| | Class 3 Device Recall mC123 |  |
| Date Initiated by Firm | July 03, 2025 |
| Date Posted | August 22, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2383-2025 |
| Recall Event ID |
97217 |
| 510(K)Number | K200304 |
| Product Classification |
System, imaging, pulsed doppler, ultrasonic - Product Code IYN
|
| Product | mC12-3 Transducer Probe |
| Code Information |
Model No. 989605495721; UDI: (01)00884838093935(21)F0FKJ7, 010088483809393521F0FRHZ, (01)00884838093935(21)B3XYRY, (01)00884838093935(21)B4161X, (01)00884838093935(21)F07K8J, (01)00884838093935(21)F0JTY0, (01)00884838093935(21)F04YTZ, (01)00884838093935(21)F0550K, (01)00884838093935(21)F094Y9, (01)00884838093935(21)F0GJC9, (01)00884838093935(21)B3XZRM, (01)00884838093935(21)F0KC6Y, (01)00884838093935(21)F0HP38, (01)00884838093935(21)F0F164, (01)00884838093935(21)F0KND1, (01)00884838093935(21)F04YV6, 010088483809393521B3Y0KC, (01)00884838093935(21)F0JZQ6, (01)00884838093935(21)F0JZQ7, (01)00884838093935(21)B3XZ5Q, (01)00884838093935(21)F07WW2, (01)00884838093935(21)F07WW9, (01)00884838093935(21)F07WWF, (01)00884838093935(21)F07WWJ, (01)00884838093935(21)F07WWL, (01)00884838093935(21)F07WWN, (01)00884838093935(21)F07WWP, (01)00884838093935(21)F07WWQ, (01)00884838093935(21)F07ZZ0, (01)00884838093935(21)F081V3, (01)00884838093935(21)F08H3V, (01)00884838093935(21)F05FK6, (01)00884838093935(21)F06GWZ, (01)00884838093935(21)F07D0C, (01)00884838093935(21)F0B94X, (01)00884838093935(21)F0GJCR, (01)00884838093935(21)B3FKZJ, (01)00884838093935(21)B3FKZN, (01)00884838093935(21)F0LC8N, (01)00884838093935(21)F04QBB, (01)00884838093935(21)F04V3M, (01)00884838093935(21)F04V3Q, (01)00884838093935(21)F04XDN, (01)00884838093935(21)F04XF1, (01)00884838093935(21)F04XF6, (01)00884838093935(21)F06XQF, (01)00884838093935(21)F0F4BF, (01)00884838093935(21)F0F4BN, (01)00884838093935(21)F0F7ZJ, (01)00884838093935(21)F0FKK9, 010088483809393521B3Y0KG, (01)00884838093935(21)F06XQV, (01)00884838093935(21)F06KRF, (01)00884838093935(21)F0JPPM, (01)00884838093935(21)F0K7LQ, (01)00884838093935(21)B3FLBF, (01)00884838093935(21)B3FLDJ, 010088483809393521F0G346, (01)00884838093935(21)B3XZRN, (01)00884838093935(21)B3XZY3, (01)00884838093935(21)B3XZ1X, (01)00884838093935(21)B41GND, (01)00884838093935(21)F0KC6Z, (01)00884838093935(21)B3FLFY, (01)00884838093935(21)F04QBW, (01)00884838093935(21)F0JL20, (01)00884838093935(21)B3FL3H, 010088483809393521B3Y0LX, (01)00884838093935(21)B3FLML, (01)00884838093935(21)F04YV4, (01)00884838093935(21)F094F4, (01)00884838093935(21)F09BZY, (01)00884838093935(21)B3FMGT, (01)00884838093935(21)B3FMGV, (01)00884838093935(21)F0715Y, (01)00884838093935(21)F07166, (01)00884838093935(21)F07D0G, (01)00884838093935(21)F08H3P, (01)00884838093935(21)F0JWZ9, (01)00884838093935(21)F09GFP, (01)00884838093935(21)B3FKZM, (01)00884838093935(21)B3FL0R, (01)00884838093935(21)B3FLT4, (01)00884838093935(21)F08R38, (01)00884838093935(21)B41GNT, 010088483809393521B3FM7L, 010088483809393521B3Y0LB, 010088483809393521B3Y0LH, 010088483809393521B3Y0LJ, (01)00884838093935(21)F07K85, (01)00884838093935(21)F0JPPL, (01)00884838093935(21)F0JRVG, (01)00884838093935(21)F0JZQ0, (01)00884838093935(21)B3FL5X, (01)00884838093935(21)B3L524, (01)00884838093935(21)F04QB9, (01)00884838093935(21)F04QBC, (01)00884838093935(21)F07RY3, (01)00884838093935(21)F09BZW, (01)00884838093935(21)F06GWR, 010088483809393521F0FP12, (01)00884838093935(21)F04QC0, (01)00884838093935(21)B3FLXG, (01)00884838093935(21)B3FLXP, (01)00884838093935(21)B3FLXR, (01)00884838093935(21)B3FLXT, (01)00884838093935(21)B3FMFF, (01)00884838093935(21)B3XZMQ, (01)00884838093935(21)B3XZXP, (01)00884838093935(21)F07GKY, (01)00884838093935(21)F07RY7, (01)00884838093935(21)F07RYB, (01)00884838093935(21)F07WWY, (01)00884838093935(21)F0842B, (01)00884838093935(21)F098FT, (01)00884838093935(21)F0HFNF, (01)00884838093935(21)F0K7LM, (01)00884838093935(21)F0K7LN, (01)00884838093935(21)F0K7LP, (01)00884838093935(21)F0K7LR, (01)00884838093935(21)F0K7M7, (01)00884838093935(21)B3FLMX, 010088483809393521F0FP13, (01)00884838093935(21)F04QBK, (01)00884838093935(21)F06LJT, (01)00884838093935(21)F06TG7, (01)00884838093935(21)F06XQZ, (01)00884838093935(21)F0715M, (01)00884838093935(21)F07ZZ5, (01)00884838093935(21)F07ZZ6, (01)00884838093935(21)F08H3M, (01)00884838093935(21)F08MG7, 010088483809393521F0FRJ0, (01)00884838093935(21)F0HP34, (01)00884838093935(21)F0GJCQ, (01)00884838093935(21)F0550W, (01)00884838093935(21)F0596C, (01)00884838093935(21)F094F3, (01)00884838093935(21)F0JRVD, (01)00884838093935(21)B3Y038, (01)00884838093935(21)B3Y052, (01)00884838093935(21)B3Y0GL, (01)00884838093935(21)F0JDZK, (01)00884838093935(21)F0JH08, (01)00884838093935(21)F0JH0G, (01)00884838093935(21)F0JL28, (01)00884838093935(21)F04QBT, (01)00884838093935(21)F04V3C, (01)00884838093935(21)B3Y0GP, (01)00884838093935(21)B3FLK9, (01)00884838093935(21)F0596V, (01)00884838093935(21)F05BQR, 010088483809393521B3FM7D, (01)00884838093935(21)F0K7M4, (01)00884838093935(21)B3FLTN, (01)00884838093935(21)B3FLRX, (01)00884838093935(21)F0841G, (01)00884838093935(21)F0KQJK, (01)00884838093935(21)F09JNR, (01)00884838093935(21)F09PVK, (01)00884838093935(21)F0F4C7, (01)00884838093935(21)B3FL3G, (01)00884838093935(21)B3FL3J, (01)00884838093935(21)F07GL0, (01)00884838093935(21)B3FLFQ, 010088483809393521F0GJC5, (01)00884838093935(21)B3Y02Y, (01)00884838093935(21)B3Y050, 010088483809393521F0FP0Q, (01)00884838093935(21)F06KR9, 010088483809393521F0GFQ2, (01)00884838093935(21)B3FMBQ, (01)00884838093935(21)F081VL, (01)00884838093935(21)F0841K, (01)00884838093935(21)F09BZT, (01)00884838093935(21)B3FLT3, 010088483809393521B3FM57, 010088483809393521B3FM7T, (01)00884838093935(21)F06GWX, (01)00884838093935(21)F0KNDF, (01)00884838093935(21)F08MG9, (01)00884838093935(21)F0JTXP, (01)00884838093935(21)F07GKG, (01)00884838093935(21)F07RY8, (01)00884838093935(21)F0GR37, (01)00884838093935(21)F0HV5V, (01)00884838093935(21)B3FL47, (01)00884838093935(21)B41625, (01)00884838093935(21)B41629, (01)00884838093935(21)B4162C, (01)00884838093935(21)F0KC74, (01)00884838093935(21)B3Y05C, (01)00884838093935(21)B3XZ5B, (01)00884838093935(21)F05FK8, (01)00884838093935(21)F07D0J, (01)00884838093935(21)F07D0K, (01)00884838093935(21)F08LD9, (01)00884838093935(21)F08MH1, (01)00884838093935(21)B3FLKD, (01)00884838093935(21)F07K8T, (01)00884838093935(21)F081VJ, (01)00884838093935(21)F094Y3, (01)00884838093935(21)F07WWK, (01)00884838093935(21)B3FLDL, (01)00884838093935(21)B3FLDN, (01)00884838093935(21)B3FLDQ, (01)00884838093935(21)F0JZQB, (01)00884838093935(21)F0K1LL, (01)00884838093935(21)F07GKZ, (01)00884838093935(21)F07WWV, (01)00884838093935(21)F07ZZL, (01)00884838093935(21)F07CZZ, (01)00884838093935(21)B3XZVJ, (01)00884838093935(21)B3XZXV, (01)00884838093935(21)F04MY9, (01)00884838093935(21)F04QBM, (01)00884838093935(21)F07K86, (01)00884838093935(21)F04QBH, (01)00884838093935(21)B3FLWH, (01)00884838093935(21)B3FLX6, (01)00884838093935(21)B3FLX9, (01)00884838093935(21)B3FLXB, (01)00884838093935(21)F09PVB, 010088483809393521B3Y0JZ, (01)00884838093935(21)F04XFF, (01)00884838093935(21)B41626, (01)00884838093935(21)B3FMKX, (01)00884838093935(21)F06KRD, (01)00884838093935(21)B3FL5W, (01)00884838093935(21)B3XYRX, (01)00884838093935(21)B41GNX, (01)00884838093935(21)F07ZZD, (01)00884838093935(21)F0GKW6, (01)00884838093935(21)F0GKW7, (01)00884838093935(21)F09Z34, (01)00884838093935(21)B411HH, (01)00884838093935(21)B41GNC, (01)00884838093935(21)F0K1LR, (01)00884838093935(21)F0J4HF, (01)00884838093935(21)F0F7ZL, (01)00884838093935(21)F0L9RY, (01)00884838093935(21)B3XYRN, (01)00884838093935(21)F0GM7P, (01)00884838093935(21)F07ZYY, 010088483809393521F0FQTL, (01)00884838093935(21)F0FKK8, (01)00884838093935(21)F0KR3Z, 010088483809393521B3FM7V, (01)00884838093935(21)F0JTXV, (01)00884838093935(21)F08R3N, (01)00884838093935(21)B3FM9R, (01)00884838093935(21)F06PPC, (01)00884838093935(21)F0JWYZ, (01)00884838093935(21)B3FL0W, (01)00884838093935(21)B3FL13, (01)00884838093935(21)B3FL21, (01)00884838093935(21)B3FL44, (01)00884838093935(21)B3FL6B, (01)00884838093935(21)B3FMDC, (01)00884838093935(21)B3XZ4J, (01)00884838093935(21)B3XZ59, (01)00884838093935(21)B3XZHC, (01)00884838093935(21)F0550F, (01)00884838093935(21)F0550J, (01)00884838093935(21)F06LJD, (01)00884838093935(21)F06LJZ, (01)00884838093935(21)F06PPB, (01)00884838093935(21)F0F7ZM, (01)00884838093935(21)F0KV7N, (01)00884838093935(21)F0KV7Q, (01)00884838093935(21)F0JTXN, (01)00884838093935(21)F07WWX, (01)00884838093935(21)F0HX2P, (01)00884838093935(21)F06XQ7, (01)00884838093935(21)F06XQB, (01)00884838093935(21)F06XQJ, (01)00884838093935(21)F081V2, (01)00884838093935(21)F0JTXY, (01)00884838093935(21)F0JTXZ, (01)00884838093935(21)F05BRB, (01)00884838093935(21)F06TG5, (01)00884838093935(21)B3L51H, (01)00884838093935(21)F0K1LM, (01)00884838093935(21)B3FL49, (01)00884838093935(21)B3FLD3, (01)00884838093935(21)F0GW61, (01)00884838093935(21)F06LJP, (01)00884838093935(21)F0JL23, (01)00884838093935(21)F0JL2B, (01)00884838093935(21)F0JWZ5, (01)00884838093935(21)F0JZQ3, (01)00884838093935(21)F0JZQ4, (01)00884838093935(21)F0JZQ5, (01)00884838093935(21)B3FLBC, (01)00884838093935(21)F09FZ3, (01)00884838093935(21)F0GJCC, (01)00884838093935(21)B3L50P, (01)00884838093935(21)F09FZ1, (01)00884838093935(21)B3FMD3, (01)00884838093935(21)B3XZHG, (01)00884838093935(21)F09D24, (01)00884838093935(21)F08LDP, (01)00884838093935(21)F09GG2, (01)00884838093935(21)F0GR33, (01)00884838093935(21)F0GW64, (01)00884838093935(21)B3Y0GK, (01)00884838093935(21)F0596Y, (01)00884838093935(21)F0KND4, (01)00884838093935(21)F06TFM, (01)00884838093935(21)B3FLTW, (01)00884838093935(21)F06XQ6, (01)00884838093935(21)F07160, (01)00884838093935(21)F04XDW, (01)00884838093935(21)F09MCG, (01)00884838093935(21)F0KK1F, (01)00884838093935(21)B3FM9W, (01)00884838093935(21)B3FMG7, (01)00884838093935(21)B3FMH5, (01)00884838093935(21)B3L507, (01)00884838093935(21)B3LBBK, (01)00884838093935(21)B3LBDC, (01)00884838093935(21)B3XZYH, (01)00884838093935(21)F04V3D, (01)00884838093935(21)F0B944, (01)00884838093935(21)F07D05, (01)00884838093935(21)F07RXZ, (01)00884838093935(21)F07RYH, (01)00884838093935(21)F07WW5, (01)00884838093935(21)F09PVL, (01)00884838093935(21)B3FL0M, (01)00884838093935(21)F07RYY, (01)00884838093935(21)F0HYTG, (01)00884838093935(21)B3XZ4F, (01)00884838093935(21)B3FLMG, (01)00884838093935(21)F0GR3J, (01)00884838093935(21)F04QBR, (01)00884838093935(21)F0596K, (01)00884838093935(21)F04YTL, (01)00884838093935(21)B3LBFP, (01)00884838093935(21)F0FKK0, (01)00884838093935(21)F0FP16, (01)00884838093935(21)F0B94R, (01)00884838093935(21)B3Y0GG, 010088483809393521B3Y0K1, (01)00884838093935(21)F0GCP3, (01)00884838093935(21)F0GFPZ, (01)00884838093935(21)B3FMKY, (01)00884838093935(21)F06XQM, (01)00884838093935(21)F094XW, (01)00884838093935(21)B3Y01W, (01)00884838093935(21)B41GNP, (01)00884838093935(21)B3XZT0, (01)00884838093935(21)B3XZXY, (01)00884838093935(21)F06KR7, (01)00884838093935(21)B3FLYF, (01)00884838093935(21)F0L7Q5, 010088483809393521F0GJC7, (01)00884838093935(21)F04YTH, (01)00884838093935(21)F05GQG, (01)00884838093935(21)F07D0M, (01)00884838093935(21)F06GWM, (01)00884838093935(21)F04QBF, (01)00884838093935(21)B3FLMP, (01)00884838093935(21)B3XZXJ, (01)00884838093935(21)F07P25, (01)00884838093935(21)F0L40H, 010088483809393521B3Y0KJ, 010088483809393521F0GCP5, (01)00884838093935(21)F0JTY1, (01)00884838093935(21)F0JZQ2, (01)00884838093935(21)F0L7PZ, (01)00884838093935(21)F07P28, (01)00884838093935(21)F05V8X, (01)00884838093935(21)F04MYB, (01)00884838093935(21)B3Y0GR, (01)00884838093935(21)B4161Q, (01)00884838093935(21)F04MXZ, (01)00884838093935(21)F05GQB, (01)00884838093935(21)F0841V, (01)00884838093935(21)F0FP15, (01)00884838093935(21)F0K7M0, (01)00884838093935(21)B3XZRJ, (01)00884838093935(21)B4161P, (01)00884838093935(21)B4161Y, (01)00884838093935(21)F0B948, (01)00884838093935(21)F05GQ9, (01)00884838093935(21)F05GQN, (01)00884838093935(21)B3XYRJ, 010088483809393521B3Y0K5, (01)00884838093935(21)B3XYRR, (01)00884838093935(21)B3FLTG, (01)00884838093935(21)F06TFJ, (01)00884838093935(21)F07164, (01)00884838093935(21)F0JWZ7, (01)00884838093935(21)F0JWZC, (01)00884838093935(21)B3XZRY, (01)00884838093935(21)B3FLRT, (01)00884838093935(21)F09Z2X, (01)00884838093935(21)F09Z36, (01)00884838093935(21)F09Z3D, (01)00884838093935(21)B3FLBT, (01)00884838093935(21)B41GNL, (01)00884838093935(21)B3FLMK, (01)00884838093935(21)B3XYRW, (01)00884838093935(21)F09D1P, (01)00884838093935(21)F09PV0, (01)00884838093935(21)B3XZVH, (01)00884838093935(21)B3XZYN, (01)00884838093935(21)F0GW63, (01)00884838093935(21)F0GW66, (01)00884838093935(21)F0GPHG, (01)00884838093935(21)F0GR35, (01)00884838093935(21)F0GR3B, (01)00884838093935(21)F0GR3G, (01)00884838093935(21)F07P20, (01)00884838093935(21)F07P21, 010088483809393521F0FQTH, 010088483809393521F0FRHV, (01)00884838093935(21)F07ZYX, (01)00884838093935(21)B41GQW, (01)00884838093935(21)F04QBY, (01)00884838093935(21)F04QBP, (01)00884838093935(21)F0JZQ1, (01)00884838093935(21)F0K1LH, (01)00884838093935(21)F0K7LT, (01)00884838093935(21)F09GFX, (01)00884838093935(21)B41GPC, (01)00884838093935(21)F06XQK, (01)00884838093935(21)B3XZQR, (01)00884838093935(21)F06LK0, (01)00884838093935(21)B3Y0GT, (01)00884838093935(21)F07K8K, (01)00884838093935(21)B3Y0GH, (01)00884838093935(21)F0FP1F, (01)00884838093935(21)B3FLFL, (01)00884838093935(21)B3FML9, (01)00884838093935(21)B3FLX3, (01)00884838093935(21)F0FP17, (01)00884838093935(21)B3XZFP, (01)00884838093935(21)F06LJH, (01)00884838093935(21)F06LJM, (01)00884838093935(21)F0715D, (01)00884838093935(21)F0715X, (01)00884838093935(21)F09BZJ, (01)00884838093935(21)F0J6DH, (01)00884838093935(21)B3FLDK, (01)00884838093935(21)B3Y0GV, (01)00884838093935(21)B3XZXQ, (01)00884838093935(21)B3FMBX, (01)00884838093935(21)F09GG0, 010088483809393521B3Y0K9, 010088483809393521B3Y0KB, (01)00884838093935(21)B3FL19, (01)00884838093935(21)B3XZVV, (01)00884838093935(21)B3XZXZ, (01)00884838093935(21)F09D2B, (01)00884838093935(21)F09FZB, (01)00884838093935(21)F09GG5, (01)00884838093935(21)F09GGB, (01)00884838093935(21)F09FZ2, (01)00884838093935(21)F09GG9, (01)00884838093935(21)F09GGG, (01)00884838093935(21)F0B94L, (01)00884838093935(21)B3XZ2Q, (01)00884838093935(21)B4161M, (01)00884838093935(21)B4161R, (01)00884838093935(21)B41621, (01)00884838093935(21)B3FL0N, (01)00884838093935(21)B3FL0P, (01)00884838093935(21)F0842F, (01)00884838093935(21)F0L09L, (01)00884838093935(21)B3FLMH, (01)00884838093935(21)F08MH6, (01)00884838093935(21)B4162F, (01)00884838093935(21)B41GQL, (01)00884838093935(21)B41628, (01)00884838093935(21)F06LJW, (01)00884838093935(21)F06TFX, (01)00884838093935(21)F06TG3, (01)00884838093935(21)F06XQL, (01)00884838093935(21)F0716F, (01)00884838093935(21)B3XZY9, (01)00884838093935(21)B3Y00X, (01)00884838093935(21)F0716D, (01)00884838093935(21)F07K8Q, (01)00884838093935(21)F04YTV, (01)00884838093935(21)F07GKX, (01)00884838093935(21)F07K84, (01)00884838093935(21)F07RY1, 010088483809393521F0FP11, (01)00884838093935(21)B3XZ29, (01)00884838093935(21)F04XF7, (01)00884838093935(21)F0KK1D, (01)00884838093935(21)F08MH3, (01)00884838093935(21)F0JWZ1, (01)00884838093935(21)B3FL3F, (01)00884838093935(21)B3FL3T, (01)00884838093935(21)B3XZVK, (01)00884838093935(21)B3XZVN, (01)00884838093935(21)B3XZVZ, (01)00884838093935(21)B3XZXM, (01)00884838093935(21)B3FL3V, (01)00884838093935(21)B3FL4B, (01)00884838093935(21)B3FL4F, (01)00884838093935(21)B3FL4J, (01)00884838093935(21)B41GR3, 010088483809393521B3Y0K3, (01)00884838093935(21)F0715L, (01)00884838093935(21)F0GR31, (01)00884838093935(21)B3XYT1, (01)00884838093935(21)F0J6DQ, (01)00884838093935(21)B41GR1, (01)00884838093935(21)B41GRD, (01)00884838093935(21)B3FLG4, (01)00884838093935(21)F04V3K, (01)00884838093935(21)F0715H, (01)00884838093935(21)B3FMBH, (01)00884838093935(21)B3FMBW, (01)00884838093935(21)F07K8N, (01)00884838093935(21)B3FLKW, (01)00884838093935(21)F05513, (01)00884838093935(21)F05970, (01)00884838093935(21)F05FKK, (01)00884838093935(21)B3XZYJ, (01)00884838093935(21)B3Y01P, (01)00884838093935(21)F0GPHD, (01)00884838093935(21)B3FLD7, (01)00884838093935(21)B411HX, (01)00884838093935(21)B41GNF, (01)00884838093935(21)F04XF9, (01)00884838093935(21)B411HV, (01)00884838093935(21)B411J3, (01)00884838093935(21)B41GN0, (01)00884838093935(21)B41GPB, (01)00884838093935(21)F06PPJ, (01)00884838093935(21)F07D0D, (01)00884838093935(21)F0842G, (01)00884838093935(21)B3XZYM, (01)00884838093935(21)B3Y00P, (01)00884838093935(21)F0715K, (01)00884838093935(21)B3FLTQ, (01)00884838093935(21)F0L15K, (01)00884838093935(21)F0L15L, (01)00884838093935(21)B3FLBM, (01)00884838093935(21)B3FLXF, (01)00884838093935(21)F07D06, (01)00884838093935(21)F07D07, (01)00884838093935(21)F0L7QB, (01)00884838093935(21)B3FLDD, (01)00884838093935(21)F08H3J, (01)00884838093935(21)F06KR8, (01)00884838093935(21)F0596N, (01)00884838093935(21)F0596P, (01)00884838093935(21)F0596Z, (01)00884838093935(21)F05BRC, (01)00884838093935(21)B3XZ5C, (01)00884838093935(21)B411H0, (01)00884838093935(21)F07D09, (01)00884838093935(21)F07RYJ, (01)00884838093935(21)F07ZZ8, (01)00884838093935(21)F0GKW3, (01)00884838093935(21)F0JBPD, (01)00884838093935(21)F08R3D, (01)00884838093935(21)B3FL42, (01)00884838093935(21)F06PPQ, (01)00884838093935(21)F08MGM, (01)00884838093935(21)F0596X, (01)00884838093935(21)F05BR4, (01)00884838093935(21)F0FP1G, 010088483809393521F0GPH7, (01)00884838093935(21)F09Z32, (01)00884838093935(21)F0842D, (01)00884838093935(21)F08H3G, (01)00884838093935(21)F0596H, (01)00884838093935(21)F05HX3, (01)00884838093935(21)F0GJCT, (01)00884838093935(21)F0HYTJ, (01)00884838093935(21)F0JDZJ, (01)00884838093935(21)F0JL27, (01)00884838093935(21)F0FKJP, (01)00884838093935(21)B3Y03J, (01)00884838093935(21)F09GGF, (01)00884838093935(21)F0HX2Q, (01)00884838093935(21)F0J9YK, (01)00884838093935(21)F06GX8, (01)00884838093935(21)B3XZMK, (01)00884838093935(21)B3XZQM, (01)00884838093935(21)B3XZQX, (01)00884838093935(21)B3XZR0, (01)00884838093935(21)B3XZR1, (01)00884838093935(21)F0GT04, (01)00884838093935(21)F07158, (01)00884838093935(21)F0KNDC, (01)00884838093935(21)F0KNDH, 010088483809393521F0FQTC, 010088483809393521F0FRHW, (01)00884838093935(21)F07GL1, (01)00884838093935(21)F04MY3, (01)00884838093935(21)B411GW, (01)00884838093935(21)B411GY, (01)00884838093935(21)B41GN3, (01)00884838093935(21)F0GKWD, (01)00884838093935(21)B3XYRT, (01)00884838093935(21)B3XZ4L, (01)00884838093935(21)F07K8F, (01)00884838093935(21)F07P1X, (01)00884838093935(21)B3Y039, (01)00884838093935(21)F06PPN, (01)00884838093935(21)F0K7LV, (01)00884838093935(21)F0K7M5, (01)00884838093935(21)B3Y00H, (01)00884838093935(21)F05FKN, (01)00884838093935(21)F0L40K, (01)00884838093935(21)F07P2D, (01)00884838093935(21)B3XZRL, (01)00884838093935(21)B3FM2F, (01)00884838093935(21)B3FL20, (01)00884838093935(21)F094DW, (01)00884838093935(21)F09FZF, (01)00884838093935(21)F09JNL, (01)00884838093935(21)F09JNP, (01)00884838093935(21)F09MCJ, (01)00884838093935(21)F09PTQ, (01)00884838093935(21)F09PTV, (01)00884838093935(21)F09PV3, (01)00884838093935(21)F09PV7, (01)00884838093935(21)F09Z30, (01)00884838093935(21)B3Y01M, 010088483809393521F0FYBX, (01)00884838093935(21)B3XZRQ, (01)00884838093935(21)B3Y035, (01)00884838093935(21)F0K1LZ, (01)00884838093935(21)F07K89, (01)00884838093935(21)F081V5, (01)00884838093935(21)F081W3, (01)00884838093935(21)F0841J, (01)00884838093935(21)F0716B, (01)00884838093935(21)F04QBJ, (01)00884838093935(21)B3FLDP, (01)00884838093935(21)F08LDH, (01)00884838093935(21)F09BZK, (01)00884838093935(21)F09D1W, (01)00884838093935(21)F0GCP6, (01)00884838093935(21)F0GJCD, (01)00884838093935(21)B3FLWG, (01)00884838093935(21)F09BZP, (01)00884838093935(21)F09D1G, (01)00884838093935(21)F09MC6, (01)00884838093935(21)F09MC9, (01)00884838093935(21)F0KV7M, (01)00884838093935(21)F0KXZM, 010088483809393521F0G0M0, (01)00884838093935(21)B3FLD9, (01)00884838093935(21)B3FLDB, (01)00884838093935(21)B3FLDH, (01)00884838093935(21)B3FLFJ, (01)00884838093935(21)B3FLWC, (01)00884838093935(21)B3XZT4, (01)00884838093935(21)B3XZYL, (01)00884838093935(21)F07GKC, (01)00884838093935(21)F07K8R, (01)00884838093935(21)F07RY0, (01)00884838093935(21)F07ZZG, (01)00884838093935(21)F0GJCM, (01)00884838093935(21)F0L7PV, (01)00884838093935(21)F09MC2, (01)00884838093935(21)B3XZ21, (01)00884838093935(21)B3XZMM, 010088483809393521F0FRJ2, (01)00884838093935(21)B3FLTL, (01)00884838093935(21)F06TGB, (01)00884838093935(21)B3FKZB, (01)00884838093935(21)B3Y00M, (01)00884838093935(21)F04XDP, (01)00884838093935(21)F0HFNJ, (01)00884838093935(21)F05V8R, (01)00884838093935(21)F0GKW9, (01)00884838093935(21)F07P2R, (01)00884838093935(21)F0GPH6, (01)00884838093935(21)B3FL8K, (01)00884838093935(21)F05BR8, (01)00884838093935(21)F07GKV, (01)00884838093935(21)F04QC5, (01)00884838093935(21)F09BZG, (01)00884838093935(21)F09D1L, (01)00884838093935(21)F0715Z, (01)00884838093935(21)B3LBBX, (01)00884838093935(21)B3XZRV, 010088483809393521F0FRHT, (01)00884838093935(21)F07D0H, (01)00884838093935(21)F06XQQ, (01)00884838093935(21)F08H3T, (01)00884838093935(21)F04YV1, (01)00884838093935(21)F06GWJ, (01)00884838093935(21)B3FMGZ, (01)00884838093935(21)B3FMC3, (01)00884838093935(21)F0L7Q8, (01)00884838093935(21)F0L40M, (01)00884838093935(21)F07ZZJ, (01)00884838093935(21)F09D23, (01)00884838093935(21)F05GQQ, (01)00884838093935(21)B3FLKT, (01)00884838093935(21)B3FLMF, (01)00884838093935(21)B3FLTV, (01)00884838093935(21)F06GWK, (01)00884838093935(21)F06TFL, (01)00884838093935(21)F07RYK, (01)00884838093935(21)F07RYX, (01)00884838093935(21)F0FKJT, (01)00884838093935(21)F0FKJV, 010088483809393521B3Y0LR, (01)00884838093935(21)F0GKVY, (01)00884838093935(21)B3FMF2, (01)00884838093935(21)B3FMFJ, (01)00884838093935(21)F06TG2, 010088483809393521B3FM2N, (01)00884838093935(21)F09D27, (01)00884838093935(21)F04QBQ, (01)00884838093935(21)B3XZ1Q, (01)00884838093935(21)B3XZ45, (01)00884838093935(21)B3XZQY, (01)00884838093935(21)F07P23, (01)00884838093935(21)F07ZZH, (01)00884838093935(21)F08MGV, (01)00884838093935(21)F09BZX, (01)00884838093935(21)F09MCD, (01)00884838093935(21)F0J4YV, (01)00884838093935(21)F09FYV, (01)00884838093935(21)F09MCF, (01)00884838093935(21)F09Z35, (01)00884838093935(21)F07RYC, (01)00884838093935(21)F0KQJH, (01)00884838093935(21)F0KQJM, (01)00884838093935(21)F0KQJP, (01)00884838093935(21)F0715J, (01)00884838093935(21)F08LDQ, (01)00884838093935(21)F0JZQ9, (01)00884838093935(21)F07WWB, (01)00884838093935(21)F04V38, (01)00884838093935(21)B3FMGG, (01)00884838093935(21)B3FMH2, (01)00884838093935(21)B3FMH6, (01)00884838093935(21)F07WX0, (01)00884838093935(21)F0FKJL, (01)00884838093935(21)F0FP18, (01)00884838093935(21)F0JZQ8, (01)00884838093935(21)F09D1V, (01)00884838093935(21)F0JL2D, (01)00884838093935(21)F04XF0, (01)00884838093935(21)F06GX2, (01)00884838093935(21)B3XZFQ, (01)00884838093935(21)B3FLY6, (01)00884838093935(21)F06LJY, (01)00884838093935(21)F06LKB, (01)00884838093935(21)F04V3F, (01)00884838093935(21)F0FKJ9, (01)00884838093935(21)F0FKK7, (01)00884838093935(21)F0550R, (01)00884838093935(21)F09D29, (01)00884838093935(21)F0HP35, (01)00884838093935(21)F07D08, (01)00884838093935(21)B3XZRP, (01)00884838093935(21)B3XZY0, (01)00884838093935(21)B3FMC5, (01)00884838093935(21)F07K8B, (01)00884838093935(21)F07RY4, (01)00884838093935(21)F08H3N, (01)00884838093935(21)B3FLYH, 010088483809393521B3FM50, (01)00884838093935(21)B3XZXX, (01)00884838093935(21)F09FZ6, (01)00884838093935(21)F07K8D, (01)00884838093935(21)F07GKK, (01)00884838093935(21)F06GWB, (01)00884838093935(21)B3XZ4R, (01)00884838093935(21)F07D0B, (01)00884838093935(21)F06PPZ, (01)00884838093935(21)F06TFN, (01)00884838093935(21)F06XQY, (01)00884838093935(21)F0F4BZ, (01)00884838093935(21)F0KQJQ, (01)00884838093935(21)F0KXZX, (01)00884838093935(21)B3XZYQ, (01)00884838093935(21)F0596T, (01)00884838093935(21)F05BR5, (01)00884838093935(21)F06KRB, (01)00884838093935(21)F06KRP, 010088483809393521F0FQTN, 010088483809393521F0FYBT, 010088483809393521F0FYBZ, 010088483809393521F0FYC0, (01)00884838093935(21)F07P27, (01)00884838093935(21)F07P2N, (01)00884838093935(21)F07WWG, (01)00884838093935(21)F081V8, (01)00884838093935(21)F081VK, (01)00884838093935(21)F081VZ, (01)00884838093935(21)F081W0, (01)00884838093935(21)B3XZ51, (01)00884838093935(21)B3XZ58, (01)00884838093935(21)B3XZFM, (01)00884838093935(21)F05V8V, (01)00884838093935(21)B3FMGN, (01)00884838093935(21)B3XZ2R, (01)00884838093935(21)B411H6, (01)00884838093935(21)F07GKR, (01)00884838093935(21)B3FMGX, (01)00884838093935(21)F04XFB, (01)00884838093935(21)F07D04, (01)00884838093935(21)B3XZHJ, (01)00884838093935(21)F09D28, (01)00884838093935(21)F09GFT, (01)00884838093935(21)B3FLD6, (01)00884838093935(21)F06TGG, (01)00884838093935(21)F07CZX, (01)00884838093935(21)F0HV5T, (01)00884838093935(21)B3XZ4T, 010088483809393521F0G5YB, (01)00884838093935(21)B41GNZ, (01)00884838093935(21)B3XZYG, (01)00884838093935(21)B3XZHK, (01)00884838093935(21)F08MGX, (01)00884838093935(21)F0B94T, (01)00884838093935(21)F0L9T9, (01)00884838093935(21)F0JWZ8, (01)00884838093935(21)F07K8P, (01)00884838093935(21)B3XZ5R, (01)00884838093935(21)F05HX5, 010088483809393521F0FRJ1, (01)00884838093935(21)B411GZ, (01)00884838093935(21)B3Y01N, (01)00884838093935(21)F0JWZ4, (01)00884838093935(21)B41GR7, (01)00884838093935(21)B3Y01X, (01)00884838093935(21)B3Y0GJ, (01)00884838093935(21)F0FKKB, 010088483809393521F0FQTD, (01)00884838093935(21)F06LJF, (01)00884838093935(21)B3Y0GN, 010088483809393521B3Y0KD, 010088483809393521B3Y0K2, (01)00884838093935(21)F0KC7D, (01)00884838093935(21)F0L9T1, (01)00884838093935(21)B411J1, (01)00884838093935(21)B41GQY, (01)00884838093935(21)F0B945, 010088483809393521F0FP10, (01)00884838093935(21)F07K8H, (01)00884838093935(21)B41GN4, (01)00884838093935(21)B41GNB, (01)00884838093935(21)B41GR2, (01)00884838093935(21)B3XZXN, (01)00884838093935(21)F06LJJ, (01)00884838093935(21)F07165, (01)00884838093935(21)B3FMH0, (01)00884838093935(21)B3FMGY, (01)00884838093935(21)F0841W, (01)00884838093935(21)F0841X, (01)00884838093935(21)F05HX2, (01)00884838093935(21)F07GKT, (01)00884838093935(21)F07K8W, (01)00884838093935(21)F07RY9, (01)00884838093935(21)F0FKK2, (01)00884838093935(21)F06PPD, (01)00884838093935(21)F05GQH, (01)00884838093935(21)B3XYRK, (01)00884838093935(21)F0715G, (01)00884838093935(21)B3FLRP, (01)00884838093935(21)B3FLTD, (01)00884838093935(21)F04V39, 010088483809393521B3FM7Q, 010088483809393521B3FM7R, 010088483809393521B3Y0L9, 010088483809393521B3Y0LC, 010088483809393521B3Y0LF, 010088483809393521B3Y0LG, (01)00884838093935(21)F081VH, (01)00884838093935(21)F0K1LP, (01)00884838093935(21)F07ZZ4, (01)00884838093935(21)F09PV6, (01)00884838093935(21)F09PVF, (01)00884838093935(21)F06XQ9, (01)00884838093935(21)F0841P, (01)00884838093935(21)F09FYW, (01)00884838093935(21)F09FZ0, (01)00884838093935(21)F09FZ4, (01)00884838093935(21)F09PV2, 010088483809393521B3FM53, 010088483809393521B3FM7P, (01)00884838093935(21)B3FLFK, (01)00884838093935(21)F09PV9, (01)00884838093935(21)B3FL4C, (01)00884838093935(21)F07K8L, (01)00884838093935(21)F09GFY, (01)00884838093935(21)F07RY2, (01)00884838093935(21)F09PTP, (01)00884838093935(21)F06TGC, (01)00884838093935(21)B3FLMR, 010088483809393521B3FM4Z, 010088483809393521F0FP0Z, (01)00884838093935(21)F081W1, (01)00884838093935(21)F098FR, (01)00884838093935(21)F0KC77, (01)00884838093935(21)B3Y01L, (01)00884838093935(21)B3Y01Y, (01)00884838093935(21)F0K1LX, (01)00884838093935(21)F0GV23, (01)00884838093935(21)B3FML4, (01)00884838093935(21)B411H8, (01)00884838093935(21)B411HG, (01)00884838093935(21)F0JRVL, (01)00884838093935(21)B3Y00R, (01)00884838093935(21)B3XZ54, (01)00884838093935(21)F0L7QC, (01)00884838093935(21)B3XZMJ, (01)00884838093935(21)B3XZQW, (01)00884838093935(21)F05HX4, (01)00884838093935(21)F06GX3, (01)00884838093935(21)F0F163, (01)00884838093935(21)F08R34, (01)00884838093935(21)F07P2H, (01)00884838093935(21)F0GR2Z, (01)00884838093935(21)B3FL0V, (01)00884838093935(21)F0841Q, (01)00884838093935(21)F08MGJ, (01)00884838093935(21)F0K1LK, (01)00884838093935(21)F0GJCB, (01)00884838093935(21)B3FLTJ, (01)00884838093935(21)F05FK9, 010088483809393521F0GJC8, (01)00884838093935(21)B3FLXM, (01)00884838093935(21)B3XZMR, (01)00884838093935(21)F07GKL, (01)00884838093935(21)B4161W, (01)00884838093935(21)F0LC8Q, (01)00884838093935(21)F094Y6, (01)00884838093935(21)F09BZF, (01)00884838093935(21)B3XZRR, (01)00884838093935(21)F05BR1, (01)00884838093935(21)F0J6DJ, (01)00884838093935(21)F0JDZT, (01)00884838093935(21)B3FLRZ, (01)00884838093935(21)F04YTQ, (01)00884838093935(21)F0550M, (01)00884838093935(21)B3XZ2N, (01)00884838093935(21)B3XZ48, 010088483809393521B3FM5D, (01)00884838093935(21)B411HF, (01)00884838093935(21)B41GNQ, (01)00884838093935(21)B3Y01Q, (01)00884838093935(21)F0F4BG, (01)00884838093935(21)B3XZY2, (01)00884838093935(21)B3XZYD, (01)00884838093935(21)B3FL4H, (01)00884838093935(21)F05BQW, (01)00884838093935(21)B3FLXJ, (01)00884838093935(21)B3XZXW, 010088483809393521F0G34B, (01)00884838093935(21)F06PQ1, (01)00884838093935(21)F05HXC, (01)00884838093935(21)F0GV20, (01)00884838093935(21)F0GV25, (01)00884838093935(21)F0HFND, (01)00884838093935(21)B3XYZ2, (01)00884838093935(21)B3XZ1V, (01)00884838093935(21)B411HD, (01)00884838093935(21)B411HW, 010088483809393521B3FM2M, 010088483809393521B3FM7B, 010088483809393521F0GPH9, (01)00884838093935(21)F0GR3C, (01)00884838093935(21)B41GQN, (01)00884838093935(21)B3XZ4N, (01)00884838093935(21)B3Y033, (01)00884838093935(21)F0550Q, (01)00884838093935(21)B3FLBW, (01)00884838093935(21)B3LBBQ, (01)00884838093935(21)B3FL65, (01)00884838093935(21)F04XDM, (01)00884838093935(21)F0K7LX, (01)00884838093935(21)F0FQTG, (01)00884838093935(21)F06PPL, (01)00884838093935(21)F0KXZZ, (01)00884838093935(21)F0L02V, (01)00884838093935(21)F0L09V, (01)00884838093935(21)F0L15C, (01)00884838093935(21)F0L15H, (01)00884838093935(21)F0L15M, (01)00884838093935(21)F0L15N, (01)00884838093935(21)F0L40J, (01)00884838093935(21)F0L40P, (01)00884838093935(21)F0L40Q, 010088483809393521F0FYBY, 010088483809393521F0G34G, 010088483809393521F0G5Y5, 010088483809393521F0G5YC, 010088483809393521F0GFPX, 010088483809393521F0GFPY, (01)00884838093935(21)F0F4BL, (01)00884838093935(21)F07P2T, (01)00884838093935(21)F0715R, (01)00884838093935(21)F0841Z, (01)00884838093935(21)F08425, (01)00884838093935(21)B3FMBP, (01)00884838093935(21)B3FMF5, (01)00884838093935(21)B3LBG0, (01)00884838093935(21)F0KND6, (01)00884838093935(21)B3FL67, (01)00884838093935(21)B3FL62, (01)00884838093935(21)B411HK, 010088483809393521B3FM52, (01)00884838093935(21)B3FL5Z, (01)00884838093935(21)F07RYR, (01)00884838093935(21)F06GWL, 010088483809393521B3Y0LQ, (01)00884838093935(21)B3FLT6, (01)00884838093935(21)F0GR30, (01)00884838093935(21)B411HT, (01)00884838093935(21)F0715V, (01)00884838093935(21)F06GW9, (01)00884838093935(21)F06PPY, (01)00884838093935(21)F0J4HJ, (01)00884838093935(21)F0J4HM, (01)00884838093935(21)F0J6DG, (01)00884838093935(21)F0J6DK, (01)00884838093935(21)B41GPD, (01)00884838093935(21)F08420, (01)00884838093935(21)B3XZVM, (01)00884838093935(21)F0841F, (01)00884838093935(21)B3Y0K7, (01)00884838093935(21)F07ZZB, (01)00884838093935(21)F081VB, (01)00884838093935(21)F08H3L, 010088483809393521F0FYC1, (01)00884838093935(21)F0F4BD, (01)00884838093935(21)F0F4C3, (01)00884838093935(21)F0F4C5, (01)00884838093935(21)F0GT02, (01)00884838093935(21)F0HFNK, (01)00884838093935(21)B3XZ56, (01)00884838093935(21)B3XZVY, (01)00884838093935(21)F09D1N, (01)00884838093935(21)B41GP6, (01)00884838093935(21)B41GN2, (01)00884838093935(21)F0715N, (01)00884838093935(21)F05BQZ, (01)00884838093935(21)B41GR6, (01)00884838093935(21)F0KR40, (01)00884838093935(21)F06GWD, (01)00884838093935(21)F06TFQ, (01)00884838093935(21)F08LDG, (01)00884838093935(21)F06GWG, (01)00884838093935(21)F07WW6, (01)00884838093935(21)F06TFH, (01)00884838093935(21)F06XQG, (01)00884838093935(21)F08423, (01)00884838093935(21)F08LDN, (01)00884838093935(21)F08MGP, (01)00884838093935(21)F07155, (01)00884838093935(21)F0HFNG, (01)00884838093935(21)F07RY5, (01)00884838093935(21)F07P24, (01)00884838093935(21)F04QBZ, (01)00884838093935(21)F07RYF, (01)00884838093935(21)F0K7M2, (01)00884838093935(21)F0KK1P, (01)00884838093935(21)F09FZ7, (01)00884838093935(21)F09MC4, (01)00884838093935(21)F07D01, (01)00884838093935(21)F09BZV, (01)00884838093935(21)B3FLTT, (01)00884838093935(21)F0KK1Q, (01)00884838093935(21)F0KK1R, (01)00884838093935(21)B3LBF6, (01)00884838093935(21)F09PV8, (01)00884838093935(21)F09MCC, (01)00884838093935(21)F0GV22, (01)00884838093935(21)F0GV21, (01)00884838093935(21)F0GV28, (01)00884838093935(21)F0FKJM, (01)00884838093935(21)B3XYYZ, (01)00884838093935(21)B4161J, (01)00884838093935(21)F0841R, (01)00884838093935(21)F094Y5, (01)00884838093935(21)F04YTD, (01)00884838093935(21)F081VC, (01)00884838093935(21)F081VT, (01)00884838093935(21)F0JDZM, (01)00884838093935(21)F0JDZP, 010088483809393521F0G8CF, 010088483809393521F0GFPF, 010088483809393521F0G0LY, 010088483809393521F0G0M2, (01)00884838093935(21)F0GKW8, (01)00884838093935(21)F0HVCD, (01)00884838093935(21)F04YT8, (01)00884838093935(21)F0JPPV, (01)00884838093935(21)F06GWV, (01)00884838093935(21)B3FLXL, (01)00884838093935(21)B3FMDW; Serial No. F0FKJ7, F0FRHZ, F0G8CP, F0G8CM, B3XYRY, B4161X, F07K8J, F0JTY0, F04YTZ, F0550K, F094Y9, F0GJC9, B3XZRM, F0KC6Y, F0HP38, F0F164, F0KND1, F04YV6, B3Y0KC, F0JZQ6, F0JZQ7, B3XZ5Q, F07WW2, F07WW9, F07WWF, F07WWJ, F07WWL, F07WWN, F07WWP, F07WWQ, F07ZZ0, F081V3, F08H3V, F05FK6, F06GWZ, F07D0C, F0B94X, F0GJCR, B3FKZJ, B3FKZN, F0LC8N, F04QBB, F04V3M, F04V3Q, F04XDN, F04XF1, F04XF6, F06XQF, F0F4BF, F0F4BN, F0F7ZJ, F0FKK9, B3Y0KG, F06XQV, F06KRF, F0JPPM, F0K7LQ, B3FLBF, B3FLDJ, F0G346, B3XZRN, B3XZY3, B3XZ1X, B41GND, F0KC6Z, B3FLFY, F04QBW, F0JL20, B3FL3H, B3Y0LX, B3FLML, F04YV4, F094F4, F09BZY, B3FMGT, B3FMGV, F0715Y, F07166, F07D0G, F08H3P, F0JWZ9, F09GFP, B3FKZM, B3FL0R, B3FLT4, F08R38, B41GNT, B3FM7L, B3Y0LB, B3Y0LH, B3Y0LJ, F07K85, F0JPPL, F0JRVG, F0JZQ0, B3FL5X, B3L524, F04QB9, F04QBC, F07RY3, F09BZW, F06GWR, F0FP12, F04QC0, B3FLXG, B3FLXP, B3FLXR, B3FLXT, B3FMFF, B3XZMQ, B3XZXP, F07GKY, F07RY7, F07RYB, F07WWY, F0842B, F098FT, F0HFNF, F0K7LM, F0K7LN, F0K7LP, F0K7LR, F0K7M7, B3FLMX, F0FP13, F04QBK, F06LJT, F06TG7, F06XQZ, F0715M, F07ZZ5, F07ZZ6, F08H3M, F08MG7, F0FRJ0, F0HP34, F0GJCQ, F0550W, F0596C, F094F3, F0JRVD, B3Y038, B3Y052, B3Y0GL, F0JDZK, F0JH08, F0JH0G, F0JL28, F04QBT, F04V3C, B3Y0GP, B3FLK9, F0596V, F05BQR, B3FM7D, F0K7M4, B3FLTN, B3FLRX, F0841G, F0KQJK, F09JNR, F09PVK, F0F4C7, B3FL3G, B3FL3J, F07GL0, B3FLFQ, F0GJC5, B3Y02Y, B3Y050, F0FP0Q, F06KR9, F0GFQ2, B3FMBQ, F081VL, F0841K, F09BZT, B3FLT3, B3FM57, B3FM7T, F06GWX, F0KNDF, F08MG9, F0JTXP, F07GKG, F07RY8, F0GR37, F0HV5V, B3FL47, B41625, B41629, B4162C, F0KC74, B3Y05C, B3XZ5B, F05FK8, F07D0J, F07D0K, F08LD9, F08MH1, B3FLKD, F07K8T, F081VJ, F094Y3, F07WWK, B3FLDL, B3FLDN, B3FLDQ, F0JZQB, F0K1LL, B3FLV0, F07GKZ, F07WWV, F07ZZL, F07CZZ, B3XZVJ, B3XZXV, F04MY9, F04QBM, F07K86, F04QBH, B3FLWH, B3FLX6, B3FLX9, B3FLXB, F09PVB, B3Y0JZ, F04XFF, B41626, B3FMKX, F06KRD, B3FL5W, B3XYRX, B41GNX, F07ZZD, F0GKW6, F0GKW7, F09Z34, B411HH, B41GNC, F0K1LR, F0J4HF, F0F7ZL, F0L9RY, B3XYRN, F0GM7P, F07ZYY, F0FQTL, F0FKK8, F0KR3Z, B3FM7V, F0JTXV, F08R3N, B3FM9R, F06PPC, F0JWYZ, B3FL0W, B3FL13, B3FL21, B3FL44, B3FL6B, B3FMDC, B3XZ4J, B3XZ59, B3XZHC, F0550F, F0550J, F06LJD, F06LJZ, F06PPB, F0F7ZM, F0KV7N, F0KV7Q, F0JTXN, F07WWX, F0HX2P, F06XQ7, F06XQB, F06XQJ, F081V2, F0JTXY, F0JTXZ, F05BRB, F06TG5, B3L51H, F0K1LM, B3FL49, B3FLD3, F0GW61, F06LJP, F0JL23, F0JL2B, F0JWZ5, F0JZQ3, F0JZQ4, F0JZQ5, B3FLBC, F09FZ3, F0GJCC, B3L50P, F09FZ1, B3FMD3, B3XZHG, F09D24, F08LDP, F09GG2, F0GR33, F0GW64, B3Y0GK, F0596Y, F0KND4, F06TFM, B3FLTW, F06XQ6, F07160, F04XDW, F09MCG, F0KK1F, B3FM9W, B3FMG7, B3FMH5, B3L507, B3LBBK, B3LBDC, B3XZYH, F04V3D, F0B944, F07D05, F07RXZ, F07RYH, F07WW5, F09PVL, B3FL0M, F07RYY, F0HYTG, B3XZ4F, B3FLMG, F0GR3J, F04QBR, F0596K, F04YTL, B3LBFP, F0FKK0, F0FP16, F0B94R, B3Y0GG, B3Y0K1, F0GCP3, F0GFPZ, B3FMKY, F06XQM, F094XW, B3Y01W, B41GNP, B3XZT0, B3XZXY, F06KR7, B3FLYF, F0L7Q5, F0GJC7, F04YTH, F05GQG, F07D0M, F06GWM, B3FMGL, B3LBDB, F04QBF, B3FLMP, B3XZXJ, F07P25, F0L40H, B3Y0KJ, F0GCP5, F0JTY1, F0JZQ2, F0L7PZ, F07P28, F05V8X, F04MYB, B3Y0GR, B4161Q, F04MXZ, F05GQB, F0841V, F0FP15, F0K7M0, B3XZRJ, B4161P, B4161Y, F0B948, F05GQ9, F05GQN, B3XYRJ, B3Y0K5, B3XYRR, B3FLTG, F06TFJ, F07164, F0JWZ7, F0JWZC, B3XZRY, B3FLRT, F09Z2X, F09Z36, F09Z3D, B3FLBT, B41GNL, B3FLMK, B3XYRW, F09D1P, F09PV0, B3XZVH, B3XZYN, F0GW63, F0GW66, F0GPHG, F0GR35, F0GR3B, F0GR3G, F07P20, F07P21, F0FQTH, F0FRHV, F07ZYX, B41GQW, F04QBY, F04QBP, F0JZQ1, F0K1LH, F0K7LT, F09GFX, B41GPC, F06XQK, B3XZQR, F06LK0, B3Y0GT, F07K8K, B3Y0GH, F0FP1F, B3FLFL, B3FML9, B3FLX3, F0FP17, B3XZFP, F06LJH, F06LJM, F0715D, F0715X, F09BZJ, F0J6DH, B3FLDK, B3Y0GV, B3XZXQ, B3FMBX, F09GG0, B3Y0K9, B3Y0KB, B3FL19, B3XZVV, B3XZXZ, F09D2B, F09FZB, F09GG5, F09GGB, F09FZ2, F09GG9, F09GGG, F0B94L, B3XZ2Q, B4161M, B4161R, B41621, B3FL0N, B3FL0P, F0842F, F0L09L, B3FLMH, F08MH6, B4162F, B41GQL, B41628, F06LJW, F06TFX, F06TG3, F06XQL, F0716F, B3XZY9, B3Y00X, F0716D, F07K8Q, F04YTV, F07GKX, F07K84, F07RY1, F0FP11, B3XZ29, F04XF7, F0KK1D, F08MH3, F0JWZ1, B3FL3F, B3FL3T, B3XZVK, B3XZVN, B3XZVZ, B3XZXM, B3FL3V, B3FL4B, B3FL4F, B3FL4J, B41GR3, B3Y0K3, F0715L, F0GR31, B3XYT1, F0J6DQ, B41GR1, B41GRD, B3FLG4, F04V3K, F0715H, B3FMBH, B3FMBW, F07K8N, B3FLKW, F05513, F05970, F05FKK, B3XZYJ, B3Y01P, F0GPHD, B3FLD7, B411HX, B41GNF, F04XF9, B411HV, B411J3, B41GN0, B41GPB, F06PPJ, F07D0D, F0842G, B3XZYM, B3Y00P, F0715K, B3FLTQ, F0L15K, F0L15L, F0G8CN, F0FYC3, B3FLBM, B3FLXF, F07D06, F07D07, F0L7QB, B3FLDD, F08H3J, F06KR8, F0596N, F0596P, F0596Z, F05BRC, B3XZ5C, B411H0, F07D09, F07RYJ, F07ZZ8, F0GKW3, F0JBPD, F08R3D, B3FL42, B3FMD2, F06PPQ, F08MGM, F0596X, F05BR4, F0FP1G, F0GPH7, F09Z32, F0842D, F08H3G, F0596H, F05HX3, F0GJCT, F0HYTJ, F0JDZJ, F0JL27, F0FKJP, B3Y03J, F09GGF, F0HX2Q, F0J9YK, F06GX8, B3XZMK, B3XZQM, B3XZQX, B3XZR0, B3XZR1, F0GT04, F07158, F0KNDC, F0KNDH, F0FQTC, F0FRHW, F07GL1, F04MY3, B411GW, B411GY, B41GN3, F0GKWD, B3XYRT, B3XZ4L, F07K8F, F07P1X, B3Y039, F06PPN, F0K7LV, F0K7M5, B3Y00H, F05FKN, F0L40K, F07P2D, B3XZRL, B3FM2F, B3FL20, F094DW, F09FZF, F09JNL, F09JNP, F09MCJ, F09PTQ, F09PTV, F09PV3, F09PV7, F09Z30, B3Y01M, F0FYBX, B3XZRQ, B3Y035, F0K1LZ, F07K89, F081V5, F081W3, F0841J, F0716B, F04QBJ, B3FLDP, F08LDH, F09BZK, F09D1W, F0GCP6, F0GJCD, B3FLWG, F09BZP, F09D1G, F09MC6, F09MC9, F0KV7M, F0KXZM, F0G0M0, B3FLD9, B3FLDB, B3FLDH, B3FLFJ, B3FLWC, B3XZT4, B3XZYL, F07GKC, F07K8R, F07RY0, F07ZZG, F0GJCM, F0L7PV, F09MC2, B3XZ21, B3XZMM, F0FRJ2, B3FLTL, F06TGB, B3FKZB, B3Y00M, B3Y00T, F04XDP, F0HFNJ, F05V8R, F0GKW9, F07P2R, F0GPH6, B3FL8K, F05BR8, F07GKV, F04QC5, F09BZG, F09D1L, F0715Z, B3LBBX, B3XZRV, F0FRHT, F07D0H, F06XQQ, F08H3T, F04YV1, F06GWJ, B3FMGZ, B3FMC3, F0L7Q8, F0L40M, F07ZZJ, F09D23, F05GQQ, B3FLKT, B3FLMF, B3FLTV, F06GWK, F06TFL, F07RYK, F07RYX, F0FKJT, F0FKJV, B3Y0LR, F0GKVY, B3FMF2, B3FMFJ, F06TG2, B3FM2N, F09D27, F04QBQ, B3XZ1Q, B3XZ45, B3XZQY, F07P23, F07ZZH, F08MGV, F09BZX, F09MCD, F0J4YV, F09FYV, F09MCF, F09Z35, F07RYC, F0KQJH, F0KQJM, F0KQJP, F0715J, F08LDQ, F0JZQ9, F07WWB, F04V38, B3FMGG, B3FMH2, B3FMH6, F07WX0, F0FKJL, F0FP18, F0JZQ8, F0841T, F08LD8, F09D1V, F0JL2D, F04XF0, F06GX2, B3XZFQ, B3FLY6, F06LJY, F06LKB, F04V3F, F0FKJ9, F0FKK7, F0550R, F09D29, F0HP35, F07D08, B3XZRP, B3XZY0, F0596R, F0596W, B3FMC5, F07K8B, F07RY4, F08H3N, B3FLYH, B3FM50, B3XZXX, F09FZ6, F07K8D, F07GKK, F06GWB, B3XZ4R, F07D0B, F06PPZ, F06TFN, F06XQY, F0F4BZ, F0KQJQ, F0KXZX, B3XZYQ, F0596T, F05BR5, F06KRB, F06KRP, F0FQTN, F0FYBT, F0FYBZ, F0FYC0, F07P27, F07P2N, F07WWG, F081V8, F081VK, F081VZ, F081W0, B3XZ51, B3XZ58, B3XZFM, F05V8V, B3FMGN, B3XZ2R, B411H6, F07GKR, B3FMGX, F04XFB, F07D04, B3XZHJ, F09D28, F09GFT, B3FLD6, F06TGG, F07CZX, F0HV5T, B3XZ4T, F0G5YB, B41GNZ, B3XZYG, B3XZHK, F08MGX, F0B94T, F0L9T9, F0JWZ8, F07K8P, B3XZ5R, F05HX5, F0FRJ1, B411GZ, B3Y01N, F0JWZ4, B41GR7, B3Y01X, B3Y0GJ, F0FKKB, F0FQTD, F06LJF, B3Y0GN, B3Y0KD, B3Y0K2, F0KC7D, F0L9T1, B411J1, B41GQY, F0B945, F0FP10, F07K8H, B41GN4, B41GNB, B41GR2, B3XZXN, F06LJJ, F07165, F0JWZ6, B3FMH0, B3FMGY, F0841W, F0841X, F05HX2, F07GKT, F07K8W, F07RY9, F0FKK2, F06PPD, F05GQH, B3XYRK, F0715G, B3FLRP, B3FLTD, F04V39, B3FM7Q, B3FM7R, B3Y0L9, B3Y0LC, B3Y0LF, B3Y0LG, F081VH, F0K1LP, F07ZZ4, F09PV6, F09PVF, F06XQ9, F0841P, F09FYW, F09FZ0, F09FZ4, F09PV2, B3FM53, B3FM7P, B3FLFK, F09PV9, B3FL4C, F07K8L, F09GFY, F07RY2, F09PTP, F06TGC, B3FLMR, B3FM4Z, F0FP0Z, F081W1, F098FR, F0KC77, B3Y01L, B3Y01Y, F0K1LX, F0GV23, B3FML4, B411H8, B411HG, F0JRVL, B3Y00R, B3XZ54, F0L7QC, B3XZMJ, B3XZQW, F05HX4, F06GX3, F0F163, F08R34, F07P2H, F0GR2Z, B3FL0V, F0841Q, F08MGJ, F0K1LK, F0GJCB, B3FLTJ, F05FK9, F0GJC8, B3FLXM, B3XZMR, F07GKL, B4161W, F0LC8Q, F094Y6, F09BZF, B3XZRR, F05BR1, F0J6DJ, F0JDZT, B3FLRZ, F04YTQ, F0550M, B3XZ2N, B3XZ48, B3FMH1, B3FM5D, B411HF, B41GNQ, B3Y01Q, F0F4BG, B3XZY2, B3XZYD, B3FL4H, F05BQW, B3FLXJ, B3XZXW, F0G34B, F06PQ1, B3LBF9, F05HXC, F0GV20, F0GV25, F0HFND, B3XYZ2, B3XZ1V, B411HD, B411HW, B3FM2M, B3FM7B, F0GPH9, F0GR3C, B41GQN, B3XZ4N, B3Y033, F0550Q, B3FLBW, B3LBBQ, B3FL65, F04XDM, F0K7LX, F0FQTG, F06PPL, F0KXZZ, F0L02V, F0L09V, F0L15C, F0L15H, F0L15M, F0L15N, F0L40J, F0L40P, F0L40Q, F0FYBY, F0G34G, F0G5Y5, F0G5YC, F0GFPX, F0GFPY, F0F4BL, F07P2T, F0715R, F0841Z, F08425, B3FMBP, B3FMF5, B3LBG0, F0KND6, B3FL67, B3FL62, B411HK, B3FM52, B3FL5Z, F07RYR, F06GWL, B3Y0LQ, B3FLT6, F0GR30, B411HT, F0715V, F06GW9, F06PPY, F0J4HJ, F0J4HM, F0J6DG, F0J6DK, B41GPD, F08420, B3XZVM, F0841F, B3Y0K7, F07ZZB, F081VB, F08H3L, F0FYC1, F0F4BD, F0F4C3, F0F4C5, F0GT02, F0HFNK, B3XZ56, B3XZVY, F09D1N, B41GP6, B41GN2, F0715N, F05BQZ, B41GR6, F0KR40, F06GWD, F06TFQ, F08LDG, F06GWG, F07WW6, F06TFH, F06XQG, F08423, F08LDN, F08MGP, F07155, F0HFNG, F07RY5, F07P24, F04QBZ, F07RYF, F0K7M2, F0KK1P, F09FZ7, F09MC4, F07D01, F09BZV, B3FLTT, F0KK1Q, F0KK1R, B3LBF6, F09PV8, F09MCC, F0GV22, F0GV21, F0GV28, F0FKJM, B3XYYZ, B4161J, F0841R, F094Y5, F04YTD, F081VC, F081VT, F0JDZM, F0JDZP, F0G8CF, F0GFPF, F0G0LY, F0G0M2, F0GKW8, F0HVCD, F04YT8, F0JPPV, F06GWV, B3FLXL, B3FMDW. |
| FEI Number |
2518586
|
Recalling Firm/ Manufacturer |
Philips Ultrasound, Inc 1 Echo Dr Reedsville PA 17084-8603
|
| For Additional Information Contact | Philips Customer Care Solutions Center 1-800-722-9377 |
Manufacturer Reason for Recall | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
FDA Determined Cause 2 | Device Design |
| Action | An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377. |
| Quantity in Commerce | 171,322 units |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = IYN
|
|
|
|