Class 2 Device Recall EMPOWR VVC KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall EMPOWR VVC KNEE TIBIAL INSERTsee related information
Date Initiated by FirmDecember 22, 2022
Date PostedFebruary 27, 2023
Recall Status1 Open3, Classified
Recall NumberZ-1187-2023
Recall Event ID 91494
510(K)NumberK180930 
Product Classification Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
ProductEMPOWR VVC KNEE TIBIAL INSERT, SIZE 6, 22MM, REF: 346-22-706; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 7, 22MM, REF: 346-22-707; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 8, 22MM, REF: 346-22-708; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 9, 22MM, REF: 346-22-709; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 10, 22MM, REF: 346-22-710; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 11, 22MM, REF: 346-22-711; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 22MM, REF: 346-22-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 22MM, REF: 346-22-766; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 2, 25MM, REF: 346-25-702; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 3, 25MM, REF: 346-25-703; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 4, 25MM, REF: 346-25-704; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5, 25MM, REF: 346-25-705; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 5 BRIDGE UP, 25MM, REF: 346-25-755; EMPOWR VVC KNEE TIBIAL INSERT, SIZE 6 BRIDGE DOWN, 25MM, REF: 346-25-766
Code Information REF/UDI-DI/Lot: 346-22-706/ 00190446219378/484V1013, 484V1014, 484V1015, 484V1016, 484V1017, 484V1018; 346-22-707/ 00190446219385/485V1001, 485V1002, 485V1003, 485V1004; 346-22-708/ 00190446219392/486V1005, 486V1006, 486V1007, 486V1008; 346-22-709/ 00190446219408/487V1001, 487V1002, 487V1003, 487V1004; 346-22-710/ 00190446219415/488V1003, 488V1004, 488V1005, 488V1006; 346-22-711/ 00190446219422/489V1003, 489V1004, 489V1005, 489V1006; 346-22-755/ 00190446219439/490V1001, 490V1002, 490V1003, 490V1004; 346-22-766/ 00190446219477/491V1002, 491V1003, 491V1004; 346-25-702/ 00190446219507/492V1002, 492V1003, 492V1004, 492V1005; 346-25-703/ 00190446219569/493V1001, 493V1002, 493V1003, 493V1004; 346-25-704/ 00190446219583/494V1001, 494V1002, 494V1003, 494V1004; 346-25-705/ 00190446219590/495V1001, 495V1002, 495V1003, 495V1004; 346-25-755/ 00190446219781/ 502V1001, 502V1002, 502V1003; 346-25-766/ 00190446219798/503V1003, 503V1004, 503V1005, 503V1006, 503V1007;
FEI Number 1000116912
Recalling Firm/
Manufacturer
Encore Medical, LP
9800 Metric Blvd
Austin TX 78758-5445
For Additional Information Contact
512-834-6200
Manufacturer Reason
for Recall
Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size discrepancy between the insert label and the pin label, the surgeon may implant the incorrect pin size.
FDA Determined
Cause 2
Labeling Change Control
ActionOn 12/22/22 recall notices were emailed to surgical agents who were asked to do the following: 1) Pass on the notice to all those within their organization or to any organization where potentially affected products have been transferred. 2) Return affected products. 3) Complete and return the Consignee Reconciliation notice. The following number was provided: Customer Service, 1-800-456-8696 On 2/8/23 recall notices were emailed with an expanded scope.
Quantity in Commerce216
DistributionUS: MN, IL, IN, NY, FL
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JWH
-
-