| | Class 2 Device Recall Philips Allura Xper Fluoroscopic Xray System |  |
| Date Initiated by Firm | December 15, 2025 |
| Date Posted | January 13, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1062-2026 |
| Recall Event ID |
98109 |
| 510(K)Number | K102005 K130638 K130842 K133292 K141979 K161563 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Product Name: Allura Xper FD10C;
Model number: 722001; |
| Code Information |
Model Number: 722001;
System Product Name: Allura Xper FD10C;
UDI-DI: N/A;
Serial Numbers: 166, 181, 153, 274, 296, 267, 206, 247, 161, 282, 213, 284, 103, 199, 129, 95, 299, 30, 235, 221, 348, 283, 276, 253, 1100, 78, 229, 232, 159, 177, 219, 217, 205, 3, 89, 101, 295, 305; |
| FEI Number |
3002807880
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
|
Manufacturer Reason for Recall | Philips has identified that in a limited number of impacted systems the drip tray
beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately
documented. If it is not installed, there is a potential for coolant liquid to contact electrical
components. This could lead to electrical short-circuits, which may trigger system fuses and result in
system shutdown. |
FDA Determined Cause 2 | Other |
| Action | On December 15, 2025, URGENT Medical Device Correction letters were sent to customers. Actions to be taken:
a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU).
b. Circulate this letter to all users so that they are aware of the issue. Keep this letter with the documentation of the system until Philips inspects and, where needed, corrects your system.
c. Establish an emergency protocol prior to all applicable diagnostics, interventional and minimally invasive procedures to manage the situation should you experience the issue during a procedure.
d. If you detect a burning odor, smoke, or fire originating from the R-cabinet in the technical room, immediately switch off the system s main power. Do not operate or use the system further. Follow your institution s emergency procedures.
e. In case the affected system has been transferred to another organization, please send a copy of this letter to that organization and inform Philips about this transfer through your local Philips representative.
f. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt.
g. Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative.
|
| Quantity in Commerce | 38 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|