|
Class 2 Device Recall BRILLIANCE CT, PRIVATE PRACTICE CV CONFIGURATION, CT SCANNER |
 |
Date Initiated by Firm |
June 24, 2019 |
Date Posted |
August 06, 2019 |
Recall Status1 |
Terminated 3 on October 06, 2022 |
Recall Number |
Z-2191-2019 |
Recall Event ID |
83365 |
510(K)Number |
K042293
|
Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
Product |
Brilliance 64 Model # 728231, computed tomography x-ray system |
Code Information |
System Serial Number 335, 338, 1897, 4001, 4003, 4016, 4021, 9009, 9029-9030, 9033, 9035, 9038, 9040, 9045, 9047, 9048, 9052, 9054, 9056, 9058, 9062, 9064-9068, 9070-9077, 9080, 9086, 9087,9089, 9092, 9093, 9098, 9099, 9104- 9108, 9112, 9117 9120, 9121, 9127, 9134, 9138, 9140, 9148, 9159- 9160
9171, 9187, 9201, 9208, 9234, 9243, 9501, 9509 through 9608, 9610 through 10288, 10291 through 10327, 10330 through 10339, 10341 through 10808, 10811-10812, 29002, 29005, 29007, 29009, 29012, 29014, 29016-29018, 29020,29022, 29024, 29025-29027, 29032, 29035, 29037-29039, 29044-29045, 29047- 29049 29053- 29054, 29056- 29057, 29062, 29064-29066, 29068, 29081, 29086 29092, 29097, 29102, 29107, 29108, 29111, 29115, 29118, 29124, 29135 29138, 32065, 64045, 64051, 90029- 90031, 90041, 90046, 90047, 90054, 90056, 90058, 90062, 9006390066, 90069, 90072, 90078, 90079, 90081, 90082, 90085, 90086, 90087, 90089,90091, 90094, 90097, 90098, 90107, 90111, 90119, 90123-90124, 90128, 90131, 90135, 90136 90137, 90139, 90140, 90141, 90143, 90147, 90150, 90154, 90157, 90158, 90161,90163, 90167-90170, 90175,90177,90180, 90185, 90189,90199, 90205, 90213, 90218, 90223-90225, 90227, 90804, 91003, 95003, 95004, 95008, 95009, 95013- 95014
95017 through 95026, 95028, 95030 through 95061, 95063, 95069, 95071 through 95114, 95116-95118, 95130 through 95296, 95298 through 95476, 95478-95479, 95481 through 95529, 95531 through 95554, 95556 through 95566, 95568, 95570 through 95575, 95577 through 95584, 95586, 95588-95589, 95591through 95947, 122706, 320034, 760014, 6000861, 6001993, 6002294, 6002305, 6003273, 6003368, 6003844, 6003941, 200000618-1, 10003/11205, 10081, 10182EBW13108, 1155-16, 1157-09, 1157-10, 1158-05, 1190-02, 1339-09, 1343-02, 1347-04, 1348-01, 1349-01, 1350-02, 1OO82, 4101_45, 4272-08, 9208A, 95040, 95459, 95585/11737/80302, 9944, 95611, 95561/ISP80282 |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland OH 44143-2131
|
For Additional Information Contact |
Ms. Holly Wright Lee 440-483-5777
|
Manufacturer Reason for Recall |
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
|
FDA Determined Cause 2 |
Other |
Action |
Philips Healthcare issued a Field Safety Notification (FSN)72800720/88200522
1. Identify affected product
2. There are 3 scenarios users may experience which result in identical images that are incorrectly labeled as phase tolerance images.
A. When performing a Step and Shoot acquisition with phase tolerance
selected, images may be reconstructed identically (a single phase), but
are labeled as different phases. This occurs when Start Final Recon
is selected prior to display of the ECG wave at the top of the
acquisition window.
B. When performing a Step and Shoot acquisition with phase tolerance
selected, images may be reconstructed identically (a single phase), but
are labeled as different phases. This occurs if ECG leads become
disconnected mid-acquisition or when the acquisition is halted
prematurely due to an application crash.
C. By system design, in certain instances when a patient exhibits an
arrhythmia or varying heart rate during a Step and Shoot acquisition,
images from the acquisition following the heart rate variation, may be
labeled in the incorrect phase
4. Determine whether you have a potentially affected system, then determine your AFFECTED PRODUCTS software version. To identify the software version of your product: Click the Help button, Select About and the software version is then displayed, Software versions 4.x, 3.x, and 2.x are potentially affected.
5. Take the Action listed in the field safety notice.
Please retain a copy with the equipment Instruction for use.
If you need any further information or support concerning this issue, please
contact your local Philips representative: North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377 and follow the prompts). |
Quantity in Commerce |
582 |
Distribution |
Worldwide distribution. US nationwide including Puerto Rico, Canada, Mexico, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Cote D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Monaco, Mongolia, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, State of Palestine, Panama, Peru, Philippines, Poland, Qatar, Reunion, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Timor-Leste, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, and Yemen |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAK and Original Applicant = PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
|
|
|
|