDV01855 BBV01001 BBV01002 BBV06000 BBV06001 BBV06002 BBV06003 BBV06004 BBV06005 BBV06006 BBV06007 BBV06008 BBV06010 BDY01631 BDY01642 BDY01643 BDY01644 BDY01653 BDY01654 BDY01659 BDY01663 BDY01665 AFV02158 BAV06454 BAV06455 BCV01287 BCV01288 BAV06508 BAV01096 BEY01134 BBV03006 BBV03008 BCY01007 BCY01560 BEV02197 BFY01000 BAV04460 BAV04461 BAV04462 BAV04463 BAV06047 BAV06050 BAV06431 BAV06435 BAV06436 BAV06441 BCV02078 BCV02079 BCV02080 BCV02081 AKV02248 BCY01002 AKV02552 AKV02553 BAV04258 BAV04346 BAV04347 BCV02021 BCV02656 BCV02659 AKV02479 BAV06473 BDV01238 BDV01240 BDV01241 BEV01307 BEV01312 BEV01313 BEV01322 BEV01345 BEV01347 BEV02124 BEV02125 BEV02132 BEV02139 BEV02142 BEV02143 BEV02144 BEV02145 BEV02146 BEV02147 BEV02148 BEV02149 BEV02150 BEV02153 BEV02154 BEV02155 BEV02157 BEV02159 BEV02160 BEV02161 BEV02162 BEV02163 BEV02165 BEV02166 BEV02168 BEV02169 BEV02170 BEV02171 BEV02172 BEV02174 BEV02175 BEV02176 BEV02180 AKV01711 AKV01714 AKV02554 AKV02555 AKV02557 AKV04353 AKV04354 BAV01021 BAV01264 BAV01265 BAV01266 BAV01269 BAV01270 BAV01272 BAV01273 BAV01274 BAV01275 BAV01276 BAV01277 BBV03685 BBV03688 BBV03689 BCV01010 BCV01019 BCV01023 BCV01518 BCV01519 BCV01520 BCV01973 BCV02181 BCV02756 BCV02773 BCV02776 BCV02796 BAV06627 BAV06630 BAV06631 BAV06632 BAV06633 BAV06634 BAV06635 BAV06640 BCV02533 BDV01021 BDV01022 BDV01038 BDV01039 BDV01055 BDV01066 BDV01067 BDV01077 BDV01092 BDV02033 BDV02047 BDV02099 BEV01595 BEV01607 BEV01680 BEV01788 AKV02735 BAV01135 BAV01137 BAV01138 BAV01140 BAV01141 BAV01142 BAV01315 BAV04029 BAV04030 BAV04032 BAV04033 BAV04035 BAV04037 BAV04038 BAV04138 BAV04278 BAV04279 BAV04280 BAV04281 BAV04282 BAV04283 BAV04285 BAV04286 BAV04295 BAV04358 BCV02613 BDV01700 BDV01703 BDV01772 BDV01775 BEV01742 BEV01749 BEV01757 BEV01759 BEV01999 BEY01323 BEY01327 AEV01560 AEV02625 AEV02632 AGV02056 AGV02063 ADV05072 ADV05073 ADV05091 ADV05119 ADV05128 AKV01289 AKV01297 AKV01487 AKV01966 AKV01968 AKV01976 AKV01980 AKV01981 AKV01982 AKV01983 AKV01984 AKV01985 AKV01986 AKV01987 AKV01990 AKV01992 AKV01993 AKV01998 AKV02001 AKV02002 AKV02003 AKV02004 AKV02005 AKV02007 AKV02014 AKV02018 AKV02019 AKV02020 AKV02021 AKV02022 AKV02023 AKV02024 AKV02025 AKV02026 AKV02027 AKV02028 AKV02029 AKV02030 AKV02031 AKV02032 AKV02033 AKV02034 AKV02035 AKV02036
AKV02037 AKV02038 AKV02039 AKV02040 AKV02041 AKV02042 AKV02043 AKV02044 AKV02045 AKV02046 AKV02047 AKV02048 AKV02049 AKV02050 AKV02051 AKV02052 AKV02053 AKV02054 AKV02055 AKV02056 AKV02057 AKV02058 AKV02059 AKV02060 AKV02061 AKV02062 AKV02063 AKV02064 AKV02065 AKV02066 AKV02067 AKV02068 AKV02069 AKV02070 AKV02071 AKV02072 AKV02073 AKV02075 AKV02076 AKV02077 AKV02078 AKV02079 AKV02081 AKV02083 AKV02084 AKV02086 AKV02087 AKV02090 AKV02091 AKV02092 AKV02093 AKV02094 AKV02095 AKV02097 AKV02098 AKV02099 AKV02100 AKV02101 AKV02102 AKV02103 AKV02104 AKV02105 AKV02106 AKV02107 AKV02108 AKV02109 AKV02110 AKV02111 AKV02112 AKV02113 AKV02114 AKV02115 AKV02116 AKV02117 AKV02118 AKV02119 AKV02120 AKV02121 AKV02122 AKV02123 AKV02124 AKV02126 AKV02127 AKV02128 AKV02129 AKV02130 AKV02131 AKV02132 AKV02133 AKV02134 AKV02136 AKV02137 AKV02138 AKV02139 AKV02140 AKV02141 AKV02142 AKV02143 AKV02144 AKV02145 AKV02146 AKV02148 AKV02149 AKV02150 AKV02151 AKV02152 AKV02153 AKV02154 AKV02155 AKV02156 AKV02157 AKV02158 AKV02159 AKV02163 AKV02164 AKV02165 AKV02166 AKV02167 AKV02168 AKV02170 AKV02171 AKV02172 AKV02173 AKV02174 AKV02176 AKV02177 AKV02179 AKV02180 AKV02181 AKV02182 AKV02183 AKV02185 AKV02186 AKV02188 AKV02189 AKV02190 AKV02191 AKV02192 AKV02196 AKV02197 AKV02199 AKV02202 AKV02203 AKV02204 AKV02208 AKV02209 AKV02210 AKV02211 AKV02216 AKV02229 AKV02238 AKV02241 AKV02243 AKV02246 AKV02253 AKV02254 AKV02255 AKV02259 AKV02260 AKV02263 AKV02264 AKV02273 AKV02274 AKV02278 AKV02279 AKV02281 AKV02287 AKV02289 AKV02291 AKV02292 AKV02293 AKV02294 AKV02295 AKV02296 AKV02297 AKV02298 AKV02299 AKV02300 AKV02301 AKV02302 AKV02303 AKV02304 AKV02305 AKV02306 AKV02307 AKV02308 AKV02309 AKV02310 AKV02312 AKV02313 AKV02315 AKV02316 AKV02317 AKV02318 AKV02319 AKV02320 AKV02321 AKV02322 AKV02323 AKV02324 AKV02325 AKV02326 AKV02327 AKV02328 AKV02329 AKV02330 AKV02331 AKV02332 AKV02333 AKV02334 AKV02342 AKV02343 AKV02344 AKV02345 AKV02346 AKV02347 AKV02348 AKV02349 AKV02350 AKV02351 AKV02352 AKV02353 AKV02354 AKV02355 AKV02356 AKV02357 AKV02358 AKV02359 AKV02360 AKV02361 AKV02362 AKV02363 AKV02364 AKV02365 AKV02366 AKV02367 AKV02368 AKV02369 AKV02370 AKV02371 AKV02376 AKV02390 AKV02391 AKV02415 AKV02416 AKV02417 AKV02418 AKV02419 AKV02420 AKV02421 AKV02422 AKV02424 AKV02425 AKV02426 AKV02427 AKV02428 AKV02429 AKV02430 AKV02431 AKV02432 AKV02433 AKV02434 AKV02435 AKV02436 AKV02437 AKV02438 AKV02439 AKV02441 AKV02442 AKV02443 AKV02444 AKV02445 AKV02446 AKV02447 AKV02448 AKV02449 AKV02450 AKV02451 AKV02452 AKV02453 AKV02454 AKV02455 AKV02456 AKV02457 AKV02459 AKV02460 AKV02461 AKV02462 AKV02463 AKV02464 AKV02465 AKV02466 AKV02467 AKV02468 AKV02469 AKV02470 AKV02476 AKV02477 AKV02512 AKV02521 AKV02522 AKV02525 AKV02536 AKV02537 AKV02538 AKV02539 AKV02540 AKV02541 AKV04075 AKV04077 AKV04078 AKV04080 AKV04084 AKV04086 AKV04087 AKV04088 AKV04106 AKV04133 AKV04134 AKV04135 AKV04136 AKV04137 AKV04138 AKV04139 AKV04140 AKV04141 AKV04142 AKV04143 AKV04144 AKV04145 AKV04146 AKV04147 AKV04148 AKV04150 AKV04151 AKV04152 AKV04153 AKV04154 AKV04155 AKV04156 AKV04157 AKV04158 AKV04159 AKV04160 AKV04161 AKV04164 AKV04165 AKV04166 AKV04167 AKV04168 AKV04169 AKV04170 AKV04171 AKV04172 AKV04173 AKV04174 AKV04175 AKV04176 AKV04177 AKV04178 AKV04179 AKV04180 AKV04181 AKV04182 AKV04242 AKV04243 AKV04244 AKV04245 AKV04246 AKV04247 AKV04248 AKV04249 AKV04250 AKV04251 AKV04252 AKV04253 AKV04254 AKV04255 AKV04256 AKV04257 AKV04258 AKV04259 AKV04260 AKV04261 AKV04262 AKV04263 AKV04264 AKV04265 AKV04266 AKV04267 AKV04268 AKV04269 AKV04270 AKV04272 AKV04275 AKV04277 AKV04278 AKV04279 AKV04280 AKV04281
AKV04282 AKV04283 AKV04284 AKV04285 AKV04286 AKV04287 AKV04319 BAV01037 BAV01051 BAV01052 BAV01054 BAV01057 BAV01058 BAV01059 BAV01060 BAV01061 BAV01062 BAV01063 BAV01064 BAV01065 BAV01066 BAV01067 BAV01068 BAV01070 BAV01071 BAV01072 BAV01073 BAV01074 BAV01075 BAV01076 BAV01077 BAV01078 BAV01079 BAV01080 BAV01081 BAV01082 BAV01083 BAV01084 BAV01085 BAV01086 BAV01087 BAV01090 BAV01091 BAV01098 BAV01100 BAV01101 BAV01102 BAV01103 BAV01104 BAV01105 BAV04006 BAV04007 BAV04988 BAV06449 BAV06452 BAV06453 BBV01009 BBV01010 BDV02013 BEV01468 BEV01840 BEV01847 BEV01850 BEV01851 BEV01853 BEV01855 BEV01858 BEV01859 BEV01861 BEV01862 BEV01863 BEV01864 BEV01867 BEV01868 BEV01869 BEV01870 BEV01871 BEV01872 BEV01873 BEV01874 BEV01876 BEV01877 BEV01878 BEV02091 BEV02092 BEV02093 BEV02095 BEV02097 BEV02098 BEV02100 BEV02101 BEV02102 BEV02103 BEV02104 BEV02105 BEV02107 BEV02108 BEV02109 BEV02111 BEV02112 AKV01526 AKV01534 AKV01535 AKV01536 AKV01537 AKV01538 AKV01539 AKV01540 AKV01541 AKV01542 AKV01543 AKV01544 AKV01545 AKV01546 AKV01547 AKV01548 AKV01549 AKV01550 AKV01551 AKV01552 AKV01553 AKV01554 AKV01555 AKV01556 AKV01557 AKV01558 AKV01559 AKV01560 AKV01561 AKV01562 AKV01563 AKV01564 AKV01565 AKV01566 AKV01567 AKV01569 AKV01570 AKV01571 AKV01572 AKV01573 AKV01574 AKV01575 AKV01576 AKV01577 AKV01578 AKV01579 AKV01580 AKV01581 AKV01583 AKV01586 AKV01707 AKV01724 AKV01728 AKV01732 AKV01734 AKV01747 AKV01754 AKV01755 AKV01759 AKV01762 AKV01791 AKV01794 AKV01796 AKV01797 AKV01798 AKV01799 AKV01800 AKV01801 AKV01802 AKV01803 AKV01804 AKV01805 AKV01806 AKV01808 AKV01809 AKV01810 AKV01811 AKV01812 AKV01813 AKV01814 AKV01816 AKV01818 AKV01819 AKV01820 AKV01821 AKV01823 AKV01824 AKV01825 AKV01827 AKV01828 AKV01829 AKV01830 AKV01831 AKV01832 AKV01833 AKV01834 AKV01835 AKV01836 AKV01838 AKV01839 AKV02661 AKV02663 AKV04107 AKV04108 AKV04109 AKV04110 AKV04111 BAV04411 BAV06509 BAV06661 AKV01666 AKV01671 AKV01907 AKV01911 AKV01960 AKV01978 AKV01979 AKV02222 AKV02232 AKV02234 AKV02235 AKV04307 AKV04308 AKV04309 AKV04310 AKV04311 BAV04036 BAV04294 BAV05107 BAV05109 BAV06357 BAV06579 BAV06583 BAV06586 BAV06587 BAV06588 BAV06591 BAV06798 BAV06810 BAV06811 BAV06812 BAV06814 BAV06815 BAV06816 BAV06817 BAV06818 BAV06819 BAV06820 BAV06821 BAV06822 BAV06823 BAV06825 BAV06826 BAV06827 BAV06828 BAV06829 BAV06831 BAV06832 BAV06833 BAV06834 BAV06836 BBV01617 BBV01618 BBV01619 BBV01626 BBV01636 BBV01637 BBV01638 BBV01639 BBV01641 BBV01642 BBV01643 BBV03100 BBV03101 BBV03102 BBV03103 BBV03104 BBV03105 BBV03106 BBV03107 BBV03806 BBV03941 BBV03942 BBV03943 BBV03944 BBV03945 BBV03946 BBV03947 BBV03948 BBV03949 BBV03950 BBV03952 BBV03953 BCV01051 BCV01052 BCV01053 BCV01054 BCV01055 BCV01063 BCV01070 BCV01071 BCV01073 BCV01077 BCV01079 BCV01080 BCV01083 BCV01084 BCV01085 BCV01086 BCV01087 BCV01088 BCV01089 BCV01090 BCV01091 BCV01092 BCV01093 BCV01094 BCV01095 BCV01096 BCV01097 BCV01098 BCV01099 BCV01150 BCV01151 BCV01152 BCV01153 BCV01154 BCV01155 BCV01156 BCV01157 BCV01158 BCV01160 BCV01161 BCV01162 BCV01164 BCV01165 BCV01166 BCV01167 BCV01168 BCV01169 BCV01170 BCV01171 BCV01172 BCV01173 BCV01371 BCV01383 BCV01384 BCV01385 BCV01387 BCV01394 BCV01400 BCV01401 BCV01402 BCV01403 BCV01404 BCV01405 BCV01406 BCV01407 BCV01408 BCV01409 BCV01410 BCV01411 BCV01412 BCV01413 BCV01414 BCV01415 BCV01416 BCV01417 BCV01418 BCV01419 BCV01420 BCV01421 BCV01422 BCV01423 BCV01424 BCV01425 BCV01426 BCV01427 BCV01428 BCV01429 BCV01430 BCV01431 BCV01432 BCV01433 BCV01434 BCV01435 BCV01436 BCV01437 BCV01438 BCV01439 BCV01440 BCV01441 BCV01442 BCV01443 BCV01444 BCV01445 BCV01446 BCV01447 BCV01448 BCV01449
BCV01450 BCV01452 BCV01453 BCV01454 BCV01455 BCV01456 BCV01457 BCV01458 BCV01459 BCV01460 BCV01461 BCV01462 BCV01463 BCV01464 BCV01465 BCV01466 BCV01467 BCV01468 BCV01469 BCV01470 BCV01471 BCV01472 BCV01473 BCV01474 BCV01475 BCV01476 BCV01477 BCV01478 BCV01479 BCV01481 BCV01482 BCV01483 BCV01484 BCV01485 BCV01486 BCV01487 BCV01489 BCV01490 BCV01492 BCV01493 BCV01494 BCV01495 BCV01496 BCV01497 BCV02274 BCV02858 BCV02860 BCV02863 BCV02867 BCV02869 BCV02871 BCV02875 BCV02877 BCV02880 BDV01100 BDV01101 BDV01102 BDV01103 BDV01107 BDV01108 BDV01109 BDV01110 BDV01111 BDV01112 BDV01113 BDV01115 BDV01116 BDV01117 BDV01118 BDV01119 BDV01120 BDV01121 BDV01123 BDV01124 BDV01125 BDV01126 BDV01127 BDV01129 BDV01130 BDV01131 BDV01132 BDV01133 BDV01134 BDV01135 BDV01136 BDV01137 BDV01138 BDV01139 BDV01140 BDV01141 BDV01142 BDV01143 BDV01144 BDV01145 BDV01146 BDV01148 BDV01149 BDV01150 BDV01151 BDV01153 BDV01154 BDV01155 BDV01156 BDV01157 BDV01158 BDV01159 BDV01161 BDV01162 BDV01164 BDV01165 BDV01166 BDV01167 BDV01169 BDV01170 BDV01171 BDV01172 BDV01174 BDV01177 BDV01178 BDV01179 BDV01180 BDV01182 BDV01183 BDV01184 BDV01185 BDV01186 BDV01189 BDV01190 BDV01191 BDV01193 BDV01195 BDV01196 BDV01198 BCV01242 BCV01243 BCV01244 BCV01245 BCV01246 BCV01247 BDV01300 BDV01309 BEY01297 BEY01314 BBV01206 BBV01207 BBV01208 BBV01209 BBV01210 BBV01211 BBV01212 BBV01213 BBV03297 BBY01210 BBY01218 BBY01219 BBY01220 BBY01227 BBY01228 BBY01232 BBY01233 BBY01235 BBY01236 BBY01238 BBY01240 BBY01241 BBY01242 BBY01243 BCY01100 BCY01101 BCY01102 BCY01103 BCY01104 BCY01105 BCY01106 BCY01107 BCY01108 BCY01109 BCY01110 BCY01112 BCY01114 BCY01115 BCY01116 BCY01117 BCY01118 BCY01119 BCY01120 BCY01121 BCY01122 BCY01123 BCY01124 BCY01127 BCY01128 BCY01129 BCY01131 BCY01132 BCY01133 BCY01134 BCY01135 BCY01136 BCY01137 BCY01139 BCY01140 BCY01142 BCY01143 BCY01145 BCY01146 BCY01147 BCY01149 BCY01150 BCY01151 BCY01153 BCY01154 BCY01155 BCY01156 BCY01157 BCY01158 BCY01159 BCY01160 BCY01161 BCY01162 BCY01163 BCY01164 BCY01165 BCY01166 BCY01167 BCY01168 BCY01169 BCY01170 BCY01171 BCY01172 BCY01173 BCY01174 BCY01175 BCY01176 BCY01177 BCY01178 BCY01179 BCY01180 BCY01182 BCY01183 BCY01184 BCY01185 BCY01186 BCY01187 BCY01188 BCY01189 BCY01190 BCY01191 BCY01192 BCY01193 BCY01195 BCY01476 BDY01200 BDY01201 BDY01203 BDY01204 BDY01205 BDY01206 BDY01207 BDY01208 BDY01209 BDY01210 BDY01211 BDY01212 BDY01213 BDY01215 BDY01216 BDY01217 BDY01218 BDY01219 BDY01220 BDY01221 BDY01222 BDY01223 BDY01224 BDY01225 BDY01226 BDY01227 BDY01228 BDY01229 BDY01230 BDY01232 BDY01233 BDY01234 BDY01235 BDY01236 BDY01237 BDY01238 BDY01239 BDY01241 BDY01242 BDY01243 BDY01244 BDY01245 BDY01246 BDY01247 BDY01248 BDY01249 BDY01250 BDY01251 BDY01252 BDY01253 BDY01254 BDY01255 BDY01256 BDY01257 BDY01258 BDY01259 BDY01260 BDY01261 BDY01262 BDY01263 BDY01264 BDY01265 BDY01266 BDY01267 BDY01268 BDY01269 BDY01270 BDY01271 BDY01272 BDY01273 BDY01274 BDY01275 BDY01276 BDY01277 BDY01278 BDY01279 BDY01280 BDY01281 BDY01283 BDY01284 BDY01286 BDY01287 BDY01288 BDY01289 BDY01290 BDY01291 BDY01292 BDY01293 BDY01294 BDY01295 BDY01296 BDY01297 BDY01298 BDY01299 BDY01500 BDY01501 BDY01502 BDY01503 BDY01504 BDY01505 BDY01507 BDY01508 BDY01509 BDY01510 BDY01511 BDY01512 BDY01513 BDY01514 BDY01515 BDY01516 BDY01517 BDY01518 BDY01519 BDY01520 BDY01522 BDY01523 BDY01524 BDY01525 BDY01526 BDY01527 BDY01528 BDY01529 BDY01530 BDY01531 BDY01532 BDY01533 BDY01534 BDY01535 BDY01536 BDY01537 BDY01538 BDY01539 BDY01540 BDY01541 BDY01542 BDY01543 BDY01544 BDY01545 BDY01546 BDY01547 BDY01548 BDY01549 BDY01550 BDY01551 BDY01552 BDY01553 BDY01554 BDY01555
BDY01556 BDY01570 BDY01574 BDY01577 BDY01578 BDY01579 BDY01581 BDY01589 BDY01598 BDY01600 BEY01232 BEY01235 BEY01236 BEY01237 BEY01238 BEY01239 BEY01240 BEY01241 BEY01242 BEY01243 BEY01244 BEY01245 BEY01246 BEY01247 BEY01248 BEY01249 BEY01250 BEY01251 BEY01252 BEY01253 BEY01254 BEY01255 BEY01256 BEY01257 BEY01258 BEY01259 BEY01260 BEY01261 BEY01262 BEY01263 BEY01264 BEY01265 BEY01266 BEY01267 BEY01268 BEY01269 BEY01420 BEY01421 BEY01422 BEY01423 BEY01424 BEY01425 BEY01426 BEY01427 BEY01428 BEY01429 BEY01430 BEY01431 BEY01432 BEY01433 BEY01434 BEY01435 BEY01436 BEY01437 BEY01438 BEY01439 BEY01440 BEY01441 BEY01442 BEY01443 BEY01445 BEY01446 BEY01447 BEY01449 BEY01451 AKV01612 AKV01613 AKV01654 BBV01161 BBV01162 BBV01580 BCV02555 BCV02556 BDV01517 BDV01522 BDV01534 BAV04041 BAV04042 BAV04043 BAV04044 BAV04045 BAV04046 BAV04047 BAV04048 BAV04293 BAV06089 BAV06101 AKV01002 AKV01019 AKV01025 BAV01008 BAV01010 BAV01016 BAV01019 BAV01024 BAV01025 BDY01011 BDY01013 BEY01003 BEY01005 BCY01029 BCY01051 BDY01158 BDY01182 BEY01028 BEY01031 BEY01049 BCY01544 BCY01545 AGV01462 AJV01207 BAV01283 BEY01133 BEY01014 BEY01016 BEY01122 BEY01127 BEY01131 BEY01100 BEY01102 BEY01103 BEY01105 BEY01108 BEY01109 BEY01111 BEY01118 BEY01151 BEY01154 BEY01101 BEY01107 BEY01110 BEY01112 BEY01114 BEY01116 BEY01117 BEY01119 BEY01155 BEY01159 BEY01161 BEY01162 BEY01163 BEY01164 BEY01165 BEY01166 BEY01167 BEY01170 BEY01171 BEY01172 BEY01173 BEY01174 BEY01175 BEY01177 BEY01178 BAV06486 BAV06791 BAV06802 BAV06803 BAV06804 BAV06807 BAV06808 BAV06809 BDV01192 BEV01912 BEV01946 BEV01957 BEV01960 BEV01969 BEV01988 BEV01989 BEV01990 BEV01991 BEV01992 BEV01994 BEV01995 BEV01996 BEV01997 BEV01998 AKV01656 BCV02318 BCV02325 BCV02487 BCV02498 BCV02504 BCV02210 BCV02216 BCV02217 BDV01502 BDV01773 BEV01811 BEV01016 BEV01035 BEV01770 BEV01771 BEV01773 BEV01774 BEV01775 BEV01776 BEV01779 BEV01885 BEV01889 BDV01261 BDV01284 BDV01561 BDV01562 BDV01569 BDV01575 BDV01578 BDV01581 BDV01587 BDV01588 BDV01593 BEV02177 BEV02179 AFV02336 BAV04095 BAV04189 BAV04190 BAV04192 BAV04194 BAV04195 BAV04196 BAV04231 BAV04232 BCV02676 BCV02695 BCV02704 BCV02705 BCV02708 BCV02714 BCV02718 BCV02720 BDV01201 BDV01202 BDV01203 BDV01205 BDV01206 BDV01207 BDV01210 BDV01211 BDV01212 BDV01217 BDV01218 BDV01219 AKV01038 BAV01176 BAV01177 BAV01178 BAV01179 BAV01180 BAV01181 BAV01182 BAV01183 BAV01184 BAV01185 BAV01186 BAV01187 BAV01188 BAV01189 BAV01190 BAV01191 BAV01192 BAV01193 BAV01194 BAV01195 BAV01196 BAV01197 BAV04987 BAV04989 BAV04990 BBV01190 BBV01437 BBV03605 BBV03683 BBV03684 BBV03704 BBV03810 BBV03812 BBV03813 BBV03814 BBV03815 BBV03816 BBV03817 BBV03818 BBV03819 BBV03820 BBV03821 BBV03822 BBV03823 BBV03825 BBV03826 BBV03827 BBV03828 BBV03829 BBV03830 BBV03831 BBV03832 BBV03833 BBV03836 BBV03837 BBV03838 BBV03842 BBV03857 BCV01332 BCV01339 BCV01342 BCV01343 BCV01345 BDV01068 BDV01073 AJV01146 AKV01123 AKV01352 AKV01359 AKV01360 BBV01644 BBV01645 BBV03618 BBV03619 BBV03620 BBV03621 BBV03622 BBV03623 BCV01816 BCV01817 BCV01818 BCV01819 BCV01820 BCV01823 BCV01824 BCV01825 BCV01826 BCV01827 BEV01758 BEV02020 BEV02022 BDY01704 BDY01722 BEY01325 BEY01335 BEY01337 BEY01340 BEY01342 BEY01343 BEY01345 BEY01346 BEY01347 BEY01348 BEY01352 BEY01353 BEY01354 BEY01355 BEY01359 BEY01390 BEY01402 BFY01062 BFY01063 AKV02478
AKV04083 AKV04089 AKV04392 AKV04393 AKV04394 AKV04395 AKV04396 AKV04397 AKV04398 AKV04399 AKV04400 BCV01294 BCV01295 BCV01296 BCV01297 BCV01635 BCV01636 AKV01384 AKV01404 AKV01405 AKV01412 AKV01424 AKV01431 AKV02480 AKV02481 AKV02482 AKV02483 AKV04391 AKV04401 AKV04402 AKV04403 AKV04404 AKV04405 AKV04406 AKV04407 AKV04408 AKV04409 AKV04410 BBV01510 BBV01511 BBV01514 BBV01515 BBV01516 BBV01517 BBV01518 BBV01519 BBV01520 BBV01521 BBV01522 BBV01523 BBV01524 BBV01525 BBV01527 BBV01528 BBV01529 BBV01530 BBV01531 BBV01534 BCV01578 BCV01584 BCV01585 BCV01587 BCV01590 BCV01593 BCV01596 BCV01601 BCV01606 BDV01160 BDV01163 BDV01176 BDV01187 BDV01188 BDV01199 BDV01305 BDV01380 BDV01490 BDV01501 BDV01519 BDV01701 BDV01702 BDV01704 BDV01707 BDV01709 BDV01711 BDV01712 BDV01714 BDV01715 BDV01716 BDV01718 BDV01719 BDV01720 BDV01721 BDV01723 BDV01724 BDV01725 BDV01726 BDV01727 BDV01728 BDV01729 BDV01730 BDV01731 BDV01732 BDV01733 BDV01734 BDV01735 BDV01736 BDV01762 BDV01770 BDV01774 BDY01625 BDY01633 BDY01634 BDY01672 BDY01681 AHV01079 AKV01486 AKV01491 AKV01494 AKV01497 AKV01499 AKV01377 AKV01748 AKV02599 AKV02602 BAV04000 BAV04819 BAV04820 BAV04821 BAV04822 BAV04823 BBV03117 BBV03131 BBV03559 BBV03714 BBV03719 BBV03721 BCV02200 BDV01152 BDV01168 BDV01105 BDV01311 BDV01314 BDV01315 BDV01324 BDV01326 BDV01327 BDV01328 BDV01329 BDV01330 BDV01335 BDV01337 BEV02012 BEY01008 BBV03105 BCY01439 BCY01444 BCY01445 BCY01446 BCY01450 AKV01739 AKV01741 AKV01781 AKV01849 BAV01032 BAV01165 AKV02578 BCV01016 BCV01024 BCV01028 BCV01038 BCV01044 BCV01048 BAV05221 BAV06005 BAV06020 BBV01014 BBV01015 BDY01000 BDY01001 BBV03086 BBV03088 BBV03089 BBV03090 BBV03091 BBV03092 BBV03093 BBV03094 BBV03095 BCY01050 BCY01053 BCY01054 BCY01055 BCY01056 BCY01344 BCY01346 BCY01348 BCY01349 BCY01500 BCY01502 BCY01508 BCY01582 BCY01583 BCY01593 BCY01595 BCY01597 BCY01599 BCY01600 BCY01603 BCY01604 BCY01605 BCY01606 BCY01607 BCY01608 BCY01609 BCY01610 BCY01611 BCY01612 BCY01613 BCY01614 BCY01615 BCY01616 BCY01617 BCY01627 BCY01632 BCY01635 BCY01637 BCY01648 BDY01118 BDY01140 BDY01141 BAV01055 BAV01056 BAV01062 BAV01063 BAV01087 BAV01113 BAV01114 BAV01115 AKV01793 BCV01789 AKV01897 BAV05161 BAV06460 BAV06744 BAV06886 BAV06860 BBV01228 BBV01229 BCV01196 BCV01202 BCV01203 BAV01097 BBV06011 BCY01005 BCY01394 AEV01680 AHV01499 AHV01514 AHV01519 AHV01528 AHV01547 BAV01200 BAV01201 BAV06432 BCV01299 BCV02003 BCV02651 BCV02698 BDV01619 BEV01832 BEV01846 AKV02560 AKV02561 AKV02562 AKV02635 AKV04365 AKV04366 AKV04367 AKV04368 AKV04369 AKV04370 BAV06515 BAV06516 BAV06517 BAV06519 BAV06520 BAV06532 BBV03091 AKV01961 AKV01995 AKV02008 AKV02016 AKV02271 AKV02280 AKV02563 AKV02564 AKV02565 AKV02566 AKV02567 AKV02568 AKV02570 AKV02571 AKV02572 AKV02573 AKV02574 AKV02575 AKV02576 AKV02577 AKV02579 AKV02583 AKV02585 AKV02617 AKV02618 AKV02619 AKV02620 AKV02621 AKV02622 AKV02623 AKV02624 AKV02625 AKV02626 AKV02627 AKV02628 AKV02629 AKV02630 AKV02631 AKV02632 AKV02633 AKV02634 AKV04096 AKV04102 AKV04103 AKV04371 AKV04372 AKV04373 AKV04375 AKV04376 AKV04377 AKV04378 AKV04379 AKV04380 AKV04381 AKV04382 AKV04383 AKV04384 AKV04385 AKV04386 AKV04387 AKV04388 AKV04389 AKV04390 BAV01092 BAV01093 BAV01094 BAV01095 BAV01096 BAV01097 BAV01133 BAV01159 BAV01160 BAV04276 BAV04292 BAV04807 BAV05217 BAV05219 BAV05220 BAV05222 BAV05223 BAV05224 BAV05225 BAV05226 BAV05227 BAV05238 BAV06016 BAV06107 BAV06111 BAV06113 BAV06114 BAV06115 BAV06116 BAV06118 BAV06119 BAV06121 BAV06122 BAV06123 BAV06242 BAV06339 BAV06342 BAV06345 BAV06346 BAV06348 BAV06349 BAV06350 BAV06351 BAV06352 BAV06353 BAV06354 BAV06355 BAV06356 BAV06358 BAV06365 BAV06477 BAV06478 BAV06479 BAV06480 BAV06481 BAV06482 BAV06483 BAV06484 BAV06485 BAV06487 BAV06488 BAV06489 BAV06490 BAV06491 BAV06492 BAV06493 BAV06494 BAV06495 BAV06496 BAV06497 BAV06498 BAV06499 BAV06500 BAV06501 BAV06502 BAV06503 BAV06504 BAV06505 BAV06536 BAV06537 BAV06538 BAV06539 BAV06540 BAV06541 BAV06542 BAV06543 BAV06544 BAV06545 BAV06546 BAV06547 BAV06548 BAV06549 BAV06550 BAV06551 BAV06552 BAV06553 BAV06554 BAV06555 BAV06556 BAV06557 BAV06558 BAV06559 BAV06560 BAV06561 BAV06562 BAV06563 BAV06564 BAV06565 BAV06566 BAV06567 BAV06568 BAV06569 BAV06570 BAV06571 BAV06572 BAV06573 BAV06574 BAV06575 BAV06576 BAV06577 BAV06578 BAV06580 BAV06584 BAV06585 BBV01486 BBV03731 BBV03739 BBV03740 BBV03744 BBV03745 BBV03748 BBV03749 BBV03752 BBV03753 BBV03754 BBV03755 BBV03756 BBV03757 BBV03758 BBV03759 BBV03760 BBV03761 BBV03762 BBV03763 BBV03764 BBV03765 BBV03769 BBV03771 BBV03772 BCV01050 BCV01064 BCV01069 BCV01248 BCV01249 BCV01350 BCV01351 BCV01352 BCV01353 BCV01354 BCV01358 BCV02250 BCV02361 BCV02392 BCV02408 BCV02424 BCV02428 BCV02429 BCV02434 BCV02436 BCV02439 BCV02588 BCV02589 BEV01641 BEV01743 BEV01745 BEV01746 AKV01011 BBV06012 BBV06013 BBV06014 BBV06015 BBV06016 BBV06017 BBV06018 BBV06019 BBV06020 BBV06021 BBV06022 BBV06023 BBV06024 BBV06026 BBV06027 BBV06028 BBV06029 BBV06030 BBV06031 BBV06032 BBV06033 BBV06034 BBV06035 BBV06037 BBV06038 BBV06039 BCY01035 BDY01002 BDY01010 BDY01014 BDY01015 BDY01016 BDY01017 BDY01018 BDY01020 BDY01021 BDY01024 BDY01025 BDY01027 BDY01029 BBV03061 BBV03062 BBV03064 BBV03065 BBV03067 BBV03068 BBV03070 BBV03071 BBV03072 BBV03073 BBV03074 BBV03075 BBV03076 BBV03077 BBV03078 BBV03079 BBV03080 BBV03081 BBV03082 BBV03083 BBV03084 BBV03085 BBV03087 BBV03157 BBV03158 BBV03159 BBV03160 BBV03176 BCY01503 BCY01504 BCY01505 BCY01507 BCY01511 BEY01356 BEY01357 BEY01358 BEY01360 BEY01361 BEY01362 BEY01363 BEY01364 BEY01365 BEY01366 BEY01367 BEY01368 BEY01369 BEY01370 BEY01371 BEY01372 BEY01373 BEY01374 BEY01376 BEY01377 BEY01378 BEY01379 BEY01381 BEY01384 BB003141 BBV01204 BBV01205 BBV03101 BBV03104 BBV03106 BBV03107 BBV03108 BBV03109 BBV03110 BBV03111 BBV03112 BBV03114 BBV03115 BBV03116 BBV03117 BBV03118 BBV03119 BBV03121 BBV03122 BBV03123 BBV03124 BBV03125 BBV03126 BBV03127 BBV03128 BBV03129 BBV03130 BBV03131 BBV03132 BBV03134 BBV03135 BBV03136 BBV03137 BBV03138 BBV03139 BBV03140 BBV03142 BBV03144 BBV03145 BBV03146 BBV03147 BBV03148 BBV03149 BBV03220 BBV03221 BBV03222 BBV03224 BBV03225 BBV03226 BBV03227 BBV03228 BBV03229 BBV03230 BBV03232 BBV03233 BBV03235 BBV03236 BBV03237 BBV03238 BBV03239 BBV03240 BBV03241 BBV03242 BBV03243 BBV03245 BBV03246 BBV03249 BBV03250 BBV03251 BBV03252 BBV03253 BBV03254 BBV03255 BBV03256 BBV03257 BBV03258 BBV03259 BBV03260 BBV03261 BBV03262 BBV03263 BBV03264 BBV03265 BBV03266 BBV03267 BBV03268 BBV03270 BBV03271 BBV03272 BBV03273 BBV03274 BBV03275 BBV03276 BBV03277 BBV03278 BBV03279 BBV03280 BBV03281 BBV03282 BBV03283 BBV03284 BBV03285 BBV03286 BBV03287 BBV03288 BBV03289 BBV03290 BBV03291 BBV03292 BBV03293 BBV03294 BBV03295 BBV03296 BBV03298 BBV03299 BBV03301 BBV03302 BBV03303 BBV03304 BBV03306 BBV03307 BBV03308 BBV03557 BBY01200 BBY01201 BBY01203 BBY01204 BBY01205 BBY01206 BBY01207 BBY01212 BBY01215 BCY01111 BCY01113 BCY01125 BCY01126 BCY01130 BCY01138 BCY01148 BCY01152 BCY01196 BCY01197 BCY01198 BCY01199 BCY01200 BCY01201 BCY01202 BCY01204 BCY01206 BCY01208 BCY01209 BCY01210 BCY01211 BCY01212 BCY01213 BCY01214 BCY01215 BCY01216 BCY01217 BCY01218 BCY01219 BCY01220 BCY01221 BCY01222 BCY01223 BCY01224 BCY01225 BCY01226 BCY01227 BCY01229 BCY01231 BCY01232 BCY01233 BCY01234 BCY01235 BCY01243 BCY01244 BCY01245 BCY01246 BCY01247 BCY01248 BCY01249 BCY01250 BCY01251 BCY01252 BCY01253 BCY01254 BCY01255 BCY01256 BCY01257 BCY01258 BCY01261 BCY01262 BCY01263 BCY01264 BCY01265 BCY01266 BCY01267 BCY01269 BCY01270 BCY01271 BCY01272 BCY01273 BCY01275 BCY01276 BCY01277 BCY01278 BCY01279 BCY01280 BCY01281 BCY01282 BCY01284 BCY01285 BCY01286 BCY01287 BCY01288 BCY01289 BCY01290 BCY01291 BCY01292 BCY01293 BCY01294 BCY01296 BCY01297 BCY01298 BCY01350 BCY01354 BCY01355 BCY01357 BCY01359 BCY01363 BCY01364 BCY01365 BCY01366 BCY01367 BCY01368 BCY01369 BCY01371 BCY01372 BCY01373 BCY01374 BCY01375 BCY01376 BCY01377 BCY01378 BCY01379 BCY01380 BCY01381 BCY01382 BCY01383 BCY01384 BCY01386 BCY01387 BCY01388 BCY01389 BCY01390 BCY01391 BCY01392 BCY01393 BCY01395 BCY01396 BCY01397 BCY01399 BCY01401 BCY01402 BCY01403 BCY01404 BCY01405 BCY01406 BCY01407 BCY01408 BCY01409 BCY01410 BCY01411 BCY01412 BCY01413 BCY01414 BCY01415 BCY01416 BCY01417 BCY01418 BCY01419 BCY01420 BCY01421 BCY01422 BCY01423 BCY01424 BCY01425 BCY01426 BCY01427 BCY01428 BCY01429 BCY01430 BCY01432 BCY01433 BCY01434 BCY01435 BCY01436 BCY01437 BCY01438 BCY01440 BCY01442 BCY01443 BCY01447 BDY01214 BDY01231 BDY01282 BDY01285 BDY01506 BDY01557 BDY01558 BDY01559 BDY01560 BDY01561 BDY01562 BDY01563 BDY01564 BDY01565 BDY01566 BDY01567 BDY01568 BDY01569 BDY01571 BDY01572 BDY01573 BDY01575 BDY01576 BDY01580 BDY01582 BDY01583 BDY01584 BDY01585 BDY01586 BDY01587 BDY01588 BDY01590 BDY01591 BDY01592 BDY01593 BDY01594 BDY01595 BDY01596 BDY01599 BDY01601 BDY01602 BDY01603 BDY01604 BDY01605 BDY01607 BDY01608 BDY01609 BDY01611 BDY01612 BDY01613 BDY01614 BDY01615 BDY01616 BDY01617 BDY01619 BDY01620 BDY01621 BDY01622 BDY01623 BDY01624 BDY01626 BDY01627 BDY01628 BDY01630 BDY01632 BDY01635 BDY01638 BDY01639 BDY01641 BDY01647 BDY01648 BDY01650 BDY01655 BDY01658 BDY01662 BDY01664 BDY01666 BDY01667 BDY01668 BDY01670 BDY01673 BDY01676 BDY01678 BDY01680 BDY01685 BDY01686 BDY01696 BEY01093 BEY01094 BEY01180 BEY01188 BEY01191 BEY01195 BEY01200 BEY01201 BEY01202 BEY01203 BEY01204 BEY01209 BEY01216 BEY01217 BEY01219 BEY01220 BEY01221 BEY01222 BEY01223 BEY01224 BEY01225 BEY01226 BEY01227 BEY01228 BEY01229 BEY01230 BEY01233 BEY01234 AHV02135 AJV02887 BAV04260 BDV01231 BEV01283 BEV01323 AKV01815 AKV01817 AKV01826 BAV04785 BBV03600 BBV03601 BBV03603 BBV03609 BBV03670 BCV01036 BCV01301 BCV01516 BCV01538 BCV01539 BCV01541 BCV01543 BCV01545 BCV01546 BCV01547 BCV01548 BCV01549 BCV01550 BCV01551 BCV01552 BCV01553 BCV01554 BCV01555 BCV01556 BCV01557 BCV01559 BCV01835 BCV01836 BCV01837 BCV01838 BCV01839 BCV01840 BCV01841 BCV01842 BCV01843 BCV01844 BCV01845 BCV01846 BCV01847 BCV01848 BCV01850 BCV01851 BCV01868 BCV01950 BCV01954 BCV01957 BCV01958 BCV01959 BCV01961 BCV01980 BCV02175 BCV02177 BCV02531 BCV02770 BDV01023 BDV01035 BDV01046 BDV01047 BDV01054 BDV01064 BDV01065 BDV01069 BDV01412 BDV01427 BEV01380 BEV01387 BEV01392 BEV01402 BEV01406 BEV01407 BEV01434 BEV01437 BEV01440 BEV01441 BEV01449 BEV01450 BEV01451 BEV01452 BEV01453 BEV01454 BEV01455 BEV01456 BEV01457 BEV01564 BEV01566 BEV01567 BEV21380 AJV01179 AKV01226 AKV01228 AKV01231 AKV01265 AKV01335 AKV01340 AKV01341 AKV01342 AKV01349 AKV01353 AKV01354 AKV01370 AKV01375 AKV01378 AKV01380 AKV01381 AKV01387 AKV01388 AKV01389 AKV01390 AKV01392 AKV01394 AKV01395 AKV01396 AKV01406 AKV01416 AKV01422 AKV01426 AKV01437 AKV01448 AKV01664 AKV01687 BAV06117 BAV06178 BAV06180 BAV06182 BAV06247 BAV06248 BAV06255 BBV01202 BBV01208 BBV01209 BBV01210 BBV01211 BBV01212 BBV01214 BBV01215 BBV01217 BBV01218 BBV01219 BBV01220 BBV01221 BBV01222 BBV01225 BBV03624 BBV03794 BBV03795 BBV03799 BBV03800 BBV03801 BBV03802 BCV02188 BCV02190 BCV02191 BCV02192 BCV02194 BCV02211 BCV02219 BCV02230 BDV01479 BDV01481 BDV01491 BDV01504 BDV01512 BDV01514 BDV01518 BDV01521 BDV01525 BDV01526 BDV01528 BDV01747 BDV01827 BEV01120 BEV01122 BEV01128 BEV01133 BEV01134 BEV01153 BEV01157 BEV01158 BEV01678 BEV02003 BEV02014 BEV02018 BEV02019 BAV01102 BAV01103 BCY01305 BCY01307 AKV02193 AKV02198 AKV02206 AKV01450 BAV01042 BDY01007 BBV03139 BBV03217 BAV01060 BBV03102 BBV03103 AEV02749 AEV02758 BAV01100 BCY01020 ADV02048 AFV01470 AGV01485 AFV01276 AFV01771 AGV01062 AGV02639 AHV01278 AHV01883 AEV01901 BBV01011 AKV02738 AKV04485 BAV01118 BAV01123 BAV01124 BAV01125 BAV04008 BBV06090 BBV06091 BBV06092 BBV06093 BBV06094 BCV01298 BCV01642 BCV01665 BCV01747 BCV01752 BCV01759 BCV01761 BCV01765 BCV01768 BCV01769 BCV01772 BCV01773 BCV01774 BCV01776 BCV01777 BCV01889 BCV01896 BCV01900 BCV01901 BCV02057 BCV02060 BCV02690 BCV02692 BCV02715 BDV01200 BDV01208 BDV01209 BDV01214 BDV01215 BEV01317 BEV01462 BEV02120 ajv02050 AKV01864 AKV01867 AKV01872 AKV01875 AKV01877 AKV01878 AKV01879 AKV01880 AKV01881 AKV01882 AKV01883 AKV01884 AKV01885 AKV01886 AKV01887 AKV01888 AKV01889 AKV01890 AKV01891 AKV01894 AKV01936 AKV02239 AKV02247 AKV02249 AKV02250 AKV02252 AKV02256 AKV02258 AKV02527 AKV02528 AKV02530 AKV02594 AKV02707 AKV02708 AKV02710 AKV02711 AKV02712 AKV02713 AKV02714 AKV04056 AKV04273 AKV04274 AKV04463 AKV04464 AKV04465 AKV04466 AKV04467 AKV04468 AKV04469 AKV04470 AKV04471 AKV04473 AKV04474 AKV04475 AKV04476 AKV04477 AKV04478 BAV01219 BAV01220 BAV01221 BAV01222 BAV01223 BAV01224 BAV01225 BAV01226 BAV01227 BAV04012 BAV04022 BAV04023 BAV04024 BAV04025 BAV04627 BAV04628 BAV04640 BAV04641 BAV04642 BAV04643 BAV04644 BAV04645 BAV04646 BAV04647 BAV06225 BAV06233 BAV06260 BAV06261 BAV06264 BAV06265 BAV06266 BAV06270 BAV06271 BAV06418 BAV06419 BAV06420 BAV06421 BAV06422 BAV06423 BAV06425 BAV06506 BAV06507 BAV06654 BAV06861 BAV06862 BAV06863 BAV06864 BAV06865 BAV06868 BAV06869 BAV06870 BAV06879 BBV01171 BBV01178 BBV01180 BBV01187 BBV01191 BBV01193 BBV01194 BBV01196 BBV01361 BBV01364 BBV01367 BBV01371 BBV01373 BBV01388 BBV01391 BBV01392 BBV01394 BBV01398 BBV01399 BBV01403 BBV01404 BBV01410 BBV01411 BBV01412 BBV01413 BBV01414 BBV01415 BBV01416 BBV01418 BBV01419 BBV01420 BBV01422 BBV01424 BBV01425 BBV01428 BBV01430 BBV01431 BBV01434 BBV01436 BBV01439 BBV01440 BBV01441 BBV01442 BBV01443 BBV01445 BBV01446 BBV01448 BBV01449 BBV01562 BBV01563 BBV01564 BBV01565 BBV01566 BBV01567 BBV01568 BBV01569 BBV01570 BBV01571 BBV01573 BBV01574 BBV01576 BBV01577 BBV01578 BBV01579 BBV01581 BBV01582 BBV01583 BBV03054 BBV03059 BBV03368 BBV03372 BBV03374 BBV03375 BBV03411 BBV03413 BBV03414 BBV03415 BBV03418 BBV03419 BBV03420 BBV03421 BBV03423 BBV03424 BBV03425 BBV03426 BBV03427 BBV03429 BCV01521 BCV01526 BCV01528 BCV01533 BCV01534 BCV01544 BCV01874 BCV01935 BCV01951 BCV01952 BCV01953 BCV01960 BCV01966 BCV01970 BCV01971 BCV02173 BCV02184 BCV02285 BCV02289 BCV02291 BCV02293 BCV02294 BCV02296 BCV02299 BCV02301 BCV02310 BCV02311 BCV02313 BCV02316 BCV02322 BCV02326 BCV02748 BCV02774 BCV02788 BCV02791 BCV02802 BCV02803 BCV02804 BCV02806 BDV01004 BDV01007 BDV01008 BDV01009 BDV01012 BDV01013 BDV01014 BDV01015 BDV01018 BDV01019 BDV01057 BDV01085 BDV01410 BDV01411 BDV01415 BDV01418 BDV01431 BDV01436 BDV02039 BEV01003 BEV01005 BEV01006 BEV01015 BEV01018 BEV01020 BEV01029 BEV01036 BEV01052 BEV01054 BEV01055 BEV01062 BEV01065 BEV01075 BEV01081 BEV01086 BEV01090 BEV01092 BEV01093 BEV01098 BEV01360 BEV01363 BEV01369 BEV01372 BEV01373 BEV01377 BEV01379 BEV01384 BEV01390 BEV01391 BEV01394 BEV01397 BEV01398 BEV01399 BEV01400 BEV01403 BEV01404 BEV01405 BEV01408 BEV01411 BEV01412 BEV01415 BEV01420 BEV01422 BEV01424 BEV01425 BEV01428 BEV01438 BEV01683 BEV01687 BEV01688 BEV01690 BEV01694 BEV01695 BEV01697 BEV01698 BEV01701 BEV01702 BEV01703 BEV01705 BEV01707 BEV01708 BEV01715 AKV01004 AJV01247 AKV01003 AKV01013 AKV01022 AKV01023 AKV01024 BCY01023 BCY01024 BCY01031 BCY01033 BCY01034 BAV01051 BAV01065 BAV01066 BAV01067 BAV01068 BAV01069 BAV01070 BAV01108 BAV01109 BAV01110 BAV01111 BAV01112 BBV01104 BBV01105 BBV01106 BBV01107 BBV01108 BBV01109 BBV01110 BBV01111 BBV01112 BBV01113 BBV01114 BBV01115 BBV03059 BBV03060 BBV03063 BBV03066 BBV03069 BBV03096 BBV03097 BBV03098 BBV03099 BBV03151 BBV03152 BBV03153 BBV03154 BBV03155 BBV03161 BBV03162 BBV03163 BBV03164 BBV03165 BBV03166 BBV03167 BBV03168 BBV03169 BBV03171 BBV03172 BBV03174 BBV03175 BCY01062 BCY01063 BCY01064 BCY01065 BCY01066 BCY01070 BCY01071 BCY01072 BCY01073 BCY01076 BCY01079 BCY01080 BCY01081 BCY01083 BCY01085 BCY01093 BCY01301 BCY01302 BCY01303 BCY01304 BCY01306 BCY01308 BCY01309 BCY01310 BCY01311 BCY01312 BCY01313 BCY01314 BCY01315 BCY01316 BCY01317 BCY01318 BCY01319 BCY01320 BCY01321 BCY01322 BCY01323 BCY01325 BCY01326 BCY01327 BCY01328 BCY01332 BCY01339 BCY01565 BCY01566 BCY01567 BCY01568 BCY01569 BCY01570 BCY01571 BCY01572 BCY01573 BCY01574 BCY01601 BCY01619 BCY01620 BCY01621 BCY01622 BCY01623 BCY01624 BCY01625 BCY01628 BCY01630 BCY01631 BCY01633 BCY01634 BCY01636 BCY01638 BCY01639 BCY01640 BCY01641 BCY01642 BCY01643 BCY01644 BCY01645 BCY01646 BCY01647 BCY01649 BCY01650 BCY01651 BCY01652 BCY01653 BCY01654 BDY01167 BDY01186 BDY01193 BDY01195 BDY01196 BDY01198 BDY01199 BDY01700 BDY01703 BDY01705 BDY01723 BDY01725 BDY01808 BEY01025 BEY01027 BEY01032 BEY01040 BEY01046 BEY01050 BEY01051 BEY01052 BEY01054 BEY01056 BEY01057 BEY01058 BEY01059 BEY01060 BEY01061 BEY01062 BEY01063 BEY01064 BEY01065 BEY01066 BEY01067 BEY01068 BEY01270 BEY01271 BEY01273 BEY01274 BEY01275 BEY01276 BEY01277 BEY01278 BEY01279 BEY01280 BEY01281 BEY01282 BEY01283 BEY01284 BEY01285 BEY01286 BEY01287 BEY01288 BEY01289 BEY01290 BEY01293 BEY01295 BEY01324 BEY01326 BEY01328 BEY01331 BEY01332 BEY01333 BEY01334 BEY01336 BEY01338 BEY01339 BEY01341 AFV01155 AFV02163 AHV01103 BAV06462 BBV03709 BAV01169 AKV04302 AKV04298 AKV04299 AKV04300 AKV04301 AKV04336 AKV04337 BAV01041 BAV01045 BAV01046 BAV01047 BAV04233 BAV04235 AKV01237 AKV02592 AKV02596 AKV02662 BAV04246 BBV01589 BBV03050 BBV03051 BBV03052 BBV03053 BBV03064 BBV03686 BBV03694 BCV01025 BCV01029 BCV01524 BCV01527 BCV01858 BCV01875 BCV02767 BCV02768 BCV02785 BCV02794 BCV02801 BDV01041 BDV01060 BDV01091 BDV01099 BDV01402 BEV01024 BEV01033 BEV01429 BEV01431 BEV01433 BEV01436 BEV01459 BEV01793 BEV01805 BDY01115 BDY01122 AGV02397 AKV02498 AKV02499 AKV02501 AKV02502 AKV04162 BAV06466 BAV06471 BAV06472 BCV01289 BEV01467 BEV02182 AKV02491 AKV02492 AKV02493 AKV02494 AKV02495 AKV02496 AKV02532 AKV02533 AKV02534 AKV02535 BAV04217 BAV04265 BAV04267 BAV04269 BAV04272 BAV04287 BAV04291 BAV06072 BAV06073 BAV06075 BAV06141 BAV06145 BAV06148 BAV06149 BAV06151 BAV06152 BAV06196 BAV06207 BAV06228 BAV06234 BAV06256 BAV06258 BAV06259 BBV01188 BBV01358 BBV01390 BBV01427 BBV01435 BCV01333 BCV01334 BCV01337 BCV01340 BCV01341 BCV01344 BCV02288 BCV02290 BCV02297 BCV02317 BCV02323 BCV02330 BCV02332 BCV02333
BCV02490 BCV02492 BCV02499 BCV02501 BCV02503 BCV02751 BCV02755 BCV02761 BCV02779 BCV02782 BCV02793 BCV02795 BCV01511 BCV01976 BCV01981 BCV02524 BCV02532 BCV02750 BCV02752 BCV02759 BCV02762 BCV02769 BCV02772 BCV02784 BCV02787 BCV02789 BCV02797 BCV02798 BDV01003 BDV01005 BDV01006 BDV01010 BDV01016 BDV01025 BDV01026 BDV01027 BDV01028 BDV01031 BDV01033 BDV01042 BDV01049 BDV01050 BDV01053 BDV01071 BDV01072 BDV01076 BDV01082 BDV01083 BDV01088 BDV01093 BDV01098 BDV01406 BDV01409 BDV01413 BDV01419 BDV01420 BDV01423 BDV01653 BDV01657 BDV01665 BDV01666 BDV01667 BDV02031 BDV02044 BDV02045 BDV02090 BDV02097 BEV01023 BEV01032 BEV01040 BEV01056 BEV01059 BEV01076 BEV01077 BEV01458 BEV01576 BEV01599 BEV01603 BEV01606 BEV01725 BEV01780 BEV01820 BEV01822 BEV01823 BEV01825 BEV01826 BEV01827 BEV01828 BEV02208 BEV02209 BEV02211 AKV04303 AKV04304 BAV05104 BAV05105 BAV05113 BAV06589 BAV06623 BAV06624 BAV06625 BAV06666 BAV06667 BAV06668 BAV06669 BAV06671 BBV03750 BBV03751 BBV03766 BBV03767 BBV03768 BCV01215 BCV01219 BCV01222 BCV01226 BCV01227 BCV01229 BCV01230 BCV01231 BCV01233 BCV01235 BCV02593 BCV02612 BCV02615 BCV02847 BCV02848 BCV02850 BCV02852 BCV02853 BCV02872 BCV02873 BCV02878 BDV01509 BDV01705 BDV01708 BDV01737 BDV01754 BDV01766 BEV01612 BEV01628 BEV01631 BEV01638 BEV01639 BEV01642 BEV01644 BEV01963 BEV01966 BEV01970 BEV02008 BEV02013 BEV02015 BEV02016 BEV02017 BEV02021 BAV01043 BCY01006 BCY01584 BCY01585 BCY01587 BCY01588 BCY01590 BDY01721 BDY01724 BDY01735 BDY01803 BDY01804 BBV01201 BBV03223 BBV03231 BBV03234 BBV03247 BBV03248 BCY01194 BCY01228 BCY01230 BDY01646 BDY01651 BDY01656 BDY01657 BDY01661 AEV02536 AGV02299 AGV02323 AFV01820 AJV01604 BCV01993 BCV02026 BCV02683 BCV02689 BCV02693 BCV02697 BCV02699 BCV01005 BCV01964 BCV02510 BCV02513 BCV02514 BCV02518 BCV02766 BCV02777 BCV02778 BCV02800 BDV01036 BDV01089 BDV01096 BDV02030 BDV02032 BDV02034 BDV02035 BDV02036 BDV02037 BDV02038 BDV02040 BDV02043 BDV02091 BDV02098 BDV02100 BEV01000 BEV01001 BEV01002 BEV01004 BEV01007 BEV01009 BEV01010 BEV01011 BEV01014 BEV01019 BEV01022 BEV01027 BEV01604 BEV01605 BEV01609 BEV01681 BEV01682 BEV01684 BEV01685 BEV01686 BEV01689 BEV01696 BEV01699 BEV01700 BEV01704 BEV01706 BEV01710 BEV01711 BEV01712 BEV01721 BEV01722 BEV01723 BEV01724 BEV01726 BEV01727 BEV01729 BEV01782 BEV01783 BEV01784 BEV01787 BEV01789 BEV01790 BEV01791 BEV01792 BEV01794 BEV01795 BEV01796 BEV01797 BEV01798 BEV01800 BEV01801 BEV01803 BEV01804 BEV01806 BEV02030 BEV02032 BEV02034 BEV02035 BEV02036 BEV02037 BEV02038 BEV02040 BEV02043 AKV01657 AKV01658 AKV01659 AKV01663 AKV01665 AKV01903 AKV01913 AKV01914 AKV01919 AKV01924 AKV01937 AKV01938 AKV01939 AKV01940 AKV01941 AKV01942 AKV01943 AKV01945 AKV01946 AKV01947 AKV01948 AKV01949 AKV01950 AKV01951 AKV01952 AKV01953 AKV01954 AKV01955 AKV01956 AKV01963 AKV02212 AKV02213 AKV02218 AKV02220 AKV02221 AKV02223 AKV02228 AKV02231 AKV02233 AKV02236 AKV02267 AKV02268 AKV02269 AKV02270 AKV02272 AKV02277 AKV02282 AKV02283 AKV02284 AKV02285 AKV02286 AKV02290 AKV02650 AKV02680 AKV02681 AKV02682 AKV02683 AKV02684 AKV02685 AKV02686 AKV02688 AKV02689 AKV02691 AKV02692 AKV02693 AKV02694 AKV02695 AKV02696 AKV02697 AKV02698 AKV02699 BBV03610 BBV03611 BBV03612 BBV03613 BBV03614 BBV03615 BBV03616 BBV03617 BBV03746 BBV03770 BBV03774 BBV03775 BBV03777 BBV03778 BBV03780 BBV03782 BBV03785 BCV01362 BCV01366 BCV01369 BCV01372 BCV01376 BCV01378 BCV01379 BCV01380 BCV01381 BCV01382 BCV01808 BCV01811 BCV01813 BCV02564 BCV02568 BCV02572 BCV02579 BCV02587 BCV02591 BCV02594 BCV02595 BCV02597 BCV02605 BDV01365 BDV01390 BDV01459 BDV01462 BDV01469 BDV01488 BDV01495 BDV01503 BDV01508 BDV01523 BDV01524 BDV01529 BDV01531 BDV01535 BDV01538 BDV01540 BDV01541 BDV01543 BDV01544 BDV01547 BDV01750 BDV01753 BDV01758 BDV01759 BDV01760 BEV01924 BEV01937 BEV01939 BAV01104 BDY01004 BDY01033 BEY01128 BEY01132 BCY01576 BCY01577 BCY01578 BCY01579 BCY01580 BDY01184 BDY01706 BDY01715 BDY01719 BEY01026 BEY01034 BEY01037 BEY01041 BEY01042 BEY01047 BEY01329 BEY01330 BBV03269 BBV03300 BBY01202 BBY01209 BBY01211 BBY01213 BCY01370 BCY01385 BCY01451 BCY01452 BCY01453 BCY01454 BCY01455 BCY01456 BCY01457 BCY01458 BCY01459 BCY01460 BCY01461 BCY01462 BCY01463 BCY01464 BCY01466 BCY01468 BCY01469 BCY01470 BCY01472 BCY01475 BAV01015 BEV01808 BEY01141 BEY01313 BEY01344 BEY01375 AKV01283 AKV01301 AKV01496 AKV01850 AKV01854 AKV01858 AKV01899 BAV01173 BAV04607 BAV06437 BAV06267 BAV06867 AKV01965 BAV01032 BBV03004 BBV03005 BDY01003 AEV01000 BCY01283 BCY01295 BCY01299 BCY01353 BCY01356 AKV04306 BCV01514 BAV04376 BAV04409 BAV04425 BAV04664 BAV04665 BAV04716 BAV04775 BAV04779 BAV06801 BAV06805 BAV04480 BCV02281 BCV02337 BCV02339 BCV02341 BCV02342 BCV02349 BEV01129 BEV01138 BEV01139 BEV01141 BEV01146 BEV01198 BEV01513 BEV01525 BEV01527 BEV01532 BEV01536 BEV01550 BCY01030 BDY01173 AFV01857 BEV01159 BAV04173 BAV04156 US Government Consignee's Serial No. BAV01314 BAV04114 BAV04028 BAV06364 BAV06362 BAV06361 BAV06359 BAV06340 BEV01144 BEV01147 BEV01149 BEV01151 BEV01156 BEV01170 BEV01174 BDV01278 BDV01281 BDV01283 BDV01293 BDV01552 BDV01554 BDV01556 ADV06006 AEV02743 BAV06360 BDV01291 |
| FEI Number |
3013421741
|
Recalling Firm/ Manufacturer |
Carefusion 211 Inc dba Carefusion 22745 Savi Ranch Pkwy Yorba Linda CA 92887-4668
|
| For Additional Information Contact | Troy Kirkpatrick 858-617-2361 |
Manufacturer Reason for Recall | A pressure transducer failure can develop, activating a false Extended High Ppeak or Circuit Occlusion Alarm. The Safety Valve will open to ambient air and the unit will stop ventilating, allowing spontaneous breathing patients to breathe. The malfunction may delay initiation or cease ventilation. Nonbreathing patients will need manual ventilation or to be connected to another ventilator. |
FDA Determined Cause 2 | Process control |
| Action | A customer letter dated 4/21/15 will be sent to all customers who received the AVEA Ventilator, all models, to inform them that CareFusion has identified a potential risk associated with AVEA Ventilator caused by a malfunction of a pressure transducer. The malfunctioning pressure transducer is detected and by design ceases ventilation and opens the safety valve to atmosphere allowing spontaneous breathing of patients capable of doing so. The non-breathing patient would need to be manually ventilated.
The letter informs the customers of the issues, the potential risk, and the actions to be taken. Customers are instructed to contact CareFusion Recall Support Center at (888) 562-6018, (858) 617-5300, or email supportcenter@carefusion.com for recall related questions.
Customers are instructed to contact CareFusion Technical Support/Customer Advocacy at (800) 231-2466, (714) 283-2228, or email support.vent.us@carefusion.com for Product Technical Support Adverse Event Reporting.
CareFusion posted a press release on their website on 5/27/15 and it was sent to PRNewswire. The press release provided an update on a global voluntary recall that was initiated on April 21, 2015 to address an issue with certain units of AVEA ventilators. The press release provide the problems and the actions to be taken. The press also list the countries that are affected by the recall. |
| Quantity in Commerce | 15,905 units total (5,670 units in US) |
| Distribution | Worldwide Distribution. US Nationwide including Guam and Puerto Rico, Canada, Albania, American Samoa, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belgium, Belize, Bermuda, Bolivia, Bosnia Herzegovina, Brazil, Bulgaria, Chile, China (Peoples Rep), Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvadore, Estonia, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea (Rep), Kuwait, Latvia, Lebanon, Libyan Jamahiriya, Lithuania, Malaysia, Mexico, Moldova, Morocco, Nepal, Netherlands, Nicaragua, Norway, Occupied Palestinian Territory, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, South Africa, South America, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CBK
|
|
|
|