New Search Back to Search Results
 Class 2 Device Recall PowerMidline Cathetersee related information
Date Initiated by FirmSeptember 14, 2020
Create DateOctober 23, 2020
Recall Status1 Terminated 3 on November 09, 2021
Recall NumberZ-0295-2021
Recall Event ID 86477
510(K)NumberK162900 
Product Classification Midline catheter - Product Code PND
ProductBard 3F SL PowerMidline Full Kit CE-indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media Ref: P6153208
Code Information Lot Number:  REEQ3598 Feb-2022 and REER0518 Mar-2022 UDI: 00801741141065
FEI Number 2243072
Recalling Firm/
Manufacturer
Becton Dickinson & Company
1 Becton Dr
Franklin Lakes NJ 07417-1815
For Additional Information ContactSAME
201-847-6800
Manufacturer Reason
for Recall
Kits do not contain the stylet/T-lock assembly indicated as a component on the label, may lead to a procedural delay and possibly discard of the kit.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionBecton Dickinson (BD) issued notification letter to FSN Europe on 9/14/20 to customer stating reason for recall, health risk and action to take specific to the country: Advice on actions to be taken by the Distributor: 1. Inspect your inventory, locate and quarantine any units of the impacted devices listed in Table 1. 2. If you have further distributed the product, please identify those facilities, notify them at once of this product removal and have them return the affected product to your facility. 3. Destroy any unused units of the affected product and complete the customer response form on page 3 indicating: o the quantities destroyed OR o that your organisation does not have any impacted units left in inventory 4. Return the completed Customer Response Form to <> as soon as possible or no later than October 9th, 2020. Should you have any questions or require assistance relating to this Field Safety Notice, please contact your Local BD representative at <>
Quantity in Commerce72 units
DistributionForeign: Italy, Spain, and UK
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = PND
-
-