  Class 2 Device Recall GE Healthcare, SIGNA 3.0T see related information
Date Initiated by Firm August 31, 2015
Date Posted November 10, 2015
Recall Status1 Terminated 3 on September 07, 2018
Recall Number Z-0255-2016
Recall Event ID 72301
510(K)Number K013636  
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
Product GE Healthcare, SIGNA 3.0T.

Product Usage: Magnetic resonance system for diagnostic imaging. The indications for use for the 3.0T Signa VH/i (Signa 3.0T MR System) Transmit/Receive Body Imaging Coil expands the imaging capability of the 3.0T Signa VH/i MR Imaging System. The Transmit/Receive Body Imaging Coil is intended for imaging of the Neck, Spine, Abdomen/Thorax and the extremities.
Code Information 00000262893MR1 520694MR3 00000GE0206027 082427040020 00000243366MR2 082427040040 00000256194MR2 409740MR01 00000261205MR9 083027828659711 00000280097MR7 IE1111MR01 00000189273MR6 A5155133 00000272076MR1 YM0222 00000267372MR1 YM0333 00000000004230 YM0451 00000271322MR0 0850270094 00000227450MR7 333MR3TXXX 00000271777MR5 5606MRS3T 00000273539MR7 570844HCM3T 00000227450MR7 333MR3T 000000HDMR5684 CEN03406 00000GE0209032 0853270052IB1 00000260651MR5 10491MRS01 
Recalling Firm/
Manufacturer
GE Healthcare
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact GE Healthcare Service
800-437-1171
Manufacturer Reason
for Recall
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
FDA Determined
Cause 2
Software design
Action GE Healthcare sent an Urgent Medical Device Correction letter dated August 31, 2015. The letter was addressed to Hospital Administrators/Risk Managers, Radiology Department Managers, & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce 19
Distribution Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = LNH and Original Applicant = GE MEDICAL SYSTEMS
-
-