• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall medtronic see related information
Date Initiated by Firm May 10, 2023
Date Posted June 16, 2023
Recall Status1 Open3, Classified
Recall Number Z-1800-2023
Recall Event ID 92328
PMA Number P980016S536 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
Product ICD-VR DVBC3D4 EVERA S DF4 GLOB, Model Number DVBC3D4; Implantable Cardioverter Defibrillators

Code Information GTIN 00643169720442, Lot Serial Numbers: BWL206052H, BWL205879H, BWL206062H, BWL206106H, BWL206057H, BWL206066H, BWL206087H, BWL205882H, BWL206049H, BWL206050H, BWL206053H, BWL206055H, BWL206059H, BWL206069H, BWL206073H, BWL206074H, BWL206078H, BWL206079H, BWL206081H, BWL206082H, BWL206083H, BWL206084H, BWL206101H, BWL206104H, BWL206110H, BWL206112H, BWL206116H, BWL206118H, BWL206120H, BWL206122H, BWL206123H, BWL206126H, BWL206127H, BWL206068H, BWL206071H, BWL206092H, BWL206093H, BWL206094H, BWL206109H, BWL206117H, BWL206119H, BWL206097H, BWL205878H, BWL206098H, BWL206063H, BWL206114H, BWL206108H, BWL206048H, BWL206064H, BWL206058H, BWL206086H, BWL206099H, BWL206076H, BWL206103H, BWL206107H, BWL206124H, BWL206125H, BWL206105H, BWL206051H, BWL206054H; GTIN 00643169720459, Lot Serial Numbers: BWL606762S, BWL606799S, BWL609508S, BWL606741S, BWL607597S, BWL607510S, BWL607234S, BWL606748S, BWL607162S, BWL607513S, BWL607947S, BWL606791S, BWL607876S, BWL606995S, BWL607071S, BWL606857S, BWL606752S, BWL607088S, BWL606806S, BWL606900S, BWL606908S, BWL606738S, BWL606758S, BWL606759S, BWL607084S, BWL607199S, BWL607220S, BWL607277S, BWL607512S, BWL607519S, BWL607558S, BWL607571S, BWL607805S, BWL607806S, BWL609503S, BWL609505S, BWL609506S, BWL606452S, BWL606742S, BWL606743S, BWL606751S, BWL606757S, BWL606766S, BWL606767S, BWL606805S, BWL606840S, BWL606899S, BWL606942S, BWL607001S, BWL607013S, BWL607074S, BWL607215S, BWL607423S, BWL607542S, BWL607590S, BWL607593S, BWL607594S, BWL607613S, BWL607716S, BWL607910S, BWL608035S, BWL606781S, BWL606807S, BWL606837S, BWL606838S, BWL607609S, BWL606948S, BWL607556S, BWL607569S, BWL607581S, BWL607587S, BWL607721S, BWL607032S, BWL607034S, BWL607560S, BWL607610S, BWL607657S, BWL607902S, BWL609738S, BWL606872S, BWL607028S, BWL607030S, BWL607508S, BWL607509S, BWL607540S, BWL607635S, BWL607638S, BWL607705S, BWL607911S, BWL607958S, BWL607280S, BWL607666S, BWL607302S, BWL607961S, BWL607035S, BWL607036S, BWL607956S, BWL606746S, BWL606754S, BWL606765S, BWL607075S, BWL607073S, BWL607192S, BWL607194S, BWL607196S, BWL607536S, BWL607568S, BWL607573S, BWL607608S, BWL607630S, BWL607631S, BWL607884S, BWL607894S, BWL607900S, BWL607901S, BWL606749S, BWL606750S, BWL607043S, BWL607044S, BWL607505S, BWL607506S, BWL607595S, BWL607596S, BWL607727S, BWL607730S, BWL607010S, BWL607011S, BWL607223S, BWL607945S, BWL607269S, BWL608004S, BWL607255S, BWL606725S, BWL606726S, BWL606727S, BWL606776S, BWL606798S, BWL607959S, BWL606941S, BWL607038S, BWL607306S, BWL607903S, BWL607086S, BWL607555S, BWL607557S, BWL608138S, BWL607507S, BWL607012S, BWL607298S, BWL606718S, BWL606734S, BWL607219S, BWL606735S, BWL608113S, BWL607469S, BWL607224S, BWL606760S, BWL606761S, BWL606847S, BWL606851S, BWL606853S, BWL606876S, BWL607543S, BWL608031S, BWL607530S, BWL607531S, BWL606901S, BWL607305S, BWL607714S, BWL607718S, BWL606846S, BWL607588S, BWL606994S, BWL606848S, BWL606756S, BWL607029S, BWL607193S, BWL607951S, BWL607272S, BWL606717S, BWL609633S, BWL607500S, BWL607037S, BWL607867S, BWL607885S, BWL606728S, BWL609911S, BWL606768S, BWL606769S, BWL606789S, BWL606790S, BWL607316S, BWL607222S, BWL607575S, BWL607578S, BWL607580S, BWL607582S, BWL607586S, BWL607591S, BWL607592S, BWL607599S, BWL607601S, BWL607603S, BWL607606S, BWL607607S, BWL607618S, BWL607619S, BWL607249S, BWL607517S, BWL607662S, BWL608028S, BWL607221S, BWL607371S, BWL609557S, BWL607047S, BWL607503S, BWL607518S, BWL607798S, BWL608732S, BWL608754S, BWL609502S, BWL609658S, BWL606733S, BWL607007S, BWL606835S, BWL606810S, BWL607541S, BWL606935S, BWL606936S, BWL606987S, BWL606988S, BWL606989S, BWL606990S, BWL606993S, BWL607585S, BWL607616S, BWL607624S, BWL607649S, BWL607688S, BWL607690S, BWL608194S, BWL608251S, BWL608321S, BWL608322S, BWL608323S, BWL608324S, BWL608326S, BWL608389S, BWL608398S, BWL608400S, BWL608976S, BWL608984S, BWL608990S, BWL609019S, BWL609179S, BWL609183S, BWL607745S, BWL607754S, BWL608116S, BWL608140S, BWL608143S, BWL606755S, BWL606930S, BWL607002S, BWL607022S, BWL607366S, BWL607561S, BWL607576S, BWL607611S, BWL607614S, BWL607620S, BWL607621S, BWL607645S, BWL607650S, BWL608378S, BWL608382S, BWL608384S, BWL608391S, BWL608866S, BWL608918S, BWL608921S, BWL608963S, BWL608966S, BWL608970S, BWL608978S, BWL608994S, BWL609020S, BWL609162S, BWL609164S, BWL609171S, BWL609172S, BWL609180S, BWL606719S, BWL606934S, BWL607021S, BWL607390S, BWL608874S, BWL607516S, BWL607897S, BWL606944S, BWL607040S, BWL607640S, BWL607653S, BWL606817S, BWL606819S, BWL606820S, BWL606822S, BWL606730S, BWL606824S, BWL606829S, BWL606833S, BWL606836S, BWL606912S, BWL606924S, BWL606925S, BWL606926S, BWL606928S, BWL607077S, BWL607078S, BWL607080S, BWL607092S, BWL607093S, BWL607094S, BWL607122S, BWL607123S, BWL607124S, BWL607133S, BWL607134S, BWL607143S, BWL607145S, BWL607146S, BWL607148S, BWL607149S, BWL607169S, BWL607171S, BWL607172S, BWL607186S, BWL607187S, BWL607189S, BWL607200S, BWL607203S, BWL607204S, BWL607370S, BWL607389S, BWL607391S, BWL607395S, BWL607396S, BWL607397S, BWL607399S, BWL607400S, BWL607407S, BWL607408S, BWL607409S, BWL607410S, BWL607411S, BWL607412S, BWL607413S, BWL607414S, BWL607415S, BWL607416S, BWL607417S, BWL607969S, BWL607979S, BWL607986S, BWL608087S, BWL608088S, BWL608089S, BWL608090S, BWL608091S, BWL608092S, BWL608099S, BWL608136S, BWL608137S, BWL608139S, BWL608144S, BWL608146S, BWL608147S, BWL608148S, BWL608151S, BWL608152S, BWL608154S, BWL608155S, BWL608156S, BWL608347S, BWL608349S, BWL608366S, BWL608367S, BWL608368S, BWL608385S, BWL608386S, BWL608387S, BWL608395S, BWL608397S, BWL608401S, BWL608405S, BWL608407S, BWL608408S, BWL608410S, BWL608466S, BWL608467S, BWL608478S, BWL608479S, BWL608480S, BWL608487S, BWL608492S, BWL608516S, BWL608517S, BWL608518S, BWL608521S, BWL608525S, BWL608584S, BWL608586S, BWL608588S, BWL608590S, BWL608635S, BWL608637S, BWL608638S, BWL608715S, BWL608717S, BWL608718S, BWL608719S, BWL608720S, BWL608721S, BWL608724S, BWL608727S, BWL608739S, BWL608740S, BWL608746S, BWL608747S, BWL608756S, BWL608757S, BWL608764S, BWL608806S, BWL608807S, BWL608826S, BWL608827S, BWL608828S, BWL608829S, BWL608830S, BWL608868S, BWL608869S, BWL608870S, BWL608871S, BWL608876S, BWL608937S, BWL608938S, BWL608939S, BWL608973S, BWL608974S, BWL608975S, BWL609027S, BWL609028S, BWL609029S, BWL609030S, BWL609031S, BWL609168S, BWL609169S, BWL609170S, BWL609173S, BWL609176S, BWL609189S, BWL609190S, BWL609333S, BWL609334S, BWL609335S, BWL609336S, BWL609337S, BWL609338S, BWL609340S, BWL609341S, BWL609342S, BWL609428S, BWL609429S, BWL609430S, BWL609431S, BWL609432S, BWL609519S, BWL609520S, BWL609521S, BWL609522S, BWL609523S, BWL609622S, BWL609623S, BWL609639S, BWL609640S, BWL609641S, BWL609642S, BWL609645S, BWL609646S, BWL609647S, BWL609648S, BWL609652S, BWL609653S, BWL609654S, BWL609655S, BWL609656S, BWL609657S, BWL609659S, BWL609669S, BWL609670S, BWL609671S, BWL609672S, BWL609729S, BWL609730S, BWL609732S, BWL609733S, BWL609734S, BWL609735S, BWL609755S, BWL609756S, BWL609757S, BWL609761S, BWL609764S, BWL609766S, BWL609772S, BWL609777S, BWL609778S, BWL609786S, BWL609789S, BWL609797S, BWL609798S, BWL609799S, BWL609800S, BWL609801S, BWL609802S, BWL609803S, BWL609804S, BWL609805S, BWL609806S, BWL609807S, BWL609808S, BWL609887S, BWL609888S, BWL609889S, BWL609890S, BWL609891S, BWL609892S, BWL609893S, BWL606863S, BWL606967S, BWL606971S, BWL606965S, BWL608485S, BWL606451S, BWL606793S, BWL606800S, BWL606813S, BWL606825S, BWL606826S, BWL606827S, BWL606828S, BWL606830S, BWL606831S, BWL606832S, BWL606869S, BWL606880S, BWL606881S, BWL606882S, BWL606884S, BWL606887S, BWL606888S, BWL606889S, BWL606890S, BWL606891S, BWL606892S, BWL606893S, BWL606894S, BWL606895S, BWL606914S, BWL606915S, BWL606918S, BWL606919S, BWL606943S, BWL606945S, BWL606946S, BWL606966S, BWL606991S, BWL607015S, BWL607081S, BWL607082S, BWL607085S, BWL607087S, BWL607089S, BWL607090S, BWL607151S, BWL607153S, BWL607173S, BWL607176S, BWL607198S, BWL607201S, BWL607202S, BWL607206S, BWL607207S, BWL607208S, BWL607209S, BWL607283S, BWL607284S, BWL607285S, BWL607309S, BWL607312S, BWL607313S, BWL607315S, BWL607317S, BWL607328S, BWL607329S, BWL607330S, BWL607331S, BWL607372S, BWL607373S, BWL607376S, BWL607749S, BWL607751S, BWL607764S, BWL607779S, BWL607824S, BWL607863S, BWL607946S, BWL607950S, BWL607970S, BWL607989S, BWL607996S, BWL608003S, BWL608008S, BWL608014S, BWL608020S, BWL608104S, BWL608106S, BWL608107S, BWL608112S, BWL608121S, BWL608205S, BWL608206S, BWL608207S, BWL608208S, BWL608209S, BWL608211S, BWL608212S, BWL608213S, BWL608214S, BWL608215S, BWL608228S, BWL608229S, BWL608232S, BWL608233S, BWL608253S, BWL608254S, BWL608350S, BWL608351S, BWL608352S, BWL608355S, BWL608356S, BWL608357S, BWL608358S, BWL608359S, BWL608360S, BWL608361S, BWL608372S, BWL608373S, BWL608374S, BWL608375S, BWL608376S, BWL608377S, BWL608379S, BWL608380S, BWL608381S, BWL608383S, BWL608464S, BWL608465S, BWL608497S, BWL608499S, BWL608502S, BWL608503S, BWL608505S, BWL608506S, BWL608591S, BWL608592S, BWL608696S, BWL608698S, BWL608700S, BWL608701S, BWL608703S, BWL608706S, BWL608708S, BWL608709S, BWL608710S, BWL608728S, BWL608729S, BWL608796S, BWL608797S, BWL608798S, BWL608799S, BWL608800S, BWL608801S, BWL608802S, BWL608809S, BWL608810S, BWL608883S, BWL608884S, BWL608885S, BWL608886S, BWL608887S, BWL608888S, BWL608889S, BWL608969S, BWL608977S, BWL608979S, BWL608980S, BWL608983S, BWL608985S, BWL608986S, BWL608987S, BWL608989S, BWL608991S, BWL608992S, BWL608993S, BWL608996S, BWL609004S, BWL609018S, BWL609035S, BWL609036S, BWL609037S, BWL609038S, BWL609241S, BWL609242S, BWL609243S, BWL609244S, BWL609245S, BWL609247S, BWL609248S, BWL609319S, BWL609320S, BWL609321S, BWL609322S, BWL609324S, BWL609330S, BWL609331S, BWL609343S, BWL609344S, BWL609345S, BWL609346S, BWL609347S, BWL609348S, BWL609433S, BWL609439S, BWL609440S, BWL609441S, BWL609443S, BWL609444S, BWL609461S, BWL609462S, BWL609463S, BWL609538S, BWL609553S, BWL609559S, BWL609560S, BWL609561S, BWL609562S, BWL609563S, BWL609564S, BWL609565S, BWL609566S, BWL609567S, BWL609568S, BWL609569S, BWL609570S, BWL609571S, BWL609572S, BWL609573S, BWL609574S, BWL609575S, BWL609576S, BWL609577S, BWL609578S, BWL609595S, BWL609596S, BWL609597S, BWL609598S, BWL609666S, BWL607239S, BWL607497S, BWL607702S, BWL606861S, BWL608498S, BWL606721S, BWL607695S, BWL608388S, BWL606731S, BWL606732S, BWL606855S, BWL606862S, BWL607281S, BWL606722S, BWL606770S, BWL606973S, BWL606996S, BWL606997S, BWL606998S, BWL606999S, BWL607014S, BWL607016S, BWL607017S, BWL607018S, BWL607019S, BWL607252S, BWL607254S, BWL607672S, BWL607676S, BWL607680S, BWL607682S, BWL607687S, BWL607689S, BWL607691S, BWL607717S, BWL607866S, BWL607039S, BWL607802S, BWL607803S, BWL606720S, BWL606870S, BWL606871S, BWL607905S, BWL606453S, BWL606969S, BWL608370S, BWL608486S, BWL608504S, BWL606951S, BWL606962S, BWL607725S, BWL607944S, BWL608122S, BWL606961S, BWL607726S, BWL606771S, BWL606772S, BWL606774S, BWL606777S, BWL606976S, BWL606978S, BWL606979S, BWL606981S, BWL606982S, BWL606983S, BWL606940S, BWL607788S, BWL606909S, BWL606910S, BWL606913S, BWL606739S, BWL606744S, BWL606867S, BWL606868S, BWL606778S, BWL606788S, BWL606792S, BWL606984S, BWL606985S, BWL606986S, BWL607375S, BWL607421S, BWL607422S, BWL607869S, BWL607878S, BWL607890S, BWL607893S, BWL608168S, BWL608169S, BWL608170S, BWL608171S, BWL608172S, BWL608173S, BWL606977S, BWL607031S, BWL607063S, BWL607070S, BWL607229S, BWL607230S, BWL607233S, BWL607236S, BWL607238S, BWL607240S, BWL607243S, BWL607245S, BWL607246S, BWL607247S, BWL607264S, BWL607265S, BWL607266S, BWL607427S, BWL607429S, BWL607430S, BWL607435S, BWL607436S, BWL607437S, BWL607447S, BWL607448S, BWL607454S, BWL607538S, BWL607544S, BWL607574S, BWL607627S, BWL607825S, BWL607826S, BWL607827S, BWL607828S, BWL607829S, BWL607830S, BWL607831S, BWL607833S, BWL607834S, BWL607835S, BWL606736S, BWL606737S, BWL606740S, BWL607041S, BWL607042S, BWL607083S, BWL607261S, BWL606834S, BWL607251S, BWL606803S, BWL607046S, BWL607267S, BWL607268S, BWL607270S, BWL607271S, BWL607274S, BWL607286S, BWL607287S, BWL607288S, BWL607289S, BWL607290S, BWL607291S, BWL607292S, BWL607293S, BWL607295S, BWL607296S, BWL607297S, BWL607681S, BWL607683S, BWL607813S, BWL607818S, BWL607819S, BWL607820S, BWL607821S, BWL607822S, BWL607823S, BWL606716S, BWL606747S, BWL606877S, BWL606879S, BWL607659S, BWL607661S, BWL607899S, BWL608030S, BWL608862S, BWL606794S, BWL606949S, BWL606958S, BWL606960S, BWL606963S, BWL607237S, BWL607477S, BWL607482S, BWL607487S, BWL607490S, BWL607617S, BWL607626S, BWL607629S, BWL607723S, BWL607877S, BWL607880S, BWL607881S, BWL607952S, BWL608027S, BWL606802S, BWL606804S, BWL606896S, BWL606897S, BWL606929S, BWL606931S, BWL606932S, BWL606933S, BWL606937S, BWL606938S, BWL606939S, BWL607072S, BWL607076S, BWL607128S, BWL607129S, BWL607178S, BWL607299S, BWL607332S, BWL607333S, BWL607334S, BWL607335S, BWL607336S, BWL607337S, BWL607341S, BWL607368S, BWL607369S, BWL607377S, BWL607378S, BWL607379S, BWL607380S, BWL607394S, BWL607401S, BWL607955S, BWL607957S, BWL607960S, BWL607962S, BWL607964S, BWL607966S, BWL608199S, BWL608289S, BWL608341S, BWL608390S, BWL608402S, BWL608403S, BWL608404S, BWL608449S, BWL608509S, BWL608675S, BWL608795S, BWL608861S, BWL608863S, BWL608940S, BWL608972S, BWL608982S, BWL609023S, BWL609050S, BWL609166S, BWL609167S, BWL609231S, BWL609232S, BWL609326S, BWL609332S, BWL609422S, BWL609423S, BWL609581S, BWL609582S, BWL609590S, BWL609605S, BWL609606S, BWL609627S, BWL609628S, BWL606823S, BWL606816S, BWL606821S, BWL606898S, BWL606903S, BWL606904S, BWL606906S, BWL606907S, BWL606920S, BWL606921S, BWL606922S, BWL606923S, BWL606947S, BWL606950S, BWL606968S, BWL606970S, BWL606972S, BWL606975S, BWL606980S, BWL607104S, BWL607105S, BWL607106S, BWL607110S, BWL607111S, BWL607112S, BWL607113S, BWL607114S, BWL607116S, BWL607117S, BWL607118S, BWL607119S, BWL607121S, BWL607136S, BWL607137S, BWL607177S, BWL607300S, BWL607301S, BWL607303S, BWL607304S, BWL607387S, BWL607388S, BWL607418S, BWL607419S, BWL607444S, BWL607445S, BWL607457S, BWL607458S, BWL607459S, BWL607461S, BWL607463S, BWL607465S, BWL607466S, BWL607470S, BWL607472S, BWL607474S, BWL607479S, BWL607481S, BWL607483S, BWL607485S, BWL607492S, BWL607493S, BWL607498S, BWL607499S, BWL608094S, BWL608095S, BWL608097S, BWL608100S, BWL608101S, BWL608669S, BWL606796S, BWL606801S, BWL606954S, BWL606955S, BWL607107S, BWL607109S, BWL607138S, BWL607139S, BWL609509S, BWL606850S, BWL606854S, BWL606858S, BWL606873S, BWL606883S, BWL606859S, BWL606864S, BWL607451S, BWL607699S, BWL606729S, BWL606974S, BWL607282S, BWL608484S, BWL608500S, BWL606911S, BWL606927S, BWL607000S, BWL607008S, BWL607009S, BWL607125S, BWL607259S, BWL607393S, BWL607511S, BWL607583S, BWL607605S, BWL607612S, BWL607615S, BWL607637S, BWL608149S, BWL608319S, BWL608392S, BWL608396S, BWL608716S, BWL608725S, BWL608872S, BWL608875S, BWL608910S, BWL608924S, BWL608931S, BWL608932S, BWL608962S, BWL608988S, BWL609174S, BWL609175S, BWL606952S, BWL606953S, BWL607097S, BWL607167S, BWL607181S, BWL607402S, BWL607405S, BWL608242S, BWL608243S, BWL608363S, BWL608365S, BWL608369S, BWL608371S, BWL608482S, BWL608510S, BWL608514S, BWL608628S, BWL608670S, BWL608672S, BWL608750S, BWL608751S, BWL608752S, BWL608867S, BWL608895S, BWL608896S, BWL608897S, BWL609022S, BWL609032S, BWL609188S, BWL609239S, BWL609240S, BWL609246S, BWL609418S, BWL609420S, BWL609588S, BWL609589S, BWL609607S, BWL609635S, BWL609753S, BWL609795S, BWL606782S, BWL606783S, BWL606784S, BWL606785S, BWL606786S, BWL606787S, BWL606808S, BWL606811S, BWL606856S, BWL607868S, BWL607870S, BWL607872S, BWL607874S, BWL607875S, BWL607879S, BWL607883S, BWL607909S, BWL607912S, BWL608162S, BWL608163S, BWL608164S, BWL608165S, BWL608166S, BWL608167S, BWL608294S, BWL608393S, BWL607548S, BWL607549S, BWL607550S, BWL607551S, BWL607552S, BWL606812S, BWL606814S, BWL606815S, BWL606875S, BWL606885S, BWL606886S, BWL606956S, BWL607098S, BWL607102S, BWL607130S, BWL607131S, BWL607155S, BWL607161S, BWL607163S, BWL607166S, BWL607381S, BWL607382S, BWL607383S, BWL607384S, BWL607385S, BWL607386S, BWL607398S, BWL607403S, BWL607404S, BWL608007S, BWL608021S, BWL608024S, BWL608108S, BWL608109S, BWL608110S, BWL608111S, BWL608204S, BWL608216S, BWL608217S, BWL608219S, BWL608220S, BWL608234S, BWL608235S, BWL608236S, BWL608237S, BWL608238S, BWL608239S, BWL608240S, BWL608241S, BWL608249S, BWL608250S, BWL608252S, BWL608269S, BWL608271S, BWL608288S, BWL608362S, BWL608450S, BWL608451S, BWL608452S, BWL608453S, BWL608475S, BWL608477S, BWL608575S, BWL608576S, BWL608577S, BWL608578S, BWL608580S, BWL608581S, BWL608582S, BWL608583S, BWL608627S, BWL608666S, BWL608667S, BWL608679S, BWL608690S, BWL608691S, BWL608694S, BWL608695S, BWL608711S, BWL608712S, BWL608713S, BWL608714S, BWL608730S, BWL608731S, BWL608733S, BWL608735S, BWL608736S, BWL608737S, BWL608738S, BWL608741S, BWL608742S, BWL608743S, BWL608744S, BWL608745S, BWL608758S, BWL608759S, BWL608760S, BWL608761S, BWL608762S, BWL608763S, BWL608765S, BWL608766S, BWL608767S, BWL608768S, BWL608769S, BWL608770S, BWL608791S, BWL608792S, BWL608793S, BWL608794S, BWL608804S, BWL608805S, BWL608808S, BWL608811S, BWL608812S, BWL608813S, BWL608814S, BWL608819S, BWL608820S, BWL608821S, BWL608822S, BWL608823S, BWL608824S, BWL608825S, BWL608864S, BWL608865S, BWL608878S, BWL608879S, BWL608881S, BWL608882S, BWL608890S, BWL608891S, BWL608892S, BWL608893S, BWL608900S, BWL608901S, BWL608902S, BWL608903S, BWL608904S, BWL608905S, BWL608919S, BWL608920S, BWL608923S, BWL608925S, BWL608926S, BWL608928S, BWL608929S, BWL608930S, BWL608933S, BWL608934S, BWL608935S, BWL608936S, BWL608961S, BWL608964S, BWL608965S, BWL608967S, BWL608968S, BWL608981S, BWL609033S, BWL609034S, BWL609039S, BWL609040S, BWL609041S, BWL609042S, BWL609043S, BWL609045S, BWL609046S, BWL609047S, BWL609048S, BWL609049S, BWL609178S, BWL609181S, BWL609182S, BWL609184S, BWL609185S, BWL609186S, BWL609187S, BWL609233S, BWL609234S, BWL609339S, BWL609436S, BWL609437S, BWL609512S, BWL609513S, BWL609514S, BWL609515S, BWL609516S, BWL609517S, BWL609579S, BWL609580S, BWL609583S, BWL609584S, BWL609585S, BWL609586S, BWL609587S, BWL609591S, BWL609592S, BWL609593S, BWL609594S, BWL609599S, BWL609600S, BWL609601S, BWL609611S, BWL609612S, BWL609619S, BWL609620S, BWL609621S, BWL609625S, BWL609626S, BWL609629S, BWL609630S, BWL609631S, BWL609634S, BWL609673S, BWL609674S, BWL609675S, BWL609676S, BWL609677S, BWL609678S, BWL609679S, BWL609736S, BWL609737S, BWL609739S, BWL609740S, BWL609741S, BWL609742S, BWL609743S, BWL609744S, BWL609745S, BWL609758S, BWL609759S, BWL609762S, BWL609767S, BWL609768S, BWL609769S, BWL609770S, BWL609771S, BWL609773S, BWL609774S, BWL609775S, BWL609776S, BWL609779S, BWL609780S, BWL609781S, BWL609782S, BWL609783S, BWL609784S, BWL609792S, BWL609793S, BWL609794S, BWL609796S, BWL609898S, BWL609899S, BWL609900S, BWL609902S, BWL609903S, BWL609904S, BWL609905S, BWL609913S, BWL609914S, BWL608221S, BWL608894S, BWL609424S, BWL609874S, BWL609880S, BWL609881S, BWL609884S, BWL609906S, BWL606865S, BWL608520S, BWL606874S, BWL608103S, BWL607709S, BWL607024S, BWL607025S, BWL607020S, BWL608511S, BWL606753S, BWL606764S, BWL607226S, BWL607228S, BWL607564S, BWL607565S, BWL607660S, BWL607664S, BWL606779S, BWL606780S, BWL606842S, BWL606843S, BWL606844S, BWL607598S, BWL607600S, BWL607602S, BWL607604S, BWL608295S, BWL606916S, BWL606917S, BWL606957S, BWL607160S, BWL607365S, BWL607367S, BWL607392S, BWL607406S, BWL608198S, BWL608244S, BWL608248S, BWL607904S, BWL607026S, BWL608995S, BWL609882S, BWL608501S, BWL608507S, BWL607325S, BWL607326S, BWL607438S, BWL607439S, BWL607489S, BWL607651S, BWL608036S, BWL608190S, BWL608191S, BWL608192S, BWL608201S, BWL608202S, BWL608316S, BWL608409S, BWL608469S, BWL608663S, BWL608664S, BWL608665S, BWL608668S, BWL608671S, BWL608203S, BWL607003S, BWL607004S, BWL607005S, BWL607475S, BWL607476S, BWL607478S, BWL607480S, BWL607484S, BWL607486S, BWL607488S, BWL607652S, BWL607654S, BWL607655S, BWL607656S, BWL607715S, BWL607786S, BWL608102S, BWL608114S, BWL608117S, BWL608179S, BWL608180S, BWL608181S, BWL608183S, BWL608184S, BWL608193S, BWL608196S, BWL608197S, BWL608200S, BWL608512S, BWL608513S, BWL608515S, BWL608574S, BWL608579S, BWL608704S, BWL608705S, BWL608880S, BWL609528S, BWL609529S, BWL609530S, BWL607432S, BWL607433S, BWL607520S, BWL607527S, BWL607529S, BWL608058S, BWL608061S, BWL608476S, BWL608685S, BWL608686S, BWL608726S, BWL608873S, BWL608877S, BWL609419S, BWL609504S, BWL609540S, BWL607027S, BWL607218S, BWL607250S, BWL607256S, BWL607257S, BWL607276S, BWL607321S, BWL607322S, BWL607420S, BWL608029S, BWL608032S, BWL608115S, BWL608125S, BWL608231S, BWL608496S, BWL608687S, BWL608707S, BWL609442S, BWL607033S, BWL607045S, BWL607197S, BWL607491S, BWL607494S, BWL607641S, BWL607643S, BWL607644S, BWL607646S, BWL607972S, BWL607974S, BWL607975S, BWL607976S, BWL608175S, BWL608176S, BWL608177S, BWL608481S, BWL608483S, BWL608488S, BWL608688S, BWL608689S, BWL608692S, BWL608693S, BWL608697S, BWL608699S, BWL608702S, BWL607242S, BWL607260S, BWL607307S, BWL607308S, BWL607310S, BWL607311S, BWL607314S, BWL607318S, BWL607319S, BWL607320S, BWL607352S, BWL607353S, BWL607354S, BWL607355S, BWL607356S, BWL607424S, BWL607425S, BWL607426S, BWL607428S, BWL607431S, BWL607452S, BWL607639S, BWL608011S, BWL608012S, BWL608038S, BWL608039S, BWL608040S, BWL608042S, BWL608057S, BWL608353S, BWL608354S, BWL608364S, BWL608399S, BWL608406S, BWL609544S, BWL609545S, BWL609546S, BWL609547S, BWL609548S, BWL609549S, BWL609550S, BWL607006S, BWL607323S, BWL607324S, BWL607344S, BWL607346S, BWL607347S, BWL607348S, BWL607350S, BWL607358S, BWL607359S, BWL607442S, BWL607443S, BWL607446S, BWL607449S, BWL607450S, BWL607453S, BWL607455S, BWL607456S, BWL607460S, BWL607462S, BWL607532S, BWL607533S, BWL607547S, BWL607570S, BWL607577S, BWL607579S, BWL607625S, BWL607634S, BWL607886S, BWL607913S, BWL607914S, BWL607967S, BWL607968S, BWL607971S, BWL607981S, BWL607984S, BWL608000S, BWL608005S, BWL608010S, BWL608013S, BWL608015S, BWL608019S, BWL608023S, BWL608026S, BWL608043S, BWL608044S, BWL608045S, BWL608046S, BWL608048S, BWL608054S, BWL608055S, BWL608118S, BWL608119S, BWL608120S, BWL608123S, BWL608124S, BWL608141S, BWL608142S, BWL608145S, BWL608150S, BWL608153S, BWL608157S, BWL608158S, BWL608159S, BWL608160S, BWL608161S, BWL608178S, BWL608182S, BWL608185S, BWL608186S, BWL608187S, BWL608189S, BWL608210S, BWL608218S, BWL608224S, BWL608230S, BWL608265S, BWL608266S, BWL608267S, BWL608268S, BWL608270S, BWL608291S, BWL608317S, BWL608338S, BWL608340S, BWL608345S, BWL608348S, BWL608489S, BWL608490S, BWL608491S, BWL608493S, BWL608494S, BWL608495S, BWL608531S, BWL608585S, BWL608587S, BWL608589S, BWL608623S, BWL608636S, BWL608639S, BWL608661S, BWL608662S, BWL608673S, BWL608674S, BWL608676S, BWL608677S, BWL608678S, BWL608680S, BWL608681S, BWL608682S, BWL608683S, BWL608684S, BWL609531S, BWL609532S, BWL609533S, BWL609534S, BWL609535S, BWL609536S, BWL609537S, BWL609539S, BWL609541S, BWL609542S, BWL609543S, BWL607965S, BWL608188S, BWL607434S, BWL607963S, BWL608096S, BWL607338S, BWL607339S, BWL607340S, BWL607342S, BWL607343S, BWL607440S, BWL607441S, BWL607663S, BWL608625S, BWL607464S, BWL607327S, BWL607467S, BWL607468S, BWL607471S, BWL607473S, BWL606878S, BWL607048S, BWL607049S, BWL607050S, BWL607051S, BWL607052S, BWL607053S, BWL607054S, BWL607055S, BWL607056S, BWL607057S, BWL607058S, BWL607059S, BWL607060S, BWL607061S, BWL607062S, BWL607064S, BWL607065S, BWL607066S, BWL607091S, BWL607095S, BWL607096S, BWL607099S, BWL607100S, BWL607101S, BWL607103S, BWL607108S, BWL607115S, BWL607120S, BWL607126S, BWL607127S, BWL607132S, BWL607144S, BWL607152S, BWL607154S, BWL607158S, BWL607164S, BWL607165S, BWL607168S, BWL607170S, BWL607174S, BWL607175S, BWL607179S, BWL607183S, BWL607185S, BWL607188S, BWL607191S, BWL607210S, BWL607211S, BWL607212S, BWL607214S, BWL607217S, BWL607225S, BWL607227S, BWL607231S, BWL607235S, BWL607241S, BWL607244S, BWL607248S, BWL607253S, BWL607258S, BWL607262S, BWL607273S, BWL607275S, BWL607278S, BWL607279S, BWL607514S, BWL607515S, BWL607521S, BWL607522S, BWL607523S, BWL607524S, BWL607525S, BWL607526S, BWL607528S, BWL607534S, BWL607535S, BWL607665S, BWL607667S, BWL607668S, BWL607673S, BWL607675S, BWL607677S, BWL607678S, BWL607707S, BWL607708S, BWL607710S, BWL607711S, BWL607712S, BWL607713S, BWL607719S, BWL607720S, BWL607722S, BWL607724S, BWL607728S, BWL607729S, BWL607732S, BWL607733S, BWL607734S, BWL607735S, BWL607736S, BWL607737S, BWL607738S, BWL607739S, BWL607740S, BWL607741S, BWL607742S, BWL607743S, BWL607744S, BWL607746S, BWL607747S, BWL607748S, BWL607750S, BWL607752S, BWL607753S, BWL607755S, BWL607756S, BWL607757S, BWL607758S, BWL607759S, BWL607760S, BWL607761S, BWL607762S, BWL607763S, BWL607765S, BWL607766S, BWL607767S, BWL607768S, BWL607769S, BWL607770S, BWL607771S, BWL607772S, BWL607773S, BWL607774S, BWL607775S, BWL607776S, BWL607777S, BWL607778S, BWL607780S, BWL607781S, BWL607782S, BWL607783S, BWL607784S, BWL607785S, BWL607792S, BWL607793S, BWL607797S, BWL607799S, BWL607800S, BWL607837S, BWL607860S, BWL607861S, BWL607862S, BWL607864S, BWL607865S, BWL606839S, BWL606841S, BWL606849S, BWL606852S, BWL606860S, BWL607156S, BWL607157S, BWL607159S, BWL607182S, BWL607184S, BWL607190S, BWL607294S, BWL607345S, BWL607349S, BWL607351S, BWL607357S, BWL607360S, BWL607361S, BWL607362S, BWL607363S, BWL607364S, BWL608025S, BWL608033S, BWL608037S, BWL608041S, BWL608049S, BWL608051S, BWL608059S, BWL608062S, BWL608064S, BWL608086S, BWL608222S, BWL608223S, BWL608225S, BWL608226S, BWL608227S, BWL608318S, BWL608325S, BWL608337S, BWL608339S, BWL608411S, BWL608412S, BWL608413S, BWL608444S, BWL608445S, BWL608446S, BWL608447S, BWL608448S, BWL608468S, BWL608470S, BWL608471S, BWL608472S, BWL608473S, BWL608474S, BWL608629S, BWL608630S, BWL608631S, BWL608632S, BWL608633S, BWL608634S, BWL608817S, BWL608818S, BWL608906S, BWL608907S, BWL608908S, BWL608909S, BWL608911S, BWL608912S, BWL608913S, BWL608914S, BWL608915S, BWL608916S, BWL608917S, BWL609161S, BWL609163S, BWL609165S, BWL609235S, BWL609236S, BWL609237S, BWL609327S, BWL609328S, BWL609329S, BWL609434S, BWL609435S, BWL609438S, BWL607140S, BWL607141S, BWL607142S, BWL607150S, BWL607180S, BWL608245S, BWL608246S, BWL608247S, BWL608342S, BWL608343S, BWL608344S, BWL608346S, BWL608527S, BWL608529S, BWL608593S, BWL608619S, BWL608621S, BWL608622S, BWL608624S, BWL608626S, BWL608722S, BWL608723S, BWL608748S, BWL608749S, BWL608753S, BWL608755S, BWL608815S, BWL608816S, BWL608898S, BWL608899S, BWL609024S, BWL609025S, BWL609026S, BWL609325S, BWL609421S, BWL609518S, BWL609603S, BWL609604S, BWL609680S, BWL609681S, BWL609787S, BWL609788S, BWL609871S, BWL609872S, BWL609873S, BWL609878S, BWL609879S, BWL609885S, BWL609894S, BWL609895S, BWL609896S, BWL609897S, BWL609909S, BWL609915S, BWL609916S, BWL607674S, BWL609177S, BWL607906S, BWL608971S, BWL609425S, BWL609426S, BWL609427S, BWL609608S, BWL609609S, BWL609610S, BWL609613S, BWL609614S, BWL609615S, BWL609616S, BWL609617S, BWL609618S, BWL609624S, BWL609643S, BWL609644S, BWL609649S, BWL609650S, BWL609651S, BWL609664S, BWL609665S, BWL609667S, BWL609668S, BWL609682S, BWL609683S, BWL609760S, BWL609790S, BWL609791S, BWL609857S, BWL609858S, BWL609859S, BWL609498S, BWL609499S, BWL609500S, BWL609510S, BWL609511S, BWL609524S, BWL609525S, BWL609526S, BWL609527S, BWL609551S, BWL609552S, BWL609554S, BWL609555S, BWL609556S, BWL609558S, BWL609602S, BWL609632S, BWL609636S, BWL609637S, BWL609638S, BWL609660S, BWL609661S, BWL609662S, BWL609663S, BWL609684S, BWL609685S, BWL609686S, BWL609687S, BWL609688S, BWL609746S, BWL609747S, BWL609748S, BWL609749S, BWL609750S, BWL609751S, BWL609752S, BWL609754S, BWL609809S, BWL609810S, BWL609811S, BWL609812S, BWL609813S, BWL609814S, BWL609815S, BWL609816S, BWL609817S, BWL609861S, BWL609862S, BWL609863S, BWL609864S, BWL609865S, BWL609866S, BWL609867S, BWL609868S, BWL609869S, BWL609870S, BWL609763S, BWL609765S, BWL609860S, BWL609876S, BWL609877S, BWL609883S, BWL609901S, BWL609907S, BWL609908S, BWL609912S; GTIN 00643169980273, Lot Serial Numbers: BWL608258S, BWL608296S, BWL607988S, BWL608287S, BWL608293S, BWL608609S, BWL608859S, BWL608335S, BWL608327S, BWL608068S, BWL608002S, BWL611277S, BWL608776S, BWL608080S, BWL610179S, BWL607922S, BWL607987S, BWL608329S, BWL608001S, BWL608006S, BWL608016S, BWL608264S, BWL608430S, BWL608426S, BWL608458S, BWL608459S, BWL608460S, BWL608462S, BWL608463S, BWL608546S, BWL608565S, BWL608610S, BWL608611S, BWL607926S, BWL607982S, BWL608073S, BWL609960S, BWL609962S, BWL609973S, BWL607917S, BWL607919S, BWL607920S, BWL607927S, BWL607934S, BWL607977S, BWL607980S, BWL608078S, BWL608311S, BWL608420S, BWL608558S, BWL607924S, BWL607925S, BWL608074S, BWL608075S, BWL608134S, BWL608135S, BWL608279S, BWL608283S, BWL608544S, BWL608545S, BWL609134S, BWL609135S, BWL609147S, BWL608570S, BWL611767S, BWL607998S, BWL608433S, BWL609131S, BWL609947S, BWL608550S, BWL608281S, BWL608282S, BWL608302S, BWL608602S, BWL609144S, BWL609965S, BWL610223S, BWL609974S, BWL609982S, BWL608306S, BWL608328S, BWL608070S, BWL608567S, BWL608594S, BWL608607S, BWL608777S, BWL609963S, BWL609980S, BWL608305S, BWL608425S, BWL608568S, BWL608601S, BWL609975S, BWL610015S, BWL608284S, BWL609944S, BWL611729S, BWL611837S, BWL609979S, BWL611697S, BWL611428S, BWL611749S, BWL610989S, BWL609060S, BWL609299S, BWL609302S, BWL610013S, BWL611331S, BWL611358S, BWL611431S, BWL611433S, BWL611600S, BWL611669S, BWL611700S, BWL611765S, BWL611805S, BWL611838S, BWL611852S, BWL612125S, BWL610326S, BWL610482S, BWL610518S, BWL610975S, BWL611728S, BWL611732S, BWL612063S, BWL611353S, BWL611328S, BWL608427S, BWL607990S, BWL608457S, BWL608612S, BWL610467S, BWL609349S, BWL609951S, BWL609350S, BWL609941S, BWL610440S, BWL610535S, BWL609195S, BWL609226S, BWL609968S, BWL609969S, BWL610025S, BWL610057S, BWL610201S, BWL610204S, BWL610247S, BWL610390S, BWL607999S, BWL611332S, BWL610005S, BWL610006S, BWL610007S, BWL610008S, BWL610011S, BWL610339S, BWL610340S, BWL610341S, BWL610343S, BWL610344S, BWL610542S, BWL610878S, BWL610886S, BWL610894S, BWL610935S, BWL607941S, BWL608285S, BWL608603S, BWL608608S, BWL608290S, BWL609958S, BWL609959S, BWL610026S, BWL608128S, BWL608275S, BWL608276S, BWL608277S, BWL608299S, BWL608300S, BWL608332S, BWL608333S, BWL608334S, BWL608424S, BWL608428S, BWL608429S, BWL608455S, BWL608537S, BWL608541S, BWL608559S, BWL608560S, BWL608563S, BWL608564S, BWL608571S, BWL608572S, BWL608573S, BWL608605S, BWL608656S, BWL609065S, BWL609067S, BWL609074S, BWL609122S, BWL609123S, BWL609125S, BWL609126S, BWL609127S, BWL609128S, BWL609138S, BWL609140S, BWL609141S, BWL609142S, BWL609143S, BWL609474S, BWL609475S, BWL609818S, BWL609820S, BWL609821S, BWL609945S, BWL609946S, BWL609952S, BWL609953S, BWL609954S, BWL609964S, BWL609967S, BWL609971S, BWL609976S, BWL610020S, BWL610027S, BWL610087S, BWL610091S, BWL610175S, BWL610176S, BWL610190S, BWL610191S, BWL610192S, BWL610218S, BWL610240S, BWL610262S, BWL610266S, BWL610272S, BWL610316S, BWL610317S, BWL610318S, BWL610320S, BWL610345S, BWL610347S, BWL610443S, BWL610448S, BWL610469S, BWL610470S, BWL610495S, BWL610496S, BWL610527S, BWL610529S, BWL610531S, BWL610532S, BWL610534S, BWL610543S, BWL610861S, BWL610862S, BWL610865S, BWL610874S, BWL610876S, BWL611224S, BWL611237S, BWL611246S, BWL611300S, BWL611301S, BWL611303S, BWL611304S, BWL611315S, BWL611323S, BWL611336S, BWL611360S, BWL611366S, BWL611368S, BWL611377S, BWL611429S, BWL611614S, BWL611662S, BWL611716S, BWL611750S, BWL611752S, BWL611757S, BWL611764S, BWL611783S, BWL611784S, BWL611788S, BWL611789S, BWL611810S, BWL611839S, BWL611840S, BWL611841S, BWL611842S, BWL611849S, BWL611850S, BWL611853S, BWL611862S, BWL611866S, BWL611872S, BWL611874S, BWL611880S, BWL612005S, BWL612035S, BWL612041S, BWL612324S, BWL612329S, BWL612573S, BWL612575S, BWL612582S, BWL607994S, BWL608084S, BWL608085S, BWL608278S, BWL608280S, BWL608654S, BWL609822S, BWL609850S, BWL609856S, BWL610019S, BWL610028S, BWL610029S, BWL610030S, BWL611126S, BWL611145S, BWL611147S, BWL611297S, BWL611389S, BWL612489S, BWL608997S, BWL607933S, BWL607938S, BWL608072S, BWL608076S, BWL608077S, BWL608079S, BWL608081S, BWL608260S, BWL608261S, BWL609202S, BWL609206S, BWL609230S, BWL609258S, BWL609277S, BWL609278S, BWL609279S, BWL609281S, BWL609283S, BWL609284S, BWL609286S, BWL609287S.BWL609288S, BWL609289S, BWL609318S, BWL609351S, BWL609359S, BWL609364S, BWL609368S, BWL609371S, BWL609378S, BWL609382S, BWL609406S, BWL609407S, BWL609411S, BWL610321S, BWL610322S, BWL610517S, BWL610489S, BWL608562S, BWL607067S, BWL607068S, BWL607916S, BWL607992S, BWL608009S, BWL608421S, BWL611722S, BWL607940S, BWL607978S, BWL608069S, BWL608259S, BWL608255S, BWL611299S, BWL610111S, BWL610113S, BWL610349S, BWL608257S, BWL608542S, BWL611294S, BWL608536S, BWL609949S, BWL608422S, BWL609303S, BWL607995S, BWL608543S, BWL608600S, BWL608775S, BWL609450S, BWL609917S, BWL609918S, BWL610516S, BWL610546S, BWL611077S, BWL611109S, BWL611116S, BWL611153S, BWL612304S, BWL609280S, BWL609313S, BWL609315S, BWL609393S, BWL609394S, BWL609402S, BWL609408S, BWL609410S, BWL610463S, BWL610561S, BWL610563S, BWL610670S, BWL609294S, BWL609300S, BWL609304S, BWL609305S, BWL609309S, BWL609312S, BWL610055S, BWL610067S, BWL610076S, BWL610400S, BWL610406S, BWL610414S, BWL610426S, BWL610428S, BWL610438S, BWL610446S, BWL610449S, BWL609273S, BWL609290S, BWL609291S, BWL609358S, BWL609384S, BWL609392S, BWL609414S, BWL609457S, BWL610009S, BWL610010S, BWL610039S, BWL610041S, BWL610044S, BWL610048S, BWL610050S, BWL610052S, BWL610549S, BWL610553S, BWL610565S, BWL610570S, BWL610571S, BWL610572S, BWL610643S, BWL610658S, BWL610665S, BWL611066S, BWL611075S, BWL609455S, BWL609458S, BWL608273S, BWL608071S, BWL608083S, BWL608131S, BWL608538S, BWL608540S, BWL608643S, BWL608646S, BWL608647S, BWL608649S, BWL611766S, BWL611769S, BWL609156S, BWL609160S, BWL610863S, BWL611325S, BWL611755S, BWL611761S, BWL610100S, BWL610101S, BWL610102S, BWL610104S, BWL610131S, BWL610165S, BWL610166S, BWL610167S, BWL610168S, BWL610169S, BWL610170S, BWL610171S, BWL610172S, BWL610173S, BWL610183S, BWL610186S, BWL610188S, BWL610199S, BWL610202S, BWL610342S, BWL610346S, BWL610367S, BWL610368S, BWL610397S, BWL610402S, BWL610409S, BWL610522S, BWL610533S, BWL610538S, BWL610585S, BWL610586S, BWL610632S, BWL611320S, BWL611252S, BWL610418S, BWL610260S, BWL610261S, BWL610271S, BWL610319S, BWL610327S, BWL610351S, BWL610352S, BWL610386S, BWL610399S, BWL610401S, BWL610416S, BWL610422S, BWL610425S, BWL610427S, BWL610429S, BWL611021S, BWL611079S, BWL611321S, BWL610245S, BWL608553S, BWL607069S, BWL607923S, BWL608126S, BWL608129S, BWL608130S, BWL608066S, BWL608416S, BWL611141S, BWL611372S, BWL611436S, BWL607937S, BWL609483S, BWL609484S, BWL609485S, BWL609486S, BWL609487S, BWL609488S, BWL609489S, BWL609490S, BWL609491S, BWL609492S, BWL609493S, BWL609494S, BWL609495S, BWL609497S, BWL609955S, BWL609956S, BWL609957S, BWL610547S, BWL610548S, BWL610550S, BWL610551S, BWL610552S, BWL610554S, BWL610555S, BWL610557S, BWL610560S, BWL610566S, BWL610568S, BWL610573S, BWL610720S, BWL610725S, BWL610727S, BWL611275S, BWL611851S, BWL609136S, BWL608435S, BWL608436S, BWL608437S, BWL608438S, BWL608439S, BWL608440S, BWL608441S, BWL608442S, BWL607932S, BWL610328S, BWL610421S, BWL610444S, BWL610466S, BWL610475S, BWL610477S, BWL610478S, BWL610499S, BWL610503S, BWL610505S, BWL610506S, BWL610512S, BWL611013S, BWL611016S, BWL611026S, BWL611036S, BWL611038S, BWL611314S, BWL611343S, BWL611347S, BWL611348S, BWL611349S, BWL611350S, BWL611351S, BWL611352S, BWL611410S, BWL611417S, BWL611672S, BWL611690S, BWL611692S, BWL611727S, BWL611737S, BWL611739S, BWL607915S, BWL607918S, BWL607929S, BWL607935S, BWL607942S, BWL607983S, BWL608132S, BWL611434S, BWL611707S, BWL611741S, BWL610596S, BWL610645S, BWL610794S, BWL610983S, BWL611118S, BWL611376S, BWL611628S, BWL610014S, BWL607921S, BWL607928S, BWL609448S, BWL610178S, BWL610595S, BWL610648S, BWL610649S, BWL610668S, BWL610673S, BWL610710S, BWL610711S, BWL610745S, BWL610781S, BWL610782S, BWL610934S, BWL610946S, BWL610956S, BWL611162S, BWL611173S, BWL611232S, BWL611267S, BWL611281S, BWL611414S, BWL612283S, BWL609451S, BWL609950S, BWL611123S, BWL609076S, BWL609121S, BWL609269S, BWL609270S, BWL609271S, BWL608417S, BWL608555S, BWL610273S, BWL609272S, BWL609275S, BWL609282S, BWL609285S, BWL609361S, BWL609404S, BWL610353S, BWL610558S, BWL610562S, BWL610567S, BWL610569S, BWL611654S, BWL610090S, BWL610092S, BWL610114S, BWL610133S, BWL610134S, BWL610354S, BWL610417S, BWL610419S, BWL610474S, BWL610521S, BWL608616S, BWL608617S, BWL608618S, BWL608620S, BWL608640S, BWL609007S, BWL609008S, BWL609009S, BWL609059S, BWL609063S, BWL609068S, BWL609069S, BWL609072S, BWL609092S, BWL609093S, BWL609196S, BWL609198S, BWL609199S, BWL609203S, BWL609204S, BWL609205S, BWL609207S, BWL609264S, BWL609265S, BWL609266S, BWL609267S, BWL609268S, BWL609689S, BWL609690S, BWL609691S, BWL609692S, BWL609693S, BWL609694S, BWL609695S, BWL609696S, BWL609697S, BWL609698S, BWL609823S, BWL609824S, BWL609825S, BWL609826S, BWL609827S, BWL609995S, BWL609996S, BWL609997S, BWL609998S, BWL609999S, BWL610000S, BWL610001S, BWL610002S, BWL610003S, BWL610004S, BWL610181S, BWL610182S, BWL610184S, BWL610185S, BWL610187S, BWL610274S, BWL610275S, BWL610276S, BWL610277S, BWL610278S, BWL610279S, BWL610280S, BWL610283S, BWL610287S, BWL610288S, BWL610291S, BWL610296S, BWL610298S, BWL610299S, BWL610301S, BWL610302S, BWL610303S, BWL610305S, BWL610306S, BWL610307S, BWL610308S, BWL610309S, BWL610310S, BWL610312S, BWL610313S, BWL610350S, BWL610451S, BWL610453S, BWL610454S, BWL610455S, BWL610457S, BWL610458S, BWL610459S, BWL610460S, BWL610461S, BWL611164S, BWL611215S, BWL611216S, BWL611218S, BWL611222S, BWL611279S, BWL611310S, BWL611356S, BWL611357S, BWL610117S, BWL610118S, BWL610119S, BWL610121S, BWL610122S, BWL610174S, BWL610213S, BWL610221S, BWL610229S, BWL610230S, BWL610231S, BWL610232S, BWL610233S, BWL610234S, BWL610244S, BWL610248S, BWL610250S, BWL610256S, BWL610257S, BWL610258S, BWL610263S, BWL610264S, BWL610265S, BWL610282S, BWL610284S, BWL610285S, BWL610286S, BWL610289S, BWL610292S, BWL610293S, BWL610294S, BWL610314S, BWL610315S, BWL610336S, BWL610360S, BWL610364S, BWL610365S, BWL610366S, BWL610369S, BWL610370S, BWL610374S, BWL610375S, BWL610376S, BWL610377S, BWL610378S, BWL610392S, BWL610393S, BWL610394S, BWL610395S, BWL610410S, BWL610411S, BWL610413S, BWL610420S, BWL610423S, BWL610424S, BWL610525S, BWL610577S, BWL610594S, BWL610597S, BWL610612S, BWL610613S, BWL610619S, BWL610622S, BWL610623S, BWL610630S, BWL610633S, BWL610646S, BWL610647S, BWL610654S, BWL610655S, BWL610656S, BWL610660S, BWL610674S, BWL610677S, BWL610678S, BWL610679S, BWL610681S, BWL610699S, BWL610700S, BWL610701S, BWL610702S, BWL610703S, BWL610704S, BWL610705S, BWL610706S, BWL610712S, BWL610713S, BWL610716S, BWL610717S, BWL610726S, BWL610728S, BWL610729S, BWL610730S, BWL610732S, BWL610733S, BWL610734S, BWL610735S, BWL610737S, BWL610738S, BWL610763S, BWL610764S, BWL610766S, BWL610767S, BWL610768S, BWL610769S, BWL610770S, BWL610776S, BWL610777S, BWL610788S, BWL610789S, BWL610793S, BWL610799S, BWL610839S, BWL610847S, BWL610856S, BWL610857S, BWL610867S, BWL610873S, BWL610875S, BWL610877S, BWL610879S, BWL610880S, BWL610881S, BWL610882S, BWL610883S, BWL610890S, BWL610908S, BWL610909S, BWL610910S, BWL610917S, BWL610923S, BWL610925S, BWL610932S, BWL610959S, BWL610961S, BWL610962S, BWL610964S, BWL610967S, BWL610968S, BWL610969S, BWL610970S, BWL610971S, BWL610972S, BWL610973S, BWL610976S, BWL610977S, BWL610978S, BWL610979S, BWL610980S, BWL610981S, BWL610982S, BWL610990S, BWL610991S, BWL610992S, BWL610993S, BWL610999S, BWL611010S, BWL611014S, BWL611025S, BWL611029S, BWL611030S, BWL611031S, BWL611032S, BWL611033S, BWL611034S, BWL611043S, BWL611044S, BWL611045S, BWL611046S, BWL611047S, BWL611054S, BWL611056S, BWL611057S, BWL611060S, BWL611062S, BWL611063S, BWL611064S, BWL611065S, BWL611067S, BWL611068S, BWL611069S, BWL611073S, BWL611074S, BWL611076S, BWL611080S, BWL611084S, BWL611085S, BWL611086S, BWL611087S, BWL611088S, BWL611090S, BWL611091S, BWL611092S, BWL611093S, BWL611108S, BWL611142S, BWL611143S, BWL611144S, BWL611146S, BWL611148S, BWL611149S, BWL611150S, BWL611161S, BWL611168S, BWL611177S, BWL611179S, BWL611182S, BWL611183S, BWL611193S, BWL611194S, BWL611197S, BWL611198S, BWL611199S, BWL611201S, BWL611202S, BWL611203S, BWL611204S, BWL611205S, BWL611206S, BWL611214S, BWL611217S, BWL611220S, BWL611230S, BWL611234S, BWL611235S, BWL611236S, BWL611338S, BWL611339S, BWL611340S, BWL611342S, BWL611361S, BWL611362S, BWL611363S, BWL611364S, BWL611365S, BWL611367S, BWL611369S, BWL611378S, BWL611379S, BWL611380S, BWL611390S, BWL611391S, BWL611392S, BWL611393S, BWL611394S, BWL611395S, BWL611396S, BWL611397S, BWL611405S, BWL611406S, BWL611407S, BWL611415S, BWL611418S, BWL611422S, BWL611424S, BWL611425S, BWL611439S, BWL611461S, BWL611465S, BWL611469S, BWL611472S, BWL611474S, BWL611475S, BWL611476S, BWL611477S, BWL611478S, BWL611479S, BWL611486S, BWL611487S, BWL611496S, BWL611497S, BWL611498S, BWL611502S, BWL611503S, BWL611529S, BWL611530S, BWL611531S, BWL611540S, BWL611546S, BWL611549S, BWL611550S, BWL611551S, BWL611552S, BWL611553S, BWL611554S, BWL611555S, BWL611556S, BWL611557S, BWL611558S, BWL611559S, BWL611560S, BWL611562S, BWL611565S, BWL611566S, BWL611567S, BWL611588S, BWL611589S, BWL611590S, BWL611591S, BWL611592S, BWL611593S, BWL611616S, BWL611629S, BWL611630S, BWL611675S, BWL611676S, BWL611677S, BWL611678S, BWL611686S, BWL611687S, BWL611793S, BWL611822S, BWL611823S, BWL611824S, BWL611825S, BWL611829S, BWL611883S, BWL611884S, BWL611896S, BWL611902S, BWL611903S, BWL611906S, BWL611908S, BWL611916S, BWL611919S, BWL611920S, BWL611924S, BWL611934S, BWL611942S, BWL611943S, BWL611944S, BWL611945S, BWL611946S, BWL611947S, BWL611948S, BWL611962S, BWL611963S, BWL611964S, BWL611966S, BWL611970S, BWL611971S, BWL611975S, BWL611976S, BWL611977S, BWL611980S, BWL611987S, BWL611988S, BWL611989S, BWL611990S, BWL611991S, BWL611992S, BWL611993S, BWL611994S, BWL611995S, BWL611996S, BWL611997S, BWL612027S, BWL612053S, BWL612058S, BWL612059S, BWL612067S, BWL612071S, BWL612095S, BWL612096S, BWL612097S, BWL612108S, BWL612111S, BWL612114S, BWL612115S, BWL612116S, BWL612117S, BWL612119S, BWL612122S, BWL612123S, BWL612124S, BWL612126S, BWL612127S, BWL612128S, BWL612129S, BWL612133S, BWL612134S, BWL612135S, BWL612142S, BWL612143S, BWL612144S, BWL612145S, BWL612146S, BWL612147S, BWL612148S, BWL612149S, BWL612157S, BWL612158S, BWL612190S, BWL612191S, BWL612192S, BWL612193S, BWL612195S, BWL612199S, BWL612236S, BWL612239S, BWL612240S, BWL612241S, BWL612242S, BWL612243S, BWL612244S, BWL612245S, BWL612246S, BWL612247S, BWL612248S, BWL612252S, BWL612253S, BWL612255S, BWL612256S, BWL612257S, BWL612258S, BWL612259S, BWL612269S, BWL612270S, BWL612271S, BWL612272S, BWL612273S, BWL612274S, BWL612275S, BWL612276S, BWL612277S, BWL612280S, BWL612284S, BWL612285S, BWL612326S, BWL612332S, BWL612337S, BWL612338S, BWL612339S, BWL612340S, BWL612341S, BWL612342S, BWL612343S, BWL612387S, BWL612388S, BWL612406S, BWL612407S, BWL611114S, BWL611121S, BWL610103S, BWL610212S, BWL610241S, BWL610242S, BWL610252S, BWL610267S, BWL610373S, BWL610379S, BWL610620S, BWL610747S, BWL610750S, BWL610841S, BWL610899S, BWL611012S, BWL611035S, BWL611178S, BWL611412S, BWL611507S, BWL611523S, BWL611524S, BWL611580S, BWL611624S, BWL611634S, BWL611733S, BWL611806S, BWL611951S, BWL611974S, BWL612029S, BWL612030S, BWL612031S, BWL612072S, BWL612073S, BWL612074S, BWL612081S, BWL612082S, BWL612105S, BWL612106S, BWL612107S, BWL612136S, BWL612167S, BWL612168S, BWL612169S, BWL612196S, BWL612254S, BWL612298S, BWL612299S, BWL612331S, BWL612344S, BWL612345S, BWL612346S, BWL612347S, BWL612425S, BWL611111S, BWL610180S, BWL611024S, BWL608847S, BWL608848S, BWL608957S, BWL609052S, BWL609053S, BWL609054S, BWL609055S, BWL609056S, BWL609057S, BWL609062S, BWL609064S, BWL609066S, BWL609070S, BWL609071S, BWL609073S, BWL609112S, BWL609155S, BWL609157S, BWL609158S, BWL609159S, BWL609191S, BWL609192S, BWL609193S, BWL609219S, BWL609220S, BWL609221S, BWL609222S, BWL609453S, BWL609456S, BWL609459S, BWL609460S, BWL609464S, BWL609465S, BWL609466S, BWL609718S, BWL609719S, BWL609720S, BWL609721S, BWL609722S, BWL609723S, BWL609724S, BWL609725S, BWL609831S, BWL609832S, BWL609928S, BWL609930S, BWL610034S, BWL610035S, BWL610037S, BWL610077S, BWL610080S, BWL610081S.BWL610082S, BWL608641S, BWL608642S, BWL608644S, BWL608645S, BWL608648S, BWL608785S, BWL608786S, BWL608849S, BWL608850S, BWL608852S, BWL608855S, BWL608856S, BWL609010S, BWL609011S, BWL609012S, BWL609013S, BWL609014S, BWL609016S, BWL609017S, BWL609104S, BWL609105S, BWL609106S, BWL609107S, BWL609108S, BWL609116S, BWL609117S, BWL609118S, BWL609119S, BWL609120S, BWL609224S, BWL609225S, BWL609227S, BWL609228S, BWL609229S, BWL609307S, BWL609308S, BWL609310S, BWL609311S, BWL609314S, BWL609316S, BWL609317S, BWL609383S, BWL609385S, BWL609386S, BWL609387S, BWL609388S, BWL609389S, BWL609704S, BWL609705S, BWL609706S, BWL609707S, BWL609708S, BWL609709S, BWL609710S, BWL609711S, BWL609712S, BWL609713S, BWL609920S, BWL609921S, BWL609922S, BWL609923S, BWL609924S, BWL610042S, BWL610043S, BWL610056S, BWL610058S, BWL610059S, BWL610060S, BWL610061S, BWL610062S, BWL610064S, BWL610065S, BWL610066S, BWL610068S, BWL610069S, BWL610070S, BWL610072S, BWL610073S, BWL610074S, BWL610075S, BWL610093S, BWL610094S, BWL610095S, BWL610097S, BWL610115S, BWL608613S, BWL608614S, BWL608615S, BWL608660S, BWL608947S, BWL608948S, BWL609006S, BWL609077S, BWL609113S, BWL609114S, BWL609154S, BWL609217S, BWL609218S, BWL609374S, BWL609375S, BWL609381S, BWL609467S, BWL609468S, BWL609726S, BWL609727S, BWL609728S, BWL609983S, BWL609984S, BWL609985S, BWL610031S, BWL609082S, BWL609083S, BWL609102S, BWL609103S, BWL609115S, BWL609261S, BWL609405S, BWL609409S, BWL609417S, BWL609699S, BWL609700S, BWL609844S, BWL610125S, BWL610143S, BWL610154S, BWL610193S, BWL610194S, BWL608650S, BWL608651S, BWL608652S, BWL608653S, BWL608831S, BWL608832S, BWL608833S, BWL608834S, BWL609109S, BWL609110S, BWL609111S, BWL609214S, BWL609215S, BWL609216S, BWL609413S, BWL609415S, BWL609416S, BWL609477S, BWL609478S, BWL609479S, BWL609836S, BWL609837S, BWL609838S, BWL609841S, BWL609842S, BWL609843S, BWL609991S, BWL609992S, BWL609993S, BWL609994S, BWL610040S, BWL610054S, BWL610083S, BWL610084S, BWL610196S, BWL610197S, BWL610198S, BWL610200S, BWL608835S, BWL608836S, BWL608837S, BWL608838S, BWL608839S, BWL608857S, BWL608858S, BWL608860S, BWL608949S, BWL608950S, BWL608951S, BWL608958S, BWL608959S, BWL608960S, BWL608998S, BWL608999S, BWL609000S, BWL609001S, BWL609084S, BWL609085S, BWL609086S, BWL609087S, BWL609396S, BWL609398S, BWL609399S, BWL609839S, BWL609840S, BWL609919S, BWL608657S, BWL608658S, BWL608659S, BWL608779S, BWL608780S, BWL608781S, BWL608782S, BWL608783S, BWL608784S, BWL608840S, BWL608841S, BWL608842S, BWL608843S, BWL608844S, BWL608845S, BWL608941S, BWL608942S, BWL608943S, BWL608944S, BWL608945S, BWL608946S, BWL608952S, BWL608953S, BWL608954S, BWL608955S, BWL608956S, BWL609002S, BWL609003S, BWL609005S, BWL609078S, BWL609079S, BWL609080S, BWL609081S, BWL609096S, BWL609097S, BWL609098S, BWL609099S, BWL609100S, BWL609101S, BWL609149S, BWL609150S, BWL609151S, BWL609152S, BWL609153S, BWL609249S, BWL609250S, BWL609251S, BWL609256S, BWL609259S, BWL609260S, BWL609262S, BWL609263S, BWL609376S, BWL609377S, BWL609379S, BWL609380S, BWL609390S, BWL609391S, BWL609395S, BWL609469S, BWL609470S, BWL609471S, BWL609472S, BWL609473S, BWL609476S, BWL609481S, BWL609482S, BWL609702S, BWL609703S, BWL609717S, BWL609828S, BWL609830S, BWL609833S, BWL609834S, BWL609835S, BWL609925S, BWL609926S, BWL609927S, BWL609931S, BWL609932S, BWL609933S, BWL609934S, BWL609935S, BWL609987S, BWL610032S, BWL610033S, BWL610036S, BWL610045S, BWL610046S, BWL610047S, BWL610049S, BWL610051S, BWL610053S, BWL608846S, BWL609223S, BWL609400S, BWL609701S, BWL608851S, BWL608854S, BWL609197S, BWL609208S, BWL609209S, BWL609210S, BWL609211S, BWL609212S, BWL609213S, BWL609252S, BWL609253S, BWL609254S, BWL609255S, BWL609257S, BWL609352S, BWL609353S, BWL609354S, BWL609356S, BWL609357S, BWL609360S, BWL609362S, BWL609363S, BWL609365S, BWL609366S, BWL609367S, BWL609369S, BWL609370S, BWL609372S, BWL609373S, BWL609397S, BWL609845S, BWL609847S, BWL609848S, BWL609849S, BWL609851S, BWL609852S, BWL609853S, BWL609854S, BWL609855S, BWL608655S, BWL608778S, BWL609058S, BWL609061S, BWL609200S, BWL609401S, BWL609403S, BWL609714S, BWL609715S, BWL609716S, BWL609989S, BWL609990S, BWL609306S, BWL608787S, BWL608788S, BWL608789S, BWL608790S, BWL609088S, BWL609094S, BWL609095S, BWL609986S, BWL610195S, BWL611695S, BWL611718S, BWL611725S, BWL611730S, BWL611743S, BWL610249S, BWL611061S, BWL610085S, BWL610086S, BWL610088S, BWL610089S, BWL610109S, BWL610110S, BWL610222S, BWL610224S, BWL610253S, BWL610254S, BWL610290S, BWL610462S, BWL610464S, BWL610578S, BWL610579S, BWL610599S, BWL610605S, BWL610610S, BWL610611S, BWL610618S, BWL610667S, BWL610676S, BWL610680S, BWL610714S, BWL610718S, BWL610801S, BWL610811S, BWL610812S, BWL610820S, BWL610902S, BWL610921S, BWL610938S, BWL610943S, BWL611005S, BWL611006S, BWL611007S, BWL611008S, BWL611009S, BWL611037S, BWL611039S, BWL611040S, BWL611102S, BWL611103S, BWL611104S, BWL611129S, BWL611130S, BWL611170S, BWL611190S, BWL611191S, BWL611195S, BWL611196S, BWL611253S, BWL611259S, BWL611260S, BWL611274S, BWL611282S, BWL611341S, BWL611374S, BWL611375S, BWL611383S, BWL611398S, BWL611399S, BWL611404S, BWL611445S, BWL611446S, BWL611447S, BWL611448S, BWL611449S, BWL611450S, BWL611451S, BWL611458S, BWL611459S, BWL611460S, BWL611473S, BWL611480S, BWL611481S, BWL611488S, BWL611489S, BWL611490S, BWL611491S, BWL611492S, BWL611500S, BWL611501S, BWL611505S, BWL611506S, BWL611534S, BWL611538S, BWL611539S, BWL611618S, BWL611619S, BWL611626S, BWL611643S, BWL611644S, BWL611645S, BWL611646S, BWL611652S, BWL611653S, BWL611655S, BWL611656S, BWL611657S, BWL611661S, BWL611663S, BWL611664S, BWL611665S, BWL611666S, BWL611667S, BWL611706S, BWL611709S, BWL611712S, BWL611719S, BWL611720S, BWL611721S, BWL611731S, BWL611786S, BWL611787S, BWL611791S, BWL611827S, BWL611830S, BWL611832S, BWL611833S, BWL611834S, BWL611889S, BWL611890S, BWL611891S, BWL611892S, BWL611893S, BWL611897S, BWL611898S, BWL611899S, BWL611900S, BWL611901S, BWL611904S, BWL611905S, BWL611907S, BWL611909S, BWL611912S, BWL611932S, BWL611933S, BWL611935S, BWL611936S, BWL611937S, BWL611955S, BWL611956S, BWL611958S, BWL611982S, BWL611984S, BWL611986S, BWL612010S, BWL612012S, BWL612014S, BWL612015S, BWL612084S, BWL612085S, BWL612102S, BWL612112S, BWL612113S, BWL612130S, BWL612131S, BWL612132S, BWL612139S, BWL612140S, BWL612141S, BWL612150S, BWL612151S, BWL612152S, BWL612153S, BWL612154S, BWL612171S, BWL612172S, BWL612173S, BWL612174S, BWL612186S, BWL612194S, BWL612197S, BWL612198S, BWL612204S, BWL612205S, BWL612206S, BWL612207S, BWL612208S, BWL612223S, BWL612224S, BWL612225S, BWL612226S, BWL612227S, BWL612228S, BWL612229S, BWL612230S, BWL612262S, BWL612263S, BWL612264S, BWL612265S, BWL612266S, BWL612287S, BWL612288S, BWL612289S, BWL612316S, BWL612317S, BWL612323S, BWL612325S, BWL612334S, BWL612335S, BWL612336S, BWL612399S, BWL612400S, BWL612401S, BWL612402S, BWL612403S, BWL612404S, BWL612409S, BWL612410S, BWL612411S, BWL612422S, BWL612432S, BWL609089S, BWL609090S, BWL609091S, BWL611122S, BWL610914S, BWL611525S, BWL611581S, BWL610096S, BWL610098S, BWL610099S, BWL610105S, BWL610106S, BWL610107S, BWL610108S, BWL610135S, BWL610136S, BWL610225S, BWL610227S, BWL610268S, BWL610281S, BWL610355S, BWL610356S, BWL610357S, BWL610358S, BWL610359S, BWL610387S, BWL610388S, BWL610389S, BWL610391S, BWL610436S, BWL610447S, BWL610450S, BWL610456S, BWL610515S, BWL610519S, BWL610520S, BWL610588S, BWL610589S, BWL610600S, BWL610607S, BWL610609S, BWL610631S, BWL610640S, BWL610657S, BWL610659S, BWL610672S, BWL610715S, BWL610736S, BWL610739S, BWL610740S, BWL610746S, BWL610765S, BWL610783S, BWL610819S, BWL610821S, BWL610822S, BWL610823S, BWL610824S, BWL610825S, BWL610840S, BWL610842S, BWL610843S, BWL610850S, BWL610851S, BWL610855S, BWL610870S, BWL610871S, BWL610922S, BWL610940S, BWL610984S, BWL610985S, BWL610986S, BWL611011S, BWL611042S, BWL611048S, BWL611082S, BWL611083S, BWL611099S, BWL611100S, BWL611101S, BWL611112S, BWL611113S, BWL611117S, BWL611157S, BWL611163S, BWL611165S, BWL611176S, BWL611184S, BWL611240S, BWL611241S, BWL611242S, BWL611243S, BWL611244S, BWL611262S, BWL611263S, BWL611264S, BWL611265S, BWL611266S, BWL611386S, BWL611387S, BWL611426S, BWL611464S, BWL611466S, BWL611482S, BWL611483S, BWL611494S, BWL611515S, BWL611521S, BWL611627S, BWL611633S, BWL611641S, BWL611642S, BWL611673S, BWL611674S, BWL611685S, BWL611797S, BWL611826S, BWL611972S, BWL611973S, BWL612021S, BWL612052S, BWL612086S, BWL612087S, BWL612088S, BWL612109S, BWL612110S, BWL612120S, BWL612187S, BWL612250S, BWL612267S, BWL612268S, BWL612282S, BWL612386S, BWL612391S, BWL612392S, BWL612396S, BWL612398S, BWL612412S, BWL612430S, BWL612431S, BWL610132S, BWL610137S, BWL610138S, BWL610139S, BWL610140S, BWL610141S, BWL610142S, BWL610144S, BWL610145S, BWL610146S, BWL610147S, BWL610148S, BWL610149S, BWL610150S, BWL610151S, BWL610152S, BWL610155S, BWL610156S, BWL610157S, BWL610158S, BWL610159S, BWL610160S, BWL610161S, BWL610162S, BWL610163S, BWL610164S, BWL610203S, BWL610208S, BWL610209S, BWL610215S, BWL610216S, BWL610217S, BWL610219S, BWL610220S, BWL610228S, BWL610251S, BWL610255S, BWL610295S, BWL610297S, BWL610300S, BWL610304S, BWL610311S, BWL610380S, BWL610381S, BWL610382S, BWL610383S, BWL610384S, BWL610396S, BWL610405S, BWL610407S, BWL610408S, BWL610476S, BWL610493S, BWL610498S, BWL610501S, BWL610502S, BWL610504S, BWL610507S, BWL610508S, BWL610509S, BWL610510S, BWL610513S, BWL610514S, BWL610526S, BWL610528S, BWL610574S, BWL610575S, BWL610576S, BWL610580S, BWL610581S, BWL610582S, BWL610583S, BWL610584S, BWL610587S, BWL610593S, BWL610602S, BWL610603S, BWL610606S, BWL610617S, BWL610621S, BWL610626S, BWL610629S, BWL610634S, BWL610661S, BWL610662S, BWL610663S, BWL610664S, BWL610666S, BWL610669S, BWL610671S, BWL610682S, BWL610683S, BWL610684S, BWL610685S, BWL610686S, BWL610687S, BWL610688S, BWL610689S, BWL610690S, BWL610691S, BWL610692S, BWL610694S, BWL610695S, BWL610696S, BWL610697S, BWL610698S, BWL610709S, BWL610723S, BWL610741S, BWL610742S, BWL610748S, BWL610749S, BWL610751S, BWL610752S, BWL610771S, BWL610772S, BWL610773S, BWL610774S, BWL610775S, BWL610786S, BWL610787S, BWL610797S, BWL610798S, BWL610800S, BWL610803S, BWL610804S, BWL610805S, BWL610806S, BWL610807S, BWL610810S, BWL610813S, BWL610816S, BWL610817S, BWL610826S, BWL610828S, BWL610830S, BWL610831S, BWL610832S, BWL610844S, BWL610845S, BWL610846S, BWL610852S, BWL610853S, BWL610854S, BWL610860S, BWL610884S, BWL610885S, BWL610896S, BWL610897S, BWL610898S, BWL610903S, BWL610904S, BWL610905S, BWL610906S, BWL610907S, BWL610924S, BWL610927S, BWL610928S, BWL610933S, BWL610941S, BWL610942S, BWL610945S, BWL610950S, BWL610951S, BWL610952S, BWL610953S, BWL610957S, BWL610958S, BWL610966S, BWL610974S, BWL610987S, BWL610988S, BWL610994S, BWL610995S, BWL610996S, BWL610997S, BWL610998S, BWL611000S, BWL611001S, BWL611002S, BWL611003S, BWL611004S, BWL611017S, BWL611027S, BWL611028S, BWL611049S, BWL611050S, BWL611051S, BWL611052S, BWL611053S, BWL611070S, BWL611071S, BWL611078S, BWL611094S, BWL611095S, BWL611097S, BWL611098S, BWL611105S, BWL611106S, BWL611107S, BWL611119S, BWL611120S, BWL611128S, BWL611139S, BWL611140S, BWL611152S, BWL611154S, BWL611155S, BWL611158S, BWL611159S, BWL611160S, BWL611171S, BWL611174S, BWL611175S, BWL611185S, BWL611186S, BWL611187S, BWL611188S, BWL611189S, BWL611211S, BWL611212S, BWL611213S, BWL611227S, BWL611228S, BWL611238S, BWL611239S, BWL611276S, BWL611280S, BWL611283S, BWL611284S, BWL611286S, BWL611287S, BWL611288S, BWL611305S, BWL611306S, BWL611308S, BWL611327S, BWL611329S, BWL611330S, BWL611333S, BWL611337S, BWL611344S, BWL611345S, BWL611346S, BWL611371S, BWL611373S, BWL611384S, BWL611385S, BWL611419S, BWL611420S, BWL611421S, BWL611427S, BWL611430S, BWL611443S, BWL611444S, BWL611516S, BWL611517S, BWL611518S, BWL611519S, BWL611522S, BWL611526S, BWL611536S, BWL611537S, BWL611541S, BWL611544S, BWL611545S, BWL611563S, BWL611570S, BWL611571S, BWL611572S, BWL611574S, BWL611575S, BWL611576S, BWL611577S, BWL611578S, BWL611579S, BWL611594S, BWL611595S, BWL611613S, BWL611615S, BWL611617S, BWL611620S, BWL611632S, BWL611636S, BWL611637S, BWL611638S, BWL611639S.BWL611640S, BWL611648S, BWL611649S, BWL611650S, BWL611651S, BWL611658S, BWL611659S, BWL611660S, BWL611668S, BWL611670S, BWL611671S, BWL611693S, BWL611701S, BWL611702S, BWL611710S, BWL611711S, BWL611723S, BWL611724S, BWL611726S, BWL611778S, BWL611779S, BWL611780S, BWL611781S, BWL611782S, BWL611794S, BWL611799S, BWL611800S, BWL611801S, BWL611802S, BWL611803S, BWL611804S, BWL611813S, BWL611814S, BWL611815S, BWL611816S, BWL611817S, BWL611819S, BWL611820S, BWL611821S, BWL611856S, BWL611881S, BWL611882S, BWL611885S, BWL611886S, BWL611887S, BWL611894S, BWL611895S, BWL611910S, BWL611911S, BWL611913S, BWL611915S, BWL611917S, BWL611918S, BWL611921S, BWL611922S, BWL611923S, BWL611925S, BWL611926S, BWL611927S, BWL611928S, BWL611929S, BWL611930S, BWL611931S, BWL611938S, BWL611940S, BWL611941S, BWL611949S, BWL611950S, BWL611953S, BWL611954S, BWL611957S, BWL611959S, BWL611960S, BWL611961S, BWL611978S, BWL611979S, BWL611981S, BWL611983S, BWL611985S, BWL611998S, BWL611999S, BWL612022S, BWL612023S, BWL612024S, BWL612025S, BWL612026S, BWL612038S, BWL612039S, BWL612040S, BWL612050S, BWL612051S, BWL612054S, BWL612055S, BWL612056S, BWL612057S, BWL612060S, BWL612061S, BWL612062S, BWL612064S, BWL612070S, BWL612090S, BWL612091S, BWL612092S, BWL612101S, BWL612104S, BWL612118S, BWL612121S, BWL612137S, BWL612138S, BWL612155S, BWL612156S, BWL612162S, BWL612179S, BWL612180S, BWL612181S, BWL612182S, BWL612183S, BWL612211S, BWL612212S, BWL612213S, BWL612214S, BWL612249S, BWL612251S, BWL612281S, BWL612286S, BWL612294S, BWL612302S, BWL612303S, BWL612385S, BWL612390S, BWL612394S, BWL612395S, BWL612397S, BWL612408S, BWL612423S, BWL612424S, BWL612426S, BWL612427S, BWL610112S, BWL610116S, BWL610123S, BWL610124S, BWL610126S, BWL610127S, BWL610128S, BWL610129S, BWL610130S, BWL610235S, BWL610236S, BWL610237S, BWL610238S, BWL610361S, BWL610362S, BWL610363S, BWL610371S, BWL610372S, BWL610385S, BWL610403S, BWL610404S, BWL610590S, BWL610592S, BWL610598S, BWL610614S, BWL610615S, BWL610616S, BWL610624S, BWL610625S, BWL610627S, BWL610635S, BWL610636S, BWL610637S, BWL610638S, BWL610639S, BWL610642S, BWL610644S, BWL610650S, BWL610651S, BWL610652S, BWL610653S, BWL610675S, BWL610707S, BWL610708S, BWL610719S, BWL610721S, BWL610722S, BWL610724S, BWL610731S, BWL610743S, BWL610744S, BWL610753S, BWL610754S, BWL610756S, BWL610757S, BWL610758S, BWL610759S, BWL610760S, BWL610761S, BWL610762S, BWL610778S, BWL610779S, BWL610780S, BWL610784S, BWL610785S, BWL610790S, BWL610791S, BWL610792S, BWL610795S, BWL610796S, BWL610802S, BWL610808S, BWL610809S, BWL610814S, BWL610815S, BWL610818S, BWL610827S, BWL610829S, BWL610833S, BWL610834S, BWL610835S, BWL610836S, BWL610837S, BWL610838S, BWL610848S, BWL610849S, BWL610858S, BWL610859S, BWL610868S, BWL610887S, BWL610888S, BWL610889S, BWL610891S, BWL610892S, BWL610893S, BWL610895S, BWL610900S, BWL610901S, BWL610911S, BWL610912S, BWL610913S, BWL610915S, BWL610916S, BWL610918S, BWL610919S, BWL610920S, BWL610926S, BWL610929S, BWL610930S, BWL610931S, BWL610936S, BWL610937S, BWL610944S, BWL610947S, BWL610948S, BWL610949S, BWL610954S, BWL610955S, BWL610965S, BWL611015S, BWL611018S, BWL611019S, BWL611020S, BWL611022S, BWL611023S, BWL611041S, BWL611058S, BWL611059S, BWL611124S, BWL611127S, BWL611131S, BWL611132S, BWL611134S, BWL611135S, BWL611136S, BWL611137S, BWL611138S, BWL611151S, BWL611156S, BWL611167S, BWL611169S, BWL611172S, BWL611180S, BWL611181S, BWL611207S, BWL611208S, BWL611209S, BWL611210S, BWL611223S, BWL611225S, BWL611226S, BWL611229S, BWL611231S, BWL611233S, BWL611245S, BWL611247S, BWL611248S, BWL611249S, BWL611250S, BWL611251S, BWL611270S, BWL611271S, BWL611289S, BWL611293S, BWL611296S, BWL611298S, BWL611302S, BWL611311S, BWL611312S, BWL611313S, BWL611326S, BWL611359S, BWL611381S, BWL611382S, BWL611388S, BWL611401S, BWL611402S, BWL611403S, BWL611413S, BWL611438S, BWL611440S, BWL611441S, BWL611442S, BWL611452S, BWL611453S, BWL611454S, BWL611455S, BWL611456S, BWL611457S, BWL611462S, BWL611463S, BWL611485S, BWL611493S, BWL611499S, BWL611504S, BWL611508S, BWL611509S, BWL611510S, BWL611511S, BWL611512S, BWL611513S, BWL611514S, BWL611520S, BWL611532S, BWL611533S, BWL611535S, BWL611542S, BWL611547S, BWL611548S, BWL611568S, BWL611569S, BWL611582S, BWL611583S, BWL611584S, BWL611585S, BWL611586S, BWL611598S, BWL611599S, BWL611601S, BWL611602S, BWL611603S, BWL611607S, BWL611608S, BWL611609S, BWL611610S, BWL611611S, BWL611612S, BWL611621S, BWL611622S, BWL611623S, BWL611625S, BWL611631S, BWL611635S, BWL611647S, BWL611680S, BWL611681S, BWL611682S, BWL611683S, BWL611684S, BWL611688S, BWL611691S, BWL611694S, BWL611696S, BWL611699S, BWL611703S, BWL611704S, BWL611705S, BWL611713S, BWL611714S, BWL611715S, BWL611717S, BWL611738S, BWL611740S, BWL611744S, BWL611745S, BWL611746S, BWL611753S, BWL611754S, BWL611759S, BWL611760S, BWL611774S, BWL611785S, BWL611796S, BWL611809S, BWL611811S, BWL611812S, BWL611835S, BWL611836S, BWL611939S, BWL611965S, BWL611967S, BWL611968S, BWL612000S, BWL612001S, BWL612002S, BWL612003S, BWL612004S, BWL612006S, BWL612007S, BWL612008S, BWL612009S, BWL612011S, BWL612013S, BWL612016S, BWL612017S, BWL612018S, BWL612019S, BWL612020S, BWL612032S, BWL612033S, BWL612034S, BWL612036S, BWL612037S, BWL612042S, BWL612043S, BWL612044S, BWL612045S, BWL612046S, BWL612047S, BWL612048S, BWL612049S, BWL612065S, BWL612066S, BWL612068S, BWL612069S, BWL612076S, BWL612077S, BWL612078S, BWL612079S, BWL612080S, BWL612083S, BWL612089S, BWL612093S, BWL612094S, BWL612098S, BWL612099S, BWL612100S, BWL612103S, BWL612159S, BWL612160S, BWL612161S, BWL612163S, BWL612164S, BWL612165S, BWL612166S, BWL612170S, BWL612175S, BWL612176S, BWL612177S, BWL612178S, BWL612184S, BWL612185S, BWL612188S, BWL612189S, BWL612200S, BWL612201S, BWL612202S, BWL612203S, BWL612209S, BWL612210S, BWL612215S, BWL612216S, BWL612217S, BWL612218S, BWL612219S, BWL612220S, BWL612221S, BWL612222S, BWL612231S, BWL612232S, BWL612233S, BWL612234S, BWL612235S, BWL612237S, BWL612238S, BWL612278S, BWL612279S, BWL612290S, BWL612291S, BWL612292S, BWL612293S, BWL612295S, BWL612296S, BWL612297S, BWL612300S, BWL612301S, BWL612305S, BWL612306S, BWL612307S, BWL612308S, BWL612309S, BWL612310S, BWL612311S, BWL612312S, BWL612313S, BWL612314S, BWL612315S, BWL612318S, BWL612319S, BWL612320S, BWL612321S, BWL612322S, BWL612327S, BWL612330S, BWL612333S, BWL612348S, BWL612349S, BWL612350S, BWL612351S, BWL612352S, BWL612353S, BWL612354S, BWL612355S, BWL612389S, BWL612393S, BWL612405S, BWL612413S, BWL612414S, BWL612415S, BWL612416S, BWL612428S, BWL610324S, BWL610325S, BWL610348S, BWL610465S, BWL610864S, BWL610866S, BWL610869S, BWL610329S, BWL610330S, BWL610333S, BWL611115S, BWL611125S, BWL611818S, BWL610259S, BWL610269S, BWL611055S, BWL611072S, BWL611096S, BWL611255S, BWL611256S, BWL611257S, BWL611258S, BWL611261S, BWL611268S, BWL611269S, BWL611272S, BWL611285S, BWL611295S, BWL611307S, BWL611309S, BWL611370S, BWL611698S, BWL611708S, BWL611742S, BWL611747S, BWL611748S, BWL611751S, BWL611756S, BWL611758S, BWL611763S, BWL611768S, BWL611770S, BWL611771S, BWL611772S, BWL611773S, BWL611775S, BWL611776S, BWL611777S, BWL611790S, BWL611860S, BWL611792S, BWL611861S, BWL611876S, BWL611877S, BWL611878S, BWL611879S, BWL611408S, BWL611411S, BWL611467S, BWL611495S, BWL611527S, BWL611543S, BWL611596S, BWL611597S, BWL611604S, BWL611605S, BWL611606S, BWL611734S, BWL611736S, BWL611798S, BWL611807S, BWL611808S, BWL611952S, BWL612028S, BWL612260S, BWL612261S, BWL611863S, BWL611864S, BWL611865S, BWL611868S, BWL611869S, BWL611870S, BWL611871S, BWL612362S, BWL612469S, BWL612470S, BWL611828S, BWL612380S, BWL612381S, BWL612382S, BWL612383S, BWL612384S, BWL612417S, BWL612418S, BWL612419S, BWL612420S, BWL612421S, BWL612433S, BWL612434S; GTIN 00763000612146, Lot Serial Numbers: BWL612462S, BWL612460S
Recalling Firm/
Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
8200 Coral Sea St Ne
Mounds View MN 55112-4391
For Additional Information Contact Technical Services
800-929-4043
Manufacturer Reason
for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA Determined
Cause 2
Under Investigation by firm
Action Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Quantity in Commerce 5296 units
Distribution US Nationwide - Worldwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA Database PMAs with Product Code = LWS and Original Applicant = MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENT
PMAs with Product Code = LWS and Original Applicant = Medtronic, Inc.
-
-