Class 2 Device Recall Shuttle Select SlipCath

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Shuttle Select SlipCathsee related information
Date Initiated by FirmApril 15, 2016
Date PostedAugust 22, 2016
Recall Status1 Terminated 3 on August 04, 2017
Recall NumberZ-2613-2016
Recall Event ID 74010
510(K)NumberK122937 
Product Classification Catheter, intravascular, diagnostic - Product Code DQO
ProductShuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques
Code Information All Lots. Catalog prefixes/suffixes SCBR5.5-/-SHTL SCBR6.5-/-SHTL  Catalog/product number; Global/order number SCBR5.5-35-125-P-NS-H1-SHTL G31129 SCBR5.5-35-125-P-NS-JB1-SHTL G31130 SCBR5.5-35-125-P-NS-JB2-SHTL G36361 SCBR5.5-35-125-P-NS-MPA-SHTL G23697 SCBR5.5-35-125-P-NS-SIM1-SHTL G23699 SCBR5.5-35-125-P-NS-SIM2-SHTL G36360 SCBR5.5-35-125-P-NS-VERT-SHTL G23698 SCBR5.5-35-125-P-NS-VTK-SHTL G36359 SCBR5.5-35-135-P-NS-ANG-SHTL G48154 SCBR5.5-35-150-P-NS-0-SHTL G23705 SCBR5.5-35-150-P-NS-ANG-SHTL G23704 SCBR5.5-35-75-P-NS-ANG-SHTL G48152 SCBR6.5-35-125-P-NS-H1-SHTL G31134 SCBR6.5-35-125-P-NS-JB1-SHTL G31135 SCBR6.5-35-125-P-NS-JB2-SHTL G36364 SCBR6.5-35-125-P-NS-SIM1-SHTL G23701 SCBR6.5-35-125-P-NS-SIM2-SHTL G36363 SCBR6.5-35-125-P-NS-VERT-SHTL G23700 SCBR6.5-35-125-P-NS-VTK-SHTL G36362 
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Inc.
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Departme
800-457-4500
Manufacturer Reason
for Recall
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
FDA Determined
Cause 2
Unknown/Undetermined by firm
ActionCook Inc. sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter dated April 15, 2016, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with any medical questions or concerns, were instructed to contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. For information regarding the recall, please contact Stericycle Expert Solutions at 866.201.9067.
Quantity in Commerce60,302
DistributionWorldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DQO
-
-