| Class 2 Device Recall Philips Ingenia Elition S, Ingenia Elition X | |
Date Initiated by Firm | December 03, 2018 |
Create Date | January 08, 2019 |
Recall Status1 |
Terminated 3 on May 12, 2020 |
Recall Number | Z-0694-2019 |
Recall Event ID |
81797 |
510(K)Number | K173356 |
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product | Ingenia Elition S, Ingenia Elition X |
Code Information |
Software version R5.5.0 Serial numbers: 45027 45011 45010 45024 45032 45026 45054 45035 45051 45052 45007 45043 45044 45031 45039 45025 46001 45033 45050 45034 45049 45037 45048 46004 45046 45008 45058 45028 45029 45030 45042 46005 46011 45040 45045 |
Recalling Firm/ Manufacturer |
Philips Medical Systems Nederlands Veenpluis 4-6 Best Netherlands
|
Manufacturer Reason for Recall | During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | On December 4, 2018, the firm notified customers of the recall via an Urgent Medical Device Correction letter. The letter informed customers of the product issue.
Customers were asked to do the following:
The user must follow the Instructions for Use to verify that no part of the patient's body, hair, clothing, cables or infusion lines can get trapped or injured by any part of the equipment. Additionally, the VitalScreen shows positioning guidance regarding arm supports and padding.
The user should avoid the use of the yellow Stop Table button on the VitalScreen.
The user must always use the tumble switch or physical Emergency Table Stop button for cases which require the table movement to stop.
Philips stated they would be rolling out FCO78100490 Service Pack 1 for R5.5.0 to affected systems. The Service Pack will remove the "Stop Table" software button on the VitalScreen. |
Quantity in Commerce | 35 |
Distribution | CA, MN, NJ, NY, OH, TN, TX. Foreign distribution to Canada, Belgium, France, Germany, INdia, Japna, Netherlands, South Africa, Sweden, Switzerland, UAE. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LNH
|
|
|
|