| | Class 2 Device Recall Philips |  |
| Date Initiated by Firm | May 03, 2024 |
| Date Posted | May 17, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1849-2024 |
| Recall Event ID |
94508 |
| 510(K)Number | K193215 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | Ingenia 3.0T CX
Model Number (REF):
(1) 782105
(2) 781271 |
| Code Information |
UDI-DI:
(1) (01)00884838098312(21);
(2) (01)00884838068452(21)
Serial Number
78121
78150
78157
78096
78208
78094
78012
78074
78154
78161
78170
78171
78172
78173
78182
78183
78189
78190
78196
78198
78205
78210
78212
78213
78216
78218
78221
78225
78233
78235
78238
78240
78241
78242
78243
78244
78250
78251
78252
78253
78256
78257
78262
78263
78265
78266
78267
78268
78269
78270
78271
78272
78273
78279
78280
78281
78282
78283
78284
78288
78289
78291
78292
78293
78294
78295
78296
78301
78305
78306
78307
78308
78309
78310
78312
78313
78314
78316
78317
78319
78320
78321
78322
78323
78327
78328
78329
78330
78332
78333
78334
78335
78338
78339
78340
78341
78342
78343
78345
78346
78347
78352
78353
78354
78356
78357
78358
78359
78360
78361
78362
78363
78364
78366
78368
78370
78374
78375
78376
78377
78386
78078
78080
78224
78098
78003
78017
78065
78070
78076
78081
78087
78097
78099
78101
78104
78107
78126
78149
78151
78181
78206
78219
78264
23032
78063
78064
78106
78187
78000
78005
78007
78009
78010
78015
78016
78018
78019
78021
78023
78025
78027
78029
78031
78032
78033
78035
78038
78039
78040
78042
78043
78045
78048
78049
78050
78051
78062
78067
78068
78072
78082
78083
78084
78088
78089
78091
78095
78100
78109
78111
78113
78114
78115
78117
78127
78129
78132
78133
78139
78140
78145
78147
78148
78155
78199
78203
78207
78214
78222
78223
78004
78008
78053
78011
78013
78014
78028
78034
78044
78046
78047
78058
78060
78066
78071
78073
78075
78077
78086
78092
78102
78103
78110
78112
78118
78120
78122
78123
78128
78135
78137
78138
78141
78142
78158
78162
78164
78166
78167
78175
78180
78186
78188
78193
78194
78195
78197
78201
78202
78209
78215
78217
78227
78229
78232
78234
78239
78247
78254
78255
78259
78260
78261
78277
78286
78287
78290
78297
78298
78299
78324
78326
78336
78337
78030
78036
78059
78156
78249
78026
78022
78516
78517
78518
78519
78521
78522
78523
78524
78525
78526
78527
78528
78529
78530
78531
78536
78537
78539
78545
78546
78547
78548
78549
78550
78551
78553
78554
78556
78557
78559
78560
78561
78562
78564
78565
78567
78500
78501
78503
78504
78505
78507
78508
78512
78513
78532
78533
78535
78538
78540
78541
78542
78543
78544
78552
78566
78568
|
| FEI Number |
3016618143
|
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
| For Additional Information Contact | Customer Care Solution Center 800-722-9377 |
Manufacturer Reason for Recall | The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's
technical room, resulting in Smoke and/or fire |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Philips issued Urgent Medical Device Correction notifications to customers on 22-Apr-2024 via FedEx . Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take:
If a smoke/fire alarm is detected:
a. Immediately stop scanning and evacuate the patient and staff from the examination room.
b. If a developing fire is detected, adhere to established hospital fire emergency procedures, which may include switching off power to the complete system and/or removing the magnet field by using the Emergency Magnet Off button.
c. Do not attempt to continue scanning.
d. Immediately contact Philips Service.
" Ensure all users are aware of facility specific Emergency Procedures as outlined in Chapter 2: Safety in the Instructions for Use Emergency procedures The User is required to establish emergency procedures for the following situations:
" A medical emergency
" A fire
" An emergency that requires immediate removal of the magnetic field
" The release of helium gas into the examination room
Philips MRI systems have an Emergency Table Stop button in case there is an emergency
during tabletop movement.
" Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue.
" Post this notice near the affected MR system(s) for ease of reference.
Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt via email to: pd.cnr@philips.com. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken.
5. The actions planned by Philips to correct the problem
Philips will contact you to schedule time for a Field Service Engineer (FSE) to visit your site to inspect the g-MDU connections in the technical room and apply the proper torque to the connection if necessary.(reference FCO78100582). Philips plans t |
| Quantity in Commerce | 350 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada,
China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan,
Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia,
Spain, Sweden, Switzerland, Thailand, United Kingdom.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LNH
|
|
|
|