• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Datascope

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Datascopesee related information
Date Initiated by FirmJuly 27, 2020
Date PostedAugust 21, 2020
Recall Status1 Open3, Classified
Recall NumberZ-2844-2020
Recall Event ID 85823
Product Classification System, balloon, intra-aortic and control - Product Code DSP
ProductDatascope Intra-Aortic Ballon Catheters (IABs)-TRANS-RAY PLUS 7.5Fr 40cc IAB (Japan) Kit P/N: 0684-00-0608 - Product Usage: Indications For Use: Acute Coronary Syndrome Cardiac and Non-Cardiac Surgery Complications of Heart Failure.
Code Information Kit Lot Number:  3000045568 3000047929 3000047928 3000047930 3000053321 3000060109 3000060110 3000067699 3000062192 3000067700 3000067701 3000069728 3000069729 3000069730 3000071288 3000073187 3000076242 3000077669 3000079105 3000079318 3000080309 3000080876 3000084268 3000084967 3000084960 3000084961 3000086800 3000088372 3000090597 3000090598 3000092668 3000092666 3000092667 3000092669 3000096268 3000095468 3000096267 3000096269 3000096270 3000096271 3000096587 3000098175 3000098747 3000098748 3000102410 3000102411 3000103470 3000104519 3000106011 3000107760 3000110622 3000110623 3000112744 3000113154 
FEI Number 1000222374
Recalling Firm/
Manufacturer
Datascope Corporation
15 Law Dr
Fairfield NJ 07004-3206
For Additional Information ContactSAME
973-244-6100
Manufacturer Reason
for Recall
Potential Endotoxin Contamination
FDA Determined
Cause 2
Process change control
ActionDatascope/Getinge issued Urgent Medical Device Removal letter (customer specific) on 7/27/20, via FedEx with Signature Proof of Delivery. Letter states reason for recall, health risk and action to take: Records indicate that you received potentially affected IABs identifiable by the lot and serial numbers listed . Please complete the steps below: Monitor patients for pyrogens reaction/humoral immune response/ coagulation and complement cascades/ inflammation. Monitor and treat any signs of inflammation according to your facilitys protocols and clinical judgment. Please examine your inventory immediately, remove and quarantine any unexpired affected IAB following the steps below: Identify any unexpired IAB Kits referencing the part and lot numbers listed in Table 1. The Kit part and lot numbers can be found on the Outer Shelf Carton Using steps below, Identify any IAB Yfittings with serial numbers listed in Table 1. Remove tape on one side of the Shelf Carton. Hinge the box open, keeping the other side taped. Do not remove any items from the carton. Instead, Lift the Accessories up and locate the YFitting through the clear Mylar overwrap. Read the serial number off the IAB YFitting If you find an IAB with a serial number listed in Table 1, immediately place the entire Kit in quarantine for return and replacement or credit. If the affected serial number is not found, close the carton and reseal. Cut out and affix one of the enclosed adhesive labels to designate the kit has been checked for affected recall Serial Numbers and may be placed back in stock. 40 peeloff labels have been provided. Color copies may be made and affixed with tape if more than 40 are needed. If you have unexpired affected product to return, please contact Getinge Customer Service at 888 9GETUSA (8889438872) (option 2) between the hours of 9 AM and 6 PM Eastern Standard Time to request a return authorization number (RMA) and shipping instructions. Pack th
Quantity in Commerce11555 Kits
DistributionWorldwide distribution - US Nationwide distribution and the countries of Argentina, Austrailia Austria Belgium Brazil Canada Chile Colombia Czech Republic Denmark Finland France Germany Hong Kong India Ireland Italy Japan Korea Mexico Netherlands New Zealand Norway Paraguay Poland Portugal Russia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Kenya, South Africa, Israel, Mauritus, Nigeria, South Africa, Namibia, Uganda.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-