• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall COULTER HmX Hematology Analyzer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall COULTER HmX Hematology Analyzersee related information
Date Initiated by FirmJune 26, 2017
Create DateSeptember 07, 2017
Recall Status1 Terminated 3 on September 14, 2023
Recall NumberZ-3114-2017
Recall Event ID 77830
510(K)NumberK010765 
Product Classification Counter, differential cell - Product Code GKZ
ProductCoulter HmX CP and Coulter HmX AL Analyzers The purpose of the HmX Hematology Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, biochemical investigation or any other test that helps diagnose the abnormality
Code Information Catalog Number(s): HMX CP - 178832, 178833, 178834, A91062, A90994; HmX AutoLoader - 6605522, 6605523, 6605524; All software versions.
Recalling Firm/
Manufacturer
Beckman Coulter Inc.
250 S Kraemer Blvd
Brea CA 92821-6232
For Additional Information ContactMarwan Fathallah
800-526-7694
Manufacturer Reason
for Recall
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
FDA Determined
Cause 2
Labeling Change Control
ActionBeckman Coulter sent an Urgent Medical Device Recall letter dated June 30, 2017, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff. If they forwarded any of the affected product to another laboratory, please provide them a copy of the letter. Customers with questions should contact the Customer Support Center at 800-528-7694.
Quantity in Commerce3326 units
DistributionWorldwide Distribution - US (nationwide) Internationally to Albania, Algeria, Andorra, Angola, Antigua and Barbuda, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Bulgaria, Burundi, Canada, China, Colombia, Congo, Croatia, Czech republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, india, Indonesia, Italy, Japan, Iraq, Jordan. Kenya, Kuwait, Libya, Lithuania, Macau, Lebanon, Macedonia, Malawi, Malaysia, Mexico, Mongolia, Morocco, Mozambique, Netherlands, New Zealand, Nigeria, Pakistan, Panama, Paraguay, Philippines, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, UAE, UK, Uruguay, Venezuela, Vietnam, Virgin Islands, Yemen, Zambia, Zimbabwe.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GKZ
-
-