| Class 2 Device Recall Atellica IM 1300 and Atellica 1600 Analyzers | |
Date Initiated by Firm | October 26, 2020 |
Create Date | November 19, 2020 |
Recall Status1 |
Terminated 3 on April 19, 2024 |
Recall Number | Z-0473-2021 |
Recall Event ID |
86718 |
510(K)Number | K151792 |
Product Classification |
Radioimmunoassay, thyroid-stimulating hormone - Product Code JLW
|
Product | Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001 |
Code Information |
Software version: V1.23.1 or lower UDI: 00630414007960 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
For Additional Information Contact | SAME 914-631-8000 |
Manufacturer Reason for Recall | Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values |
FDA Determined Cause 2 | Under Investigation by firm |
Action | Siemens issued an Urgent Medical Device Correction (UMDC ASW21-01.A.US) via Fed'X to US Customers on 10/26/20 and Urgent Field Safety Notice, (UFSN ASW21-01.A.OUS) via electronic distribution to regional countries for implementation Outside US (OUS)
The UMDC and UFSN explain the behaviors that can occur and instructs customers on actions to be taken for the issues described in the communication. Customers are requested to return the attached effectiveness check form via fax or email.
A software version correcting the issues will be released when available and all systems will be updated. |
Quantity in Commerce | US 392 units; OUS 449 units |
Distribution | Worldwide distribution - US Nationwide distribution and the countries of Algeria, Argentina, Australia, Austria, Bangladesh,
Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia,
Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany,
Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel,
Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Libya, Lithuania,
Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman,
P.R. China, Pakistan, Peru, Poland, Portugal, Qatar Republic, Korea, Romania, Russian Fed., Saudi Arabia, Singapore,
Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland,
Taiwan, Thailand, Turkey, U.A.E., Ukraine, United Kingdom,
Uruguay, Vatikancity, Vietnam. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JLW
|
|
|
|