• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall ACIST Kodama Intravascular Ultrasound Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall ACIST Kodama Intravascular Ultrasound Cathetersee related information
Date Initiated by FirmJanuary 22, 2021
Date PostedMarch 04, 2021
Recall Status1 Terminated 3 on October 30, 2023
Recall NumberZ-1161-2021
Recall Event ID 87254
510(K)NumberK193183 
Product Classification Catheter, ultrasound, intravascular - Product Code OBJ
ProductACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a component of the ACIST HDi System. The ACIST HDi System is intended to be used for ultrasound examination of coronary and peripheral intravascular pathology. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. The ACIST Kodama Intravascular Ultrasound Catheter is intended for use with the ACIST HDi System.
Code Information Model Number: 017788, 018125 (Japan only); Lot codes: 00233370 (100 units), 00233371 (90 units), 00233372 (100 units), 00233373 (100 units), 00233374 (100 units), 00233380 (100 units), 00233384 (60 units), 00233385 (100 units), 00233393 (100 units), 00233394 (100 units), 00237604 (35 units), 00237613 (100 units), 03012517 (100 units)
FEI Number 3001726808
Recalling Firm/
Manufacturer
Acist Medical Systems
7905 Fuller Rd
Eden Prairie MN 55344-2137
For Additional Information ContactKristen Knox
952-374-9083
Manufacturer Reason
for Recall
Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-ring had the potential to be flushed out of the catheter. ACIST is confirming the source of the failure mode to assure the quality and reliability of the Kodama catheter. There have been no related field reports related to this incident, nor any evidence or report of patient injury or adverse health consequence.
FDA Determined
Cause 2
Device Design
ActionThe firm, ACIST, sent an, "URGENT: MEDICAL DEVICE RECALL" letter and response form dated Jan 22, 2021 to customers on Jan. 22, 2021. The letter describe the product, problem and actions to be taken. The customers were instructed to do the following: to complete all of the steps outlined below and return the completed Recall Response Form to Stericycle by e-mail: acistmedical8961@stericycle.com or fax to 877-576-9366. 1. Check your inventory of Kodama HD-IVUS Catheter 2. Record quantities of each lot in the Response Form 3. Remove the affected lots from your inventory. 4. Use the enclosed, prepaid return label to return your affected product including a copy of the response form with the product. If you need additional labels, please contact Stericycle at 877-576-9382. If you have received any reports of illness, injury or other health consequence related to the use of product please contact Customer Support: Customer.Support@acistmedical.com Please forward this notice to those who need to be aware within your organization. If you have any further questions or concerns, please contact Stericycle at 877-576-9382.
Quantity in Commerce1185 units
DistributionWorldwide - US Nationwide Distribution in the states of AL, AR, CA, CO, FL, IL, KS, KY, LA, MD, MI, MO, NC, NJ, NY, OK, PA, RI, TN; In the countries of India, Italy, Japan, Poland, and United Arab Emirates.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = OBJ
-
-