| Date Initiated by Firm | March 21, 2025 |
| Date Posted | June 09, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1820-2025 |
| Recall Event ID |
96902 |
| 510(K)Number | K203346 K233669 |
| Product Classification |
Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
|
| Product | OEC 3D |
| Code Information |
Two populations are potentially affected by this notification.
1. Listed in Appendix A - OEC Elite Mobile C-arm Systems (GTIN: 00001234567895) and OEC 3D Mobile C-arm Systems (GTIN: 00195278096944) that potentially have an X-ray field size greater than the 4% SID limit and do not contain the acceptance limit for the X-ray field size measurement in the System Service Manual.
2. Systems NOT listed in Appendix A - OEC Elite Mobile C-arm Systems (GTIN: 00001234567895) and OEC 3D Mobile C-arm Systems (GTIN: 00195278096944) that are within the allowed SID limit. However, the acceptance limit for the X-ray field size measurement is not stated in the System Service Manual.
Appendix A
Serial Numbers
FAHXTX00002
FAHXTX00005
FAHXTX00014
FAHXTX00018
FAHXTX00019
FAHXTX00020
FAHXTX00024
FAHXTX00028
FAHXTX00034
FAHXTX00037
FAHXTX00043
FAHXTX00052
FAHXTX00063
FAHXTX00066
FAHXTX00068
FAHXTX00070
FAHXTX00075
FAHXTX00078
FAHXTX00079
FAHXTX00080
FAHXTX00081
FAHXTX00083
FAHXTX00092
FAHXTX00093
FAHXTX00097
FAHXTX00110
FAHXTX00111
FAHXTX00112
FAHXTX00114
FAHXTX00115
FAHXTX00116
FAHXTX00117
FAHXTX00118
FAHXTX00119
FAHXTX00120
FAHXTX00121
FAHXTX00122
FAHXTX00124
FAHXTX00125
FAHXTX00135
FAHXTX00137
FAHXTX00138
FAHXTX00151
FAHXTX00152
FAHXTX00155
FAHXTX00156
FAHXTX00162
FAHXTX00163
FAHXTX00166
FAHXTX00170
FAHXTX00171
FAHXTX00177
FAHXTX00178
FAHXTX00188
FAHXTX00192
FAHXTX00193
FAHXTX00202
FAHXTX00204
FAHXTX00211
FAHXTX00212
FAHXTX00217
FAHXTX00219
FAHXTX00220
FAHXTX00221
FAHXTX00222
FAHXTX00223
FAHXTX00224
FAHXTX00226
FAHXTX00227
FAHXTX00229
FAHXTX00230
FAHXTX00236
FAHXTX00240
FAHXTX00244
FAHXTX00261
FAHXTX00265
FAHXTX00266
FAHXTX00269
FAHXTX00271
FAHXTX00272
FAHXTX00273
FAHXTX00274
FAHXTX00280
FAHXTX00283
FAHXTX00284
FAHXTX00296
FAHXTX00299
FAHXTX00300
FAHXTX00304
FAHXTX00305
FAHXTX00308
FAHXTX00312
FAHXTX00313
FAHXTX00320
FAHXTX00323
FAHXTX00325
FAHXTX00332
FAHXTX00340
FAHXTX00342
FAHXTX00343
FAHXTX00352
FAHXTX00353
FAHXTX00354
FAHXTX00357
FAHXTX00358
FAHXTX00359
FAHXTX00360
FAHXTX00363
FAHXTX00376
FAHXTX00379
FAHXTX00381
FAHXTX00382
FAHXTX00390
FAHXTX00392
FAHXTX00399
FAHXTX00400
FAHXTX00410
FAHXTX00413
FAHXTX00414
FAHXTX00415
FAHXTX00420
FAHXTX00421
FAHXTX00429
FAHXTX00439
FAHXTX00453
FAHXTX00455
FAHXTX00456
FAHXTX00463
FAHXTX00465
FAHXTX00466
FAHXTX00479
FAHXTX00481
FAHXTX00483
FAHXTX00484
FAHXTX00486
FAHXTX00488
FAHXTX00495
FAHXTX00497
FAHXTX00498
FAHXTX00507
FAHXTX00509
FAHXTX00513
FAHXTX00523
FAHXTX00530
FAHXTX00534
FAHXTX00538
FAHXTX00539
FAHXTX00540
FAHXTX00547
FAHXTX00549
FAHXTX00555
FAHXTX00557
FAHXTX00560
FAHXTX00565
FAHXTX00569
FAHXTX00570
FAHXTX00571
FAHXTX00572
FAHXTX00575
FAHXTX00576
FAHXTX00579
FAHXTX00592
FAHXTX00595
FAHXTX00610
FAHXTX00611
FAHXXX00029
FAHXXX00030
FAHXXX00038
FAHXXX00040
FAHXXX00056
FAHXXX00057
FAHXXX00059
FAHXXX00060
FAHXXX00061
FAHXXX00069
FAHXXX00082
FAHXXX00085
FAHXXX00086
FAHXXX00089
FAHXXX00091
FAHXXX00093
FAHXXX00097
FAHXXX00098
FAHXXX00099
FAHXXX00105
FAHXXX00106
FAHXXX00112
FAHXXX00113
FAHXXX00114
FAHXXX00116
FAHXXX00117
FAHXXX00119
FAHXXX00123
FAHXXX00142
FAHXXX00144
FAHXXX00145
FAHXXX00147
FAHXXX00150
FAHXXX00151
FAHXXX00153
FAHXXX00155
FAHXXX00164
FAHXXX00173
FAHXXX00176
FAHXXX00178
FAHXXX00179
FAHXXX00188
FAHXXX00190
FAHXXX00193
FAHXXX00196
FAHXXX00199
FAHXXX00200
FAHXXX00207
FAHXXX00208
FAHXXX00209
FAMHTX00001
FAMHTX00005
FAMHTX00007
FAMHTX00012
FAMHTX00014
FAMHTX00041
FAMHTX00051
FAMHTX00055
FAMHTX00064
FAMHTX00069
FAMHTX00081
FAMHTX00085
FAMHXX00003
FAMHXX00011
FAMHXX00019
FAMHXX00023
FAXXTE00003
FAXXTE00006
FAXXTE00009
FAXXTE00011
FAXXTE00013
FAXXTE00014
FAXXTE00015
FAXXTE00017
FAXXTE00021
FAXXTE00022
FAXXTE00024
FAXXTE00025
FAXXTE00028
FAXXTE00031
FAXXTE00032
FAXXTE00033
FAXXTE00034
FAXXTE00035
FAXXTE00037
FAXXTE00038
FAXXTE00039
FAXXTE00040
FAXXTE00044
FAXXTE00046
FAXXTE00047
FAXXTE00049
FAXXTE00050
FAXXTE00051
FAXXTE00052
FAXXTE00056
FAXXTE00062
FAXXTE00063
FAXXTE00064
FAXXTE00067
FAXXTE00071
FAXXTE00072
FAXXTE00073
FAXXTE00076
FAXXTE00084
FAXXTE00085
FAXXTE00091
FAXXTE00092
FAXXTE00094
FAXXTE00095
FAXXTE00099
FAXXTE00100
FAXXTE00101
FAXXTE00104
FAXXTE00105
FAXXTE00106
FAXXTE00107
FAXXTE00109
FAXXTE00110
FAXXTE00114
FAXXTE00116
FAXXTE00117
FAXXTE00120
FAXXTE00121
FAXXTE00128
FAXXTE00129
FAXXTE00134
FAXXTE00136
FAXXTE00137
FAXXTE00141
FAXXTE00144
FAXXTE00145
FAXXTE00147
FAXXTE00149
FAXXTE00163
FAXXTE00164
FAXXTE00170
FAXXTE00173
FAXXTE00175
FAXXTE00179
FAXXTE00180
FAXXTE00182
FAXXTE00184
FAXXTE00185
FAXXTE00195
FAXXTE00197
FAXXTE00205
FAXXTE00212
FAXXTE00220
FAXXTE00224
FAXXTE00225
FAXXTE00227
FAXXTE00228
FAXXTE00229
FAXXTE00230
FAXXTE00231
FAXXTE00234
FAXXTE00236
FAXXTE00237
FAXXTE00238
FAXXTE00240
FAXXTE00241
FAXXTE00243
FAXXTE00246
FAXXTE00248
FAXXTE00253
FAXXTE00256
FAXXTE00257
FAXXTE00258
FAXXTE00259
FAXXTE00263
FAXXTE00267
FAXXTE00269
FAXXTE00270
FAXXTE00271
FAXXTE00272
FAXXTE00277
FAXXTE00278
FAXXTE00279
FAXXTE00280
FAXXTE00282
FAXXTE00286
FAXXTE00291
FAXXTE00293
FAXXTE00295
FAXXTE00301
FAXXTE00317
FAXXTE00319
FAXXTE00320
FAXXTE00321
FAXXTE00323
FAXXTE00328
FAXXTE00329
FAXXTE00330
FAXXTE00331
FAXXTE00334
FAXXTE00336
FAXXTE00337
FAXXTE00338
FAXXTE00339
FAXXTE00340
FAXXTE00342
FAXXTE00344
FAXXTE00352
FAXXTE00353
FAXXTE00354
FAXXTE00357
FAXXTE00358
FAXXTE00361
FAXXTE00362
FAXXTE00363
FAXXTE00372
FAXXTE00373
FAXXTE00376
FAXXTE00379
FAXXTE00380
FAXXTE00383
FAXXTE00389
FAXXTE00392
FAXXTE00393
FAXXTE00396
FAXXTE00407
FAXXTE00413
FAXXTE00418
FAXXTE00422
FAXXTE00424
FAXXTE00425
FAXXTE00429
FAXXTE00431
FAXXTE00433
FAXXTE00434
FAXXTE00436
FAXXTE00439
FAXXTE00442
FAXXTE00443
FAXXTE00451
FAXXTE00452
FAXXTE00454
FAXXTE00455
FAXXTE00460
FAXXTE00461
FAXXTE00472
FAXXTE00475
FAXXTE00477
FAXXTE00481
FAXXTE00482
FAXXTE00485
FAXXTE00486
FAXXTE00487
FAXXTE00489
FAXXTE00490
FAXXTE00491
FAXXTE00492
FAXXTE00495
FAXXTE00499
FAXXTE00506
FAXXTE00508
FAXXTE00511
FAXXTE00515
FAXXTE00516
FAXXTE00517
FAXXTE00518
FAXXTE00519
FAXXTE00522
FAXXTE00523
FAXXTE00525
FAXXTE00526
FAXXTE00527
FAXXTE00532
FAXXTE00533
FAXXTE00536
FAXXTE00539
FAXXTE00540
FAXXTE00543
FAXXTE00544
FAXXTE00545
FAXXTE00546
FAXXTE00547
FAXXTE00549
FAXXTE00550
FAXXTE00551
FAXXTE00553
FAXXTE00555
FAXXTE00560
FAXXTE00561
FAXXTE00562
FAXXTE00566
FAXXTE00570
FAXXTE00571
FAXXTE00572
FAXXTE00573
FAXXTE00582
FAXXTE00591
FAXXTE00593
FAXXTE00594
FAXXTE00596
FAXXTE00604
FAXXTE00607
FAXXTE00608
FAXXTE00611
FAXXTE00615
FAXXTE00616
FAXXTE00617
FAXXTE00618
FAXXTE00624
FAXXTE00631
FAXXTE00638
FAXXTE00639
FAXXTE00643
FAXXTE00646
FAXXTE00650
FAXXTE00651
FAXXTE00652
FAXXTE00653
FAXXTE00657
FAXXTE00661
FAXXTE00664
FAXXTE00667
FAXXTE00669
FAXXTE00672
FAXXTE00680
FAXXTE00681
FAXXTE00684
FAXXTE00685
FAXXTE00687
FAXXTE00688
FAXXTE00701
FAXXTE00706
FAXXTE00707
FAXXTE00710
FAXXTE00712
FAXXTE00713
FAXXTE00716
FAXXTE00718
FAXXTE00720
FAXXTE00724
FAXXTE00725
FAXXTE00728
FAXXTE00731
FAXXTE00734
FAXXTE00735
FAXXTE00737
FAXXTE00738
FAXXTE00747
FAXXTE00748
FAXXTE00749
FAXXTE00752
FAXXTE00753
FAXXTE00762
FAXXTE00766
FAXXTE00772
FAXXTE00789
FAXXTE00794
FAXXTE00799
FAXXTE00801
FAXXTE00805
FAXXTE00809
FAXXTE00810
FAXXTE00811
FAXXTE00814
FAXXTE00819
FAXXTE00820
FAXXTE00829
FAXXTE00831
FAXXTE00843
FAXXTE00853
FAXXTE00854
FAXXTE00874
FAXXTE00881
FAXXTE00893
FAXXTE00894
FAXXTE00896
FAXXTE00897
FAXXTE00899
FAXXTE00902
FAXXTE00903
FAXXTE00906
FAXXTE00909
FAXXTE00914
FAXXTE00915
FAXXTE00916
FAXXTE00917
FAXXTE00918
FAXXTE00919
FAXXTE00922
FAXXTE00925
FAXXTE00930
FAXXTE00932
FAXXTE00936
FAXXTE00939
FAXXTE00940
FAXXTE00945
FAXXTE00958
FAXXTE00961
FAXXTE00964
FAXXTE00970
FAXXTE00973
FAXXTE00979
FAXXTE00980
FAXXTE00983
FAXXTE00985
FAXXTE00987
FAXXTE00990
FAXXTE00993
FAXXTE00995
FAXXTE00996
FAXXTE00999
FAXXTE01006
FAXXTE01029
FAXXTE01030
FAXXTE01048
FAXXTE01053
FAXXTE01057
FAXXTE01064
FAXXTE01068
FAXXTE01084
FAXXTE01085
FAXXTE01086
FAXXTE01088
FAXXTE01089
FAXXTE01092
FAXXTE01114
FAXXTE01124
FAXXTE01139
FAXXTE01140
FAXXTE01141
FAXXTE01145
FAXXTE01154
FAXXTE01156
FAXXTE01159
FAXXTE01163
FAXXTE01166
FAXXTE01167
FAXXTE01168
FAXXTE01169
FAXXTE01171
FAXXTE01180
FAXXTE01183
FAXXTE01191
FAXXTE01193
FAXXTE01194
FAXXTE01195
FAXXTE01204
FAXXTE01205
FAXXTE01206
FAXXTE01210
FAXXTE01212
FAXXTE01219
FAXXTE01222
FAXXTP00003
FAXXTP00004
FAXXTP00007
FAXXTP00009
FAXXTP00010
FAXXTP00011
FAXXTP00012
FAXXTP00014
FAXXTP00018
FAXXTP00019
FAXXTP00020
FAXXXE00003
FAXXXE00004
FAXXXE00005
FAXXXE00007
FAXXXE00011
FAXXXE00012
FAXXXE00013
FAXXXE00017
FAXXXE00020
FAXXXE00021
FAXXXE00023
FAXXXE00024
FAXXXE00025
FAXXXE00026
FAXXXE00030
FAXXXE00032
FAXXXE00033
FAXXXE00036
FAXXXE00037
FAXXXE00038
FAXXXE00039
FAXXXE00043
FAXXXE00044
FAXXXE00045
FAXXXE00046
FAXXXE00047
FAXXXE00049
FAXXXE00053
FAXXXE00054
FAXXXE00056
FAXXXE00062
FAXXXE00071
FAXXXE00072
FAXXXE00076
FAXXXE00081
FAXXXE00082
FAXXXE00083
FAXXXE00085
FAXXXE00087
FAXXXE00091
FAXXXE00092
FAXXXE00093
FAXXXE00095
FAXXXE00096
FAXXXE00097
FAXXXE00098
FAXXXE00100
FAXXXE00105
FAXXXE00107
FAXXXE00109
FAXXXE00117
FAXXXE00120
FAXXXE00121
FAXXXE00124
FAXXXE00141
FAXXXE00147
FAXXXE00152
FAXXXE00155
FAXXXE00159
FAXXXE00164
FAXXXE00165
FAXXXE00167
FAXXXE00172
FAXXXE00174
FAXXXE00176
FAXXXE00180
FAXXXE00181
FAXXXE00183
FAXXXE00184
FAXXXE00185
FAXXXE00188
FAXXXE00196
FAXXXE00197
FAXXXE00198
FAXXXE00199
FAXXXE00200
FAXXXE00204
FAXXXE00206
FAXXXE00208
FAXXXE00210
FAXXXE00215
FAXXXE00217
FAXXXE00218
FAXXXE00222
FAXXXE00225
FAXXXE00227
FAXXXE00228
FAXXXE00229
FAXXXE00230
FAXXXE00231
FAXXXE00232
FAXXXE00235
FAXXXE00236
FAXXXE00238
FAXXXE00240
FAXXXE00244
FAXXXE00247
FAXXXE00251
FAXXXE00253
FAXXXE00254
FAXXXE00255
FAXXXE00256
FAXXXE00257
FAXXXE00261
FAXXXE00269
FAXXXE00270
FAXXXE00273
FAXXXE00275
FAXXXE00278
FAXXXE00279
FAXXXE00280
FAXXXE00282
FAXXXE00283
FAXXXE00287
FAXXXE00288
FAXXXE00290
FAXXXE00291
FAXXXE00292
FAXXXE00296
FAXXXE00297
FAXXXE00302
FAXXXE00303
FAXXXE00306
FAXXXE00312
FAXXXE00314
FAXXXE00315
FAXXXE00318
FAXXXE00319
FAXXXE00320
FAXXXE00321
FAXXXE00324
FAXXXE00325
FAXXXE00327
FAXXXE00328
FAXXXE00329
FAXXXE00330
FAXXXE00332
FAXXXE00335
FAXXXE00336
FAXXXE00339
FAXXXE00340
FAXXXE00343
FAXXXE00345
FAXXXE00346
FAXXXE00348
FAXXXE00354
FAXXXE00356
FAXXXE00359
FAXXXX00014
FAXXXX00020
FBHVXX00037
FBHVXX00040
FBHXTX00002
FBHXTX00006
FBHXTX00007
FBHXTX00009
FBHXTX00010
FBHXTX00011
FBHXTX00013
FBHXTX00019
FBHXTX00020
FBHXTX00021
FBHXTX00022
FBHXTX00023
FBHXTX00024
FBHXTX00025
FBHXTX00026
FBHXTX00031
FBHXTX00032
FBHXTX00033
FBHXTX00035
FBHXTX00038
FBHXTX00045
FBHXTX00047
FBHXTX00048
FBHXTX00051
FBHXTX00057
FBHXTX00059
FBHXTX00062
FBHXTX00068
FBHXTX00070
FBHXTX00071
FBHXTX00074
FBHXTX00079
FBHXTX00080
FBHXTX00081
FBHXTX00095
FBHXTX00098
FBHXTX00102
FBHXTX00103
FBHXTX00104
FBHXTX00106
FBHXTX00114
FBHXTX00115
FBHXTX00125
FBHXTX00126
FBHXTX00128
FBHXTX00130
FBHXTX00137
FBHXTX00144
FBHXTX00149
FBHXTX00151
FBHXTX00154
FBHXTX00159
FBHXTX00161
FBHXTX00163
FBHXTX00172
FBHXTX00173
FBHXTX00177
FBHXTX00184
FBHXTX00198
FBHXTX00199
FBHXTX00201
FBHXTX00203
FBHXTX00204
FBHXTX00205
FBHXTX00208
FBHXTX00210
FBHXTX00218
FBHXTX00223
FBHXTX00231
FBHXTX00233
FBHXTX00234
FBHXTX00235
FBHXTX00237
FBHXTX00241
FBHXTX00261
FBHXTX00266
FBHXTX00274
FBHXTX00275
FBHXTX00276
FBHXTX00277
FBHXTX00278
FBHXTX00279
FBHXTX00281
FBHXTX00282
FBHXTX00287
FBHXTX00290
FBHXTX00295
FBHXTX00316
FBHXTX00317
FBHXTX00319
FBHXTX00320
FBHXTX00324
FBHXTX00330
FBHXTX00335
FBHXTX00336
FBHXTX00337
FBHXTX00338
FBHXTX00339
FBHXTX00343
FBHXTX00344
FBHXTX00345
FBHXTX00348
FBHXTX00349
FBHXTX00350
FBHXTX00351
FBHXTX00352
FBHXTX00353
FBHXTX00354
FBHXTX00355
FBHXTX00356
FBHXTX00357
FBHXTX00360
FBHXTX00361
FBHXTX00362
FBHXTX00365
FBHXTX00366
FBHXTX00368
FBHXTX00369
FBHXTX00371
FBHXTX00375
FBHXTX00379
FBHXTX00381
FBHXTX00411
FBHXTX00415
FBHXTX00416
FBHXTX00417
FBHXTX00429
FBHXTX00431
FBHXTX00432
FBHXTX00433
FBHXTX00436
FBHXTX00438
FBHXTX00442
FBHXTX00444
FBHXTX00447
FBHXTX00448
FBHXTX00451
FBHXTX00458
FBHXTX00462
FBHXTX00468
FBHXTX00479
FBHXTX00482
FBHXTX00490
FBHXTX00491
FBHXTX00493
FBHXTX00498
FBHXTX00499
FBHXTX00500
FBHXTX00501
FBHXTX00503
FBHXTX00504
FBHXTX00505
FBHXTX00507
FBHXTX00508
FBHXTX00509
FBHXTX00510
FBHXTX00514
FBHXTX00516
FBHXTX00520
FBHXTX00524
FBHXTX00529
FBHXTX00535
FBHXTX00537
FBHXTX00541
FBHXTX00543
FBHXTX00546
FBHXTX00549
FBHXTX00551
FBHXTX00555
FBHXTX00558
FBHXTX00561
FBHXTX00565
FBHXTX00568
FBHXTX00570
FBHXTX00571
FBHXTX00580
FBHXTX00583
FBHXTX00584
FBHXTX00586
FBHXTX00587
FBHXTX00589
FBHXTX00593
FBHXTX00597
FBHXTX00599
FBHXTX00600
FBHXTX00601
FBHXTX00605
FBHXTX00606
FBHXTX00613
FBHXTX00614
FBHXTX00616
FBHXTX00617
FBHXTX00624
FBHXTX00630
FBHXTX00641
FBHXTX00642
FBHXTX00644
FBHXTX00649
FBHXTX00650
FBHXTX00651
FBHXTX00655
FBHXTX00664
FBHXTX00665
FBHXTX00667
FBHXTX00669
FBHXTX00671
FBHXTX00672
FBHXTX00673
FBHXTX00679
FBHXTX00680
FBHXTX00682
FBHXTX00684
FBHXTX00686
FBHXTX00687
FBHXTX00689
FBHXTX00690
FBHXTX00691
FBHXTX00693
FBHXTX00698
FBHXTX00700
FBHXTX00703
FBHXTX00713
FBHXTX00719
FBHXTX00721
FBHXTX00727
FBHXTX00737
FBHXTX00745
FBHXTX00748
FBHXTX00753
FBHXTX00755
FBHXTX00757
FBHXTX00758
FBHXTX00760
FBHXTX00762
FBHXTX00764
FBHXTX00769
FBHXTX00770
FBHXTX00771
FBHXTX00774
FBHXTX00783
FBHXTX00785
FBHXTX00798
FBHXTX00800
FBHXTX00803
FBHXTX00804
FBHXTX00821
FBHXTX00823
FBHXTX00826
FBHXTX00834
FBHXTX00837
FBHXTX00839
FBHXTX00841
FBHXTX00843
FBHXTX00844
FBHXTX00847
FBHXTX00848
FBHXTX00849
FBHXTX00866
FBHXTX00867
FBHXTX00894
FBHXTX00896
FBHXTX00897
FBHXTX00898
FBHXTX00899
FBHXTX00901
FBHXTX00909
FBHXTX00912
FBHXTX00913
FBHXTX00914
FBHXTX00916
FBHXTX00920
FBHXTX00921
FBHXTX00922
FBHXTX00924
FBHXTX00925
FBHXTX00926
FBHXTX00930
FBHXTX00933
FBHXTX00934
FBHXTX00936
FBHXTX00937
FBHXTX00938
FBHXTX00942
FBHXTX00943
FBHXTX00946
FBHXTX00952
FBHXTX00954
FBHXTX00955
FBHXTX00964
FBHXTX00971
FBHXTX00972
FBHXTX00974
FBHXTX00977
FBHXTX00980
FBHXTX00982
FBHXTX00986
FBHXTX00991
FBHXTX00995
FBHXTX00996
FBHXTX01002
FBHXTX01006
FBHXTX01013
FBHXTX01014
FBHXTX01016
FBHXTX01031
FBHXTX01040
FBHXTX01045
FBHXTX01048
FBHXTX01049
FBHXTX01050
FBHXTX01062
FBHXTX01063
FBHXTX01070
FBHXTX01074
FBHXTX01075
FBHXTX01076
FBHXTX01077
FBHXTX01078
FBHXTX01079
FBHXTX01080
FBHXTX01100
FBHXTX01101
FBHXTX01102
FBHXTX01104
FBHXTX01108
FBHXTX01109
FBHXTX01110
FBHXTX01111
FBHXTX01119
FBHXTX01123
FBHXTX01126
FBHXTX01127
FBHXTX01130
FBHXTX01131
FBHXTX01132
FBHXTX01135
FBHXTX01138
FBHXTX01139
FBHXTX01140
FBHXTX01145
FBHXTX01147
FBHXTX01150
FBHXTX01153
FBHXTX01154
FBHXTX01156
FBHXTX01157
FBHXTX01158
FBHXTX01165
FBHXTX01170
FBHXTX01176
FBHXTX01180
FBHXTX01184
FBHXTX01189
FBHXTX01193
FBHXTX01194
FBHXTX01200
FBHXTX01201
FBHXTX01202
FBHXTX01217
FBHXTX01222
FBHXTX01223
FBHXTX01228
FBHXTX01229
FBHXTX01230
FBHXTX01231
FBHXTX01232
FBHXTX01233
FBHXTX01234
FBHXTX01239
FBHXTX01240
FBHXTX01241
FBHXTX01244
FBHXTX01246
FBHXTX01251
FBHXTX01252
FBHXTX01264
FBHXTX01265
FBHXTX01267
FBHXTX01268
FBHXTX01269
FBHXTX01270
FBHXTX01271
FBHXTX01273
FBHXTX01286
FBHXTX01287
FBHXTX01289
FBHXTX01291
FBHXTX01303
FBHXTX01306
FBHXTX01311
FBHXTX01319
FBHXTX01322
FBHXTX01323
FBHXTX01326
FBHXTX01330
FBHXTX01331
FBHXTX01348
FBHXTX01356
FBHXTX01368
FBHXTX01371
FBHXTX01373
FBHXTX01374
FBHXTX01381
FBHXTX01382
FBHXTX01387
FBHXTX01388
FBHXTX01389
FBHXTX01397
FBHXTX01400
FBHXTX01405
FBHXTX01410
FBHXTX01421
FBHXXX00016
FBHXXX00027
FBHXXX00032
FBHXXX00036
FBHXXX00041
FBHXXX00042
FBHXXX00045
FBHXXX00050
FBHXXX00054
FBHXXX00056
FBHXXX00060
FBHXXX00061
FBHXXX00063
FBHXXX00065
FBHXXX00067
FBHXXX00069
FBHXXX00074
FBHXXX00077
FBHXXX00078
FBHXXX00081
FBHXXX00085
FBHXXX00088
FBHXXX00091
FBHXXX00092
FBHXXX00093
FBHXXX00094
FBHXXX00097
FBHXXX00099
FBHXXX00100
FBHXXX00101
FBHXXX00102
FBHXXX00104
FBHXXX00105
FBHXXX00106
FBHXXX00107
FBHXXX00108
FBHXXX00109
FBHXXX00110
FBHXXX00111
FBHXXX00113
FBHXXX00117
FBHXXX00118
FBHXXX00119
FBHXXX00120
FBHXXX00121
FBHXXX00122
FBHXXX00126
FBHXXX00127
FBHXXX00132
FBHXXX00133
FBHXXX00134
FBHXXX00135
FBHXXX00143
FBHXXX00148
FBHXXX00156
FBHXXX00158
FBHXXX00159
FBHXXX00160
FBHXXX00166
FBHXXX00169
FBHXXX00170
FBHXXX00171
FBHXXX00179
FBHXXX00180
FBHXXX00182
FBHXXX00185
FBHXXX00186
FBHXXX00187
FBHXXX00189
FBHXXX00192
FBHXXX00196
FBHXXX00202
FBHXXX00204
FBHXXX00205
FBHXXX00206
FBHXXX00207
FBHXXX00208
FBHXXX00211
FBHXXX00213
FBHXXX00218
FBHXXX00225
FBHXXX00231
FBHXXX00238
FBHXXX00239
FBHXXX00240
FBHXXX00243
FBHXXX00246
FBHXXX00247
FBHXXX00248
FBHXXX00249
FBHXXX00251
FBHXXX00256
FBHXXX00257
FBHXXX00265
FBHXXX00266
FBHXXX00269
FBHXXX00270
FBHXXX00273
FBHXXX00275
FBHXXX00277
FBHXXX00280
FBHXXX00281
FBHXXX00285
FBHXXX00286
FBHXXX00288
FBHXXX00290
FBHXXX00291
FBHXXX00292
FBHXXX00293
FBHXXX00298
FBHXXX00300
FBHXXX00301
FBHXXX00303
FBHXXX00304
FBHXXX00305
FBHXXX00309
FBHXXX00311
FBHXXX00313
FBHXXX00319
FBHXXX00320
FBHXXX00321
FBHXXX00322
FBHXXX00326
FBHXXX00327
FBHXXX00330
FBHXXX00331
FBHXXX00332
FBHXXX00333
FBHXXX00338
FBHXXX00339
FBHXXX00342
FBHXXX00352
FBHXXX00353
FBHXXX00356
FBHXXX00361
FBHXXX00362
FBHXXX00365
FBHXXX00367
FBHXXX00368
FBHXXX00371
FBHXXX00372
FBHXXX00374
FBHXXX00389
FBHXXX00390
FBHXXX00391
FBHXXX00393
FBHXXX00394
FBHXXX00395
FBHXXX00401
FBHXXX00402
FBMHDX00013
FBMHDX00015
FBMHDX00018
FBMHDX00019
FBMHDX00020
FBMHDX00021
FBMHDX00024
FBMHDX00025
FBMHDX00026
FBMHDX00028
FBMHDX00029
FBMHDX00030
FBMHDX00043
FBMHDX00044
FBMHDX00046
FBMHDX00047
FBMHDX00049
FBMHDX00054
FBMHDX00056
FBMHDX00058
FBMHDX00059
FBMHDX00061
FBMHDX00065
FBMHDX00066
FBMHDX00067
FBMHDX00071
FBMHDX00073
FBMHDX00078
FBMHDX00080
FBMHDX00081
FBMHDX00082
FBMHDX00083
FBMHDX00085
FBMHDX00086
FBMHDX00087
FBMHDX00089
FBMHDX00090
FBMHDX00092
FBMHDX00095
FBMHDX00097
FBMHDX00098
FBMHDX00100
FBMHDX00101
FBMHDX00104
FBMHDX00107
FBMHDX00108
FBMHDX00110
FBMHDX00111
FBMHDX00112
FBMHDX00118
FBMHDX00119
FBMHDX00120
FBMHDX00122
FBMHDX00124
FBMHDX00125
FBMHDX00126
FBMHDX00127
FBMHDX00129
FBMHDX00130
FBMHDX00133
FBMHDX00134
FBMHDX00136
FBMHDX00137
FBMHDX00139
FBMHDX00140
FBMHDX00141
FBMHDX00143
FBMHDX00144
FBMHDX00147
FBMHDX00151
FBMHDX00152
FBMHDX00153
FBMHDX00155
FBMHDX00157
FBMHDX00163
FBMHDX00166
FBMHDX00168
FBMHDX00169
FBMHDX00170
FBMHDX00173
FBMHDX00177
FBMHDX00180
FBMHDX00182
FBMHDX00184
FBMHDX00186
FBMHDX00187
FBMHDX00188
FBMHDX00189
FBMHDX00193
FBMHDX00195
FBMHDX00196
FBMHDX00197
FBMHDX00198
FBMHDX00200
FBMHDX00201
FBMHDX00203
FBMHDX00208
FBMHDX00210
FBMHDX00212
FBMHDX00216
FBMHDX00218
FBMHDX00220
FBMHDX00221
FBMHDX00224
FBMHDX00226
FBMHDX00227
FBMHDX00231
FBMHDX00237
FBMHDX00238
FBMHDX00239
FBMHDX00240
FBMHDX00245
FBMHDX00250
FBMHDX00252
FBMHDX00253
FBMHDX00255
FBMHDX00262
FBMHDX00264
FBMHDX00266
FBMHDX00271
FBMHDX00273
FBMHDX00282
FBMHDX00284
FBMHDX00286
FBMHDX00287
FBMHDX00290
FBMHDX00292
FBMHDX00294
FBMHDX00296
FBMHDX00297
FBMHDX00301
FBMHDX00302
FBMHDX00310
FBMHDX00313
FBMHDX00314
FBMHDX00318
FBMHDX00321
FBMHDX00326
FBMHDX00328
FBMHDX00333
FBMHDX00335
FBMHDX00346
FBMHDX00350
FBMHDX00352
FBMHDX00363
FBMHDX00366
FBMHDX00369
FBMHDX00374
FBMHDX00375
FBMHDX00379
FBMHDX00386
FBMHDX00387
FBMHDX00388
FBMHDX00389
FBMHDX00395
FBMHDX00397
FBMHDX00398
FBMHDX00399
FBMHDX00400
FBMHDX00401
FBMHDX00403
FBMHDX00406
FBMHDX00409
FBMHDX00410
FBMHDX00412
FBMHDX00413
FBMHDX00416
FBMHDX00423
FBMHDX00424
FBMHDX00425
FBMHDX00430
FBMHDX00431
FBMHDX00437
FBMHDX00439
FBMHDX00440
FBMHDX00441
FBMHDX00445
FBMHDX00449
FBMHDX00451
FBMHDX00452
FBMHDX00453
FBMHDX00458
FBMHDX00459
FBMHDX00460
FBMHDX00461
FBMHDX00462
FBMHDX00463
FBMHDX00464
FBMHDX00466
FBMHDX00468
FBMHDX00471
FBMHDX00477
FBMHDX00480
FBMHDX00487
FBMHDX00490
FBMHDX00491
FBMHDX00500
FBMHDX00509
FBMHDX00510
FBMHDX00516
FBMHDX00518
FBMHDX00521
FBMHDX00522
FBMHDX00523
FBMHDX00524
FBMHDX00525
FBMHDX00527
FBMHDX00528
FBMHDX00531
FBMHDX00532
FBMHDX00534
FBMHDX00542
FBMHDX00543
FBMHDX00544
FBMHDX00552
FBMHDX00554
FBMHTX00006
FBMHTX00007
FBMHTX00008
FBMHTX00017
FBMHTX00019
FBMHTX00020
FBMHTX00021
FBMHTX00024
FBMHTX00028
FBMHTX00030
FBMHTX00033
FBMHTX00041
FBMHTX00044
FBMHTX00046
FBMHTX00052
FBMHTX00053
FBMHTX00054
FBMHTX00057
FBMHTX00059
FBMHTX00060
FBMHTX00065
FBMHTX00066
FBMHTX00068
FBMHTX00070
FBMHTX00074
FBMHTX00077
FBMHTX00078
FBMHTX00081
FBMHTX00084
FBMHTX00088
FBMHTX00091
FBMHTX00094
FBMHTX00104
FBMHTX00106
FBMHTX00107
FBMHTX00113
FBMHTX00115
FBMHTX00116
FBMHTX00117
FBMHTX00118
FBMHTX00122
FBMHTX00126
FBMHTX00127
FBMHTX00129
FBMHTX00130
FBMHTX00133
FBMHTX00139
FBMHTX00140
FBMHTX00141
FBMHTX00143
FBMHTX00147
FBMHTX00151
FBMHTX00155
FBMHTX00161
FBMHTX00162
FBMHTX00163
FBMHTX00164
FBMHTX00166
FBMHTX00167
FBMHTX00169
FBMHTX00170
FBMHTX00171
FBMHTX00172
FBMHTX00173
FBMHTX00179
FBMHTX00180
FBMHTX00185
FBMHTX00186
FBMHTX00187
FBMHTX00188
FBMHTX00191
FBMHTX00195
FBMHTX00200
FBMHTX00203
FBMHTX00205
FBMHTX00210
FBMHTX00220
FBMHTX00221
FBMHTX00222
FBMHTX00223
FBMHTX00225
FBMHTX00230
FBMHTX00238
FBMHTX00245
FBMHTX00246
FBMHTX00250
FBMHTX00252
FBMHTX00255
FBMHTX00259
FBMHTX00275
FBMHTX00276
FBMHTX00278
FBMHTX00279
FBMHTX00280
FBMHTX00282
FBMHTX00283
FBMHTX00285
FBMHTX00290
FBMHTX00292
FBMHTX00293
FBMHTX00298
FBMHTX00306
FBMHTX00309
FBMHTX00310
FBMHTX00313
FBMHTX00329
FBMHTX00330
FBMHTX00333
FBMHTX00343
FBMHTX00344
FBMHTX00347
FBMHTX00353
FBMHTX00354
FBMHTX00358
FBMHTX00363
FBMHTX00369
FBMHTX00372
FBMHTX00383
FBMHTX00389
FBMHTX00392
FBMHTX00393
FBMHTX00399
FBMHTX00404
FBMHTX00405
FBMHTX00407
FBMHTX00408
FBMHTX00409
FBMHTX00410
FBMHTX00432
FBMHTX00433
FBMHTX00436
FBMHTX00437
FBMHTX00444
FBMHTX00446
FBMHTX00462
FBMHTX00467
FBMHTX00471
FBMHTX00473
FBMHTX00477
FBMHTX00488
FBMHTX00492
FBMHXX00011
FBMHXX00016
FBMHXX00017
FBMHXX00019
FBMHXX00024
FBMHXX00026
FBMHXX00032
FBMHXX00034
FBMHXX00037
FBMHXX00038
FBMHXX00043
FBMHXX00047
FBMHXX00049
FBMHXX00051
FBMHXX00052
FBMHXX00055
FBMHXX00057
FBMHXX00060
FBMHXX00065
FBMHXX00067
FBMHXX00072
FBMHXX00073
FBMHXX00081
FBMHXX00082
FBMHXX00083
FBMHXX00085
FBMHXX00086
FBMHXX00087
FBMHXX00089
FBMHXX00096
FBMHXX00099
FBMHXX00101
FBMHXX00102
FBMHXX00103
FBMHXX00104
FBXXTE00005
FBXXTE00010
FBXXTE00012
FBXXTE00013
FBXXTE00015
FBXXTE00016
FBXXTE00018
FBXXTE00025
FBXXTE00028
FBXXTE00029
FBXXTE00030
FBXXTE00032
FBXXTE00033
FBXXTE00035
FBXXTE00038
FBXXTE00039
FBXXTE00040
FBXXTE00043
FBXXTE00044
FBXXTE00047
FBXXTE00050
FBXXTE00051
FBXXTE00053
FBXXTE00054
FBXXTE00055
FBXXTE00056
FBXXTE00057
FBXXTE00064
FBXXTE00067
FBXXTE00075
FBXXTE00079
FBXXTE00082
FBXXTE00085
FBXXTE00088
FBXXTE00091
FBXXTE00092
FBXXTE00093
FBXXTE00098
FBXXTE00101
FBXXTE00102
FBXXTE00104
FBXXTE00105
FBXXTE00108
FBXXTE00110
FBXXTE00113
FBXXTE00114
FBXXTE00115
FBXXTE00117
FBXXTE00119
FBXXTE00120
FBXXTE00122
FBXXTE00123
FBXXTE00124
FBXXTE00125
FBXXTE00126
FBXXTE00129
FBXXTE00130
FBXXTE00131
FBXXTE00134
FBXXTE00136
FBXXTE00137
FBXXTE00138
FBXXTE00139
FBXXTE00148
FBXXTE00155
FBXXTE00158
FBXXTE00159
FBXXTE00160
FBXXTE00161
FBXXTE00163
FBXXTE00164
FBXXTE00165
FBXXTE00167
FBXXTE00172
FBXXTE00173
FBXXTE00175
FBXXTE00177
FBXXTE00180
FBXXTE00181
FBXXTE00182
FBXXTE00186
FBXXTE00187
FBXXTE00190
FBXXTE00196
FBXXTE00199
FBXXTE00203
FBXXTE00204
FBXXTE00206
FBXXTE00208
FBXXTE00209
FBXXTE00216
FBXXTE00217
FBXXTE00222
FBXXTE00223
FBXXTE00228
FBXXTE00229
FBXXTE00234
FBXXTE00237
FBXXTE00238
FBXXTE00249
FBXXTE00251
FBXXTE00252
FBXXTE00254
FBXXTE00256
FBXXTE00262
FBXXTE00263
FBXXTE00266
FBXXTE00267
FBXXTE00272
FBXXTE00279
FBXXTE00282
FBXXTE00283
FBXXTE00285
FBXXTE00288
FBXXTE00294
FBXXTE00300
FBXXTE00306
FBXXTE00316
FBXXTE00317
FBXXTE00318
FBXXTE00319
FBXXTE00321
FBXXTE00325
FBXXTE00326
FBXXTE00328
FBXXTE00329
FBXXTE00332
FBXXTE00339
FBXXTE00342
FBXXTE00346
FBXXTE00347
FBXXTE00353
FBXXTE00356
FBXXTE00357
FBXXTE00358
FBXXTE00359
FBXXTE00360
FBXXTE00384
FBXXTE00386
FBXXTE00389
FBXXTE00390
FBXXTE00391
FBXXTE00393
FBXXTE00401
FBXXTE00403
FBXXTE00406
FBXXTE00407
FBXXTE00409
FBXXTE00415
FBXXTE00420
FBXXTE00422
FBXXTE00430
FBXXTE00431
FBXXTE00432
FBXXTE00433
FBXXTE00434
FBXXTE00436
FBXXTE00443
FBXXTE00450
FBXXTE00454
FBXXTE00467
FBXXTE00480
FBXXTE00485
FBXXTE00496
FBXXTE00497
FBXXTE00502
FBXXTE00504
FBXXTE00506
FBXXTE00507
FBXXTE00509
FBXXTE00512
FBXXTE00513
FBXXTE00514
FBXXTE00515
FBXXTE00519
FBXXTE00520
FBXXTE00525
FBXXTE00528
FBXXTE00533
FBXXTE00534
FBXXTE00535
FBXXTE00537
FBXXTE00539
FBXXTE00543
FBXXTE00550
FBXXTE00553
FBXXTE00555
FBXXTE00557
FBXXTE00558
FBXXTE00559
FBXXTE00560
FBXXTE00561
FBXXTE00562
FBXXTE00563
FBXXTE00564
FBXXTE00565
FBXXTE00566
FBXXTE00567
FBXXTE00569
FBXXTE00571
FBXXTE00572
FBXXTE00575
FBXXTE00576
FBXXTE00577
FBXXTE00578
FBXXTE00579
FBXXTE00581
FBXXTE00582
FBXXTE00584
FBXXTE00585
FBXXTE00586
FBXXTE00587
FBXXTE00588
FBXXTE00589
FBXXTE00593
FBXXTE00595
FBXXTE00596
FBXXTE00597
FBXXTE00600
FBXXTE00602
FBXXTE00603
FBXXTE00613
FBXXTE00614
FBXXTE00617
FBXXTE00623
FBXXTE00625
FBXXTE00630
FBXXTE00631
FBXXTE00636
FBXXTE00638
FBXXTE00639
FBXXTE00640
FBXXTE00641
FBXXTE00642
FBXXTE00643
FBXXTE00644
FBXXTE00645
FBXXTE00646
FBXXTE00652
FBXXTE00654
FBXXTE00657
FBXXTE00659
FBXXTE00662
FBXXTE00663
FBXXTE00665
FBXXTE00668
FBXXTE00669
FBXXTE00681
FBXXTE00682
FBXXTE00684
FBXXTE00685
FBXXTE00692
FBXXTE00694
FBXXTE00700
FBXXTE00701
FBXXTE00702
FBXXTE00704
FBXXTE00708
FBXXTE00709
FBXXTE00712
FBXXTE00713
FBXXTE00714
FBXXTE00717
FBXXTE00720
FBXXTE00723
FBXXTE00725
FBXXTE00728
FBXXTE00735
FBXXTE00736
FBXXTE00737
FBXXTE00740
FBXXTE00742
FBXXTE00744
FBXXTE00745
FBXXTE00746
FBXXTE00748
FBXXTE00749
FBXXTE00755
FBXXTE00769
FBXXTE00771
FBXXTE00772
FBXXTE00774
FBXXTE00775
FBXXTE00776
FBXXTE00777
FBXXTE00779
FBXXTE00783
FBXXTE00787
FBXXTE00790
FBXXTE00792
FBXXTE00801
FBXXTE00802
FBXXTE00805
FBXXTE00807
FBXXTE00811
FBXXTE00812
FBXXTE00813
FBXXTE00814
FBXXTE00815
FBXXTE00825
FBXXTE00828
FBXXTE00829
FBXXTE00831
FBXXTE00832
FBXXTE00833
FBXXTE00834
FBXXTE00837
FBXXTE00839
FBXXTE00841
FBXXTE00842
FBXXTE00845
FBXXTE00846
FBXXTE00851
FBXXTE00852
FBXXTE00856
FBXXTE00859
FBXXTE00860
FBXXTE00861
FBXXTE00862
FBXXTE00863
FBXXTE00864
FBXXTE00865
FBXXTE00867
FBXXTE00868
FBXXTE00877
FBXXTE00878
FBXXTE00880
FBXXTE00883
FBXXTE00886
FBXXTE00887
FBXXTE00889
FBXXTE00895
FBXXTE00896
FBXXTE00897
FBXXTE00900
FBXXTE00904
FBXXTE00915
FBXXTE00917
FBXXTE00919
FBXXTE00925
FBXXTE00926
FBXXTE00929
FBXXTE00936
FBXXTE00942
FBXXTE00952
FBXXTE00953
FBXXTE00954
FBXXTE00957
FBXXTE00958
FBXXTE00960
FBXXTE00962
FBXXTE00963
FBXXTE00975
FBXXTE00979
FBXXTE00980
FBXXTE00981
FBXXTE00985
FBXXTE00994
FBXXTE00999
FBXXTE01003
FBXXTE01004
FBXXTE01005
FBXXTE01013
FBXXTE01014
FBXXTE01015
FBXXTE01018
FBXXTE01020
FBXXTE01027
FBXXTE01030
FBXXTE01033
FBXXTE01034
FBXXTE01035
FBXXTE01037
FBXXTE01038
FBXXTE01039
FBXXTE01041
FBXXTE01047
FBXXTE01048
FBXXTE01049
FBXXTE01052
FBXXTE01054
FBXXTE01055
FBXXTE01058
FBXXTE01059
FBXXTE01060
FBXXTE01061
FBXXTE01065
FBXXTE01067
FBXXTE01068
FBXXTE01070
FBXXTE01073
FBXXTE01074
FBXXTE01075
FBXXTE01077
FBXXTE01080
FBXXTE01083
FBXXTE01084
FBXXTE01085
FBXXTE01088
FBXXTE01089
FBXXTE01093
FBXXTE01096
FBXXTE01097
FBXXTE01098
FBXXTE01100
FBXXTE01102
FBXXTE01103
FBXXTE01104
FBXXTE01106
FBXXTE01108
FBXXTE01109
FBXXTE01112
FBXXTE01113
FBXXTE01117
FBXXTE01119
FBXXTE01121
FBXXTE01123
FBXXTE01126
FBXXTE01129
FBXXTE01130
FBXXTE01131
FBXXTE01132
FBXXTE01143
FBXXTE01146
FBXXTE01148
FBXXTE01153
FBXXTE01154
FBXXTE01160
FBXXTE01165
FBXXTE01166
FBXXTE01167
FBXXTE01168
FBXXTE01169
FBXXTE01173
FBXXTE01177
FBXXTE01181
FBXXTE01183
FBXXTE01184
FBXXTE01185
FBXXTE01186
FBXXTE01190
FBXXTE01193
FBXXTE01194
FBXXTE01197
FBXXTE01198
FBXXTE01200
FBXXTE01201
FBXXTE01203
FBXXTE01207
FBXXTE01209
FBXXTE01210
FBXXTE01211
FBXXTE01216
FBXXTE01218
FBXXTE01222
FBXXTE01224
FBXXTE01226
FBXXTE01227
FBXXTE01228
FBXXTE01230
FBXXTE01231
FBXXTE01232
FBXXTE01233
FBXXTE01235
FBXXTE01238
FBXXTE01241
FBXXTE01249
FBXXTE01253
FBXXTE01257
FBXXTE01260
FBXXTE01262
FBXXTE01263
FBXXTE01267
FBXXTE01268
FBXXTE01272
FBXXTE01273
FBXXTE01274
FBXXTE01275
FBXXTE01276
FBXXTE01277
FBXXTE01278
FBXXTE01281
FBXXTE01282
FBXXTE01286
FBXXTE01287
FBXXTE01288
FBXXTE01289
FBXXTE01290
FBXXTE01291
FBXXTE01292
FBXXTE01293
FBXXTE01294
FBXXTE01296
FBXXTE01298
FBXXTE01301
FBXXTE01305
FBXXTE01306
FBXXTE01307
FBXXTE01310
FBXXTE01312
FBXXTE01313
FBXXTE01314
FBXXTE01315
FBXXTE01317
FBXXTE01320
FBXXTE01325
FBXXTE01326
FBXXTE01331
FBXXTE01333
FBXXTE01337
FBXXTE01343
FBXXTE01344
FBXXTE01349
FBXXTE01353
FBXXTE01355
FBXXTE01356
FBXXTE01363
FBXXTE01371
FBXXTE01377
FBXXTE01382
FBXXTE01384
FBXXTE01385
FBXXTE01387
FBXXTE01397
FBXXTE01399
FBXXTE01404
FBXXTE01407
FBXXTE01410
FBXXTE01411
FBXXTE01412
FBXXTE01414
FBXXTE01416
FBXXTE01417
FBXXTE01418
FBXXTE01419
FBXXTE01421
FBXXTE01422
FBXXTE01430
FBXXTE01432
FBXXTE01436
FBXXTE01437
FBXXTE01438
FBXXTE01441
FBXXTE01444
FBXXTE01446
FBXXTE01448
FBXXTE01449
FBXXTE01451
FBXXTE01452
FBXXTE01453
FBXXTE01454
FBXXTE01457
FBXXTE01458
FBXXTE01460
FBXXTE01461
FBXXTE01462
FBXXTE01463
FBXXTE01465
FBXXTE01466
FBXXTE01467
FBXXTE01469
FBXXTE01472
FBXXTE01474
FBXXTE01476
FBXXTE01477
FBXXTE01478
FBXXTE01479
FBXXTE01480
FBXXTE01482
FBXXTE01483
FBXXTE01485
FBXXTE01486
FBXXTE01487
FBXXTE01488
FBXXTE01489
FBXXTE01491
FBXXTE01494
FBXXTE01496
FBXXTE01497
FBXXTE01500
FBXXTE01501
FBXXTE01502
FBXXTE01504
FBXXTE01505
FBXXTE01507
FBXXTE01508
FBXXTE01509
FBXXTE01510
FBXXTE01512
FBXXTE01513
FBXXTE01514
FBXXTE01515
FBXXTE01516
FBXXTE01517
FBXXTE01518
FBXXTE01519
FBXXTE01520
FBXXTE01521
FBXXTE01522
FBXXTE01525
FBXXTE01526
FBXXTE01528
FBXXTE01529
FBXXTE01532
FBXXTE01535
FBXXTE01537
FBXXTE01538
FBXXTE01539
FBXXTE01540
FBXXTE01542
FBXXTE01543
FBXXTE01544
FBXXTE01546
FBXXTE01547
FBXXTE01549
FBXXTE01551
FBXXTE01553
FBXXTE01555
FBXXTE01557
FBXXTE01559
FBXXTE01563
FBXXTE01566
FBXXTE01569
FBXXTE01570
FBXXTE01571
FBXXTE01572
FBXXTE01574
FBXXTE01579
FBXXTE01585
FBXXTE01586
FBXXTE01587
FBXXTE01590
FBXXTE01593
FBXXTE01602
FBXXTE01604
FBXXTE01613
FBXXTE01617
FBXXTE01618
FBXXTE01619
FBXXTE01620
FBXXTE01630
FBXXTE01632
FBXXTE01634
FBXXTE01644
FBXXTE01650
FBXXTE01651
FBXXTE01654
FBXXTE01663
FBXXTE01667
FBXXTE01672
FBXXTE01673
FBXXTE01677
FBXXTE01678
FBXXTE01681
FBXXTE01686
FBXXTE01687
FBXXTE01688
FBXXTE01690
FBXXTE01691
FBXXTE01692
FBXXTE01702
FBXXTE01704
FBXXTE01708
FBXXTE01718
FBXXTE01719
FBXXTE01720
FBXXTE01722
FBXXTE01723
FBXXTE01725
FBXXTE01727
FBXXTE01728
FBXXTE01731
FBXXTE01732
FBXXTE01734
FBXXTE01740
FBXXTE01748
FBXXTE01750
FBXXTE01754
FBXXTE01756
FBXXTE01758
FBXXTE01760
FBXXTE01761
FBXXTE01762
FBXXTE01765
FBXXTE01766
FBXXTE01767
FBXXTE01768
FBXXTE01769
FBXXTE01770
FBXXTE01771
FBXXTE01775
FBXXTE01779
FBXXTE01786
FBXXTE01787
FBXXTE01789
FBXXTE01801
FBXXTE01803
FBXXTE01805
FBXXTE01806
FBXXTE01811
FBXXTE01812
FBXXTE01820
FBXXTE01822
FBXXTE01826
FBXXTE01829
FBXXTE01830
FBXXTE01832
FBXXTE01835
FBXXTE01837
FBXXTE01839
FBXXTE01841
FBXXTE01845
FBXXTE01846
FBXXTE01847
FBXXTE01851
FBXXTE01852
FBXXTE01853
FBXXTE01856
FBXXTE01858
FBXXTE01860
FBXXTE01861
FBXXTE01862
FBXXTE01863
FBXXTE01865
FBXXTE01874
FBXXTE01875
FBXXTE01876
FBXXTE01877
FBXXTE01881
FBXXTE01884
FBXXTE01885
FBXXTE01889
FBXXTE01890
FBXXTE01892
FBXXTE01901
FBXXTE01903
FBXXTE01906
FBXXTE01908
FBXXTE01909
FBXXTE01913
FBXXTE01919
FBXXTE01921
FBXXTE01922
FBXXTE01923
FBXXTE01926
FBXXTE01929
FBXXTE01930
FBXXTE01931
FBXXTE01932
FBXXTE01933
FBXXTE01935
FBXXTE01936
FBXXTE01937
FBXXTE01940
FBXXTE01948
FBXXTE01950
FBXXTE01951
FBXXTE01967
FBXXTE01973
FBXXTE01975
FBXXTE01976
FBXXTE01978
FBXXTE01979
FBXXTE01980
FBXXTE01981
FBXXTE01987
FBXXTE01991
FBXXTE01992
FBXXTE01993
FBXXTE01994
FBXXTE01995
FBXXTE01996
FBXXTE02003
FBXXTE02004
FBXXTE02007
FBXXTE02012
FBXXTE02013
FBXXTE02014
FBXXTE02016
FBXXTE02018
FBXXTE02020
FBXXTE02021
FBXXTE02022
FBXXTE02024
FBXXTE02025
FBXXTE02026
FBXXTE02039
FBXXTE02040
FBXXTE02045
FBXXTE02046
FBXXTE02048
FBXXTE02049
FBXXTE02053
FBXXTE02054
FBXXTE02056
FBXXTE02059
FBXXTE02061
FBXXTE02064
FBXXTE02065
FBXXTE02066
FBXXTE02067
FBXXTE02070
FBXXTE02074
FBXXTE02075
FBXXTE02085
FBXXTE02092
FBXXTE02096
FBXXTE02097
FBXXTE02104
FBXXTE02105
FBXXTE02108
FBXXTE02109
FBXXTE02113
FBXXTE02115
FBXXTE02117
FBXXTE02118
FBXXTE02119
FBXXTE02121
FBXXTE02122
FBXXTE02126
FBXXTE02133
FBXXTE02139
FBXXTE02140
FBXXTE02141
FBXXTE02142
FBXXTE02148
FBXXTE02149
FBXXTE02150
FBXXTE02151
FBXXTE02153
FBXXTE02154
FBXXTE02157
FBXXTE02162
FBXXTE02163
FBXXTE02165
FBXXTE02168
FBXXTE02179
FBXXTE02187
FBXXTE02197
FBXXTE02199
FBXXTE02200
FBXXTE02201
FBXXTE02202
FBXXTE02207
FBXXTE02208
FBXXTE02213
FBXXTE02216
FBXXTE02244
FBXXTE02250
FBXXTE02251
FBXXTE02252
FBXXTE02257
FBXXTE02262
FBXXTE02279
FBXXTE02280
FBXXTE02281
FBXXTE02295
FBXXTE02300
FBXXTE02303
FBXXTE02304
FBXXTE02311
FBXXTE02312
FBXXTE02314
FBXXTE02318
FBXXTE02320
FBXXTE02323
FBXXTE02325
FBXXTE02331
FBXXTE02332
FBXXTE02333
FBXXTE02334
FBXXTE02337
FBXXTE02338
FBXXTE02340
FBXXTE02341
FBXXTE02344
FBXXTE02345
FBXXTE02350
FBXXTE02355
FBXXTE02356
FBXXTE02357
FBXXTE02358
FBXXTE02359
FBXXTE02365
FBXXTE02374
FBXXTE02385
FBXXTE02387
FBXXTE02388
FBXXTE02389
FBXXTE02390
FBXXTE02391
FBXXTE02401
FBXXTE02414
FBXXTE02418
FBXXTE02421
FBXXTE02425
FBXXTE02426
FBXXTE02428
FBXXTE02429
FBXXTE02430
FBXXTE02431
FBXXTE02432
FBXXTE02433
FBXXTE02434
FBXXTE02435
FBXXTE02437
FBXXTE02438
FBXXTE02439
FBXXTE02442
FBXXTE02448
FBXXTE02453
FBXXTE02454
FBXXTE02455
FBXXTE02458
FBXXTE02462
FBXXTE02463
FBXXTE02464
FBXXTE02468
FBXXTE02469
FBXXTE02470
FBXXTE02471
FBXXTE02472
FBXXTE02475
FBXXTE02476
FBXXTE02477
FBXXTE02478
FBXXTE02481
FBXXTE02488
FBXXTE02490
FBXXTE02491
FBXXTE02492
FBXXTE02493
FBXXTE02494
FBXXTE02504
FBXXTE02508
FBXXTE02509
FBXXTE02510
FBXXTE02511
FBXXTE02512
FBXXTE02515
FBXXTE02517
FBXXTE02521
FBXXTE02524
FBXXTE02525
FBXXTE02526
FBXXTE02528
FBXXTE02529
FBXXTE02530
FBXXTE02535
FBXXTE02539
FBXXTE02541
FBXXTE02543
FBXXTE02544
FBXXTE02548
FBXXTE02549
FBXXTE02550
FBXXTE02553
FBXXTE02571
FBXXTE02580
FBXXTE02582
FBXXTE02583
FBXXTE02585
FBXXTE02610
FBXXTE02624
FBXXTE02625
FBXXTE02626
FBXXTE02628
FBXXTE02629
FBXXTE02640
FBXXTE02646
FBXXTE02647
FBXXTE02655
FBXXTE02656
FBXXTE02662
FBXXTE02663
FBXXTE02664
FBXXTE02665
FBXXTE02667
FBXXTE02669
FBXXTE02670
FBXXTE02673
FBXXTE02674
FBXXTE02680
FBXXTE02685
FBXXTE02693
FBXXTE02695
FBXXTE02696
FBXXTE02697
FBXXTE02698
FBXXTE02699
FBXXTE02701
FBXXTE02702
FBXXTE02703
FBXXTE02704
FBXXTE02705
FBXXTE02707
FBXXTE02708
FBXXTE02711
FBXXTE02713
FBXXTE02727
FBXXTE02728
FBXXTE02731
FBXXTE02732
FBXXTE02737
FBXXTE02739
FBXXTE02744
FBXXTE02745
FBXXTE02749
FBXXTE02752
FBXXTE02755
FBXXTE02756
FBXXTE02757
FBXXTE02758
FBXXTE02759
FBXXTE02760
FBXXTE02766
FBXXTE02773
FBXXTE02774
FBXXTE02775
FBXXTE02776
FBXXTE02777
FBXXTE02780
FBXXTE02781
FBXXTE02782
FBXXTE02783
FBXXTE02784
FBXXTE02785
FBXXTE02808
FBXXTE02816
FBXXTE02817
FBXXTE02819
FBXXTE02823
FBXXTE02824
FBXXTE02825
FBXXTE02827
FBXXTE02831
FBXXTE02832
FBXXTE02833
FBXXTE02839
FBXXTE02842
FBXXTE02845
FBXXTE02846
FBXXTE02848
FBXXTE02851
FBXXTE02853
FBXXTE02854
FBXXTE02855
FBXXTE02857
FBXXTE02858
FBXXTE02859
FBXXTE02860
FBXXTE02863
FBXXTE02865
FBXXTE02867
FBXXTE02868
FBXXTE02871
FBXXTE02873
FBXXTE02874
FBXXTE02875
FBXXTE02876
FBXXTE02881
FBXXTE02886
FBXXTE02887
FBXXTE02889
FBXXTE02893
FBXXTE02895
FBXXTE02896
FBXXTE02897
FBXXTE02899
FBXXTE02900
FBXXTE02902
FBXXTE02904
FBXXTE02909
FBXXTE02914
FBXXTE02915
FBXXTE02916
FBXXTE02917
FBXXTE02918
FBXXTE02919
FBXXTE02931
FBXXTE02932
FBXXTE02933
FBXXTE02947
FBXXTE02948
FBXXTE02950
FBXXTE02951
FBXXTE02958
FBXXTE02963
FBXXTE02965
FBXXTE02966
FBXXTE02970
FBXXTE02973
FBXXTE02975
FBXXTE02976
FBXXTE03000
FBXXTE03002
FBXXTE03012
FBXXTE03015
FBXXTE03019
FBXXTE03020
FBXXTE03030
FBXXTE03032
FBXXTE03037
FBXXTE03038
FBXXTE03040
FBXXTE03041
FBXXTE03042
FBXXTE03043
FBXXTE03045
FBXXTE03046
FBXXTE03047
FBXXTE03048
FBXXTE03051
FBXXTE03053
FBXXTE03059
FBXXTE03060
FBXXTE03062
FBXXTE03063
FBXXTE03064
FBXXTE03065
FBXXTE03068
FBXXTE03070
FBXXTE03074
FBXXTE03076
FBXXTE03078
FBXXTE03079
FBXXTE03083
FBXXTE03084
FBXXTE03086
FBXXTE03088
FBXXTE03089
FBXXTE03091
FBXXTE03092
FBXXTE03094
FBXXTE03097
FBXXTE03098
FBXXTE03099
FBXXTE03101
FBXXTE03102
FBXXTE03104
FBXXTE03105
FBXXTE03106
FBXXTE03108
FBXXTE03109
FBXXTE03112
FBXXTE03113
FBXXTE03114
FBXXTE03115
FBXXTE03118
FBXXTE03119
FBXXTE03120
FBXXTE03122
FBXXTE03123
FBXXTE03124
FBXXTE03125
FBXXTE03126
FBXXTE03128
FBXXTE03129
FBXXTE03132
FBXXTE03133
FBXXTE03136
FBXXTE03141
FBXXTE03145
FBXXTE03148
FBXXTE03149
FBXXTE03151
FBXXTE03153
FBXXTE03155
FBXXTE03157
FBXXTE03161
FBXXTE03162
FBXXTE03165
FBXXTE03167
FBXXTE03169
FBXXTE03173
FBXXTE03176
FBXXTE03177
FBXXTE03178
FBXXTE03179
FBXXTE03180
FBXXTE03186
FBXXTE03188
FBXXTE03189
FBXXTE03190
FBXXTE03191
FBXXTE03192
FBXXTE03193
FBXXTE03194
FBXXTE03195
FBXXTE03196
FBXXTE03201
FBXXTE03204
FBXXTE03207
FBXXTE03208
FBXXTE03209
FBXXTE03218
FBXXTE03219
FBXXTE03221
FBXXTE03225
FBXXTE03232
FBXXTE03235
FBXXTE03236
FBXXTE03245
FBXXTE03250
FBXXTE03255
FBXXTE03264
FBXXTE03276
FBXXTE03277
FBXXTE03279
FBXXTE03283
FBXXTE03286
FBXXTE03288
FBXXTE03293
FBXXTE03305
FBXXTE03306
FBXXTE03307
FBXXTE03308
FBXXTE03310
FBXXTE03318
FBXXTP00002
FBXXTP00008
FBXXTP00009
FBXXTP00010
FBXXTP00011
FBXXTP00012
FBXXTP00013
FBXXTP00014
FBXXTP00015
FBXXTP00018
FBXXTP00020
FBXXTP00022
FBXXTP00023
FBXXTP00025
FBXXTP00026
FBXXTP00027
FBXXTP00028
FBXXTP00030
FBXXTP00032
FBXXTP00033
FBXXTP00034
FBXXTP00035
FBXXTP00036
FBXXTP00039
FBXXTP00040
FBXXTP00041
FBXXTP00042
FBXXTP00043
FBXXTP00044
FBXXTP00045
FBXXXE00004
FBXXXE00007
FBXXXE00012
FBXXXE00015
FBXXXE00016
FBXXXE00017
FBXXXE00020
FBXXXE00023
FBXXXE00024
FBXXXE00030
FBXXXE00032
FBXXXE00033
FBXXXE00034
FBXXXE00035
FBXXXE00036
FBXXXE00039
FBXXXE00047
FBXXXE00048
FBXXXE00050
FBXXXE00051
FBXXXE00053
FBXXXE00055
FBXXXE00056
FBXXXE00057
FBXXXE00064
FBXXXE00066
FBXXXE00072
FBXXXE00073
FBXXXE00077
FBXXXE00078
FBXXXE00079
FBXXXE00083
FBXXXE00085
FBXXXE00087
FBXXXE00088
FBXXXE00089
FBXXXE00090
FBXXXE00091
FBXXXE00094
FBXXXE00099
FBXXXE00107
FBXXXE00109
FBXXXE00113
FBXXXE00115
FBXXXE00116
FBXXXE00125
FBXXXE00126
FBXXXE00127
FBXXXE00131
FBXXXE00133
FBXXXE00135
FBXXXE00136
FBXXXE00139
FBXXXE00142
FBXXXE00143
FBXXXE00144
FBXXXE00147
FBXXXE00148
FBXXXE00149
FBXXXE00157
FBXXXE00162
FBXXXE00163
FBXXXE00164
FBXXXE00166
FBXXXE00170
FBXXXE00171
FBXXXE00172
FBXXXE00173
FBXXXE00177
FBXXXE00178
FBXXXE00180
FBXXXE00183
FBXXXE00189
FBXXXE00190
FBXXXE00193
FBXXXE00199
FBXXXE00200
FBXXXE00204
FBXXXE00205
FBXXXE00206
FBXXXE00210
FBXXXE00213
FBXXXE00215
FBXXXE00217
FBXXXE00218
FBXXXE00219
FBXXXE00220
FBXXXE00221
FBXXXE00222
FBXXXE00224
FBXXXE00225
FBXXXE00227
FBXXXE00229
FBXXXE00230
FBXXXE00231
FBXXXE00239
FBXXXE00243
FBXXXE00244
FBXXXE00247
FBXXXE00252
FBXXXE00253
FBXXXE00256
FBXXXE00257
FBXXXE00258
FBXXXE00259
FBXXXE00261
FBXXXE00262
FBXXXE00263
FBXXXE00266
FBXXXE00267
FBXXXE00268
FBXXXE00269
FBXXXE00271
FBXXXE00272
FBXXXE00283
FBXXXE00284
FBXXXE00287
FBXXXE00288
FBXXXE00296
FBXXXE00298
FBXXXE00299
FBXXXE00300
FBXXXE00303
FBXXXE00304
FBXXXE00306
FBXXXE00307
FBXXXE00308
FBXXXE00309
FBXXXE00312
FBXXXE00313
FBXXXE00315
FBXXXE00317
FBXXXE00320
FBXXXE00321
FBXXXE00323
FBXXXE00329
FBXXXE00330
FBXXXE00340
FBXXXE00342
FBXXXE00343
FBXXXE00345
FBXXXE00346
FBXXXE00351
FBXXXE00356
FBXXXE00359
FBXXXE00362
FBXXXE00363
FBXXXE00365
FBXXXE00370
FBXXXE00375
FBXXXE00384
FBXXXE00386
FBXXXE00391
FBXXXE00393
FBXXXE00394
FBXXXE00397
FBXXXE00399
FBXXXE00402
FBXXXE00403
FBXXXE00404
FBXXXE00405
FBXXXE00407
FBXXXE00408
FBXXXE00413
FBXXXE00414
FBXXXE00415
FBXXXE00417
FBXXXE00418
FBXXXE00419
FBXXXE00421
FBXXXE00422
FBXXXE00423
FBXXXE00424
FBXXXE00425
FBXXXE00426
FBXXXE00427
FBXXXE00430
FBXXXE00433
FBXXXE00437
FBXXXE00438
FBXXXE00439
FBXXXE00440
FBXXXE00441
FBXXXE00442
FBXXXE00444
FBXXXE00446
FBXXXE00449
FBXXXE00456
FBXXXE00457
FBXXXE00458
FBXXXE00459
FBXXXE00460
FBXXXE00462
FBXXXE00465
FBXXXE00467
FBXXXE00468
FBXXXE00470
FBXXXE00472
FBXXXE00474
FBXXXE00479
FBXXXE00480
FBXXXE00481
FBXXXE00483
FBXXXE00484
FBXXXE00485
FBXXXE00487
FBXXXE00488
FBXXXE00490
FBXXXE00491
FBXXXE00492
FBXXXE00494
FBXXXE00495
FBXXXE00496
FBXXXE00497
FBXXXE00498
FBXXXE00500
FBXXXE00502
FBXXXE00503
FBXXXE00504
FBXXXE00505
FBXXXE00507
FBXXXE00508
FBXXXE00511
FBXXXE00513
FBXXXE00514
FBXXXE00515
FBXXXE00516
FBXXXE00517
FBXXXE00518
FBXXXE00529
FBXXXE00530
FBXXXE00531
FBXXXE00532
FBXXXE00533
FBXXXE00534
FBXXXE00535
FBXXXE00536
FBXXXE00540
FBXXXE00541
FBXXXE00542
FBXXXE00543
FBXXXE00545
FBXXXE00546
FBXXXE00547
FBXXXE00549
FBXXXE00551
FBXXXE00552
FBXXXE00553
FBXXXE00554
FBXXXE00555
FBXXXE00557
FBXXXE00560
FBXXXE00563
FBXXXE00565
FBXXXE00566
FBXXXE00567
FBXXXE00572
FBXXXE00573
FBXXXE00574
FBXXXE00576
FBXXXE00578
FBXXXE00582
FBXXXE00584
FBXXXE00585
FBXXXE00586
FBXXXE00587
FBXXXE00588
FBXXXE00591
FBXXXE00592
FBXXXE00593
FBXXXE00594
FBXXXE00595
FBXXXE00596
FBXXXE00597
FBXXXE00598
FBXXXE00599
FBXXXE00600
FBXXXE00601
FBXXXE00603
FBXXXE00604
FBXXXE00605
FBXXXE00611
FBXXXE00612
FBXXXE00615
FBXXXE00616
FBXXXE00618
FBXXXE00622
FBXXXE00623
FBXXXE00625
FBXXXE00633
FBXXXE00634
FBXXXE00635
FBXXXE00639
FBXXXE00642
FBXXXE00646
FBXXXE00647
FBXXXE00648
FBXXXE00649
FBXXXE00651
FBXXXE00653
FBXXXE00654
FBXXXE00655
FBXXXE00659
FBXXXE00661
FBXXXE00662
FBXXXE00663
FBXXXE00665
FBXXXE00666
FBXXXE00667
FBXXXE00668
FBXXXE00669
FBXXXE00675
FBXXXE00677
FBXXXE00682
FBXXXE00684
FBXXXE00685
FBXXXE00689
FBXXXE00690
FBXXXE00691
FBXXXE00693
FBXXXE00695
FBXXXE00699
FBXXXE00700
FBXXXE00702
FBXXXE00703
FBXXXE00704
FBXXXE00707
FBXXXE00708
FBXXXE00714
FBXXXE00715
FBXXXE00716
FBXXXE00723
FBXXXE00725
FBXXXE00726
FBXXXE00727
FBXXXE00728
FBXXXE00729
FBXXXE00730
FBXXXE00731
FBXXXE00739
FBXXXE00741
FBXXXE00742
FBXXXE00746
FBXXXE00747
FBXXXE00748
FBXXXE00749
FBXXXE00750
FBXXXE00751
FBXXXE00752
FBXXXE00753
FBXXXE00754
FBXXXE00757
FBXXXE00762
FBXXXE00764
FBXXXE00765
FBXXXE00767
FBXXXE00772
FBXXXE00773
FBXXXE00777
FBXXXE00786
FBXXXE00787
FBXXXE00790
FBXXXE00793
FBXXXE00794
FBXXXE00796
FBXXXE00798
FBXXXE00802
FBXXXE00803
FBXXXE00805
FBXXXE00806
FBXXXE00807
FBXXXE00808
FBXXXE00816
FBXXXE00817
FBXXXE00820
FBXXXE00821
FBXXXE00822
FBXXXE00823
FBXXXE00826
FBXXXE00827
FBXXXE00829
FBXXXE00837
FBXXXE00840
FBXXXE00842
FBXXXE00844
|
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | 262-544-3011 |
Manufacturer Reason for Recall | OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field. |
FDA Determined Cause 2 | Process control |
| Action | All impacted customers with receive a customer letter notifying them of the potential
issue. GE HealthCare will send a field service engineer to perform measurements of the Xray
field on the identified systems (listed in Appendix A of the customer letter). |
| Quantity in Commerce | Total for both products: 3472 (2,919 US; 553 OUS) |
| Distribution | US and OUS |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = OXO
|