Date Initiated by Firm |
June 04, 2020 |
Date Posted |
July 10, 2020 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-2506-2020 |
Recall Event ID |
85816 |
PMA Number |
P890003 |
Product Classification |
Pulse generator, permanent, implantable - Product Code NVZ
|
Product |
Patient Connector, Model Number 24967 |
Code Information |
serial numbers: RFA001342A RFA001412A RFA001748A RFA001760A RFA001802A RFA001845A RFA001868A RFA001885A RFA002011A RFA002039A RFA002152A RFA003642A RFA003644A RFA003652A RFA003658A RFA003669A RFA003671A RFA003673A RFA003677A RFA003684A RFA003690A RFA003691A RFA003692A RFA003703A RFA003704A RFA003705A RFA004075A RFA004079A RFA004082A RFA004208A RFA004218A RFA004232A RFA004277A RFA004386A RFA004390A RFA004391A RFA004681A RFA004706A RFA005006A RFA005009A RFA005011A RFA005023A RFA005025A RFA005028A RFA005029A RFA005049A RFA005079A RFA005081A RFA005093A RFA005095A RFA005097A RFA005103A RFA005105A RFA005125A RFA005249A RFA005989A RFA006832A RFA008087A RFA008119A RFA008908A RFA009323A RFA010721A RFA010722A RFA010895A RFA011167A RFA011196A RFA011348A RFA011359A RFA011456A RFA011493A RFA011519A RFA011725A RFA011764A RFA011770A RFA011774A RFA011780A RFA011795A RFA011796A RFA011797A RFA011824A RFA011827A RFA011840A RFA011846A RFA011859A RFA011860A RFA011862A RFA011872A RFA011879A RFA011885A RFA011898A RFA011919A RFA011927A RFA011928A RFA011931A RFA011933A RFA011938A RFA011952A RFA011958A RFA011959A RFA011966A RFA011973A RFA011976A RFA011978A RFA012031A RFA012034A RFA012048A RFA012124A RFA012126A RFA012129A RFA012133A RFA012149A RFA012155A RFA012158A RFA012159A RFA012160A RFA012164A RFA012167A RFA012174A RFA012176A RFA012188A RFA012198A RFA012199A RFA012200A RFA012201A RFA012202A RFA012205A RFA012223A RFA012227A RFA012230A RFA012233A RFA012234A RFA012235A RFA012260A RFA012273A RFA012274A RFA012276A RFA012277A RFA012279A RFA012290A RFA012301A RFA012309A RFA012328A RFA012369A RFA012411A RFA012412A RFA012423A RFA012433A RFA012441A RFA012449A RFA012450A RFA012469A RFA012476A RFA012496A RFA012506A RFA012521A RFA012527A RFA012530A RFA012531A RFA012532A RFA012535A RFA012539A RFA012542A RFA012554A RFA012556A RFA012562A RFA012571A RFA012580A RFA012586A RFA012587A RFA012590A RFA012598A RFA012603A RFA012611A RFA012624A RFA012633A RFA012636A RFA012639A RFA012643A RFA012647A RFA012654A RFA012658A RFA012661A RFA012667A RFA012676A RFA012678A RFA012679A RFA012693A RFA012696A RFA012707A RFA012708A RFA012719A RFA012721A RFA012727A RFA012754A RFA012757A RFA012759A RFA012763A RFA012768A RFA012771A RFA012781A RFA012786A RFA012791A RFA012802A RFA012805A RFA012807A RFA012812A RFA012813A RFA012831A RFA012882A RFA012976A RFA013056A RFA013060A RFA013063A RFA013066A RFA013067A RFA013069A RFA013070A RFA013074A RFA013086A RFA013090A RFA013093A RFA013099A RFA013105A RFA013121A RFA013127A RFA013129A RFA013162A RFA013163A RFA013164A RFA013165A RFA013167A RFA013168A RFA013169A RFA013170A RFA013171A RFA013172A RFA013174A RFA013175A RFA013201A RFA013203A RFA013204A RFA013207A RFA013208A RFA013209A RFA013210A RFA013211A RFA013213A RFA013227A RFA013228A RFA013233A RFA013235A RFA013237A RFA013239A RFA013251A RFA013255A RFA013264A RFA013265A RFA013267A RFA013269A RFA013271A RFA013274A RFA013275A RFA013277A RFA013278A RFA013279A RFA013301A RFA013303A RFA013314A RFA013328A RFA013330A RFA013388A RFA013412A RFA013414A RFA013435A RFA013453A RFA013458A RFA013465A RFA013468A RFA013490A RFA013494A RFA013516A RFA013537A RFA013557A RFA013562A RFA013566A RFA013576A RFA013584A RFA013612A RFA013719A RFA013721A RFA013730A RFA013732A RFA013734A RFA013735A RFA013745A RFA013746A RFA013754A RFA013780A RFA013787A RFA013794A RFA013800A RFA013830A RFA013832A RFA013833A RFA013888A RFA013895A RFA013901A RFA013904A RFA013905A RFA013946A RFA013952A RFA013962A RFA013981A RFA014454A RFA014479A RFA014508A RFA014527A RFA014693A RFA014710A RFA014762A RFA014772A RFA014778A RFA014781A RFA014785A RFA014787A RFA014809A RFA014810A RFA014813A RFA014816A RFA014829A RFA014837A RFA014848A
RFA014854A RFA015059A RFA015060A RFA015068A RFA015074A RFA015076A RFA015099A RFA015103A RFA015107A RFA015112A RFA015133A RFA015145A RFA015149A RFA015156A RFA015160A RFA015165A RFA015178A RFA015179A RFA015186A RFA015187A RFA015198A RFA015201A RFA015202A RFA015204A RFA015207A RFA015218A RFA015223A RFA015231A RFA015271A RFA015272A RFA015276A RFA015281A RFA015290A RFA015292A RFA015297A RFA015298A RFA015299A RFA015304A RFA015311A RFA015313A RFA015320A RFA015322A RFA015326A RFA015338A RFA015339A RFA015347A RFA015348A RFA015349A RFA015352A RFA015356A RFA015357A RFA015361A RFA015364A RFA015401A RFA015402A RFA015403A RFA015410A RFA015415A RFA015417A RFA015418A RFA015419A RFA015423A RFA015427A RFA015428A RFA015429A RFA015430A RFA015438A RFA015439A RFA015443A RFA015448A RFA015455A RFA015459A RFA015465A RFA015467A RFA015468A RFA015470A RFA015471A RFA015472A RFA015480A RFA015486A RFA015493A RFA015497A RFA015498A RFA015499A RFA015501A RFA015502A RFA015567A RFA015568A RFA015573A RFA015574A RFA015576A RFA015577A RFA015579A RFA015581A RFA015582A RFA015587A RFA015588A RFA015590A RFA015592A RFA015595A RFA015598A RFA015600A RFA015601A RFA015610A RFA015611A RFA015614A RFA015616A RFA015620A RFA015621A RFA015622A RFA015624A RFA015630A RFA015631A RFA015633A RFA015635A RFA015638A RFA015647A RFA015648A RFA015651A RFA015654A RFA015659A RFA015660A RFA015661A RFA015664A RFA015708A RFA015725A RFA015726A RFA015746A RFA015747A RFA015748A RFA015751A RFA015758A RFA015759A RFA015760A RFA015761A RFA015763A RFA015765A RFA015766A RFA015771A RFA015772A RFA015776A RFA015779A RFA015783A RFA015785A RFA015790A RFA015841A RFA015842A RFA015854A RFA015856A RFA015857A RFA015873A RFA015910A RFA015956A RFA015959A RFA015962A RFA015964A RFA015965A RFA015967A RFA015968A RFA015969A RFA015975A RFA015976A RFA015981A RFA016026A RFA016029A RFA016031A RFA016032A RFA016036A RFA016041A RFA016043A RFA016056A RFA016057A RFA016061A RFA016063A RFA016065A RFA016070A RFA016072A RFA016073A RFA016079A RFA016081A RFA016082A RFA016087A RFA016103A RFA016104A RFA016107A RFA016118A RFA016123A RFA016126A RFA016128A RFA016139A RFA016142A RFA016146A RFA016194A RFA016196A RFA016197A RFA016201A RFA016204A RFA016207A RFA016210A RFA016214A RFA016216A RFA016220A RFA016227A RFA016236A RFA016239A RFA016242A RFA016244A RFA016249A RFA016251A RFA016255A RFA016256A RFA016258A RFA016259A RFA016265A RFA016272A RFA016275A RFA016276A RFA016279A RFA016288A RFA016291A RFA016294A RFA016307A RFA016312A RFA016347A RFA016351A RFA016388A RFA016389A RFA016489A RFA016490A RFA016491A RFA016494A RFA016506A RFA016510A RFA016513A RFA016514A RFA016518A RFA016520A RFA016524A RFA016528A RFA016533A RFA016534A RFA016541A RFA016544A RFA016545A RFA016547A RFA016555A RFA016556A RFA016560A RFA016563A RFA016566A RFA016567A RFA016571A RFA016574A RFA016577A RFA016586A RFA016588A RFA016596A RFA016636A RFA016641A RFA016647A RFA016762A RFA016766A RFA016767A RFA016769A RFA016772A RFA016776A RFA016777A RFA016779A RFA016780A RFA016781A RFA016783A RFA016793A RFA016795A RFA016816A RFA016825A RFA016834A RFA016838A RFA016839A RFA016848A RFA016852A RFA016870A RFA016872A RFA016878A RFA016885A RFA016889A RFA016897A RFA016905A RFA016911A RFA016920A RFA016923A RFA016926A RFA016928A RFA016931A RFA016932A RFA016933A RFA016934A RFA016937A RFA016940A RFA016942A RFA016945A RFA016955A RFA016956A RFA016958A RFA017028A RFA017030A RFA017039A RFA017040A RFA017045A RFA017054A RFA017060A RFA017063A RFA017074A RFA017091A RFA017116A RFA017119A RFA017125A RFA017138A RFA017149A RFA017166A RFA017183A RFA017185A RFA017186A RFA017196A RFA017197A RFA017203A RFA017205A RFA017206A RFA017207A RFA017252A RFA017260A RFA017266A RFA017270A RFA017272A RFA017273A RFA017277A RFA017600A RFA017603A RFA017607A RFA017611A RFA017620A RFA017626A RFA017627A RFA017632A RFA017634A RFA017635A RFA017643A RFA017647A RFA017650A RFA017652A RFA017655A RFA017658A RFA017674A RFA017676A RFA017685A RFA017686A RFA017687A RFA017688A RFA017694A RFA017698A RFA017704A RFA017719A RFA017720A RFA017721A RFA017955A RFA017956A RFA017959A RFA017964A RFA017965A RFA017971A RFA017976A RFA017980A RFA017981A RFA017984A RFA017994A RFA017996A RFA018004A RFA018005A RFA018007A RFA018010A RFA018011A RFA018020A RFA018023A RFA018024A RFA018026A RFA018029A RFA018030A RFA018033A RFA018035A RFA018036A RFA018039A RFA018042A RFA018044A RFA018048A RFA018049A RFA018054A RFA018058A RFA018059A RFA018060A RFA018062A RFA018064A RFA018083A RFA018089A RFA018091A RFA018092A RFA018093A RFA018122A RFA018128A RFA018133A RFA018137A RFA018141A RFA018144A RFA018146A RFA018148A RFA018166A RFA018173A RFA018176A RFA018188A RFA018245A RFA018252A RFA018257A RFA018258A RFA018261A RFA018262A RFA018271A RFA018274A RFA018278A RFA018283A RFA018294A RFA018300A RFA018301A RFA018305A RFA018307A RFA018312A RFA018316A RFA018319A RFA018320A RFA018322A RFA018323A RFA018326A RFA018327A RFA018332A RFA018333A RFA018340A RFA018346A RFA018399A RFA018400A RFA018401A RFA018408A RFA018409A RFA018410A RFA018416A RFA018417A RFA018425A RFA018454A RFA018459A RFA018461A RFA018470A RFA018483A RFA018486A RFA018488A RFA018524A RFA018525A RFA018526A RFA018527A RFA018600A RFA018602A RFA018604A RFA018607A RFA018614A RFA018619A RFA018623A RFA018628A RFA018631A RFA018633A RFA018654A RFA018659A RFA018663A RFA018673A RFA018805A RFA018807A RFA018897A RFA018898A RFA018899A RFA018907A RFA018908A RFA018909A RFA018913A RFA018914A RFA018932A RFA018933A RFA018934A RFA018941A RFA018945A RFA018947A RFA018950A RFA018954A RFA018963A RFA018969A RFA018970A RFA018972A RFA018975A RFA018978A RFA018979A RFA018982A RFA018983A RFA018990A RFA018992A RFA018993A RFA018995A RFA018998A RFA019005A RFA019006A RFA019007A RFA019009A RFA019012A RFA019013A RFA019015A RFA019017A RFA019020A RFA019023A RFA019028A RFA019031A RFA019032A RFA019033A RFA019034A RFA019035A RFA019042A RFA019044A RFA019045A RFA019046A RFA019047A RFA019049A RFA019084A RFA019088A RFA019189A RFA019200A RFA019219A RFA019236A RFA019247A RFA019248A RFA019262A RFA019313A RFA019361A RFA019377A RFA019387A RFA019415A RFA019417A RFA019418A RFA019446A RFA019448A RFA019453A RFA019466A RFA019501A RFA019504A RFA019651A RFA019668A RFA019727A RFA019758A RFA019957A RFA019959A RFA019960A RFA019970A RFA019971A RFA019972A RFA019973A RFA019974A RFA019975A RFA019976A RFA019977A RFA019978A RFA019982A RFA019985A RFA020006A RFA020025A RFA020027A RFA020037A RFA020043A RFA020046A RFA020051A RFA020061A RFA020062A RFA020080A RFA020104A RFA020118A RFA020120A RFA020127A RFA020139A RFA020164A RFA020176A RFA020182A RFA020188A RFA020189A RFA020194A RFA020231A RFA020234A RFA020236A RFA020248A RFA020269A RFA020277A RFA020289A RFA020290A RFA020293A RFA020300A RFA020301A RFA020308A RFA020315A RFA020325A RFA020329A RFA020342A RFA020349A RFA020356A RFA020367A RFA020369A RFA020374A RFA020383A RFA020390A RFA020396A RFA020399A RFA020401A RFA020405A RFA020439A RFA020444A RFA020448A RFA020459A RFA020460A RFA020463A RFA020469A RFA020470A RFA020472A RFA020475A RFA020481A RFA020491A RFA020507A RFA020508A RFA020513A RFA020516A RFA020518A RFA020553A RFA020555A RFA020556A RFA020557A RFA020558A RFA020735A RFA020736A RFA020737A RFA020738A RFA020739A RFA020740A RFA020742A RFA020743A RFA020775A RFA020777A RFA020787A RFA020791A RFA020792A RFA020912A RFA020913A RFA020914A RFA020915A RFA020916A RFA020917A RFA022213A RFA022393A RFA022398A RFA022400A RFA022401A RFA022416A RFA022423A RFA022426A RFA022433A RFA022435A RFA022436A RFA022439A RFA022503A RFA022508A RFA022625A RFA022650A RFA022683A RFA022687A RFA022713A RFA022727A RFA022818A RFA022820A RFA022824A RFA022825A RFA022827A RFA022828A RFA022829A RFA022831A RFA022839A RFA022841A RFA022842A RFA022846A |
Recalling Firm/ Manufacturer |
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) 8200 Coral Sea St Ne Mounds View MN 55112-4391
|
For Additional Information Contact |
Technical Services 800-638-1991
|
Manufacturer Reason for Recall |
There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.
|
FDA Determined Cause 2 |
Software design |
Action |
The firm initiated their medical device correction on 06/04/2020 by letter. The letter explained the problem and informed the consignee of the available software update (version 3.2.01) along with download instructions. |
Quantity in Commerce |
1025 devices |
Distribution |
worldwide |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
PMA Database |
PMAs with Product Code = NVZ and Original Applicant = Medtronic, Inc.
|