U.S. Department of Health and Human Services

Class 2 Device Recall CARESAFE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CARESAFEsee related information
Date Initiated by FirmOctober 29, 2025
Date PostedNovember 26, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0623-2026
Recall Event ID 97816
510(K)NumberK191658 
Product Classification Set, administration, intravascular - Product Code FPA
ProductCARESAFE utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. CARESAFE IV ADMIN SET w/AIRSTOP, 104 IN.; Catalog Number: 354302. 2. CARESAFE ADMIN SET w/15um FILTER, 104 IN; Catalog Number: 354303. 3. CARESAFE IV ADMIN SET w/AIRSTOP, 114 IN.; Catalog Number: 354304. 4. CARESAFE ADMIN SET w/15um FILTER, 114 IN; Catalog Number: 354305. 5. CARESAFE IV ADMIN SET w/AIRSTOP, 110 IN.; Catalog Number: 354306. 6. CARESAFE ADMIN SET w/15um FILTER, 110 IN; Catalog Number: 354307. 7. CARESAFE" IV Administration Set with AirStop Filter, Not made with PVC or DEHP, 2 CARESITE Injection Sites, 2 ULTRAPORT High-FlowFour-Way Stopcocks; CARESAFE ANESTH SET W/AIRSTOP, 134 IN.; Catalog Number: 354308. 8. CARESAFE ANESTH SET w/15um FILTER,134IN.; Catalog Number: 354310.
Code Information 1. Catalog Number: 354302; Primary UDI-DI: 04046955259946; Unit of Dose UDI-DI: 04046955259939. 2. Catalog Number: 354303; Primary UDI-DI: 04046955052424; Unit of Dose UDI-DI: 04046955052417. 3. Catalog Number: 354304; Primary UDI-DI: 04046955259960; Unit of Dose UDI-DI: 04046955259953. 4. Catalog Number: 354305; Primary UDI-DI: 04046955052448; Unit of Dose UDI-DI: 04046955052431. 5. Catalog Number: 354306; Primary UDI-DI: 04046955259984; Unit of Dose UDI-DI: 04046955259977. 6. Catalog Number: 354307; Primary UDI-DI: 04046955052462; Unit of Dose UDI-DI: 04046955052455. 7. Catalog Number: 354308; Primary UDI-DI: 04046955260003; Unit of Dose UDI-DI: 04046955259991. 8. Catalog Number: 354310; Primary UDI-DI: 04046955052929; Unit of Dose UDI-DI: 04046955052905.
FEI Number 2521402
Recalling Firm/
Manufacturer
B Braun Medical Inc
824 12th Ave
Bethlehem PA 18018-3524
For Additional Information ContactAllison Longenhagen
001-484-2408373
Manufacturer Reason
for Recall
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA Determined
Cause 2
Process control
ActionB Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."
Quantity in Commerce4,438 units
DistributionWorldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.