Class 2 Device Recall Philips

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philipssee related information
Date Initiated by FirmAugust 11, 2021
Create DateSeptember 03, 2021
Recall Status1 Completed
Recall NumberZ-2414-2021
Recall Event ID 88491
510(K)NumberK162931 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
Product Ingenia 1.5T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781261 781262
Code Information Model: 781261 Serial Number Unique Device Identifier (UDI#) 77008 77000 77004 77007 77003 77001 77002 77018 77016 77019 77010 77014 Model:781262 77131 (01)00884838068445(21)77131 77031 N/A 77046 N/A 77070 N/A 77124 (01)00884838068445(21)77124 77144 (01)00884838068445(21)77144 77172 (01)00884838068445(21)77172 77074 (01)00884838068445(21)77074 77058 N/A 77114 N/A 77127 (01)00884838068445(21)77127 77160 (01)00884838068445(21)77160 77159 (01)00884838068445(21)77159 77182 (01)00884838068445(21)77182 77180 (01)00884838068445(21)77180 77171 (01)00884838068445(21)77171 77179 (01)00884838068445(21)77179 77048 N/A 77040 N/A 77034 N/A 77039 N/A 77035 N/A 77092 N/A 77140 (01)00884838068445(21)77140 77177 (01)00884838068445(21)77177 77181 (01)00884838068445(21)77181 77178 (01)00884838068445(21)77178 77041 N/A 77103 N/A 77193 (01)00884838068445(21)77193 77203 (01)00884838068445(21)77203 77134 (01)00884838068445(21)77134 77047 N/A 77051 N/A 77050 N/A 77060 N/A 77194 (01)00884838068445(21)77194 77111 N/A 77119 (01)00884838068445(21)77119 77167 (01)00884838068445(21)77167 77183 (01)00884838068445(21)77183 77185 (01)00884838068445(21)77185 77187 (01)00884838068445(21)77187 77188 (01)00884838068445(21)77188 77198 (01)00884838068445(21)77198 77195 (01)00884838068445(21)77195 77196 (01)00884838068445(21)77196 77199 N/A 77202 N/A 77037 N/A 77057 N/A 77072 N/A 77075 N/A 77084 N/A 77085 N/A 77078 N/A 77086 N/A 77101 N/A 77110 (01)00884838068445(21)77110 77038 (01)00884838068445(21)77038 77157 (01)00884838068445(21)77157 77156 (01)00884838068445(21)77156 77056 N/A 77052 N/A 77054 N/A 77053 N/A 77055 N/A 77064 N/A 77065 N/A 77067 N/A 77066 N/A 77068 N/A 77069 N/A 77105 N/A 77126 (01)00884838068445(21)77126 77125 (01)00884838068445(21)77125 77137 (01)00884838068445(21)77137 77139 (01)00884838068445(21)77139 77136 (01)00884838068445(21)77136 77135 (01)00884838068445(21)77135 77138 (01)00884838068445(21)77138 77148 (01)00884838068445(21)77148 77141 (01)00884838068445(21)77141 77030 N/A 77033 N/A 77036 N/A 77042 N/A 77044 N/A 77043 N/A 77032 N/A 77049 N/A 77061 N/A 77059 N/A 77071 N/A 77090 N/A 77088 N/A 77093 N/A 77091 N/A 77087 N/A 77097 N/A 77083 N/A 77106 N/A 77112 N/A 77107 (01)00884838068445(21)77107 77115 N/A 77100 (01)00884838068445(21)77100 77118 (01)00884838068445(21)77118 77121 (01)00884838068445(21)77121 77129 (01)00884838068445(21)77129 77120 (01)00884838068445(21)77120 77128 (01)00884838068445(21)77128 77151 (01)00884838068445(21)77151 77143 (01)00884838068445(21)77143 77158 (01)00884838068445(21)77158 77162 (01)00884838068445(21)77162 77161 (01)00884838068445(21)77161 77155 (01)00884838068445(21)77155 77190 (01)00884838068445(21)77190 77191 (01)00884838068445(21)77191 77150 (01)00884838068445(21)77150 77081 N/A 77117 (01)00884838068445(21)77117 77123 (01)00884838068445(21)77123 77166 (01)00884838068445(21)77166 77045 N/A 77063 N/A 77109 N/A 77201 (01)00884838068445(21)77201 77113 N/A 77147 (01)00884838068445(21)77147 77122 (01)00884838068445(21)77122 77154 (01)00884838068445(21)77154 77169 (01)00884838068445(21)77169 77168 (01)00884838068445(21)77168 77130 (01)00884838068445(21)77130 77094 N/A 77062 N/A
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
Manufacturer Reason
for Recall
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
FDA Determined
Cause 2
Device Design
ActionPhilips notified U.S. customers an Urgent Medical Device Correction letter on 8/11/21. Letter states reason for recall, health risk and action to take: Philips advises that customers continue to utilize the primary motorized function for the MR system since the identified issue with the HDU does not occur in the motorized mode. If you utilize the manual function and it is not functioning properly, then you must perform table movement using the motorized function. A Philips Field Service Engineer (FSE) will perform a HDU during the next scheduled preventive maintenance visit. If the issue is detected in the HDU, the Philips FSE will replace the necessary mechanical component(s) to resolve the issue. If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377) and reference FCO78100527.
Quantity in Commerce12 units, 137 units
DistributionWorldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.