• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall JS Series SterilContainer S2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall JS Series SterilContainer S2 see related information
Date Initiated by Firm September 08, 2017
Date Posted February 15, 2018
Recall Status1 Terminated 3 on April 04, 2019
Recall Number Z-0624-2018
Recall Event ID 79089
Product Classification Sterilization wrap containers, trays, cassettes & other accessories - Product Code KCT
Product JS Series SterilContainer S2 System
Code Information JS440 S2 1/1CONT.BOTTON PERFOR.H88MM H202 JS441 S2 1/1 CONT.BOTTOMPERFOR.H1118MM H202 JS442 S2 1/1-CONT.BOTTOM PERFOR.H133MM H202 JS444 S2 1/1 CONT.BOTTOM PEFOR.H185MM H202 JS489 S2 1/1-CONTAINER LID H202 JS740 S2 3/4 CONT.BOTTOM PERFOR.H88MMH202 JS741 S2 3/4 CONT.BOTTOM PERFOR.H118MM H202 JS742 S2 3/4 CONT.BOTTOM PERFOR.H133MM H202 JS789 S2 3/4 CONTAINER LID H202 JS340 S2 1/2 CONT.BOTTOM PERFOR.H88MM H202 JS341 S2 1/2 CONT.BOTTOM PERFOR.H118MM H202 JS342 S2 1/2 CONT.BOTTOM PERFOR.H133MM H202 JS389 S2 1/2 CONTAINER LID H202  
Recalling Firm/
Manufacturer
Aesculap Implant Systems LLC
3773 Corporate Pkwy
Center Valley PA 18034-8217
For Additional Information Contact Val Strawn
800-258-1946
Manufacturer Reason
for Recall
The JS Series SterilContainer S2 System was released to the market place prior to an authorized FDA 510K clearance. No other Aesculap SterilContainer System products are impacted by this recall.
FDA Determined
Cause 2
No Marketing Application
Action On August 24, 2017 a Medical Device Recall(Reference 2916714-08/24/2017-010R) was issued requesting that all customers take the following steps: 1.Review the Urgent Medical Device Recall Notification Letter in its entirety and ensure that all users in your organization and other concerned persons are informed about this recall and immediately discontinue use. 2. Determine your current inventory of affected containers within your facility. Immediately quarantine the affected products. A sales representative will come to your facility and remove your affected product and immediately return it to Aesculap Inc. They will assist you with ordering an appropriate replacement container. Do not destroy any affected products. 3. Please contact Customer Service and they will provide you a Returned Good number and a return label. Customer Service: Phone: 1-800-282-9000 aesculap_emailorders.bbmus_service@bbraun.com
Quantity in Commerce 292 units
Distribution Nationally
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-