| | Class 2 Device Recall Datascope |  |
| Date Initiated by Firm | July 27, 2020 |
| Date Posted | August 21, 2020 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2842-2020 |
| Recall Event ID |
85823 |
| Product Classification |
System, balloon, intra-aortic and control - Product Code DSP
|
| Product | Datascope Intra-Aortic Ballon Catheters (IABs)-Trans-Ray 7Fr 40cc IAB (Japan) - Product Usage: Indications For Use: Acute Coronary Syndrome Cardiac and Non-Cardiac Surgery Complications of Heart Failure.
Kit P/N: 0684-00-0546-01 |
| Code Information |
Kit Lot Number: 3000051381 3000053316 3000057969 3000062074 3000066987 3000062557 3000062558 3000066988 3000064524 3000066989 3000066990 3000067890 3000069733 3000074397 3000074398 3000077670 3000077671 3000078857 3000080306 3000080310 3000083334 3000083335 3000084272 3000084969 3000087490 3000087491 3000092779 3000092780 3000095470 3000096586 3000098053 3000098054 3000098055 3000098056 3000099590 3000099591 3000112273 3000107761 3000107762 3000112274 |
| FEI Number |
1000222374
|
Recalling Firm/ Manufacturer |
Datascope Corporation 15 Law Dr Fairfield NJ 07004-3206
|
| For Additional Information Contact | SAME 973-244-6100 |
Manufacturer Reason for Recall | Potential Endotoxin Contamination |
FDA Determined Cause 2 | Process change control |
| Action | Datascope/Getinge issued Urgent Medical Device Removal letter (customer specific) on 7/27/20, via FedEx with Signature Proof of Delivery. Letter states reason for recall, health risk and action to take: Records indicate that you received potentially affected IABs identifiable by the lot and serial numbers listed . Please complete the steps below:
Monitor patients for pyrogens reaction/humoral immune response/ coagulation and complement cascades/ inflammation. Monitor and treat any signs of inflammation according to your facilitys protocols and clinical judgment.
Please examine your inventory immediately, remove and quarantine any unexpired affected IAB following the steps below:
Identify any unexpired IAB Kits referencing the part and lot numbers listed in Table 1.
The Kit part and lot numbers can be found on the Outer Shelf Carton Using steps below, Identify any IAB Yfittings with serial numbers listed in Table 1.
Remove tape on one side of the Shelf Carton.
Hinge the box open, keeping the other side taped.
Do not remove any items from the carton. Instead, Lift the Accessories up and locate the YFitting through the clear Mylar overwrap.
Read the serial number off the IAB YFitting If you find an IAB with a serial number listed in Table 1, immediately place the entire Kit in
quarantine for return and replacement or credit.
If the affected serial number is not found, close the carton and reseal. Cut out and affix one of the enclosed adhesive labels to designate the kit has been checked for affected recall Serial Numbers and may be placed back in stock. 40 peeloff labels have been provided. Color copies may be made and affixed with tape if more than 40 are needed.
If you have unexpired affected product to return, please contact Getinge Customer Service at 888 9GETUSA (8889438872) (option 2) between the hours of 9 AM and 6 PM Eastern Standard Time to request a return authorization number (RMA) and shipping instructions.
Pack th |
| Quantity in Commerce | 3059 Kits |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of Argentina, Austrailia Austria Belgium Brazil Canada Chile Colombia Czech Republic Denmark Finland France Germany Hong Kong India Ireland Italy Japan Korea Mexico Netherlands New Zealand Norway Paraguay Poland Portugal Russia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Kenya, South Africa, Israel, Mauritus, Nigeria, South Africa, Namibia, Uganda. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|