Class 2 Device Recall IV Administration Set

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall IV Administration Setsee related information
Date Initiated by FirmJuly 06, 2020
Create DateOctober 22, 2020
Recall Status1 Terminated 3 on May 11, 2023
Recall NumberZ-0270-2021
Recall Event ID 86054
Product Classification Set, administration, intravascular - Product Code FPA
Product165cm (65 IN)APPX 2.2ml, PUR EXT SET, 3-PORT NanoClave MANIFOLD, BCV, NanoClave. 1 unit per pouch, 50 pouches per case. UDI:(01)10840619062761(17)241201(30)50(10)4553703. Administration sets and their accessories are single use, pre-sterilized devices that provide access for the administration of fluids from a container to a patients vascular system through the administrations set needle or catheter, which is inserted into the vein
Code Information Lot Number:4553703.                                                                                        
FEI Number 2025816
Recalling Firm/
Manufacturer
ICU Medical, Inc.
951 Calle Amanecer
San Clemente CA 92673-6212
For Additional Information Contact
844-654-7780
Manufacturer Reason
for Recall
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
FDA Determined
Cause 2
Process control
ActionOn July 6, 2020, the firm, icumedical, sent an "URGENT: MEDICAL DEVICE RECALL" notification of the market action to direct and indirect consignees of record via UPS. Stericycle, Inc. was contracted to send letters to consignees. The following is included in the "Required Actions for Users" section of the medical device recall notification: 1) Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with the letter. Call 1-866-203-5570 (M F, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, you will be credited for any product returned. You will only receive credit for product that you return. NOTE: Credits for product purchased through distributor will be credited by the distributor. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product I Table 1 sections of this notification and ask them to call 1-866-203-5570 (M-F, 8am-5pm ET) to obtain a response form. Follow up Actions: Contact Customer Service using the information provided below for assistance reordering replacement product and for further inquiries: Global Complaint Management -1-844-654-7780 or ProductCompla
Quantity in CommerceTotal of all products (Listed #1 thru 101) = 304735 units
DistributionWorldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Total Product Life CycleTPLC Device Report