| | Class 2 Device Recall Zimmer Tourniquet Systems |  |
| Date Initiated by Firm | December 24, 2025 |
| Date Posted | January 28, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1201-2026 |
| Recall Event ID |
98235 |
| 510(K)Number | K202919 |
| Product Classification |
Tourniquet, pneumatic - Product Code KCY
|
| Product | Brand Name: Zimmer Tourniquet Systems
Product Name: A.T.S. 5000TS Tourniquet Systems
Model/Catalog Number: 60500010100
Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S. 5000TS
Product Description: A.T.S 5000TS Tourniquet Systems
Component: N/A |
| Code Information |
Lot Code: Model No. 60500010100 ; UDI (01)00889024626454(11)240410(10)66722609(21)50066722609001; Lot No. 66722609 ; Serial No. 50066722609001
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609002; Lot No. 66722609 ; Serial No. 50066722609002
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609003" ; Lot No. 66722609 ; Serial No. 50066722609003
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609004" ; Lot No. 66722609 ; Serial No. 50066722609004
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722609(21)50066722609005" ; Lot No. 66722609 ; Serial No. 50066722609005
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610001" ; Lot No. 66722610 ; Serial No. 50066722610001
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610002" ; Lot No. 66722610 ; Serial No. 50066722610002
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610003" ; Lot No. 66722610 ; Serial No. 50066722610003
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610004" ; Lot No. 66722610 ; Serial No. 50066722610004
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722610(21)50066722610005" ; Lot No. 66722610 ; Serial No. 50066722610005
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611001" ; Lot No. 66722611 ; Serial No. 50066722611001
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611002" ; Lot No. 66722611 ; Serial No. 50066722611002
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611003" ; Lot No. 66722611 ; Serial No. 50066722611003
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611004" ; Lot No. 66722611 ; Serial No. 50066722611004
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611005" ; Lot No. 66722611 ; Serial No. 50066722611005
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611006" ; Lot No. 66722611 ; Serial No. 50066722611006
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611007" ; Lot No. 66722611 ; Serial No. 50066722611007
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611008" ; Lot No. 66722611 ; Serial No. 50066722611008
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611009" ; Lot No. 66722611 ; Serial No. 50066722611009
Model No. 60500010100 ; UDI "(01)00889024626454(11)240410(10)66722611(21)50066722611010" ; Lot No. 66722611 ; Serial No. 50066722611010
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892001" ; Lot No. 66742892 ; Serial No. 50066742892001
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892002" ; Lot No. 66742892 ; Serial No. 50066742892002
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892003" ; Lot No. 66742892 ; Serial No. 50066742892003
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892004" ; Lot No. 66742892 ; Serial No. 50066742892004
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892005" ; Lot No. 66742892 ; Serial No. 50066742892005
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892006" ; Lot No. 66742892 ; Serial No. 50066742892006
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892007" ; Lot No. 66742892 ; Serial No. 50066742892007
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892008" ; Lot No. 66742892 ; Serial No. 50066742892008
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892009" ; Lot No. 66742892 ; Serial No. 50066742892009
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892010" ; Lot No. 66742892 ; Serial No. 50066742892010
Model No. 60500010100 ; UDI "(01)00889024626454(11)240422(10)66742892(21)50066742892016" ; Lot No. 66742892 ; Serial No. 50066742892016
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893001" ; Lot No. 66742893 ; Serial No. 50066742893001
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893002" ; Lot No. 66742893 ; Serial No. 50066742893002
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893003" ; Lot No. 66742893 ; Serial No. 50066742893003
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893004" ; Lot No. 66742893 ; Serial No. 50066742893004
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893005" ; Lot No. 66742893 ; Serial No. 50066742893005
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893006" ; Lot No. 66742893 ; Serial No. 50066742893006
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893007" ; Lot No. 66742893 ; Serial No. 50066742893007
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893008" ; Lot No. 66742893 ; Serial No. 50066742893008
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893009" ; Lot No. 66742893 ; Serial No. 50066742893009
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893010" ; Lot No. 66742893 ; Serial No. 50066742893010
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893011" ; Lot No. 66742893 ; Serial No. 50066742893011
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893012" ; Lot No. 66742893 ; Serial No. 50066742893012
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893013" ; Lot No. 66742893 ; Serial No. 50066742893013
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893014" ; Lot No. 66742893 ; Serial No. 50066742893014
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893015" ; Lot No. 66742893 ; Serial No. 50066742893015
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893016" ; Lot No. 66742893 ; Serial No. 50066742893016
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893017" ; Lot No. 66742893 ; Serial No. 50066742893017
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893018" ; Lot No. 66742893 ; Serial No. 50066742893018
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893019" ; Lot No. 66742893 ; Serial No. 50066742893019
Model No. 60500010100 ; UDI "(01)00889024626454(11)240423(10)66742893(21)50066742893020" ; Lot No. 66742893 ; Serial No. 50066742893020
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087002" ; Lot No. 66751087 ; Serial No. 50066751087002
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087003" ; Lot No. 66751087 ; Serial No. 50066751087003
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087004" ; Lot No. 66751087 ; Serial No. 50066751087004
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087007" ; Lot No. 66751087 ; Serial No. 50066751087007
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087008" ; Lot No. 66751087 ; Serial No. 50066751087008
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087009" ; Lot No. 66751087 ; Serial No. 50066751087009
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087010" ; Lot No. 66751087 ; Serial No. 50066751087010
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087011" ; Lot No. 66751087 ; Serial No. 50066751087011
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087012" ; Lot No. 66751087 ; Serial No. 50066751087012
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087013" ; Lot No. 66751087 ; Serial No. 50066751087013
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087014" ; Lot No. 66751087 ; Serial No. 50066751087014
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087016" ; Lot No. 66751087 ; Serial No. 50066751087016
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087017" ; Lot No. 66751087 ; Serial No. 50066751087017
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087018" ; Lot No. 66751087 ; Serial No. 50066751087018
Model No. 60500010100 ; UDI "(01)00889024626454(11)240528(10)66751087(21)50066751087019" ; Lot No. 66751087 ; Serial No. 50066751087019
Model No. 60500010100 ; UDI "(01)00889024626454(11)240723(10)66848797(21)50066848797005" ; Lot No. 66848797 ; Serial No. 50066848797005
Model No. 60500010100 ; UDI "(01)00889024626454(11)240723(10)66848797(21)50066848797007" ; Lot No. 66848797 ; Serial No. 50066848797007
Model No. 60500010100 ; UDI "(01)00889024626454(11)240723(10)66848797(21)50066848797008" ; Lot No. 66848797 ; Serial No. 50066848797008
Model No. 60500010100 ; UDI "(01)00889024626454(11)240723(10)66848797(21)50066848797014" ; Lot No. 66848797 ; Serial No. 50066848797014
Model No. 60500010100 ; UDI "(01)00889024626454(11)240723(10)66848797(21)50066848797015" ; Lot No. 66848797 ; Serial No. 50066848797015
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799001" ; Lot No. 66848799 ; Serial No. 50066848799001
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799002" ; Lot No. 66848799 ; Serial No. 50066848799002
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799003" ; Lot No. 66848799 ; Serial No. 50066848799003
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799004" ; Lot No. 66848799 ; Serial No. 50066848799004
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799005" ; Lot No. 66848799 ; Serial No. 50066848799005
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799006" ; Lot No. 66848799 ; Serial No. 50066848799006
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799007" ; Lot No. 66848799 ; Serial No. 50066848799007
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799008" ; Lot No. 66848799 ; Serial No. 50066848799008
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799009" ; Lot No. 66848799 ; Serial No. 50066848799009
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799010" ; Lot No. 66848799 ; Serial No. 50066848799010
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799011" ; Lot No. 66848799 ; Serial No. 50066848799011
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799012" ; Lot No. 66848799 ; Serial No. 50066848799012
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799013" ; Lot No. 66848799 ; Serial No. 50066848799013
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799014" ; Lot No. 66848799 ; Serial No. 50066848799014
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799015" ; Lot No. 66848799 ; Serial No. 50066848799015
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799016" ; Lot No. 66848799 ; Serial No. 50066848799016
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799017" ; Lot No. 66848799 ; Serial No. 50066848799017
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799018" ; Lot No. 66848799 ; Serial No. 50066848799018
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799019" ; Lot No. 66848799 ; Serial No. 50066848799019
Model No. 60500010100 ; UDI "(01)00889024626454(11)240730(10)66848799(21)50066848799020" ; Lot No. 66848799 ; Serial No. 50066848799020
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330010" ; Lot No. 66870330 ; Serial No. 50066870330010
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330011" ; Lot No. 66870330 ; Serial No. 50066870330011
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330012" ; Lot No. 66870330 ; Serial No. 50066870330012
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330013" ; Lot No. 66870330 ; Serial No. 50066870330013
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330014" ; Lot No. 66870330 ; Serial No. 50066870330014
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330015" ; Lot No. 66870330 ; Serial No. 50066870330015
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330016" ; Lot No. 66870330 ; Serial No. 50066870330016
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330017" ; Lot No. 66870330 ; Serial No. 50066870330017
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330018" ; Lot No. 66870330 ; Serial No. 50066870330018
Model No. 60500010100 ; UDI "(01)00889024626454(11)240808(10)66870330(21)50066870330019" ; Lot No. 66870330 ; Serial No. 50066870330019
Model No. 60500010100 ; UDI "(01)00889024626454(11)240905(10)66959543(21)50066959543009" ; Lot No. 66959543 ; Serial No. 50066959543009
Model No. 60500010100 ; UDI "(01)00889024626454(11)240910(10)66969025(21)50066969025001" ; Lot No. 66969025 ; Serial No. 50066969025001
Model No. 60500010100 ; UDI "(01)00889024626454(11)240910(10)66969025(21)50066969025006" ; Lot No. 66969025 ; Serial No. 50066969025006
Model No. 60500010100 ; UDI "(01)00889024626454(11)240910(10)66969025(21)50066969025010" ; Lot No. 66969025 ; Serial No. 50066969025010
Model No. 60500010100 ; UDI "(01)00889024626454(11)240910(10)66969025(21)50066969025011" ; Lot No. 66969025 ; Serial No. 50066969025011
Model No. 60500010100 ; UDI "(01)00889024626454(11)240910(10)66969025(21)50066969025016" ; Lot No. 66969025 ; Serial No. 50066969025016
Model No. 60500010100 ; UDI "(01)00889024626454(11)240918(10)66969027(21)50066969027008" ; Lot No. 66969027 ; Serial No. 50066969027008
Model No. 60500010100 ; UDI "(01)00889024626454(11)241021(10)66995371(21)50066995371013" ; Lot No. 66995371 ; Serial No. 50066995371013
Model No. 60500010100 ; UDI "(01)00889024626454(11)241021(10)66995371(21)50066995371014" ; Lot No. 66995371 ; Serial No. 50066995371014
Model No. 60500010100 ; UDI "(01)00889024626454(11)241106(10)67046297(21)50067046297014" ; Lot No. 67046297 ; Serial No. 50067046297014
Model No. 60500010100 ; UDI "(01)00889024626454(11)241106(10)67046298(21)50067046298014" ; Lot No. 67046298 ; Serial No. 50067046298014
Model No. 60500010100 ; UDI "(01)00889024626454(11)241106(10)67046298(21)50067046298015" ; Lot No. 67046298 ; Serial No. 50067046298015
Model No. 60500010100 ; UDI "(01)00889024626454(11)241106(10)67046298(21)50067046298016" ; Lot No. 67046298 ; Serial No. 50067046298016
Model No. 60500010100 ; UDI "(01)00889024626454(11)241106(10)67046298(21)50067046298018" ; Lot No. 67046298 ; Serial No. 50067046298018
Model No. 60500010100 ; UDI "(01)00889024626454(11)241106(10)67046298(21)50067046298019" ; Lot No. 67046298 ; Serial No. 50067046298019
Model No. 60500010100 ; UDI "(01)00889024626454(11)241106(10)67046298(21)50067046298020" ; Lot No. 67046298 ; Serial No. 50067046298020
Model No. 60500010100 ; UDI "(01)00889024626454(11)241115(10)67071240(21)50067071240003" ; Lot No. 67071240 ; Serial No. 50067071240003
Model No. 60500010100 ; UDI "(01)00889024626454(11)241115(10)67071240(21)50067071240005" ; Lot No. 67071240 ; Serial No. 50067071240005
Model No. 60500010100 ; UDI "(01)00889024626454(11)241115(10)67071240(21)50067071240008" ; Lot No. 67071240 ; Serial No. 50067071240008
Model No. 60500010100 ; UDI "(01)00889024626454(11)241115(10)67071240(21)50067071240011" ; Lot No. 67071240 ; Serial No. 50067071240011
Model No. 60500010100 ; UDI "(01)00889024626454(11)241115(10)67071240(21)50067071240012" ; Lot No. 67071240 ; Serial No. 50067071240012
Model No. 60500010100 ; UDI "(01)00889024626454(11)241115(10)67071240(21)50067071240014" ; Lot No. 67071240 ; Serial No. 50067071240014
Model No. 60500010100 ; UDI "(01)00889024626454(11)241115(10)67071240(21)50067071240018" ; Lot No. 67071240 ; Serial No. 50067071240018
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765001" ; Lot No. 67117765 ; Serial No. 50067117765001
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765002" ; Lot No. 67117765 ; Serial No. 50067117765002
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765003" ; Lot No. 67117765 ; Serial No. 50067117765003
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765004" ; Lot No. 67117765 ; Serial No. 50067117765004
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765005" ; Lot No. 67117765 ; Serial No. 50067117765005
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765007" ; Lot No. 67117765 ; Serial No. 50067117765007
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765008" ; Lot No. 67117765 ; Serial No. 50067117765008
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765009" ; Lot No. 67117765 ; Serial No. 50067117765009
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765010" ; Lot No. 67117765 ; Serial No. 50067117765010
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765011" ; Lot No. 67117765 ; Serial No. 50067117765011
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765019" ; Lot No. 67117765 ; Serial No. 50067117765019
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117765(21)50067117765020" ; Lot No. 67117765 ; Serial No. 50067117765020
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117766(21)50067117766002" ; Lot No. 67117766 ; Serial No. 50067117766002
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117766(21)50067117766007" ; Lot No. 67117766 ; Serial No. 50067117766007
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117766(21)50067117766008" ; Lot No. 67117766 ; Serial No. 50067117766008
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117766(21)50067117766013" ; Lot No. 67117766 ; Serial No. 50067117766013
Model No. 60500010100 ; UDI "(01)00889024626454(11)241207(10)67117766(21)50067117766014" ; Lot No. 67117766 ; Serial No. 50067117766014
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768002" ; Lot No. 67117768 ; Serial No. 50067117768002
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768004" ; Lot No. 67117768 ; Serial No. 50067117768004
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768005" ; Lot No. 67117768 ; Serial No. 50067117768005
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768006" ; Lot No. 67117768 ; Serial No. 50067117768006
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768007" ; Lot No. 67117768 ; Serial No. 50067117768007
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768008" ; Lot No. 67117768 ; Serial No. 50067117768008
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768009" ; Lot No. 67117768 ; Serial No. 50067117768009
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768010" ; Lot No. 67117768 ; Serial No. 50067117768010
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768011" ; Lot No. 67117768 ; Serial No. 50067117768011
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768012" ; Lot No. 67117768 ; Serial No. 50067117768012
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768013" ; Lot No. 67117768 ; Serial No. 50067117768013
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768014" ; Lot No. 67117768 ; Serial No. 50067117768014
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768015" ; Lot No. 67117768 ; Serial No. 50067117768015
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768017" ; Lot No. 67117768 ; Serial No. 50067117768017
Model No. 60500010100 ; UDI "(01)00889024626454(11)241220(10)67117768(21)50067117768018" ; Lot No. 67117768 ; Serial No. 50067117768018
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667001" ; Lot No. 67482667 ; Serial No. 50067482667001
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667002" ; Lot No. 67482667 ; Serial No. 50067482667002
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667003" ; Lot No. 67482667 ; Serial No. 50067482667003
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667004" ; Lot No. 67482667 ; Serial No. 50067482667004
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667005" ; Lot No. 67482667 ; Serial No. 50067482667005
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667006" ; Lot No. 67482667 ; Serial No. 50067482667006
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667007" ; Lot No. 67482667 ; Serial No. 50067482667007
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667008" ; Lot No. 67482667 ; Serial No. 50067482667008
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667009" ; Lot No. 67482667 ; Serial No. 50067482667009
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667010" ; Lot No. 67482667 ; Serial No. 50067482667010
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667012" ; Lot No. 67482667 ; Serial No. 50067482667012
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667013" ; Lot No. 67482667 ; Serial No. 50067482667013
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667014" ; Lot No. 67482667 ; Serial No. 50067482667014
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667015" ; Lot No. 67482667 ; Serial No. 50067482667015
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667016" ; Lot No. 67482667 ; Serial No. 50067482667016
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667017" ; Lot No. 67482667 ; Serial No. 50067482667017
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667018" ; Lot No. 67482667 ; Serial No. 50067482667018
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667019" ; Lot No. 67482667 ; Serial No. 50067482667019
Model No. 60500010100 ; UDI "(01)00889024626454(11)250729(10)67482667(21)50067482667020" ; Lot No. 67482667 ; Serial No. 50067482667020
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336002" ; Lot No. 67533336 ; Serial No. 50067533336002
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336003" ; Lot No. 67533336 ; Serial No. 50067533336003
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336004" ; Lot No. 67533336 ; Serial No. 50067533336004
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336005" ; Lot No. 67533336 ; Serial No. 50067533336005
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336006" ; Lot No. 67533336 ; Serial No. 50067533336006
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336007" ; Lot No. 67533336 ; Serial No. 50067533336007
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336008" ; Lot No. 67533336 ; Serial No. 50067533336008
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336009" ; Lot No. 67533336 ; Serial No. 50067533336009
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336010" ; Lot No. 67533336 ; Serial No. 50067533336010
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336012" ; Lot No. 67533336 ; Serial No. 50067533336012
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336013" ; Lot No. 67533336 ; Serial No. 50067533336013
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336016" ; Lot No. 67533336 ; Serial No. 50067533336016
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336017" ; Lot No. 67533336 ; Serial No. 50067533336017
Model No. 60500010100 ; UDI "(01)00889024626454(11)250808(10)67533336(21)50067533336020" ; Lot No. 67533336 ; Serial No. 50067533336020
Model No. 60500010100 ; UDI "(01)00889024626454(11)250902(10)67542546(21)50067542546012" ; Lot No. 67542546 ; Serial No. 50067542546012
Model No. 60500010100 ; UDI "(01)00889024626454(11)250902(10)67542546(21)50067542546018" ; Lot No. 67542546 ; Serial No. 50067542546018
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349001" ; Lot No. 67553349 ; Serial No. 50067553349001
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349004" ; Lot No. 67553349 ; Serial No. 50067553349004
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349006" ; Lot No. 67553349 ; Serial No. 50067553349006
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349007" ; Lot No. 67553349 ; Serial No. 50067553349007
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349008" ; Lot No. 67553349 ; Serial No. 50067553349008
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349009" ; Lot No. 67553349 ; Serial No. 50067553349009
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349010" ; Lot No. 67553349 ; Serial No. 50067553349010
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349011" ; Lot No. 67553349 ; Serial No. 50067553349011
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349012" ; Lot No. 67553349 ; Serial No. 50067553349012
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349014" ; Lot No. 67553349 ; Serial No. 50067553349014
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349016" ; Lot No. 67553349 ; Serial No. 50067553349016
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349017" ; Lot No. 67553349 ; Serial No. 50067553349017
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349018" ; Lot No. 67553349 ; Serial No. 50067553349018
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349019" ; Lot No. 67553349 ; Serial No. 50067553349019
Model No. 60500010100 ; UDI "(01)00889024626454(11)250822(10)67553349(21)50067553349020" ; Lot No. 67553349 ; Serial No. 50067553349020
|
| FEI Number |
1526350
|
Recalling Firm/ Manufacturer |
Zimmer Surgical Inc 200 W Ohio Ave Dover OH 44622-9642
|
| For Additional Information Contact | 411 Technical Services 800-6136131 |
Manufacturer Reason for Recall | Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On December 30, 2025, the firm began notifying consignees through Urgent Medical Device Recall letters.
The firm is conducting this recall in two phases. In Phase 1, which began on 12/30/2025, consignees are informed that the screen may freeze or stop responding to touch when used in certain international languages. Users may continue to use the device by utilizing the English language mode. Use of the device in an affected language is not advised, and may cause the issue to occur.
ATS5000 Affected Languages:
Czech, Danish, Finnish, French, Italian, Lithuanian, Norwegian, Romanian, Slovak, Slovenian, and Turkish
Unaffected Languages:
English, Chinese (simplified), Korean, Japanese, Greek, Bulgarian, and Russian
All devices manufactured between April 2024 - September 2025 with software version v2.04 are within scope of the recall. Affected devices will require an upgrade to the new software version v2.05 when it is available in early 2026.
The firm will issue a removal notice (Phase 2) once the new software version is available. Customers are instructed to not return affected devices before the initiation of Phase 2. |
| Quantity in Commerce | 206 |
| Distribution | International distribution to the countries of Canada and EMEA only. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KCY
|
|
|
|