 Class 2 Device Recall Zimmer Tourniquet Systemssee related information
Date Initiated by FirmDecember 24, 2025
Date PostedJanuary 28, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1202-2026
Recall Event ID 98235
510(K)NumberK050411 
Product Classification Tourniquet, pneumatic - Product Code KCY
ProductBrand Name: Zimmer Tourniquet Systems Product Name: A.T.S. 3200TS Tourniquet Systems Model/Catalog Number: 60320010100 Software Version: Software version v2.06 or prior are within scope of the recall for A.T.S. 3200TS Product Description: A.T.S 3200TS Tourniquet Systems Component: N/A
Code Information Lot Code: Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739001" ; Lot No. 67505739 ; Serial No. 32067505739001 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739003" ; Lot No. 67505739 ; Serial No. 32067505739003 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739004" ; Lot No. 67505739 ; Serial No. 32067505739004 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739005" ; Lot No. 67505739 ; Serial No. 32067505739005 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739006" ; Lot No. 67505739 ; Serial No. 32067505739006 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739007" ; Lot No. 67505739 ; Serial No. 32067505739007 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739008" ; Lot No. 67505739 ; Serial No. 32067505739008 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739010" ; Lot No. 67505739 ; Serial No. 32067505739010 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739013" ; Lot No. 67505739 ; Serial No. 32067505739013 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739014" ; Lot No. 67505739 ; Serial No. 32067505739014 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739015" ; Lot No. 67505739 ; Serial No. 32067505739015 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739017" ; Lot No. 67505739 ; Serial No. 32067505739017 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739019" ; Lot No. 67505739 ; Serial No. 32067505739019 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739020" ; Lot No. 67505739 ; Serial No. 32067505739020 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816001" ; Lot No. 67508816 ; Serial No. 32067508816001 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816002" ; Lot No. 67508816 ; Serial No. 32067508816002 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816003" ; Lot No. 67508816 ; Serial No. 32067508816003 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816004" ; Lot No. 67508816 ; Serial No. 32067508816004 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816005" ; Lot No. 67508816 ; Serial No. 32067508816005 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816006" ; Lot No. 67508816 ; Serial No. 32067508816006 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816007" ; Lot No. 67508816 ; Serial No. 32067508816007 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816008" ; Lot No. 67508816 ; Serial No. 32067508816008 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816009" ; Lot No. 67508816 ; Serial No. 32067508816009 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816010" ; Lot No. 67508816 ; Serial No. 32067508816010 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816012" ; Lot No. 67508816 ; Serial No. 32067508816012 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816013" ; Lot No. 67508816 ; Serial No. 32067508816013 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816014" ; Lot No. 67508816 ; Serial No. 32067508816014 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816016" ; Lot No. 67508816 ; Serial No. 32067508816016 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816017" ; Lot No. 67508816 ; Serial No. 32067508816017 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816019" ; Lot No. 67508816 ; Serial No. 32067508816019 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816020" ; Lot No. 67508816 ; Serial No. 32067508816020 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820001" ; Lot No. 67508820 ; Serial No. 32067508820001 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820002" ; Lot No. 67508820 ; Serial No. 32067508820002 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820003" ; Lot No. 67508820 ; Serial No. 32067508820003 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820004" ; Lot No. 67508820 ; Serial No. 32067508820004 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820005" ; Lot No. 67508820 ; Serial No. 32067508820005 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820006" ; Lot No. 67508820 ; Serial No. 32067508820006 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820007" ; Lot No. 67508820 ; Serial No. 32067508820007 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820008" ; Lot No. 67508820 ; Serial No. 32067508820008 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820009" ; Lot No. 67508820 ; Serial No. 32067508820009 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820010" ; Lot No. 67508820 ; Serial No. 32067508820010 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820011" ; Lot No. 67508820 ; Serial No. 32067508820011 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820012" ; Lot No. 67508820 ; Serial No. 32067508820012 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820013" ; Lot No. 67508820 ; Serial No. 32067508820013 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820014" ; Lot No. 67508820 ; Serial No. 32067508820014 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820015" ; Lot No. 67508820 ; Serial No. 32067508820015 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820016" ; Lot No. 67508820 ; Serial No. 32067508820016 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820017" ; Lot No. 67508820 ; Serial No. 32067508820017 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820018" ; Lot No. 67508820 ; Serial No. 32067508820018 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820019" ; Lot No. 67508820 ; Serial No. 32067508820019 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820020" ; Lot No. 67508820 ; Serial No. 32067508820020
FEI Number 1526350
Recalling Firm/
Manufacturer
Zimmer Surgical Inc
200 W Ohio Ave
Dover OH 44622-9642
For Additional Information Contact411 Technical Services
800-6136131
Manufacturer Reason
for Recall
Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn December 30, 2025, the firm began notifying consignees through Urgent Medical Device Recall letters. The firm is conducting this recall in two phases. In Phase 1, which began on 12/30/2025, consignees are informed that the screen may freeze or stop responding to touch when used in certain international languages. Users may continue to use the device by utilizing the English language mode. Use of the device in an affected language is not advised, and may cause the issue to occur. ATS5000 Affected Languages: Czech, Danish, Finnish, French, Italian, Lithuanian, Norwegian, Romanian, Slovak, Slovenian, and Turkish Unaffected Languages: English, Chinese (simplified), Korean, Japanese, Greek, Bulgarian, and Russian All devices manufactured between April 2024 - September 2025 with software version v2.04 are within scope of the recall. Affected devices will require an upgrade to the new software version v2.05 when it is available in early 2026. The firm will issue a removal notice (Phase 2) once the new software version is available. Customers are instructed to not return affected devices before the initiation of Phase 2.
Quantity in Commerce51
DistributionInternational distribution to the countries of Canada and EMEA only.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KCY
-
-