• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Hydroline Trumpet Valve, ASU1220

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Hydroline Trumpet Valve, ASU1220see related information
Date Initiated by FirmJune 11, 2018
Create DateJuly 18, 2018
Recall Status1 Terminated 3 on August 17, 2020
Recall NumberZ-2494-2018
Recall Event ID 80288
Product Classification Catheter, irrigation - Product Code GBX
ProductHydroline Trumpet Valve, Trumpet Valve, No Probe, Single Spike, Ref ASU1220, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
Code Information Lot numbers 71725, 81743, 91722, and 101723.
FEI Number 3001236905
Recalling Firm/
Manufacturer
Cardinal Health 200, LLC
3651 Birchwood Dr
Waukegan IL 60085-8337
For Additional Information ContactCustomer Advocacy
800-292-9332
Manufacturer Reason
for Recall
The suction valve may not close properly which could cause continuous suction.
FDA Determined
Cause 2
Device Design
ActionThe recalling firm issued letters dated 6/11/2018 via FedEx Next Day Priority Mail requesting return of the product.
Quantity in CommerceA total of 63,020 valves were distributed. Specific product quantities were not provided.
DistributionDistribution was made to AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MT, NC, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, and WY. Government and military distribution was also made. Foreign distribution was made to Belgium, France, Germany, Israel, Italy, Netherlands, Singapore, and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-