• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall ISH Protease 3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall ISH Protease 3see related information
Date Initiated by FirmAugust 02, 2018
Date PostedSeptember 14, 2018
Recall Status1 Terminated 3 on November 22, 2021
Recall NumberZ-3005-2018
Recall Event ID 80807
Product Classification Trypsin - Product Code IBG
ProductISH Protease 3, Catalog Number 05273331001, model 780-4149 Immunohistochemistry (IHC) for in vitro diagnostic use.
Code Information Lot Numbers: Y13927, Y18872, Y22569, Y25883, UDI: 4015630971886
FEI Number 1000125183
Recalling Firm/
Manufacturer
Ventana Medical Systems Inc
1910 E Innovation Park Dr
Oro Valley AZ 85755-1962
For Additional Information Contact
800-227-2155
Manufacturer Reason
for Recall
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm disseminated their field safety notices by letter on 08/02/2018. The firm requested the destruction of the product.
Quantity in Commerce2231 units
Distributionworldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-