 Class 2 Device Recall Intera 1.5T Power/Pulsarsee related information
Date Initiated by FirmMarch 12, 2024
Date PostedApril 17, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1582-2024
Recall Event ID 94233
510(K)NumberK193215 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
ProductIntera 1.5T Power/Pulsar, Magnetic Resonance System.
Code Information Model No. 781105; UDI-DI n/a; Serial No. 8342, 8104, 8899, 8557, 8797, 8665, 8154, 8590, 8333, 8418, 8615, 8963, 8846, 8216, 8112, 8296, 8384, 8626, 8093, 8217, 10209, 10093, 8738, 10212, 8622, 8040, 8748, 8695, 8188, 8326, 8684, 8854, 8878, 8936, 8974, 8935, 10156, 10140, 8337, 10510, 10920, 10488, 10517, 8588, 8158, 8405, 8610, 8017, 10479, 8381, 9003, 10163, 10470, 9027, 8032, 8207, 8280, 8283, 8293, 8439, 8618, 8688, 10232, 8838, 8605, 10120, 10129, 8420, 10080, 8546, 8548, 8434, 10500, 10187, 9029, 8676, 8836, 8685, 8469, 8505, 8470, 8277, 8991, 8510, 8633, 8357, 8048, 10159, 8643, 8414, 8708, 8925, 8130, 8617, 8818, 8027, 10039, 8335, 8952, 8663, 8549, 8034, 10108, 8956, 8806, 8916, 8791, 8393, 8058, 8455, 8033, 8558, 8579, 10178, 10526.
FEI Number 3016618143
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
For Additional Information ContactCustomer Care Solution Center
1-800-722-9377
Manufacturer Reason
for Recall
Patient support table floor plate may be incorrectly installed.
FDA Determined
Cause 2
Process design
ActionConsignees were mailed an URGENT Medical Device Correction notice, dated 3/12/24. Consignees are instructed by the notice to identify affected systems, which can continue to be used with the provided instruction. If a patient's weight is at or near the maximum load of the device take care that they do not sit on the end of the tabletop opposite the bore entrance and that they do not hop down from their sitting position while the support is at its highest position. If the affected system has any unexpected movement or becomes unstable immediately stop use and contact your Philips representative for interim support. The recall notice is to be circulated to all users and the provided advisory displayed. Completed response forms are to be emailed to pd.cnr@philips.com.
Quantity in Commerce115 units
DistributionWorldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, Turkey, T'rkiye, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.