| Class 2 Device Recall HeartWare |  |
Date Initiated by Firm | November 29, 2022 |
Date Posted | December 21, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0745-2023 |
Recall Event ID |
91281 |
Product Classification |
Ventricular (assist) bypass - Product Code DSQ
|
Product | HeartWare HVAD Pump Kit, REF 1103 |
Code Information |
UDI/DI 00888707003261, Serial Numbers: HW11011, HW11013, HW11019, HW11033, HW11036, HW11047, HW11060, HW11064, HW11067, HW11074, HW11075, HW11078, HW11086, HW11088, HW11094, HW11106, HW11107, HW11112, HW11113, HW11116, HW11120, HW11123, HW11126, HW11129, HW11140, HW11141, HW11146, HW11155, HW11157, HW11168, HW11171, HW11174, HW11181, HW11183, HW11187, HW11193, HW11201, HW11209, HW11210, HW11222, HW11224, HW11244, HW11263, HW11277, HW11299, HW11318, HW11332, HW11335, HW11338, HW11339, HW11347, HW11353, HW11356, HW11357, HW11364, HW11368, HW11383, HW11384, HW11385, HW11386, HW11388, HW11405, HW11413, HW11427, HW11442, HW11453, HW11463, HW11471, HW11478, HW11482, HW11488, HW11493, HW11494, HW11501, HW11504, HW11510, HW11514, HW11515, HW11516, HW11525, HW11530, HW11532, HW11537, HW11557, HW11558, HW11561, HW11565, HW11586, HW11598, HW11600, HW11609, HW11610, HW11611, HW11616, HW11624, HW11630, HW11631, HW11638, HW11639, HW11647, HW11648, HW11650, HW11654, HW11655, HW11658, HW11664, HW11669, HW11670, HW11671, HW11678, HW11680, HW11685, HW11686, HW11690, HW11691, HW11695, HW11699, HW11700, HW11707, HW11714, HW11718, HW11722, HW11724, HW11727, HW11734, HW11736, HW11740, HW11748, HW11753, HW11756, HW11774, HW11775, HW11780, HW11785, HW11792, HW11795, HW11802, HW11812, HW11823, HW11830, HW11834, HW11846, HW11853, HW11854, HW11860, HW11863, HW11864, HW11886, HW11899, HW11902, HW11903, HW11904, HW11915, HW11932, HW11938, HW11943, HW11945, HW11950, HW11954, HW11959, HW11972, HW11973, HW11977, HW11982, HW12013, HW12017, HW12024, HW12026, HW12061, HW12065, HW12067, HW12072, HW12073, HW12080, HW12086, HW12092, HW12097, HW12113, HW12116, HW12118, HW12121, HW12127, HW12130, HW12131, HW12133, HW12134, HW12142, HW12155, HW12157, HW12163, HW12172, HW12174, HW12175, HW12177, HW12180, HW12183, HW12207, HW12213, HW12216, HW12223, HW12224, HW12225, HW12231, HW12232, HW12233, HW12240, HW12241, HW12244, HW12250, HW12260, HW12262, HW12264, HW12279, HW12283, HW12296, HW12307, HW12315, HW12317, HW12321, HW12322, HW12332, HW12345, HW12347, HW12350, HW12355, HW12356, HW12358, HW12359, HW12367, HW12369, HW12380, HW12383, HW12385, HW12389, HW12392, HW12399, HW12402, HW12404, HW12406, HW12407, HW1998, HW2012, HW2017, HW2068, HW2070, HW2093, HW2129, HW2130, HW2137, HW2142, HW21507, HW21528, HW21529, HW21530, HW21542, HW21552, HW21556, HW21562, HW21576, HW21578, HW21580, HW21590, HW21592, HW21599, HW21609, HW2161, HW21614, HW21621, HW21625, HW21628, HW21630, HW21638, HW21663, HW21673, HW21678, HW21688, HW21695, HW21699, HW21701, HW21702, HW21703, HW21704, HW21705, HW21707, HW21711, HW21714, HW21719, HW21726, HW21731, HW21732, HW2177, HW2185, HW21943, HW21946, HW21950, HW21952, HW21962, HW21963, HW21972, HW21973, HW21980, HW21981, HW21985, HW21986, HW22000, HW22005, HW22009, HW22011, HW22015, HW22016, HW22017, HW22037, HW22038, HW22042, HW22044, HW22045, HW22046, HW22050, HW22051, HW22052, HW22062, HW22065, HW22067, HW22072, HW22074, HW22077, HW2208, HW22081, HW22083, HW22086, HW22088, HW22089, HW2209, HW22091, HW22095, HW22097, HW22106, HW22107, HW22112, HW22116, HW22117, HW22121, HW22130, HW22135, HW22136, HW22148, HW22151, HW22152, HW22154, HW22156, HW22175, HW22177, HW22184, HW22204, HW22205, HW22206, HW22212, HW22213, HW22275, HW22281, HW22282, HW22285, HW22286, HW22291, HW22293, HW22320, HW22326, HW22339, HW22341, HW22344, HW22346, HW22351, HW22354, HW22355, HW22358, HW22360, HW2245, HW22453, HW22459, HW22460, HW22462, HW22472, HW22473, HW22474, HW22479, HW22480, HW22484, HW22494, HW22497, HW2256, HW22565, HW22572, HW2258, HW2260, HW22612, HW22614, HW22624, HW22626, HW22629, HW22644, HW22658, HW22661, HW22663, HW22667, HW22670, HW22720, HW22723, HW22726, HW22730, HW22741, HW22746, HW22750, HW22757, HW22763, HW22774, HW22778, HW22804, HW22805, HW22809, HW22814, HW22816, HW22823, HW22876, HW22879, HW22885, HW22927, HW22928, HW22932, HW22934, HW22937, HW22978, HW22985, HW22991, HW22992, HW22993, HW22996, HW22997, HW22998, HW23001, HW23074, HW23075, HW23086, HW23088, HW23089, HW23090, HW23091, HW23104, HW23115, HW23149, HW23150, HW23151, HW23159, HW23161, HW23167, HW23171, HW23172, HW23179, HW23183, HW23184, HW23207, HW23222, HW23229, HW23232, HW23236, HW23237, HW23240, HW23241, HW23249, HW23290, HW23291, HW23292, HW23302, HW23310, HW23312, HW23320, HW23321, HW23389, HW23390, HW23394, HW23397, HW23411, HW23414, HW23416, HW23420, HW23434, HW23441, HW2345, HW23459, HW23465, HW23467, HW23470, HW23488, HW23489, HW23494, HW23496, HW23498, HW23500, HW23503, HW23528, HW23529, HW23530, HW23531, HW23534, HW23538, HW23543, HW23546, HW2355, HW23554, HW23558, HW2356, HW23562, HW23566, HW23568, HW23571, HW23572, HW23600, HW23606, HW23620, HW23624, HW23626, HW23678, HW23691, HW23696, HW23697, HW23700, HW23701, HW23703, HW23709, HW23726, HW23729, HW23746, HW23756, HW23761, HW23765, HW23771, HW23772, HW23774, HW23781, HW23787, HW23795, HW23801, HW23805, HW23807, HW23815, HW23816, HW23821, HW23826, HW23828, HW23850, HW23855, HW23859, HW23860, HW23877, HW23942, HW23944, HW23948, HW23950, HW23959, HW23964, HW23988, HW23989, HW23991, HW24008, HW24009, HW24011, HW24034, HW24045, HW24048, HW24051, HW24053, HW24070, HW24071, HW24111, HW24116, HW24118, HW24119, HW24120, HW24121, HW24169, HW24194, HW24195, HW24199, HW24210, HW24216, HW24245, HW24251, HW24255, HW24260, HW24263, HW24314, HW24318, HW24331, HW24339, HW24343, HW2443, HW24434, HW2450, HW24500, HW24502, HW24504, HW24516, HW24518, HW24522, HW24529, HW24534, HW2456, HW24595, HW24597, HW24601, HW24609, HW24615, HW24618, HW24683, HW24685, HW24724, HW24726, HW24727, HW24738, HW24739, HW24751, HW24754, HW24769, HW24771, HW24782, HW24790, HW24813, HW24816, HW24818, HW24820, HW24821, HW24822, HW24824, HW24828, HW24832, HW24837, HW24847, HW24852, HW24854, HW24862, HW24870, HW24872, HW24873, HW24874, HW24876, HW24877, HW24882, HW24896, HW24902, HW24916, HW24917, HW24919, HW24922, HW24943, HW24944, HW24946, HW24947, HW24948, HW24956, HW24961, HW24974, HW24976, HW24983, HW25007, HW25011, HW25019, HW25020, HW25029, HW25034, HW25038, HW25040, HW25041, HW25045, HW25046, HW25049, HW25054, HW25055, HW25057, HW25063, HW25099, HW25118, HW25123, HW25125, HW25127, HW25129, HW25134, HW25153, HW25188, HW25189, HW25190, HW25191, HW25197, HW25208, HW2521, HW2522, HW25256, HW25263, HW25265, HW25275, HW25276, HW25284, HW25285, HW25286, HW25291, HW2532, HW25321, HW25328, HW25331, HW25350, HW25418, HW2542, HW25423, HW25424, HW25425, HW25442, HW25443, HW25449, HW25458, HW25477, HW25482, HW25485, HW25494, HW25499, HW25500, HW25507, HW25508, HW25509, HW25512, HW25543, HW2555, HW25670, HW25679, HW25683, HW25684, HW25714, HW25717, HW25725, HW25737, HW25760, HW25773, HW25782, HW25783, HW25786, HW25787, HW25856, HW25863, HW25868, HW25869, HW25923, HW25937, HW25942, HW25943, HW25946, HW25947, HW25949, HW25955, HW25961, HW25966, HW25969, HW25970, HW25972, HW25974, HW25978, HW25983, HW25985, HW25997, HW26001, HW26019, HW26024, HW26025, HW26027, HW26028, HW26029, HW26030, HW26035, HW26042, HW26062, HW26066, HW26072, HW26081, HW26082, HW26087, HW26088, HW26089, HW26090, HW26117, HW26118, HW26123, HW26131, HW26132, HW26139, HW26142, HW26176, HW26177, HW26179, HW26181, HW26190, HW26197, HW2629, HW26338, HW26339, HW26341, HW26342, HW26346, HW26348, HW26350, HW26351, HW26353, HW26354, HW26356, HW26357, HW26361, HW26374, HW26376, HW26402, HW26407, HW26408, HW26412, HW26415, HW26416, HW26418, HW26422, HW26423, HW26429, HW26435, HW26463, HW26469, HW26475, HW26481, HW26482, HW26486, HW26487, HW26493, HW26518, HW26525, HW2653, HW26531, HW26538, HW2654, HW26540, HW26542, HW26545, HW26564, HW26578, HW26579, HW26580, HW26584, HW26587, HW2659, HW26591, HW26596, HW26597, HW26598, HW26605, HW26607, HW26609, HW26627, HW26628, HW26633, HW26634, HW26666, HW2667, HW26670, HW26680, HW26681, HW26682, HW26686, HW26687, HW26688, HW26694, HW26710, HW26711, HW26712, HW26713, HW26718, HW26723, HW26726, HW26729, HW26745, HW26752, HW26757, HW26759, HW26760, HW26767, HW26774, HW26775, HW26777, HW26779, HW26783, HW26799, HW26801, HW26803, HW26805, HW26808, HW26810, HW26811, HW26812, HW26816, HW26817, HW26818, HW26819, HW26820, HW26823, HW26826, HW26827, HW26828, HW26829, HW26832, HW26834, HW26840, HW26845, HW26848, HW26851, HW26862, HW26863, HW26864, HW26865, HW26866, HW26869, HW26870, HW26873, HW26874, HW26875, HW26878, HW26880, HW26881, HW26882, HW26885, HW26886, HW26889, HW26890, HW26891, HW26897, HW26898, HW26899, HW26903, HW26905, HW26907, HW26909, HW26927, HW26931, HW26932, HW26937, HW26943, HW26944, HW26961, HW26965, HW26972, HW26975, HW26978, HW2698, HW26991, HW26993, HW26995, HW26996, HW27008, HW27009, HW27010, HW27014, HW27017, HW27057, HW27081, HW27082, HW27083, HW27090, HW27098, HW27141, HW27218, HW27221, HW27228, HW27229, HW2723, HW27240, HW27250, HW27251, HW27255, HW27257, HW27262, HW27263, HW27265, HW27267, HW27269, HW27273, HW27274, HW27276, HW27278, HW27279, HW27282, HW27288, HW27290, HW27296, HW27297, HW27298, HW27299, HW27302, HW27305, HW27306, HW27308, HW27309, HW27314, HW27316, HW27317, HW27319, HW27322, HW27324, HW27326, HW27333, HW27335, HW27336, HW27338, HW27342, HW27352, HW27353, HW27363, HW27369, HW27371, HW27374, HW27376, HW27377, HW27381, HW27382, HW27385, HW27386, HW27405, HW2741, HW27421, HW27422, HW27425, HW27426, HW27427, HW27435, HW27440, HW27447, HW27450, HW27459, HW27464, HW27466, HW27467, HW27478, HW27482, HW27484, HW27486, HW27490, HW27491, HW27493, HW27495, HW27497, HW27506, HW27508, HW27512, HW27514, HW27515, HW27519, HW27521, HW27522, HW27543, HW27544, HW27545, HW27546, HW27547, HW27550, HW27554, HW27558, HW27562, HW27565, HW27567, HW27569, HW27607, HW27609, HW27614, HW27615, HW27621, HW27623, HW27624, HW27626, HW27642, HW27653, HW27654, HW27657, HW27691, HW2771, HW27717, HW27720, HW27727, HW27749, HW27754, HW27757, HW27758, HW27762, HW27765, HW27768, HW27772, HW27801, HW27802, HW27803, HW27829, HW27830, HW27835, HW27838, HW27839, HW27842, HW27843, HW27855, HW27870, HW27895, HW27896, HW27911, HW27918, HW27922, HW27927, HW27930, HW27932, HW27944, HW27945, HW27949, HW27950, HW27951, HW27953, HW2798, HW28011, HW28012, HW28016, HW28018, HW28027, HW28074, HW28086, HW28087, HW28088, HW28090, HW28094, HW28110, HW28112, HW28113, HW28119, HW2812, HW28128, HW28130, HW28132, HW28134, HW28136, HW28138, HW28140, HW2815, HW28178, HW28184, HW28185, HW28188, HW2819, HW28192, HW28193, HW28213, HW28219, HW28225, HW28240, HW28243, HW28251, HW28252, HW28253, HW28255, HW28257, HW28259, HW28260, HW28261, HW28262, HW28267, HW28270, HW28271, HW28274, HW28275, HW28278, HW28288, HW28289, HW28300, HW28304, HW28305, HW28318, HW28322, HW28323, HW28325, HW28327, HW28358, HW28360, HW28365, HW28366, HW28367, HW28381, HW28396, HW28397, HW28398, HW28401, HW28406, HW28408, HW28426, HW28427, HW28428, HW28431, HW28432, HW28434, HW28437, HW28438, HW28440, HW28449, HW28452, HW28454, HW28456, HW28459, HW28460, HW28487, HW28488, HW28493, HW28495, HW28496, HW28497, HW28509, HW28514, HW28519, HW28526, HW28528, HW28530, HW28532, HW28533, HW28534, HW28535, HW28592, HW28603, HW28607, HW28614, HW28615, HW28641, HW28645, HW28646, HW28648, HW28649, HW28652, HW28655, HW28664, HW28682, HW28686, HW28687, HW28695, HW28696, HW28699, HW28702, HW28703, HW28704, HW28706, HW28708, HW28711, HW28714, HW28716, HW28718, HW28728, HW28729, HW2873, HW28731, HW28732, HW28735, HW28738, HW28739, HW28749, HW28762, HW28765, HW28769, HW28773, HW28775, HW28779, HW28785, HW28788, HW28793, HW28794, HW28797, HW28799, HW28803, HW28810, HW28813, HW28832, HW28836, HW28840, HW28841, HW28843, HW28871, HW28883, HW28889, HW28903, HW28904, HW28905, HW28906, HW28907, HW28909, HW28910, HW28914, HW28915, HW28919, HW28920, HW28929, HW28938, HW28939, HW28940, HW28941, HW28944, HW28945, HW28946, HW28950, HW28953, HW28954, HW28956, HW28969, HW28972, HW28974, HW28979, HW28983, HW28991, HW28995, HW29005, HW29012, HW29033, HW29034, HW29035, HW29040, HW29042, HW29045, HW29048, HW29050, HW29053, HW29055, HW29056, HW29058, HW29060, HW29062, HW29064, HW29066, HW29077, HW29082, HW29083, HW29085, HW29088, HW29091, HW29105, HW29108, HW29113, HW29115, HW29120, HW29121, HW29123, HW29125, HW29127, HW29141, HW29142, HW29143, HW29144, HW29145, HW29148, HW29150, HW29151, HW29152, HW29155, HW29156, HW29157, HW29162, HW29164, HW29165, HW29166, HW29171, HW29172, HW29175, HW29179, HW29180, HW29181, HW29182, HW29183, HW29185, HW29187, HW29188, HW29194, HW29198, HW29202, HW29215, HW29259, HW29260, HW29270, HW29272, HW29277, HW29280, HW29294, HW29295, HW29319, HW29324, HW29327, HW29328, HW29330, HW29331, HW29335, HW29358, HW29364, HW29372, HW29375, HW29376, HW29377, HW29378, HW29382, HW29383, HW29385, HW29386, HW29391, HW29397, HW29399, HW29400, HW29402, HW29403, HW29404, HW29405, HW29407, HW29408, HW29412, HW29417, HW29420, HW29436, HW29437, HW29445, HW29448, HW29449, HW29455, HW29457, HW29459, HW29488, HW29499, HW29500, HW29506, HW29508, HW29509, HW29511, HW29513, HW29514, HW29529, HW29544, HW29554, HW2956, HW29565, HW29584, HW29599, HW29606, HW29607, HW29609, HW29611, HW29613, HW29622, HW29623, HW29624, HW29625, HW29628, HW29630, HW29631, HW29632, HW29637, HW29643, HW29648, HW29651, HW29654, HW29695, HW29703, HW29704, HW29706, HW29707, HW29710, HW29718, HW29719, HW29722, HW29723, HW29725, HW29726, HW29730, HW29740, HW29742, HW29743, HW29745, HW29751, HW29753, HW29756, HW29766, HW29772, HW29773, HW29777, HW29778, HW29801, HW29807, HW29808, HW29809, HW29812, HW29813, HW29817, HW29821, HW29822, HW29826, HW29831, HW29835, HW29849, HW29852, HW29853, HW29854, HW29857, HW29860, HW29862, HW29868, HW29871, HW29874, HW29878, HW29882, HW29884, HW29891, HW29906, HW29924, HW29929, HW29930, HW29934, HW29937, HW29943, HW29944, HW29945, HW29946, HW2995, HW29952, HW29953, HW29955, HW29956, HW29959, HW29962, HW29969, HW29971, HW29973, HW30000, HW30001, HW30004, HW30005, HW30006, HW30008, HW30010, HW30012, HW30015, HW30016, HW30018, HW30019, HW30025, HW30026, HW30037, HW30041, HW30049, HW30050, HW30055, HW30057, HW30059, HW30061, HW30063, HW30065, HW30071, HW30075, HW30076, HW30077, HW3008, HW3009, HW30097, HW30122, HW30125, HW30126, HW30128, HW30131, HW30132, HW3016, HW30161, HW30162, HW30165, HW30171, HW30172, HW30173, HW30177, HW30179, HW30184, HW30185, HW30187, HW3019, HW30191, HW30192, HW30203, HW30206, HW3021, HW30215, HW30218, HW30219, HW30235, HW30236, HW30238, HW30241, HW30244, HW30246, HW30248, HW30249, HW30250, HW30257, HW30264, HW30273, HW30279, HW30281, HW30283, HW30287, HW30289, HW30290, HW30295, HW30296, HW30301, HW30302, HW30305, HW30307, HW30308, HW30310, HW30311, HW30318, HW30326, HW30327, HW30328, HW30329, HW30334, HW30335, HW30336, HW30339, HW30340, HW30343, HW30344, HW30346, HW30348, HW30349, HW30377, HW30379, HW30388, HW30403, HW30417, HW30418, HW30419, HW30420, HW30422, HW30426, HW30427, HW30429, HW30431, HW30432, HW30433, HW30434, HW30436, HW30438, HW30440, HW30441, HW30444, HW30448, HW30450, HW30451, HW30452, HW30457, HW30460, HW30465, HW30470, HW30476, HW30478, HW30481, HW30482, HW30487, HW30488, HW30492, HW30493, HW30495, HW30507, HW30527, HW30532, HW30534, HW30536, HW30539, HW30542, HW30544, HW30545, HW30547, HW30550, HW30552, HW30553, HW30554, HW30555, HW30556, HW30557, HW30563, HW30569, HW30570, HW30572, HW30574, HW30576, HW30577, HW30580, HW30581, HW30583, HW30584, HW30585, HW30587, HW30592, HW30612, HW30615, HW30617, HW30618, HW30620, HW30621, HW30622, HW30623, HW30624, HW30625, HW30626, HW30627, HW30629, HW30630, HW30636, HW30641, HW30644, HW30648, HW30653, HW30665, HW30668, HW30671, HW30672, HW30678, HW30681, HW30682, HW30683, HW30687, HW30688, HW30689, HW30690, HW30691, HW30693, HW30695, HW30698, HW30699, HW30701, HW30704, HW30709, HW30713, HW30714, HW30716, HW30718, HW30720, HW30724, HW30725, HW30727, HW30730, HW30732, HW30742, HW30744, HW30745, HW30749, HW30750, HW30751, HW30753, HW30754, HW30755, HW30756, HW30757, HW30758, HW30762, HW30772, HW30775, HW30777, HW30779, HW30781, HW30784, HW30785, HW30786, HW30788, HW30790, HW30792, HW30882, HW30884, HW30888, HW30889, HW30891, HW30892, HW30893, HW30896, HW30897.HW30898, HW30899, HW30902, HW30905, HW30906, HW30907, HW30908, HW30909, HW30912, HW30913, HW30915, HW30918, HW30920, HW30925, HW30926, HW30928, HW30931, HW30932, HW30933, HW30936, HW30938, HW30941, HW30942, HW30945, HW30947, HW30965, HW30975, HW30989, HW30991, HW30992, HW30994, HW30999, HW3100, HW31002, HW31005, HW31006, HW31011, HW31022, HW31023, HW31027, HW31028, HW31029, HW31030, HW31031, HW31033, HW31034, HW31041, HW31042, HW31043, HW31044, HW31047, HW31049, HW31050, HW31054, HW31055, HW31056, HW31057, HW31062, HW31066, HW31067, HW31068, HW31069, HW31073, HW31078, HW31097, HW31099, HW31102, HW31122, HW31131, HW31139, HW31142, HW31152, HW31159, HW31169, HW31174, HW31177, HW31181, HW31182, HW31186, HW31189, HW31190, HW31191, HW31193, HW31195, HW31203, HW31204, HW31207, HW31211, HW31212, HW31213, HW31215, HW31218, HW31220, HW31221, HW31224, HW31231, HW31241, HW31243, HW31245, HW31246, HW3126, HW31269, HW31270, HW31274, HW31278, HW31279, HW31280, HW31281, HW31289, HW31302, HW31306, HW31308, HW31311, HW31312, HW31313, HW31315, HW31316, HW31318, HW31320, HW31323, HW31324, HW31326, HW31327, HW31329, HW31330, HW31331, HW31333, HW31334, HW31336, HW31344, HW31345, HW31347, HW31349, HW31361, HW31363, HW31367, HW31369, HW31389, HW31396, HW31397, HW31401, HW31403, HW31405, HW31462, HW31464, HW31465, HW31466, HW31469, HW31470, HW31474, HW31475, HW31476, HW31477, HW31478, HW31479, HW31480, HW31484, HW31498, HW31529, HW31535, HW31536, HW31540, HW31542, HW31562, HW31567, HW31568, HW31575, HW31576, HW31591, HW31595, HW31599, HW31606, HW31608, HW31609, HW31611, HW31612, HW31613, HW31615, HW31617, HW31623, HW31624, HW31626, HW31634, HW31643, HW31647, HW31648, HW31649, HW31652, HW31653, HW31671, HW31674, HW31677, HW31678, HW31689, HW31690, HW31692, HW31695, HW31705, HW31709, HW31716, HW31719, HW31720, HW31730, HW31738, HW31740, HW31741, HW31747, HW31762, HW31764, HW31765, HW31775, HW31776, HW31782, HW31783, HW31785, HW31797, HW31798, HW31804, HW31807, HW31829, HW31862, HW31864, HW31866, HW31869, HW31870, HW31871, HW31880, HW31897, HW31898, HW31900, HW31914, HW31918, HW31923, HW31942, HW31951, HW31971, HW31990, HW32003, HW32005, HW32019, HW32020, HW32044, HW32045, HW32063, HW32069, HW32074, HW32083, HW32087, HW32089, HW32107, HW32117, HW32119, HW32126, HW32130, HW32136, HW32140, HW32145, HW32146, HW32151, HW32176, HW32178, HW32184, HW32188, HW32249, HW32251, HW32262, HW32267, HW32271, HW32280, HW32283, HW32284, HW32300, HW32312, HW32313, HW32318, HW32323, HW32325, HW32327, HW32332, HW32334, HW32339, HW32345, HW32346, HW32351, HW32352, HW32361, HW32362, HW32363, HW32370, HW32375, HW32376, HW32379, HW32380, HW32381, HW32382, HW32390, HW32395, HW32399, HW32405, HW32409, HW32417, HW32418, HW32422, HW32424, HW32425, HW32431, HW32438, HW32439, HW32442, HW32449, HW32451, HW32456, HW32463, HW32469, HW32480, HW32485, HW32486, HW32487, HW32491, HW32498, HW32500, HW32527, HW32532, HW32533, HW32535, HW32545, HW32550, HW3285, HW3298, HW3334, HW3336, HW3342, HW3349, HW3369, HW3385, HW34014, HW34016, HW34064, HW34067, HW34080, HW34081, HW34083, HW34105, HW34116, HW34118, HW34119, HW34127, HW34137, HW34143, HW34153, HW34155, HW34156, HW34158, HW34162, HW34163, HW34164, HW34166, HW34167, HW34168, HW34171, HW34174, HW34175, HW34183, HW34188, HW34189, HW34194, HW34195, HW34212, HW34216, HW34218, HW34219, HW34221, HW34230, HW34240, HW34242, HW34245, HW34258, HW34263, HW34264, HW34265, HW34282, HW34289, HW34298, HW34299, HW34304, HW34306, HW34311, HW34312, HW34314, HW34315, HW34321, HW34322, HW34323, HW34333, HW34353, HW34391, HW34394, HW34397, HW34399, HW34406, HW34407, HW34413, HW34416, HW34441, HW34451, HW34453, HW34463, HW34466, HW34471, HW34472, HW34473, HW34476, HW34477, HW34483, HW34485, HW34486, HW34492, HW34496, HW34499, HW34505, HW34506, HW34525, HW34532, HW34537, HW34545, HW34549, HW34562, HW34566, HW34572, HW34574, HW34575, HW34576, HW34580, HW34585, HW34589, HW34593, HW34599, HW34601, HW34602, HW34608, HW34613, HW34635, HW34645, HW34646, HW34650, HW34653, HW34654, HW34657, HW34658, HW34660, HW34662, HW34671, HW34676, HW34677, HW34679, HW34683, HW34689, HW34690, HW34699, HW34702, HW34709, HW34714, HW34715, HW34721, HW34728, HW34729, HW34735, HW34736, HW34741, HW34763, HW34770, HW34775, HW34781, HW34787, HW34790, HW34795, HW34798, HW34814, HW34823, HW34826, HW34828, HW34831, HW34832, HW34856, HW34860, HW34865, HW34872, HW34874, HW34883, HW34896, HW34921, HW34931, HW34937, HW34952, HW34956, HW34967, HW34970, HW34976, HW34984, HW34987, HW3499, HW34999, HW35003, HW35008, HW35010, HW35011, HW35014, HW35015, HW3502, HW35020, HW35026, HW35045, HW35048, HW35051, HW35058, HW35066, HW35069, HW35072, HW35076, HW35077, HW35082, HW35089, HW35098, HW35110, HW35111, HW35115, HW35116, HW3512, HW35126, HW35129, HW35130, HW35135, HW35166, HW35167, HW35168, HW35169, HW3517, HW35176, HW35180, HW35182, HW35183, HW35190, HW35192, HW35193, HW35195, HW35219, HW35221, HW35226, HW35232, HW35235, HW35236, HW35238, HW3524, HW35243, HW35249, HW3525, HW35255, HW35257, HW35263, HW35264, HW35273, HW35280, HW35287, HW35288, HW35293, HW35294, HW35295, HW35299, HW35403, HW35413, HW35415, HW35422, HW35425, HW35431, HW35433, HW35434, HW35436, HW35437, HW35438, HW35442, HW35448, HW35449, HW35450, HW35452, HW35456, HW35461, HW35462, HW35471, HW35473, HW3665, HW3684, HW3718, HW3765, HW3768, HW3770, HW3771, HW3772, HW3775, HW3785, HW3788, HW3789, HW3795, HW3803, HW3813, HW3825, HW3829, HW3842, HW3845, HW3847, HW3849, HW3855, HW3860, HW3862, HW3865, HW3877, HW3882, HW3892, HW3899, HW3914, HW3918, HW3919, HW3925, HW3930, HW3934, HW3944, HW3946, HW3948, HW3952, HW3953, HW3960, HW3963, HW3964, HW3967, HW3970, HW3971, HW3973, HW3975, HW3981, HW3984, HW3995, HW40002, HW40004, HW40006, HW40009, HW40013, HW40017, HW40025, HW4003, HW40033, HW40037, HW40045, HW40046, HW40048, HW40049, HW40051, HW40054, HW40061, HW40067, HW40068, HW40073, HW40074, HW40075, HW40079, HW40085, HW40087, HW40089, HW40090, HW40094, HW40097, HW40099, HW40100, HW40109, HW40112, HW40120, HW40121, HW40122, HW40138, HW40139, HW40142, HW40143, HW40145, HW40147, HW40149, HW40152, HW40153, HW40154, HW40157, HW40159, HW40166, HW40168, HW40169, HW40176, HW40177, HW40178, HW4018, HW40186, HW40189, HW40190, HW40192, HW40197, HW40198, HW40199, HW4020, HW40200, HW40201, HW40202, HW40206, HW40209, HW4021, HW40210, HW40211, HW40212, HW40223, HW40226, HW40230, HW40232, HW40238, HW40240, HW40244, HW40245, HW40247, HW40255, HW40258, HW40259, HW40260, HW40262, HW40263, HW40270, HW40271, HW40272, HW40273, HW40279, HW40281, HW40283, HW40288, HW40291, HW40292, HW40293, HW40294, HW40295, HW40298, HW40299, HW40301, HW40302, HW40303, HW40305, HW40307, HW40317, HW40322, HW40324, HW40326, HW40329, HW40334, HW40337, HW40340, HW4035, HW40359, HW40371, HW40373, HW40376, HW4038, HW40381, HW40382, HW40384, HW40389, HW40390, HW40391, HW40392, HW40395, HW40397, HW40399, HW4040, HW40402, HW40404, HW40405, HW40433, HW40436, HW40437, HW40440, HW40441, HW40443, HW40444, HW40445, HW40448, HW40451, HW40453, HW40455, HW40460, HW40466, HW40481, HW40482, HW40483, HW40488, HW40489, HW40490, HW40491, HW40499, HW40501, HW40503, HW40504, HW40509, HW40510, HW40512, HW40515, HW40517, HW40518, HW40520, HW40523, HW40524, HW40527, HW40528, HW40529, HW40530, HW40533, HW40534, HW40537, HW40546, HW40558, HW40563, HW40567, HW40568, HW40569, HW40572, HW40576, HW40583, HW40585, HW4059, HW40590, HW40592, HW40593, HW40595, HW40601, HW40605, HW40606, HW40616, HW40620, HW40621, HW40622, HW40627, HW40631, HW40633, HW40635, HW40636, HW40643, HW40644, HW40646, HW40647, HW40672, HW40678, HW40682, HW40683, HW40684, HW40687, HW40690, HW40694, HW40696, HW40700, HW40703, HW40705, HW40708, HW40710, HW40715, HW40717, HW40718, HW40719, HW40723, HW40728, HW40732, HW40734, HW40737, HW40738, HW40739, HW40744, HW40745, HW40746, HW40749, HW40750, HW40753, HW40755, HW40757, HW40759, HW40762, HW40764, HW40766, HW40767, HW40768, HW40769, HW40772, HW40776, HW40777, HW40778, HW40780, HW40783, HW40794, HW4080, HW40800, HW40801, HW40805, HW40807, HW40808, HW40809, HW40810, HW40813, HW40815, HW40819, HW40821, HW40822, HW40829, HW40831, HW40832, HW40834, HW40838, HW40841, HW40857, HW40861, HW40862, HW40863, HW40865, HW40868, HW40870, HW40874, HW40875, HW40876, HW40882, HW40883, HW40884, HW40887, HW40891, HW40892, HW40894, HW40897, HW40902, HW40903, HW40905, HW40907, HW40908, HW40913, HW40916, HW40917, HW40918, HW40922, HW40924, HW40925, HW40927, HW40928, HW40930, HW40935, HW40939, HW40941, HW40944, HW40948, HW40962, HW40965, HW40967, HW40968, HW4097, HW40976, HW40977, HW40986, HW40987, HW40988, HW40991, HW40992, HW40995, HW40996, HW40998, HW41001, HW41005, HW41006, HW41007, HW41008, HW41012, HW41017, HW41022, HW41024, HW41025, HW41032, HW41035, HW41038, HW41042, HW41054, HW41057, HW41058, HW41059, HW41060, HW41064, HW41066, HW41067, HW41069, HW41072, HW41073, HW41076, HW41077, HW4108, HW41081, HW41084, HW41091, HW41092, HW41093, HW41097, HW41098, HW41100, HW41102, HW41104, HW41105, HW41107, HW41111, HW41112, HW41115, HW4112, HW41124, HW41131, HW41132, HW41137, HW41138, HW41141, HW41146, HW41148, HW41154, HW41157, HW41158, HW41165, HW41167, HW41172, HW41173, HW41176, HW41179, HW41180, HW41182, HW41185, HW41186, HW41188, HW41194, HW41195, HW41207, HW41208, HW41212, HW41224, HW41225, HW41227, HW41228, HW41229, HW41232, HW41234, HW41237, HW41238, HW41240, HW41247, HW41248, HW41250, HW41275, HW41276, HW41278, HW41280, HW41283, HW41288, HW41289, HW41291, HW41295, HW41298, HW41300, HW41310, HW41318, HW41321, HW41327, HW41330, HW41331, HW41334, HW41338, HW41339, HW4134, HW41341, HW41345, HW41351, HW41360, HW41361, HW41369, HW41378, HW41381, HW41382, HW41385, HW41388, HW41389, HW41393, HW41394, HW41396, HW41400, HW41404, HW41410, HW41412, HW41418, HW41419, HW41421, HW41423, HW41424, HW41425, HW41427, HW41431, HW41432, HW41435, HW41438, HW41441, HW41443, HW41444, HW41446, HW41451, HW41453, HW41456, HW41457, HW41459, HW41460, HW41461, HW41462, HW41463, HW41464, HW41468, HW41470, HW41478, HW41479, HW41480, HW41481, HW41489, HW41490, HW41492, HW41493, HW41497, HW41503, HW41504, HW41507, HW41508, HW41509, HW41511, HW41514, HW41517, HW41519, HW4152, HW41522, HW41524, HW41525, HW41526, HW41529, HW41531, HW41537, HW41538, HW41541, HW41548, HW41551, HW41556, HW41573, HW41577, HW41578, HW41584, HW41585, HW41586, HW41588, HW41589, HW41590, HW41591, HW41594, HW41596, HW41598, HW41603, HW41605, HW41606, HW41608, HW41614, HW41616, HW41618, HW41626, HW41627, HW41630, HW41631, HW41635, HW41636, HW4164, HW41642, HW41650, HW41657, HW41659, HW41662, HW41664, HW41666, HW41668, HW41678, HW41679, HW41680, HW41681, HW41686, HW41688, HW4169, HW41692, HW41702, HW41703, HW41707, HW41708, HW41711, HW41712, HW41719, HW41729, HW41730, HW41731, HW41733, HW41735, HW41747, HW41748, HW41749, HW41756, HW41772, HW41787, HW41789, HW41795, HW41797, HW41807, HW41812, HW41814, HW41820, HW41821, HW41822, HW41823, HW41836, HW41838, HW41841, HW41842, HW41844, HW41861, HW41866, HW41867, HW41873, HW41880, HW41884, HW4190, HW4192, HW4193, HW4201, HW4207, HW42121, HW42125, HW42129, HW42134, HW42139, HW42142, HW42144, HW42146, HW42147, HW42210, HW42220, HW42222, HW42242, HW4226, HW42260, HW42277, HW42279, HW42281, HW42283, HW4231, HW42317, HW4232, HW42323, HW42324, HW42338, HW42343, HW42389, HW4247, HW4254, HW4255, HW4261, HW4272, HW42727, HW42736, HW42765, HW42768, HW42804, HW42810, HW42819, HW42824, HW4307, HW4308, HW4329, HW4339, HW4340, HW4343, HW4347, HW4353, HW4355, HW4362, HW4365, HW4366, HW4370, HW4377, HW4378, HW4383, HW4401, HW4402, HW4415, HW4425, HW4436, HW4453, HW4456, HW4458, HW4462, HW4492, HW4493, HW28725, HW28807, HW28961, HW29336, HW29337, HW29338, HW29340, HW29398, HW29443, HW29497, HW29498, HW29501, HW29530, HW29531, HW29598, HW29626, HW29627, HW29634, HW29636, HW29662, HW29663, hw29709, HW29716, HW29720, HW29735, HW29739, HW29799, HW29806, HW29810, HW29818, HW29819, HW29823, hw29828, HW29829, HW29855, HW29856, HW29858, HW29861, HW29863, HW29865, HW29872, HW29879, HW29899, HW29905, HW29908.HW29910, HW29913, HW29914, HW29925, HW29935, HW29938, HW29941, HW29966, HW29977, HW29979, HW29981, HW29984, HW29987, HW30009, HW30014, HW30017, HW30021, HW30023, HW30052, HW30056, HW30058, HW30064, HW30078, HW30119, HW30129, HW30130, HW30160, HW30220, HW30230, HW30240, HW30242, HW30258, HW30259, HW30260, HW30261, HW30262, HW30263, HW30265, HW30266, HW30268, HW30269, HW30270, HW30271, HW30272, HW30274, HW30276, HW30277, HW30280, HW30284, HW30286, HW30288, HW30291, HW30294, HW30299, HW30309, HW30331, HW30347, HW30435, HW30464, HW30480, HW30489, HW30496, HW30546, HW30610, HW30673, HW30686, HW30710, HW30752, HW30922, HW30923, hw30943, hw30944, hw30946, hw30998, HW31118, hw31210, hw31236, HW31300, HW31459, HW31460, HW31490, HW31501, HW31528, HW10991, HW11017, HW11020, HW11022, HW11023, HW11026, HW11030, HW11032, HW11034, HW11035, HW11037, HW11038, HW11042, HW11044, HW11050, HW11051, HW11053, HW11054, HW11057, HW11062, HW11063, HW11071, HW11072, HW11077, HW11082, HW11085, HW11087, HW11089, HW11091, HW11093, HW11096, HW11100, HW11101, HW11111, HW11114, HW11117, HW11118, HW11119, HW11121, HW11122, HW11133, HW11134, HW11137, HW11138, HW11139, HW11142, HW11147, HW11149, HW11151, HW11152, HW11160, HW11164, HW11165, HW11166, HW11176, HW11177, HW11178, HW11180, HW11184, HW11185, HW11186, HW11190, HW11191, HW11192, HW11194, HW11195, HW11196, HW11197, HW11198, HW11199, HW11202, HW11204, HW11205, HW11217, HW11218, HW11219, HW11221, HW11223, HW11227, HW11228, HW11229, HW11230, HW11231, HW11233, HW11234, HW11235, HW11236, HW11239, HW11240, HW11241, HW11242, HW11246, HW11247, HW11248, HW11250, HW11251, HW11252, HW11255, HW11256, HW11257, HW11258, HW11259, HW11260, HW11262, HW11264, HW11266, HW11267, HW11274, HW11276, HW11279, HW11281, HW11283, HW11286, HW11287, HW11288, HW11289, HW11290, HW11291, HW11292, HW11294, HW11295, hw11296, HW11298, HW11302, HW11303, HW11305, HW11306, HW11307, HW11309, HW11310, HW11311, HW11312, HW11314, HW11316, HW11317, HW11321, HW11324, HW11326, HW11328, HW11330, HW11331, HW11333, HW11336, HW11337, HW11342, HW11343, HW11348, HW11350, HW11351, HW11355, HW11359, HW11360, HW11365, HW11366, HW11369, HW11370, HW11372, HW11373, HW11374, HW11378, HW11379, HW11380, HW11387, HW11391, HW11392, HW11393, HW11394, HW11398, HW11401, HW11402, HW11403, HW11404, HW11410, HW11411, HW11412, HW11419, HW11422, HW11424, HW11426, HW11430, HW11433, HW11434, HW11437, HW11446, HW11448, HW11454, HW11456, HW11459, HW11460, HW11465, HW11466, HW11468, HW11470, HW11473, HW11474, HW11477, HW11485, HW11487, HW11489, HW11497, HW11498, HW11500, HW11503, HW11505, HW11506, HW11509, HW11517, HW11518, HW11519, HW11521, HW11524, HW11526, HW11527, HW11528, HW11534, HW11540, HW11542, HW11543, HW11545, HW11547, HW11548, HW11549, HW11554, HW11555, HW11563, HW11572, HW11576, HW11577, HW11579, HW11580, HW11581, HW11582, HW11585, HW11588, HW11589, HW11593, HW11595, HW11596, HW11597, HW11599, HW11601, HW11602, HW11604, HW11605, HW11606, HW11612, HW11613, HW11614, HW11619, HW11620, HW11626, HW11627, HW11628, HW11632, HW11633, HW11634, HW11637, HW11641, HW11643, HW11645, HW11651, HW11653, HW11659, HW11660, HW11662, HW11665, HW11666, HW11668, HW11672, HW11677, HW11681, HW11682, HW11683, HW11684, HW11688, HW11689, HW11694, HW11701, HW11702, HW11705, HW11708, HW11709, HW11711, HW11716, HW11717, HW11719, HW11721, HW11725, HW11728, HW11729, HW11737, HW11742, HW11743, HW11744, HW11745, HW11746, HW11749, HW11750, HW11751, HW11752, HW11757, HW11760, HW11762, HW11766, HW11768, HW11772, HW11773, HW11776, HW11777, HW11778, HW11779, HW11783, HW11788, HW11789, HW11790, HW11796, HW11797, HW11798, HW11799, HW11801, HW11804, HW11805, HW11807, HW11809, HW11811, HW11813, HW11814, HW11816, HW11817, HW11820, HW11824, HW11825, HW11827, HW11828, HW11836, HW11848, HW11855, HW11858, HW11859, HW11865, HW11866, HW11870, HW11873, HW11875, HW11876, HW11877, HW11879, HW11880, HW11882, HW11884, HW11887, HW11888, HW11889, HW11891, HW11892, HW11894, HW11895, HW11896, HW11897, HW11898, HW11901, HW11908, HW11911, HW11912, HW11913, HW11916, HW11917, HW11919, HW11920, HW11921, HW11922, HW11923, HW11924, HW11925, HW11926, HW11928, HW11930, HW11931, HW11934, HW11935, HW11939, HW11941, HW11946, HW11947, HW11951, HW11953, HW11955, HW11956, HW11961, HW11963, HW11964, HW11965, HW11968, HW11970, HW11975, HW11976, HW11980, HW11981, HW11985, HW11986, HW11987, HW11988, HW11989, HW11991, HW11994, HW11995, HW11996, HW11997, HW11998, HW12000, HW12001, HW12002, HW12005, HW12006, HW12007, HW12008, HW12009, HW12010, HW12011, HW12012, HW12014, HW12015, HW12016, HW12018, HW12023, HW12025, HW12027, HW12029, HW12030, HW12031, HW12033, HW12036, HW12038, HW12039, HW12040, HW12041, HW12042, HW12043, HW12046, HW12047, HW12048, HW12049, HW12050, HW12053, HW12054, HW12056, HW12057, HW12058, HW12060, HW12064, HW12066, HW12076, HW12077, HW12078, HW12079, HW12081, HW12082, HW12084, HW12085, HW12087, HW12089, HW12091, HW12093, HW12094, HW12096, HW12098, HW12099, HW12101, HW12105, HW12108, HW12110, HW12111, HW12114, HW12124, HW12126, HW12128, HW12129, HW12136, HW12138, HW12140, HW12141, HW12145, HW12146, HW12147, HW12150, HW12151, HW12153, HW12156, HW12159, HW12161, HW12162, HW12165, HW12167, HW12169, HW12171, HW12176, HW12184, HW12186, HW12188, HW12190, HW12191, HW12192, HW12197, HW12200, HW12204, HW12206, HW12208, HW12209, HW12214, HW12217, HW12218, HW12220, HW12226, HW12228, HW12229, HW12234, HW12235, HW12236, HW12237, HW12238, HW12239, HW12243, HW12246, HW12247, HW12252, HW12253, HW12254, HW12256, HW12257, HW12258, HW12261, HW12266, HW12270, HW12276, HW12277, HW12280, HW12282, HW12284, HW12285, HW12288, HW12289, HW12292, HW12293, HW12294, HW12295, HW12297, HW12298, HW12300, HW12301, HW12302, HW12303, HW12304, HW12305, HW12306, HW12308, HW12310, HW12312, HW12314, HW12320, HW12323, HW12325, HW12326, HW12327, HW12328, HW12330, HW12331, HW12333, HW12334, HW12335, HW12337, HW12338, HW12339, HW12340, HW12342, HW12343, HW12344, HW12346, HW12351, HW12353, HW12354, HW12357, HW12361, HW12362, HW12363, HW12364, HW12366, HW12368, HW12371, HW12374, HW12376, HW12377, HW12378, HW12382, HW12386, HW12387, HW12390, HW12393, HW12395, HW12397, HW12398, HW12403, HW12409, HW1904, HW2001, HW2005, HW2010, HW2011, HW2016, HW2050, HW2063, HW2072, HW2074, HW2077, HW2121, HW2123, HW2131, HW2133, HW21447, HW2150, HW2151, HW21532, HW21533, HW21534, HW21535, HW2154, HW21544, HW21548, HW21549, HW21550, HW21553, HW21560, HW21564, HW21565, HW21567, HW21573, HW21581, HW21582, HW21589, HW2160, HW21608, HW21610, HW21611, HW21615, HW21617, HW21618, HW21619, HW21627, HW21631, HW21632, HW21633, HW21634, HW21662, HW21685, HW21687, HW21694, HW21706, HW21708, HW21712, HW21713, HW21716, HW21717, HW21718, HW21720, HW21730, HW21737, HW21738, HW21741, HW21744, HW21948, HW21949, HW21954, HW21956, HW21957, HW21960, HW21961, HW21965, HW21966, HW21967, HW21969, HW21970, HW21971, HW21976, HW21977, HW21979, HW21983, HW21984, HW21988, HW21989, HW21990, HW21991, HW21993, HW21994, HW21995, HW21996, HW21997, HW22001, HW22004, HW22006, HW22008, HW22010, HW22012, HW22014, HW22019, HW22020, HW22021, HW22022, HW22023, HW22026, HW22027, HW22028, HW2203, HW22031, HW22033, HW22034, HW22041, HW22047, HW22048, HW22049, HW22053, HW22054, HW22055, HW22057, HW22059, HW22061, HW22063, HW22064, HW22070, HW22071, HW22075, HW22079, HW22080, HW22085, HW22087, HW22090, HW22093, HW22098, HW2210, HW22100, HW22102, HW22103, HW22108, HW22109, HW2211, HW22110, HW22111, HW22118, HW22119, HW22120, HW22125, HW22126, HW22127, HW22129, HW22131, HW22132, HW22138, HW22141, HW22142, HW22143, HW22146, HW22153, HW22158, HW22160, HW22161, HW22162, HW22163, HW22164, HW22165, HW22166, HW22167, HW22169, HW22170, HW22172, HW22173, HW22174, HW22202, HW22207, HW22210, HW22214, HW22215, HW22216, HW22217, HW22220, HW22223, HW22224, HW22228, HW2225, HW22262, HW22263, HW22266, HW22268, HW22270, HW22271, HW22272, HW22273, HW22278, HW22279, HW22283, HW22284, HW22290, HW22292, HW22318, HW22319, HW22340, HW22342, HW22343, HW22345, HW22350, HW22352, HW22353, HW22359, HW2240, HW22405, HW22409, HW2241, HW22428, HW22449, HW22454, HW22456, HW22463, HW22467, HW22476, HW22478, HW22481, HW22482, HW22485, HW22488, HW22489, HW22491, HW22492, HW22514, HW22528, HW22531, HW22535, HW22536, HW22538, HW22539, HW22540, HW22546, HW22547, HW22552, HW22553, HW22554, HW22556, HW22557, HW22558, HW22559, HW22560, HW22561, HW22562, HW22563, HW22566, HW22567, HW22568, HW22569, HW22570, HW22571, HW22574, HW22575, HW22576, HW22577, HW22578, HW22579, HW22584, HW22585, HW22606, HW22610, HW22611, HW22613, HW22615, HW22617, HW22619, HW22621, HW22622, HW22623, HW22637, HW22639, HW2264, HW22643, HW2265, HW22653, HW22655, HW22660, HW22662, HW22669, HW22673, HW22675, HW2268, HW2269, HW22704, HW22711, HW22716, HW22728, HW22732, HW22733, HW22734, HW22736, HW22737, HW22738, HW22743, HW22748, HW22749, HW22751, HW22752, HW22754, HW22755, HW22759, HW22760, HW22768, HW22769, HW22771, HW22796, HW22800, HW22803, HW22807, HW2281, HW22813, HW22818, HW22821, HW22873, HW22874, HW22875, HW22877, HW22878, HW22880, HW22882, HW22883, HW22887, HW22894, HW22895, HW22896, HW22897, HW22898, HW22924, HW22926, HW22933, HW22936, HW22951, HW22980, HW22981, HW22982, HW22984, HW22986, HW22989, HW22990, HW22994, HW22995, HW22999, HW23000, HW23002, HW23005, HW23008, HW23010, HW23014, HW23063, HW23065, HW23066, HW23069, HW23072, HW23073, HW23076, HW23079, HW23082, HW23087, HW23092, HW23096, HW23098, HW23099, HW23100, HW23102, HW23103, HW23108, HW23113, HW23118, HW23120, HW23121, HW23122, HW23123, HW23139, HW23148, HW23152, HW23153, HW23154, HW23163, HW23164, HW23165, HW23173, HW23175, HW23178, HW23181, HW23182, HW23185, HW2319, HW23193, HW2322, HW23221, HW23225, HW23227, HW23233, HW23234, HW23246, HW23247, HW23251, HW23294, HW23299, HW23300, HW23308, HW23313, HW23314, HW23316, HW23318, HW23319, HW23383, HW23385, HW23391, HW23392, HW23393, HW23398, HW23399, HW23407, HW23412, HW23421, HW23422, HW23423, HW23425, HW23426, HW23427, HW23428, HW23430, HW23431, HW23432, HW23436, HW23437, HW23448, HW23461, HW23462, HW23463, HW23468, HW23469, HW23474, HW23476, HW23478, HW23480, HW23482, HW23487, HW23491, HW23493, HW23497, HW23499, HW23502, HW23504, HW2351, HW23532, HW23533, HW23535, HW23536, HW23540, HW23545, HW23555, HW23557, HW23559, HW23563, HW23567, HW23570, HW23602, HW23603, HW23605, HW2361, HW23621, HW23623, HW23627, HW23629, HW23630, HW23635, HW23637, HW23641, HW23675, HW23685, HW23686, HW23689, HW23694, HW23698, HW23723, HW23730, HW23732, HW23755, HW23757, HW23758, HW23762, HW23770, HW23773, HW23775, HW23777, HW23779, HW23782, HW23784, HW23786, HW23790, HW23792, HW23799, HW23802, HW23804, HW23808, HW23809, HW23810, HW23812, HW23813, HW23817, HW23823, HW23827, HW23831, HW23836, HW23845, HW23846, HW23847, HW23851, HW23853, HW23854, HW23857, HW23858, HW23861, HW23879, HW23880, HW23881, HW23882, HW23884, HW23885, HW23886, HW23887, HW23888, HW23889, HW23893, HW23897, HW23900, HW23901, HW23904, HW23908, HW23909, HW23933, HW23939, HW23941, HW23943, HW23946, HW23951, HW23952, HW23954, HW23956, HW23962, HW23992, HW23993, HW23999, HW24002, HW24005, HW24006, HW24010, HW24012, HW24049, HW24052, HW24054, HW24058, HW24059, HW24064, HW24067, HW24072, HW24073, HW24075, HW24076, HW24077, HW24105, HW24106, HW24107, HW24109, HW24127, HW24131, HW24132, HW24134, HW24144, HW2415, HW24197, HW24200, HW24201, HW24202, HW24203, HW24204, HW24205, HW24206, HW24207, HW24208, HW24212, HW24214, HW24215, HW24218, HW24244, HW24248, HW24250, HW24253, HW24256, HW24262, HW24294, HW24308, HW24309, HW24310, HW24312, HW24315, HW24336, HW24337, HW2434, HW24341, HW24342, HW24344, HW24347, HW24348, HW24349, HW24352, HW2444, HW2446, HW2447, HW24496, HW24501, HW24503, HW24508, HW24514, HW24515, HW24517, HW24519, HW24520, HW24521, HW24523, HW24530, HW24531, HW24532, HW24535, HW24594, HW24599, HW24602, HW24603, HW24608, HW24611, HW24612, HW24613, HW2466, HW24678, HW24686, HW24690, HW24700, HW24730, HW24732, HW24734, HW24735, HW24737, HW24740, HW24742, HW24743, HW24745, HW24746, HW24750, HW24752, HW24764, HW24765, HW24766, HW24767, HW24770, HW24773, HW24775, HW24776, HW24779, HW24781, HW24785, HW24806, HW24814, HW24819, HW24823, HW24825, HW24827, HW24829, HW24830, HW24853, HW24855, HW24858, HW24860, HW24861, HW24863, HW24864, HW24869, HW24880, HW24881, HW24884, HW24885, HW24887, HW24888, HW24893, HW24899, HW24903, HW24914, HW24915, HW24918, HW24920, HW24921, HW24923, HW24949, HW24950, HW24960, HW24962, HW2497, HW24977, HW24979, HW24982, HW24993, HW25000, HW25006, HW25009, HW25010, HW25012, HW25017, HW2502, HW25022, HW25023, HW25025, HW25026, HW25028, HW25031, HW25033, HW25035, HW2504, HW25047, HW25048, HW25050, HW25052, HW25053, HW25056, HW25058, HW25060, HW25064, HW25066, HW25068, HW25069, HW25070, HW25071, HW25072, HW25073, HW25075, HW25077, HW25078, HW2508, HW25098, HW25119, HW25120, HW25130, HW25132, HW25135, HW25136, HW25151, HW25164, HW25171, HW25179, HW25181, HW25184, HW25187, HW25192, HW25194, HW25199, HW25200, HW25203, HW25205, HW25206, HW25207, HW25213, HW25216, HW25217, HW25219, HW25220, HW25223, HW25224, HW25226, HW25227, HW25229, HW25231, HW25239, HW25240, HW25242, HW25253, HW25254, HW25264, HW25268, HW25288, HW25316, HW25318, HW25320, HW25322, HW25323, HW25324, HW25327, HW25330, HW25333, HW25349, HW25351, HW25403, HW25408, HW25410, HW25414, HW25416, HW25417, HW25422, HW25426, HW25430, HW25432, HW25435, HW25436, HW25437, HW25438, HW25440, HW25441, HW25444, HW25445, HW25446, HW25456, HW25459, HW2546, HW25463, HW25479, HW25480, HW25484, HW25496, HW25503, HW25504, HW25506, HW25511, HW25532, HW25645, HW25646, HW25647, HW25648, HW25649, HW25650, HW25652, HW25654, HW25656, HW25658, HW25659, HW25660, HW25663, HW25713, HW25716, HW25719, HW25722, HW25726, HW25756, HW25777, HW25781, HW25784, HW25810, HW25824, HW25838, HW25839, HW25858, HW25860, HW25866, HW25870, HW25894, HW25912, HW25913, HW25915, HW25936, HW25939, HW25945, HW25948, HW25952, HW25953, HW25957, HW25959, HW25963, HW25965, HW25967, HW25975, HW25976, HW25979, HW25986, HW25991, HW25992, HW25995, HW25998, HW25999, HW26005, HW26009, HW26010, HW26013, HW26014, HW26016, HW26017, HW26033, HW26036, HW26057, HW26063, HW26064, HW26065, HW26068, HW26071, HW26075, HW26083, HW26085, HW26091, HW26092, HW26127, HW26130, HW26137, HW26141, HW26144, HW26154, HW26175, HW26178, HW26182, HW26184, HW26185, HW26187, HW26188, HW26191, HW26192, HW26247, HW2625, HW2630, HW26300, HW26311, HW26312, HW26313, HW26314, HW26316, HW26317, HW26318, HW26319, HW26320, HW26322, HW26323, HW26324, HW26326, HW26327, HW26328, HW2633, HW26343, HW26345, HW26347, HW2635, HW26352, HW26355, HW26358, HW26360, HW26367, HW26370, HW26372, HW26375, HW26377, HW2641, HW26410, HW26413, HW26417, HW26420, HW26424, HW26425, HW26431, HW26434, HW2646, HW26460, HW26461, HW26465, HW26466, HW26479, HW2648, HW26483, HW26488, HW26489, HW26508, HW26516, HW26521, HW26522, HW26523, HW26524, HW26526, HW26534, HW26539, HW26546, HW26548, HW26576, HW26577, HW26581, HW26585, HW26586, HW26589, HW26592, HW26593, HW26594, HW26599, HW26601, HW26603, HW26606, HW26629, HW26631, HW26632, HW26639, HW26641, HW26643, HW26645, HW26646, HW26654, HW26672, HW26673, HW26674, HW26677, HW26679, HW26685, HW26693, HW26705, HW26716, HW26717, HW26725, HW26727, HW26737, HW26738, HW26756, HW26758, HW26764, HW26771, HW26772, HW26776, HW26782, HW26785, HW26786, HW26804, HW26809, HW26814, HW26822, HW26824, HW26830, HW26837, HW26842, HW26867, HW26879, HW26894, HW26896, HW26902, HW26904, HW26910, HW26915, HW26916, HW26917, HW26920, HW26921, HW26923, HW26924, HW26928, HW26929, HW26930, HW26934, HW26945, HW26954, HW26957, HW26963, HW26970, HW26973, HW26986, HW26989, HW26992, HW26994, HW27006, HW27007, HW27013, HW27016, HW27020, HW27022, HW2704, HW27058, HW2706, HW27085, HW27086, HW27087, HW27089, HW27091, HW27092, HW27097, HW27100, HW27102, HW27103, HW27149, HW27152, HW27160, HW2718, HW27237, HW27245, HW27248, HW27252, HW27253, HW27256, HW27258, HW27259, HW27260, HW27264, HW27275, HW27277, HW27285, HW27286, HW27289, HW27291, HW27292, HW27295, HW27304, HW27312, HW27313, HW27315, HW27321, HW27325, HW27331, HW27334, HW27339, HW27341, HW27347, HW27365, HW27373, HW27378, HW27380, HW27394, HW2740, HW2742, HW27420, HW27428, HW27429, HW27430, HW27431, HW27432, HW27437, HW27441, HW27457, HW2746, HW27462, HW27465, HW27470, HW27471, HW27494, HW27498, HW27502, HW27503, HW27505, HW27507, HW27511, HW27513, HW27516, HW27524, HW27525, HW27526, HW27539, HW27549, HW27552, HW27555, HW27556, HW27559, HW27585, HW27597, HW27603, HW27606, HW27616, HW2762, HW27622, HW27627, HW27633, HW27637, HW27643, HW27648, HW27655, HW27656, HW27658, HW2769, HW27702, HW27709, HW27710, HW27712, HW27716, HW27724, HW27760, HW27764, HW27769, HW27795, HW27798, HW27810, HW27813, HW27815, HW27816, HW27818, HW27819, HW27820, HW27821, HW27822, HW27827, HW27831, HW27844, HW27845, HW27889, HW27891, HW27892, HW27919, HW27920, HW27921, HW27933, HW27941, HW27948, HW27952, HW28017, HW28019, HW2804, HW28040, HW2805, HW28096, HW28098, HW28099, HW28100, HW28101, HW28102, HW28103, HW28104, HW28105, HW28108, HW28111, HW28116, HW28123, HW28124, HW28129, HW28131, HW28141, HW28143, HW28171, HW28172, HW28174, HW28183, HW28186, HW28187, HW28191, HW28226, HW28241, HW28246, HW28249, HW28264, HW28268, HW28281, HW28295, HW28302, HW28320, HW28326, HW28329, HW28333, HW28334, HW28357, HW28359, HW28361, HW28368, HW28392, HW28393, HW28395, HW2840, HW28400, HW28409, HW28424, HW28433, HW28453, HW28464, HW28476, HW28492, HW28498, HW28499, HW28504, HW28516, HW28517, HW28518, HW28522, HW28525, HW28565, HW28593, HW28595, HW28596, HW28597, HW28604, HW28613, HW28647, HW28651, HW28657, HW28663, HW28684, HW2869, HW28700, HW28712, HW28713, HW28715, HW28719, HW28721, HW28742, HW28746, HW28758, HW28768, HW28771, HW28781, HW28786, HW28795, HW28812, HW28869, HW28872, HW28877, HW28878, HW28881, HW28898, HW28899, HW28902, HW28908, HW28936, HW28937, HW28948, HW28949, HW28952, HW28970, HW28976, HW28985, HW28989, HW28996, HW28997, HW29010, HW29030, HW29031, HW29032, HW29036, HW29037, HW29039, HW29043, HW29052, HW29079, HW29081, HW29106, HW29117, HW29122, HW29126, HW29137, HW29139, HW29149, HW29153, HW29190, HW29191, HW29193, HW29200, HW29204, HW29261, HW29262, HW29263, HW29269, HW29291, HW29316, HW29323, HW29326, HW29332, HW29334, HW29341, HW29342, HW29343, HW29345, HW29346, HW29348, HW29349, HW29350, HW29379, HW29389, HW29394, HW29395, HW29401, HW29413, HW29431, HW29467, HW29468, HW29469, HW29479, HW29481, HW29484, HW29489, HW29490, HW29491, HW29492, HW29493, HW29494, HW29496, HW29504, HW29510, HW29545, HW29557, HW2960, HW2963, HW2964, HW29641, HW2967, HW29681, HW2969, HW29775, HW2987, HW2990, HW2992, HW29954, HW29968, HW2997, HW2998, HW29980, HW29986, HW3000, HW30027, HW30029, HW3003, HW30030, HW30031, HW30032, HW30034, HW3005, HW3013, HW3014, HW3023, HW3025, HW3027, HW3035, HW3036, HW3037, HW3040, HW3072, HW3075, HW3076, HW3078, HW3080, HW3081, HW3082, HW3083, HW3086, HW3104, HW3276, HW3277, HW3278, HW3281, HW3287, HW3290, HW3293, HW3294, HW3295, HW3296, HW3297, HW3299, HW3300, HW3303, HW3304, HW3305, HW3306, HW3309, HW3310, HW3311, HW3313, HW3315, HW3316, HW3318, HW3319, HW3322, HW3323, HW3326, HW3328, HW3329, HW3332, HW3337, HW3341, HW3343, HW3346, HW3350, HW3352, HW3353, HW3355, HW3356, HW3361, HW3364, HW3365, HW3368, HW3370, HW3371, HW3372, HW3373, HW3375, HW3376, HW3377, HW3380, HW3382, HW3392, HW3394, HW3455, HW3456, HW3458, HW3459, HW3460, HW3461, HW3462, HW3463, HW3464, HW3465, HW3469, HW3471, HW3500, HW3501, HW3504, HW3505, HW3506, HW3508, HW3509, HW3511, HW3515, HW3516, HW3518, HW3519, HW3522, HW3523, HW3529, HW3532, HW3536, HW3537, HW3538, HW3539, HW3540, HW3542, HW3543, HW3545, HW3546, HW3547, HW3552, HW3554, HW3557, HW3560, HW3563, HW3667, HW3672, HW3674, HW3676, HW3678, HW3679, HW3681, HW3682, HW3683, HW3685, HW3687, HW3688, HW3690, HW3692, HW3694, HW3695, HW3696, HW3697, HW3698, HW3699, HW3700, HW3701, HW3702, HW3703, HW3705, HW3706, HW3707, HW3710, HW3712, HW3713, HW3715, HW3716, HW3719, HW3722, HW3726, HW3727, HW3728, HW3730, HW3735, HW3736, HW3737, HW3739, HW3741, HW3742, HW3744, HW3748, HW3751, HW3752, HW3758, HW3767, HW3769, HW3773, HW3774, HW3776, HW3777, HW3778, HW3779, HW3780, HW3781, HW3782, HW3783, HW3784, HW3786, HW3793, HW3794, HW3796, HW3799, HW3802, HW3804, HW3805, HW3807, HW3809, HW3810, HW3811, HW3812, HW3814, HW3815, HW3817, HW3818, HW3822, HW3823, HW3828, HW3831, HW3833, HW3836, HW3841, HW3846, HW3848, HW3854, HW3856, HW3859, HW3861, HW3863, HW3868, HW3869, HW3870, HW3872, HW3876, HW3879, HW3880, HW3883, HW3884, HW3885, HW3888, HW3890, HW3891, HW3893, HW3894, HW3895, HW3898, HW3900, HW3901, HW3902, HW3904, HW3906, HW3907, HW3908, HW3910, HW3911, HW3913, HW3915, HW3917, HW3920, HW3921, HW3922, HW3923, HW3924, HW3926, HW3929, HW3931, HW3932, HW3937, HW3938, HW3939, HW3941, HW3943, HW3947, HW3949, HW3950, HW3951, HW3954, HW3956, HW3957, HW3959, HW3962, HW3965, HW3966, HW3968, HW3972, HW3980, HW3982, HW3985, HW3987, HW3990, HW3991, HW3992, HW3994, HW3996, HW3997, HW3998, HW3999, HW4000, HW4001, HW4007, HW4008, HW4009, HW4010, HW4012, HW4013, HW4014, HW4016, HW4025, HW4029, HW4030, HW4031, HW4034, HW4036, HW4041, HW4044, HW4045, HW4046, HW4047, HW4050, HW4053, HW4054, HW4055, HW4056, HW4057, HW4060, HW4061, HW4063, HW4065, HW4068, HW4069, HW4070, HW4072, HW4073, HW4074, HW4075, HW4076, HW4077, HW4078, HW4081, HW4084, HW4085, HW4086, HW4087, HW4088, HW4093, HW4096, HW4099, HW4100, HW4103, HW4105, HW4107, HW4109, HW4110, HW4114, HW4117, HW4118, HW4121, HW4124, HW4125, HW4126, HW4127, HW4128, HW4130, HW4131, HW4133, HW4135, HW4136, HW4137, HW4138, HW4139, HW4140, HW4143, HW4150, HW4156, HW4162, HW4165, HW4168, HW4171, HW4173, HW4175, HW4176, HW4177, HW4178, HW4180, HW4181, HW4182, HW4194, HW4195, HW4197, HW4199, HW4202, HW4203, HW4204, HW4208, HW4210, HW4214, HW4216, HW4218, HW4220, HW4221, HW4222, HW4224, hw4225, HW4229, HW4236, HW4238, HW4243, HW4246, HW4249, HW4251, HW4258, HW4260, HW4280, HW4285, HW4294, HW4296, HW4298, HW4299, HW4301, HW4309, HW4316, HW4317, HW4319, HW4322, HW4325, HW4327, HW4328, HW4330, HW4331, HW4332, HW4334, HW4336, HW4337, HW4338, HW4341, HW4342, HW4344, HW4346, HW4348, HW4349, HW4354, HW4356, HW4361, HW4363, HW4364, HW4368, HW4373, HW4374, HW4376, HW4381, HW4384, HW4387, HW4388, HW4391, HW4393, HW4395, HW4398, HW4399, HW4400, HW4403, HW4406, HW4408, HW4409, HW4410, HW4412, HW4413, HW4414, HW4416, HW4419, HW4421, HW4423, HW4428, HW4430, HW4431, HW4433, HW4434, HW4435, HW4438, HW4439, HW4442, HW4444, HW4445, HW4446, HW4449, HW4451, HW4454, HW4455, HW4457, HW4459, HW4460, HW4461, HW4463, HW4464, HW4467, HW4472, HW4473, HW4475, HW4476, HW4477, HW4480, HW4481, HW4484, HW4486, HW4488, HW4489 |
Recalling Firm/ Manufacturer |
Heartware, Inc. 14400 Nw 60th Ave Miami Lakes FL 33014-2807
|
For Additional Information Contact | MCS Customer Service 877-367-4823 |
Manufacturer Reason for Recall | Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing. |
FDA Determined Cause 2 | Device Design |
Action | The firm issued an URGENT: MEDICAL DEVICE COMMUNICATION to the healthcare providers on 11/29/2022 by UPS 2-day delivery. The letter explained the issue and provided the following:
"A. Patient Management Recommendations:
In consultation with our Independent Practitioner Quality Panel, Medtronic is providing the following patient management recommendations:
"Prior to exchanging a controller in the clinic, check to ensure the driveline cover is not stuck or hardened and can be pulled back from the driveline connector, per the instructions below.This may allow for early detection of the issue if the driveline cover is noticeably harder to manipulate but can still be moved.
"During routine clinic visits, check to ensure that the driveline cover is not stuck or hardened and can be pulled back from the driveline connector.
Instructions for Evaluating Driveline Cover:
While immobilizing the controller, push the driveline towards the controller and pull the driveline cover back from the flange with a side-to-side wiggle motion (this may require support from a second person). DO NOT pull the driveline cover with excessive force or use a twisting motion as this may damage the driveline connector or dislodge the driveline connection.
"If it is determined that a driveline cover is hardened or stuck, please contact your Medtronic field representative to evaluate the need to perform a field service procedure to remove the driveline cover.
"As a part of the servicing procedure, risks will be discussed to determine the servicing plan, including deciding if the driveline cover will be replaced if removed. If the driveline cover is removed and not replaced, care and maintenance of the driveline will remain the same without a driveline cover, however care should be taken to avoid the ringed, grooved area of the driveline connector unless you intend to disconnect the driveline from the controller. Supplemental labeling information will be provided by the Medtronic field representati |
Quantity in Commerce | 5704 units |
Distribution | Worldwide distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|