Class 2 Device Recall FUSION Dual Pump Packs |
|
Date Initiated by Firm |
December 07, 2022 |
Date Posted |
January 20, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1006-2023 |
Recall Event ID |
91290 |
510(K)Number |
K160236
|
Product Classification |
Unit, phacofragmentation - Product Code HQC
|
Product |
FUSION Dual Pump Packs, REF: OPO73 |
Code Information |
UDI-DI/Lots: 15050474602080/60373238, 60394037, 60308197 |
Recalling Firm/ Manufacturer |
Johnson & Johnson Surgical Vision, Inc. 31 Technology Dr Ste 200 Irvine CA 92618-2302
|
Manufacturer Reason for Recall |
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
|
FDA Determined Cause 2 |
Process control |
Action |
On 12/07/22, recall notices were sent to customers who were asked to do the following:
1) Discontinue using affected product.
2) Return product to the recalling firm.
3) Complete and return the Customer Reply form via email to RegCompliOne@its.jnj.com
4) Distributors, distribute the recall notice to your customers.
Customers with questions or concerns can call the firm at 1-877-266-4543 Option 2. |
Quantity in Commerce |
7644 |
Distribution |
US: FL, MS, NY, TN, TX, IL, MN, OH.
OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom
|
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = HQC and Original Applicant = Abbott Medical Optics Inc.
|
|
|
|